K131586 · Palo Alto Health Sciences, Inc. · HCC · Dec 10, 2013 · Neurology
Device Facts
Record ID
K131586
Device Name
CANARY BREATHING SYSTEM
Applicant
Palo Alto Health Sciences, Inc.
Product Code
HCC · Neurology
Decision Date
Dec 10, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Canary Breathing System is intended for use as a relaxation treatment for the reduction of stress by leading the user through guided and monitored breathing exercises. The device is indicated as an adjunctive treatment of symptoms associated with panic disorder, to be used under the direction of a healthcare professional, together with other pharmacological and/or non-pharmacological interventions.
Device Story
Canary Breathing System (CBS) is a biofeedback device for stress reduction and panic disorder symptom management. System components include a capnometer (CO2 sensor) with nasal cannula and a mobile application running on a tablet. The capnometer measures end-tidal CO2 (EtCO2) and respiratory rate (RR), transmitting data via Bluetooth to the mobile app. The app provides tone-guided breathing exercises and displays real-time physiological data to the patient. By observing the relationship between breathing mechanics and EtCO2 levels, patients learn to regulate their respiration. Used under the direction of a healthcare professional, the device serves as an adjunct to other therapies. Clinical benefits include reduced panic attack frequency/severity, decreased anxiety, and improved quality of life, mediated by normalized CO2 levels.
Clinical Evidence
Evidence based on three peer-reviewed clinical studies of the Capnometry-Assisted Respiratory Training (CART) protocol. Studies demonstrated that CART training leads to long-lasting reductions in panic attack frequency, severity, anxiety symptoms, and avoidance behaviors. Causal analysis linked normalization of CO2 to clinical improvements. No adverse events reported. Clinical results from CART studies are applicable to CBS as both use identical exercises and capnometry-based feedback, with CBS substituting a tablet-based interface for the tape recorder used in original studies.
Technological Characteristics
System comprises a capnometer (LoFlo C5 module) and mobile application on a tablet. Measures EtCO2 and respiratory rate via nasal cannula. Connectivity via Bluetooth. Electrical safety/EMC testing performed per IEC 60601-1:1988, IEC 60601-1-2:2007, IEC 60601-1-4:2000, and IEC 60601-1-11:2010. Software level of concern: Minor.
Indications for Use
Indicated for patients with panic disorder as an adjunctive treatment for symptoms, to be used under the direction of a healthcare professional alongside other pharmacological or non-pharmacological interventions.
Regulatory Classification
Identification
A biofeedback device is an instrument that provides a visual or auditory signal corresponding to the status of one or more of a patient's physiological parameters (e.g., brain alpha wave activity, muscle activity, skin temperature, etc.) so that the patient can control voluntarily these physiological parameters.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter when it is a prescription battery powered device that is indicated for relaxation training and muscle reeducation and prescription use, subject to § 882.9.
K233580 — A Breathing System (ABS) · Deepwell Dtx, Inc. · Aug 1, 2024
K201590 — EMMA Capnograph · Masimo Corporation · Mar 29, 2021
K062115 — CARBON DIOXIDE MONITOR, MODEL OLG-2800A · Nihon Kohden America, Inc. · Dec 27, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
#### 510(k) Summary
#### 510(k) Notification K131586
#### GENERAL INFORMATION
#### Applicant:
Palo Alto Health Sciences, Inc. 555 Bryant Street #312 · Palo Alto, CA 94301-1704 U.S.A. Phone: 650-269-4633
#### Contact Person:
Lori Adels, PhD Regulatory Consultant for Palo Alto Health Sciences Experien Group, LLC 755 N. Mathilda Avenue, Suite 100 Sunnyvale, CA 94085 U.S.A. Phone: 408-400-0856 ext. 112 FAX: 408-400-0865
#### Date Prepared: December 1, 2013
#### DEVICE INFORMATION
#### Trade Name:
Canary Breathing™ System
#### Generic/Common Name:
Biofeedback device Carbon dioxide gas analyzer
#### Classification :
21 CFR8882.5050, Class II 21 CFR§868.1400, Class II
#### Product Code:
HCC ССК
{1}------------------------------------------------
#### PREDICATE DEVICES
- . InterCure, Inc. RESPeRATE (K020399)
- . Respironics Novametrix, LLC, LoFlo C5 CO2 Sensor (K053174)
#### INDICATIONS FOR USE
The Canary Breathing" System is intended for use as a relaxation treatment for the reduction of stress by leading the user through guided and monitored breathing exercises. The device is indicated as an adjunctive treatment of symptoms associated with panic disorder, to be used under the direction of a healthcare professional, together with other pharmacological and/or non-pharmacological interventions.
2
#### INTENDED USE
The Canary Breathing™ System ("CBS") is a device used to provide biofeedback based respiratory training to patients with panic disorder. It is intended to train the patient to control their respiratory rate and end-tidal CO2 levels, which may allow the patient to alleviate abnormal breathing and therefore stress associated with panic attacks.
#### PRODUCT DESCRIPTION
The CBS is a biofeedback device that provides the user with a series of tone-guided breathing exercises and an awareness of his or her physiological data. The CBS uses standard biofeedback concepts to teach the patient to regulate their end-tidal CO2 (EtCO2) and respiratory rate (RR). The user's physiological data display allows the patient to see l ) the actual rate of their breathing and 2) how changes in breathing mechanics (depth and volume) affect EtCO2 levels. The CBS consists of a biofeedback training software program (mobile app) and an EtCO2 sensor (capnometer) used with a nasal cannula. The mobile app guides the user through an exercise and displays physiological data, while the sensor collects physiological data and feeds it to the mobile application for biofeedback. The patient's EtCO2 levels and RR are relaved from the capnometer to the mobile application via Bluetooth and are displayed on a tablet device, through the mobile application.
#### SUBSTANTIAL EQUIVALENCE
The technological characteristics and intended use of the CBS are substantially similar to an existing biofeedback device, the InterCure. Inc. RESPERATE. The RESPeRATE trains the patient to synchronize his or her breathing to the guiding tones and biofeedback displays on the device. The RESPeRATE gradually prolongs the exhalation tone to slow down breathing as an adjunctive treatment for high blood pressure, with other pharmacological and/or non-pharmacological interventions. The difference in the specific intended use (reduction in blood pressure vs. reduction in symptoms of panic disorder) does not raise new issues of safety and effectiveness as in both cases the devices use normalized breathing to help alleviate an abnormal condition caused in part by stress and in both cases the biofeedback device is used as an adjunct to other pharmacological and/or non-pharmacological interventions.
{2}------------------------------------------------
Both the CBS and the RESPeRATE provide a respiratory training protocol that uses biofeedback for relaxation (reduction of stress), and both devices are intended as adjunctive devices to other forms of pharmacological and/or non-pharmacological therapy. Both devices include a computerized control unit and a sensor. The RESPeRATE sensor is a band that goes around the patient's upper abdomen to measure breathing rate and depth. The CBS uses a capnometer to measure breathing rate and EtCO2.
The capnometer is substantially equivalent to an existing portable capnometer (LoFlo C5 CO2 capnometer manufactured by Respironics Novametrix, LLC) as determined by bench testing. The LoFlo C5 CO2 capnometer, per the predicate IFU, is used to monitor carbon dioxide in recovery and emergency medicine or respiratory care. Its use in measuring carbon dioxide and respiratory rate in the CBS system for the adjunctive treatment of panic attacks does not raise new issues of safety and effectiveness. Thus, the CBS is substantially equivalent to the predicate devices.
{3}------------------------------------------------
#### TESTING IN SUPPORT OF SUBSTANTIAL EQUIVALENCE DETERMINATION
All necessary non-clinical and clinical testing was conducted to support a determination of substantial equivalence to the predicate devices.
#### NON-CLINICAL DATA
The following table lists the non-clinical testing performed and the results for each test.
| Testing Type | Test Description | Result |
|-----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| CBS Design Verification | Canary CO₂ Sensor Functional<br>Testing | The CBS passed all functional<br>testing and met all product<br>specification requirements. |
| | Canary CO₂ Cannula Functional<br>Testing | |
| | Nexus 7 tablet Functional Testing | |
| | Canary Mobile Application<br>Functional Testing | |
| CBS Integration Testing | The integration testing assessed the<br>integrity of RS232 serial output data<br>generated by an OEM purchased<br>Respironics LoFlo capnometer<br>module integrated into a Canary<br>Breathing System (CBS)<br>capnometer. | RR, ETCO₂ and error information<br>were successfully and accurately<br>transmitted to the Nexus tablet<br>running the CBS software. |
| Software Verification<br>and Validation | Canary CO₂ Sensor Software Testing<br>(The Level of Concern for the CBS<br>software was determined to be<br>"Minor".) | The Canary CO₂ Sensor Software<br>met all requirements of the SRS. |
| Capnometer<br>Equivalency<br>Verification | Canary CO₂ Sensor Accuracy<br>Testing | The performance of the Canary CO₂<br>Sensor is equivalent to theLoFlo<br>predicate device, for EtCO₂ and<br>respiratory rate measurements. |
| Electrical Safety and<br>Electromagnetic<br>Compatibility | Testing in accordance with the<br>following standards:<br>● IEC 60601-1:1988<br>● IEC 60601-1-2:2007<br>● IEC 60601-1-4:2000<br>● IEC 60601-1-11:2010 | The CBS met all acceptance criteria<br>in accordance with:<br>● IEC 60601-1:1988<br>● IEC 60601-1-2:2007<br>● IEC 60601-1-4:2000<br>● IEC 60601-1-11:2010 |
#### CLINICAL DATA
The CBS is based on the capnometry-assisted respiratory training (CART) protocol developed by Alicia Meuret, Ph.D., a clinical psychologist and Associate Professor in the Department of Psychology at Southern Methodist University (Dallas, Texas) in collaboration with colleagues at Stanford University, Boston University and Southern Methodist University. Dr. Meuret and colleagues have published three peer reviewed articles describing the results of two randomized clinical trials using the CART protocol to treat patients with panic disorder. The references are listed below. Findings of these empirical studies have shown that completing the CART training led to long-lasting
{4}------------------------------------------------
reductions in panic attack frequency and severity, anxiety symptoms, avoidance behaviors, along with improvements in mood and quality of life in the majority of patients. Respiratory training using the CART protocol is safe, and no adverse events associated with the use of the device have been reported. Causal analysis linked the normalization in CO2 to the observed improvements.
5
The CBS and CART protocol lead the user through identical exercises, however, the user interface for the CBS device is different from that used in the CART protocol studies. .The CBS uses a tablet based mobile application, in place of the tape recorder used in the CART protocol. Both devices use a capnometer to measure EtCO2 levels and respiratory rate and to provide the patient with feedback on these parameters. Therefore, the clinical results obtained by following the CART protocol in the above mentioned studies can be directly applied to the CBS device.
#### Clinical Study References:
Meuret AE, et al. Feedback of end-tidal pCQ2 as a therapeutic approach for panic disorder. Journal of Psychiatric Research 2008;42;560-8.
Meuret AE, et al. Changes in respiration mediate changes in fear of bodily sensations in panic disorder. Journal of Psychiatric Research 2009;43;634-41.
Meuret AE. et al. Respiratory and Cognitive Mediators of Treatment Change in Panic Disorder: Evidence for Intervention Specificity. Journal of Consulting and Clinical Psychology 2010:78:691-704.
The collective performance testing demonstrates that the CBS does not raise new questions of safety or effectiveness when compared to the predicate devices. The results of the performance testing demonstrate that the CBS performs as intended and that its biofeedback protocol is safe and effective for reducing symptoms associated with panic disorder.
#### CONCLUSION
The CBS has the same intended use and similar technological characteristics as the predicate biofeedback devices as well as the same physiological parameter measurements as the predicate capnometer. The differences in the means for collecting physiological data for biofeedback does not change the intended use or raise any new issues of safety or effectiveness as compared to the predicate biofeedback devices.
#### SUMMARY
The CBS is substantially equivalent to the predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 10, 2013
Palo Alto Health Sciences, Inc. c/o Dr. Lori Adels Regulatory Consultant 755 N Mathilda Ave., Suite 100 Sunnyvale, CA 94085
Re: K131586
Trade/Device Name: Canary Breathing System Regulation Number: 21 CFR 882.5050 Regulation Name: Biofeedback Device Regulatory Class: Class II Product Code: HCC and CCK Dated: November 1, 2013 Received: November 4, 2013
Dear Dr. Adels:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may. therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA is issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{6}------------------------------------------------
Page 2 - Dr. Adels
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm.
Sincerely yours,
## Victor Krauthamer -A
Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### Indications for Use
510(k) Number (if known): K131586
Device Name: Canary Breathing System
Indications For Use:
The Canary Breathing System is intended for use as a relaxation treatment for the reduction of stress by leading the user through quided and monitored breathing exercises. The device is indicated as an adjunctive treatment of symptoms associated with panic disorder, to be used under the direction of a healthcare professional, together with other pharmacological and/or non-pharmacological interventions.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
# Victor Krauthar 2013.12.10 14
Page 1 of .1
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.