NEUROGEN NERVE GUIDE

K011168 · Integra Lifesciences Corp. · JXI · Jun 22, 2001 · Neurology

Device Facts

Record IDK011168
Device NameNEUROGEN NERVE GUIDE
ApplicantIntegra Lifesciences Corp.
Product CodeJXI · Neurology
Decision DateJun 22, 2001
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5275
Device ClassClass 2
AttributesTherapeutic

Indications for Use

NeuroGen" Nerve Guide is indicated for repair of peripheral nerve discontinuities where gap closure can be achieved by flexion of the extremity.

Device Story

NeuroGen™ Nerve Guide is a sterile, porous, collagen-based tubular implant designed for peripheral nerve repair. It acts as a conduit between nerve ends and surrounding tissue, providing a protective environment to facilitate axonal growth across nerve gaps. The device is flexible, allowing for joint and tendon movement while resisting occlusive forces from surrounding tissue. It is intended for surgical use by physicians to bridge nerve discontinuities. By maintaining a stable, protective environment, the device supports natural nerve regeneration, potentially improving functional recovery for patients with peripheral nerve injuries.

Clinical Evidence

Evidence includes a long-term (3.5 years) primate study and rodent animal models. Results demonstrated performance substantially equivalent to nerve graft, direct suture, and silicone tubes. Biocompatibility testing confirmed the device is non-cytotoxic, non-pyrogenic, non-irritating, non-sensitizing, non-toxic, and non-mutagenic. Mechanical and physical testing verified suture retention, compression resistance, and appropriate porosity/permeability for axonal growth.

Technological Characteristics

Tubular, porous collagen conduit; bioresorbable material. Designed for flexibility and resistance to occlusive forces. Meets ISO 10993 biocompatibility standards. Provided sterile in double blister packaging.

Indications for Use

Indicated for repair of peripheral nerve discontinuities in patients where gap closure is achievable via extremity flexion.

Regulatory Classification

Identification

A nerve cuff is a tubular silicone rubber sheath used to encase a nerve for aid in repairing the nerve (e.g., to prevent ingrowth of scar tissue) and for capping the end of the nerve to prevent the formation of neuroma (tumors).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # NeuroGen™ Nerve Guide JUN 2 2 2001 ## 510(K) SUMMARY #### Submitter's name and address: Integra LifeSciences Corporation 105 Morgan Lane Plainsboro, NJ 08536 USA #### Contact person and telephone number: Judith E. O'Grady Senior Vice President, Regulatory Affairs, Quality Assurance and Clinical Research (609) 275-0500 #### Date Summary was prepared: April 16, 2001 #### Name of the device: | Proprietary Name: | NeuroGen™ | |----------------------|--------------------------------| | Common Name: | Nerve Guide | | Classification Name: | Nerve Cuff, Product Code 84JXI | #### Substantial Equivalence: NeuroGen™ Nerve Guide is substantially equivalent in function and intended use to the following products which have been cleared to market under Premarket Notifications 510(k): Salumedica" Nerve Cuff, NeuroTube and Fastube" Nerve Cuff. ## Device Description: NeuroGen" Nerve Guide is an implant designed for repair of peripheral nerve discontinuities. NeuroGen™ Nerve Guide provides a protective environment for peripheral nerve repair after injury. NeuroGen" Nerve Guide is designed to be an interface between the nerve and surrounding tissue and to create a conduit for axonal growth across a nerve gap. NeuroGen" Nerve Guide is flexible to accommodate movement of joint and associated tendons while retaining its shape and is resistant to occlusive forces from surrounding tissue. When hydrated, NeuroGen" is an easy to handle, soft, pliable, nonfriable, porous collagen tube. NeuroGen™ Nerve Guides are provided sterile in double blister packages in a variety of sizes. ## Intended Use: NeuroGen" Nerve Guide is indicated for repair of peripheral nerve discontinuities where gap closure can be achieved by flexion of the extremity. {1}------------------------------------------------ ## Safety Biocompatibility studies have demonstrated NeuroGen™ Nerve Guides to be: noncytotoxic, nonpyrogenic, nonirritating, and nonsensitizing. The following studies were conducted: - a) Cytotoxicity - b) Irritation / Intracutaneous Reactivity - Sensitization c) - d) Acute Systemic Toxicity - e) Subchronic Toxicity - Chronic Toxicity f) - Genotoxicity g) - h) Implantation - Hemolysis i) ## Performance Characteristics: The effectiveness of NeuroGen™ Nerve Guide to repair peripheral nerve discontinuities was studied in a long-term (3.5 years) primate study and in rodent animal models. The studies demonstrated that NeuroGen™ Nerve Guides are substantially equivalent to nerve graft, direct suture and silicone tubes. The mechanical and physical characteristics of the NeuroGen™ Nerve Guides were evaluated in a series of tests. These tests were conducted to ensure that the NeuroGen™ Nerve Guides possess the mechanical properties (suture retention and mechanical compression) as well as physical properties (porosity and permeability) that determine their suitability for use in the human body. Testing has demonstrated that the nerve guides are able to hold a suture, resist repeated compression from surrounding tissues, have a porous outer surface and tube wall, and allow the passage of molecules of specific size through the tube wall. ### Technological Characteristics Compared to Predicate Devices: NeuroGen™ Nerve Guide is a tubular device which is equivalent to the predicate devices, Salumedica™ Nerve Cuff. NeuroTube® and Fastube™ Nerve Cuff in its design for repair of peripheral nerve discontinuities. Like the predicate devices, NeuroGen" is provided sterile, for single use only. The NeuroGen™ Nerve Guide is manufactured from a bioresorbable material, as is one of the predicate devices, NeuroTube®. NeuroGen™ Nerve Guide meets ISO 10993 requirements for Biocompatibility testing. ### Conclusion NeuroGen" Nerve Guide is indicated for the repair of nerve discontinuities where gap closure can be achieved by flexion of the extremity. NeuroGen™ Nerve Guide is flexible to accommodate movement of joint and associated tendons while retaining its shape and is resistant to occlusive forces from surrounding tissue. Biocompatibility studies have demonstrated NeuroGen" Nerve Guide to be non-cytotoxic, nonsensitizing, non-toxic and non-mutagenic. Extensive, long-term evaluations in primates demonstrates NeuroGen™ Nerve Guide to biocompatible and provides an environment for axonal growth. Valid scientific evidence through substantial testing of descriptive characteristics, Biocompatibility, mechanical and physical property testing and extensive performance testing in a primate model, provide reasonable assurance that NeuroGen™ Nerve Guide is safe and effective under the proposed conditions of use, and substantially equivalent to its predicate devices. {2}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three overlapping human profiles or a bird in flight, composed of curved lines. #### Public Health Service JUN 2 2 2001 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Judith E. O'Grady, RN, MSN Senior Vice President, Regulatory Affairs, Quality Assurance and Clinical Affairs Integra Life Sciences Corporation 105 Morgan Lane Plainsboro, New Jersey 08536 Re: K011168 Trade/Device Name: NeuroGen Nerve Guide Regulation Number: 882.5275 Regulatory Class: II Product Code: JXI Dated: April 16, 2001 Received: April 17, 2001 Dear Ms. O'Grady: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ Page 2 - Ms. Judith E. O'Grady, RN, MSN This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours. Marlen Millenor Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 510(k) Number: K011168 Device Name: NeuroGen™ Nerve Guide ## Indications for Use NeuroGen™ Nerve Guide is indicated for repair of peripheral nerve discontinuities where gap closure can be achieved by flexion of the extremity. ## PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) (Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use OR Over-the-Counter Use X (Per 21 CFR 801.109) Mark N. Milhusan (Division Sign-Off) (Division of General, Restorative and Neurological Devices 510(k) Number -
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...