THE PAIN PILOT (A.K.A PAIN PILOT)

K120500 · Hollywog, LLC · NUH · Aug 20, 2012 · Neurology

Device Facts

Record IDK120500
Device NameTHE PAIN PILOT (A.K.A PAIN PILOT)
ApplicantHollywog, LLC
Product CodeNUH · Neurology
Decision DateAug 20, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5890
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The intended use is to provide approximately thirty minutes of analgesic electrical stimulus to reduce the perception of pain by electrically stimulating peripheral nerves across healthy intact skin of the user's lower back.

Device Story

Battery-powered transcutaneous electrical nerve stimulator (TENS) for OTC use; applies electrical stimulus to lower back via integral carbon rubber electrodes with replaceable hydrogel pads. Device features wireless remote control (key-fob) using RF transceiver for power on/off and intensity adjustment. User applies device directly to healthy intact skin of lower back. Provides 30-minute analgesic sessions. Benefits include pain reduction through peripheral nerve stimulation. Designed for patient self-use; eliminates wires between electrodes and generator found in traditional TENS units.

Clinical Evidence

Bench testing only. No clinical data provided. Testing included electrical performance characterization (current/power density), biocompatibility (ISO 10993-1, -5, -10), and adherence to safety standards (IEC 60601 series).

Technological Characteristics

Battery-powered (2x AAA) TENS unit. ABS plastic housing. Carbon rubber electrodes with hydrogel pads. Wireless RF remote control. Dimensions: 7.5" x 3.5" x 0.7". Max power density < 0.25 W/cm². Complies with IEC 60601-1, 60601-1-2, 60601-1-4, 60601-2-10, and ISO 10993 standards. Microprocessor-controlled with automatic 30-minute shut-off.

Indications for Use

Indicated for temporary relief of pain associated with sore and aching muscles in the lower back due to strain from exercise or normal household and work activities in adults.

Regulatory Classification

Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### 510(k) # K120500 Page 1 of 4 | 510(k) SUBMITTER: | Hollywog, LLC<br>2830 Amnicola Highway<br>Chattanooga, TN 37406 | |-----------------------------|-------------------------------------------------------------------------------------------------| | ESTABLISHMENT REGISTRATION: | 3008585473 | | CONTACT: | Michael W. Treas<br>Chief Compliance Officer | | DATE PREPARED: | August 16, 2012 | | PROPRIETARY NAME: | The Pain Pilot™ (a.k.a. Pain Pilot™) / WiTouch™ | | PANEL: | Neurology | | REGULATION NUMBER: | CFR Title 21, 882.5890 | | CLASSIFICATION: | Class II | | PRODUCT CODE: | NUH | | COMMON NAME: | Transcutaneous electrical nerve stimulator for pain<br>relief intended for over the counter use | ### Description: The device is a battery powered over the counter transcutaneous electrical nerve stimulator with a wireless remote control feature. The general purpose of the device is to apply an electrical stimulus to integral electrodes that are applied to a user's lower back to relieve pain. The device includes a current generator and permanently-affixed electrodes with user replaceable single patient use hydrogel pads (gel-pads) that may be reused for a limited number of reuses. The number of reuses of the gel-pads varv person-to-person dependent upon skin type, oils and pH levels. One side of the adhesive gel-pad adheres to the electrode, and the other the device to the healthy intact skin of the user's lower back to provide an analgesic electrical stimulus to the painful area. The user controls are located on a hand-held wireless remote control commonly referred to as a Key-fob. The user controls are simple straightforward power on/off and electrical stimulus intensity up/down. ### Indication for Use: The device is for temporary relief of pain associated with sore and aching muscles in the lower back due to strain from exercise or normal household and work activities. ### Intended Use: The intended use is to provide approximately thirty minutes of analgesic electrical stimulus to reduce the perception of pain by electrically stimulating peripheral nerves across healthy intact skin of the user's lower back. {1}------------------------------------------------ ### 510(k) # K120500 Page 2 of 4 ### Accessories: The device utilizes hydrogel pads) for achieving the indications for use and intended use. The composition of the gel-pads is common materials found in the electrode industry. The uniqueness of the gel-pads is in the shape. The maximum average power density of the electrodes with the gel-pads applied is less than 0.25 watts per square centimeter of electrode conductive surface area to reduce the risk of thermal burns which is consistent with the referenced predicate devices. | Predicate<br>Devices | 510(k)# | Proprietary Name | Device Classification Name(s), Regulation Number(s),<br>Classification(s) and Product Code(s). | |----------------------|---------|------------------------|------------------------------------------------------------------------------------------------| | #1 | K110068 | PM3030TM | CFR Title 21, Sec. 882.5890, Class II, NUH | | #2 | K090042 | PainmasterTM MCT Patch | CFR Title 21, Sec. 882.5890, Class II, NUH | | #3 | K060846 | T1040TM | CFR Title 21, Sec. 882.5890, Class II, NUH, NGX | ### Substantially Equivalent Predicate Devices The predicate devices PM3030" and T1040" utilize flexible wires between the electrodes and the electrical stimulus generator; thus, increases the indications for use to the lower back and to body surfaces with greater ranges of motion (e.g., knee, shoulder, elbow, and hip). The indications for use for the predicate device Painmaster™ MCT Patch is solely for application to the lower back; same as The Pain Pilot™ / WiTouch™ device. The intended design of The Pain Pilot™ / WiTouch" device limits the application for use to the anatomical site of the back. The design includes carbon rubber electrodes that are intended for reuse and are permanently-affixed to a rigid surface of the electrical stimulus generator. The unique connection makes the electrodes integral to the generator. The connection and shape of the electrodes limit application of the device to the contours of the back. These unique characteristic of the integral electrodes are insignificant as it relates to safety and effectiveness, and is not critical to the intended use between the device and the referenced predicate devices. The referenced predicate devices utilize affixed buttons as the sole method to control the electrical stimulus generator on/off and intensity up/down. The Pain Pilot" / WiTouch " device utilizes a wireless remote control radio frequency transceiver to control the electrical stimulus generator on/off and intensity up/down. The transceiver operates in the ISM radio frequency band for wireless medical technology. This uniqueness of controlling the electrical stimulus generator by utilizing a radio frequency transceiver is insignificant as it relates to safety and is not critical to the intended use between the device and the referenced predicate devices. The characteristics of the analgesic electrical stimulus between the device and the referenced predicate devices are substantially equivalent as it relates to safety and effectiveness. {2}------------------------------------------------ ### 510(k) Summary per 21 CFR 807.92(c) ## 510(k) # K120500 Page 3 of 4 | Substantial Equivalence | | | | | | |-------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|-------------------------------------|--------------------------------------------------------|------------------------------------|---------------| | 510(k) Number | K120500 | K110068 | K090042 | K060846 | | | Device Name and Model | The Pain Pilot™ | PM3030™ | Painmaster MCT Patch | T1040™ | | | Manufacturer | Hollywog, LLC | Omron Healthcare, Inc. | Newmark, Inc. | Endurance Therapeutics | | | TENS Device Power Source (DC battery) | Battery 1.5VDC<br>(2-Alkaline AAA) | Battery 1.5VDC<br>(3-Alkaline AAA) | No replaceable bonded battery to<br>micro circuit chip | Battery 1.5VDC<br>(2-Alkaline AAA) | | | Number of Output Modes | 4 | 3 | 1 | 10 | | | Number of Output Channels: | 1 Channel, Asynchronous - Biphasic | 1 Channel, Asynchronous - Biphasic | 1 Channel, Monophasic Microcurrent | 1 Channel, Asynchronous - Biphasic | | | Software/Firmware/Microprocessor Control? | Yes | Yes | Yes | Yes | | | Automatic Shut Off? | Yes | Yes | Yes | Yes | | | User Override Control? | Yes | Yes | Yes | Yes | | | Indicator Display: | On/Off Status? | Yes | Yes | Yes | | | | Low Battery? | Yes | Yes | No | Yes | | | Voltage/Current Level? | No | Yes | No | Yes | | Timer Range (minutes) | Nonadjustable<br>30 minutes | Nonadjustable<br>30 minutes | Nonadjustable<br>Continuous up to<br>4 - 5 days | Nonadjustable<br>15 minutes | | | Weight (lbs., oz.) | 4.8 oz.<br>w/ batteries<br>included | 2.1 oz.<br>w/ batteries<br>included | 2.0 oz.<br>w/ battery<br>included | 3.1 oz. | | | Dimensions (in.) [W x H x D]. | 7.5"(W) x 3.5(H)" x<br>0.7"(D) | 2.17" x 3.74" x<br>0.74 | unspecified | 5.91" x 2.68" x<br>1.02" | | | Housing Material and Construction | ABS plastic | ABS plastic | unspecified | unspecified | | | Compliance with U.S. FDA Title 21 CFR<br>898 Electrode Lead Wire Performance<br>Standard? | Yes | Yes | Yes | Yes | | | For multiphasic | Symmetrical phases? | No | No | Yes | No | | | Phase Duration<br>(include units), (state<br>range, if applicable),<br>(both phases, if<br>asymmetrical) | 120μs - 480μs | unspecified | unspecified | 4.1μs - 500μs | | Maximum Current Density, (mA/cm², r.m.s.) | 0.12mA @500Ω | 0.095mA @500Ω | unspecified | 2.71mA @500Ω | | | Maximum Average Current (average absolute value), mA | 1.6mA @500 Ω | 3.5mA @500 Ω | unspecified | 2.71mA @500Ω | | | Maximum Average Power Density, (W/cm²),<br>(using | 0.69mW/cm²<br>@500Ω | 89mW/cm²<br>@500Ω | unspecified | 5.35mW/cm²<br>@500Ω | | {3}------------------------------------------------ ### 510(k) # K120500 Page 4 of 4 ### Declarations of Conformity The device complies with the following FDA recognized standards: FDA Recognized Number 5-4, IEC 60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety, 1988 Amendment 1, 1991-11, Amendment 2, 1995. (General) FDA Recognized Number 5-60, IEC 60601-1-2 Int. 1 Third Edition/I-SH 01:2007, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests, Interpretation Sheet. (General) FDA Recognized Number 5-41. Medical electrical equipment - Part 1-4: General requirements for safety- Collateral standard: Programmable electrical medical systems, edition 1.1. (General) FDA Recognized Number 17-5, IEC 60601-2-10 1987/Amendment 1 2001, Medical electrical equipment - Part 2-10: Particular requirements for the safety of nerve and muscle stimulators. (Neurology) FDA Recognition Number 2-156: AAMI/AMSI/ ISO 10993-1:2009, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. (Biocompatibility) FDA Recognition Number 2-153 (Electrodes) ISO 10993-5:2009, Biological evaluation of medical devices - Part 5: Tests for In Vitro cytotoxicity. (Biocompatibility) FDA Recognized Standard 2-173 (Electrodes) Recognition Number 2-173: AAMI / ANSI / ISO 10993-10:2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization. (Biocompatibility) ### Truthful and Accurate Statement A statement was included in the Premarket Notification attesting to the truthfulness and accuracy of the information provided. ### Further Information In the event that additional information is required, please contact: Michael W. Treas Chief Compliance Officer Hollywog, LLC 2830 Amnicola Highway Chattanooga, TN 37406 Telephone: (423) 305-7778 ext. 108 Fax: (423) 305-7867 E-mail: mike.treas@hollywog.com {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002 AUG 20 2012 Hollywog, LLC c/o Mr. Michael Treas Chief Compliance Officer 2830 Amnicola Highway Chattanooga, TN 37406 Re: K120500 Trade/Device Name: The Pain Pilot™ (a.k.a. Pain Pilot™) / WiTouch™ Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator for Pain Relief Regulatory Class: Class II Product Code: NUH Dated: July 31, 2012 Received: August 1, 2012 Dear Mr. Treas: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be fray of cable of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ Page 2 - Mr. Michael Treas Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Kisia Alexander Malvina B. Evdelman, M.D. Director Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indications for Use 510(k) Number: K120500 Device Names: The Pain Pilot™ (a.k.a. Pain Pilot™) / WiTouch™ Indications for Use: The device is for temporary relief of pain associated with sore and aching muscles in the lower back due to strain from exercise or normal household and work activities. | Prescription Use<br>(Part 21 CFR 801 Subpart D) | AND/OR | Over-The-Counter Use <span style="text-decoration: overline;">√</span><br>(21 CFR 801 Subpart C) | |-------------------------------------------------|--------|--------------------------------------------------------------------------------------------------| |-------------------------------------------------|--------|--------------------------------------------------------------------------------------------------| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Callaway --- (Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices 510(k) Number K120500
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...