Nerbridge™ is intended for the repair of peripheral nerve injuries in which there is no gap or where a gap closure can be achieved by flexion of the extremity.
Device Story
Nerbridge™ is a resorbable, semipermeable tubular nerve conduit composed of polyglycolic acid (PGA) and porcine-derived Type I and III collagen. It acts as a non-constricting, protective interface between injured peripheral nerves and surrounding tissue. The device is hydrated and placed around the nerve site by a surgeon to facilitate healing. It provides a porous environment for nerve regeneration while maintaining structural integrity and flexibility. The device is provided sterile for single use. Clinical benefit includes nerve repair and protection of the neural environment during the healing process. It is intended for use in clinical settings by surgeons.
Clinical Evidence
Multi-center, randomized, evaluator-blinded comparative clinical study (n=60 subjects with hand injuries; 58 in Full Analysis Set). Compared Nerbridge™ to autogenous free nerve graft (n=6). Primary endpoint: sensory function improvement via Semmes-Weinstein method. Results showed higher mean percentage improvement in sensory function and lower incidence of adverse events for Nerbridge™ compared to the control group. Supporting animal study (rabbit sciatic nerve) confirmed resorption by day 72 and nerve regeneration.
Technological Characteristics
Materials: Polyglycolic acid (PGA) and porcine-derived Type I and III collagen. Form: Flexible, porous, resorbable tubular membrane matrix. Dimensions: ID 0.5-4.0mm, Length 30-50mm. Sterilization: Ethylene oxide. Biocompatibility: ISO 10993 compliant. Performance standards: ISO 7198 (suture retention), JIS T0401 (compression/stiffness), ISO 527-1 (tensile strength), ISO 15814 (degradation).
Indications for Use
Indicated for repair of peripheral nerve injuries in patients where no gap exists or gap closure is achievable via extremity flexion.
Regulatory Classification
Identification
A nerve cuff is a tubular silicone rubber sheath used to encase a nerve for aid in repairing the nerve (e.g., to prevent ingrowth of scar tissue) and for capping the end of the nerve to prevent the formation of neuroma (tumors).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is an abstract symbol that resembles a stylized caduceus or a representation of the human profile in triplicate.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 22, 2016
Toyobo Co., Ltd. % James A. Boiani, MS, JD Official Correspondent Epstein Becker & Green, P.C. 1227 25th St. NW, Suite 700 Washington, DC 20037
Re: K152967
Trade/Device Name: Nerbridge™ Regulation Number: 21 CFR 882.5275 Regulation Name: Nerve Cuff Regulatory Class: Class II Product Code: JXI Dated: May 20, 2016 Received: May 23, 2016
Dear Mr. Boiani:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena -S/A
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152967
Device Name Nerbridge™
Indications for Use (Describe)
Nerbridge™ is intended for the repair of peripheral nerve injuries in which there is no gap or where a gap closure can be achieved by flexion of the extremity.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY
The assigned 510(k) number is: K152967
#### 1. Submitter's Identification:
Tovobo Co., Ltd. 2-8 Dojima Hama 2-chome, Kita-ku, Osaka 530-8230 Japan Tel: 81-6-6348-3336 Fax: 81-6-6348-3696 Contact: Mr. Yuta Kawakatsu
Date Summary Revised: June 21, 2016
#### 2. Name of the Device:
Nerbridge™ Device Common Name: Cuff, Nerve Device Classification Name: Nerve Cuff Product Regulation Number: 21 CFR Part 882.5275 Product Code: JXI Requlatory Class: II Classification Panel: Neurology
#### 3. Predicate Device Information:
- 1) K103081 Biom'Up SA. Cova™ ORTHO-NERVE
- 2) K983007 Neuroregen LLC Neurotube™
#### 4. Device Description:
Nerbridge™ is a product composed of polyglycolic acid and collagen derived from porcine skin. Nerbridge™ is a flexible, resorbable and semipermeable tubular membrane matrix filled with porous collagen that provides a non-constricting encasement for injured peripheral nerves for protection of the neural environment. Nerbridge™ is designed to be an interface between the nerve and the surrounding tissue. When hydrated, Nerbridge™ is a pliable, soft, non-friable, porous conduit. The resilience of Nerbridge allows the product to recover and maintain closure without constricting the nerve once the device is placed around the nerve. Nerbridge™ is manufactured using validated viral inactivation and removal processes for the collagen. The product is provided in a foil pouch, sterile, nonpyrogenic, for single use only, in a variety of sizes, and placed in an outer Tyvek header bag for added protection.
#### 5. Indications for Use:
Nerbridge™ is intended for the repair of peripheral nerve injuries in which there is no gap or where a gap closure can be achieved by flexion of the extremity.
{4}------------------------------------------------
#### 6. Comparison to Predicate Devices:
| | Device Name | Nerbridge™ | Neurotube™ | Cova™ORTHO-NERVE | |
|-----------------------------|----------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|
| | Manufacturer | TOYOBO CO., LTD. | Neuroregen, LLC | Biom'Up S.A. | |
| | 510(k) No. | K152967 | K983007 | K103081 | |
| | Intended Use and Indications<br>for Use | Nerbridge™ is<br>intended for the<br>repair of peripheral<br>nerve injuries in<br>which there is no<br>gap or where a gap<br>closure can be<br>achieved by flexion<br>of the extremity. | The Neurotube™ is<br>intended for single<br>use in patients with<br>an injury to a<br>peripheral nerve in<br>which the nerve gap<br>is more than or<br>equal to 8mm, but<br>less than or equal to<br>3 cm. The nerve<br>gap may be created<br>primarily at the time<br>of injury or created<br>secondarily at the<br>time of exploration<br>of failed primary<br>repair. | Cova™ORTHO-<br>NERVE is indicated<br>for the repair of<br>peripheral nerve<br>injuries in which<br>there is no gap or<br>where a gap<br>closure can be<br>achieved by flexion<br>of the extremity | |
| Material | | Synthetic<br>polyester | Polyglycolic acid<br>(PGA) | PGA | Contains no<br>Synthetic Polyester<br>material |
| | Collagen | Type | I and III | None | I |
| | | Origin | Porcine | None | Porcine |
| | Shape | | Cylindrical | Cylindrical | Membrane, rollable<br>if needed |
| | Physical structure | | PGA conduit whose<br>outer surface is<br>coated by collagen<br>and with inner<br>porous collagen | PGA conduit | Collagen<br>membrane |
| Device Name | NerbridgeTM | NeurotubeTM | CovaTMORTHO-<br>NERVE | | |
| Sizes | ID: 0.5, 1.0, 1.5, 2.0,<br>2.5, 3.0, 3.5, 4.0mm<br>Length: 30, 50 mm | ID: 2.0,3.0, 4.0,<br>8.0mm<br>Length: 20, 40mm | Rectangular sheets<br>of 15 x 25 mm, 20 x<br>30 mm, 30 x 40 mm<br>and 40 x 60 mm | | |
| Sterilization | Ethylene oxide | Ethylene oxide | Gamma irradiation | | |
| Clinical testing | Yes | Yes | No | | |
| Non-clinical animal testing | Yes | Unknown | Yes | | |
| Biocompatibility | ISO 10993 | Unknown | ISO 10993 | | |
| | Suture<br>retention<br>strength | Yes | Unknown | Yes | |
| | Mechanical<br>compression<br>and rebound | Yes | Yes | Yes | |
| | Porosity | Yes | Unknown | Unknown | |
| | Permeability | Yes | Yes | Unknown | |
| | In vitro<br>degradation | Yes | Yes | Yes | |
| Performance | Tensile<br>strength | Yes | Unknown | Yes | |
| | pH | Yes | Unknown | Yes | |
| | Swelling rate | Yes | Unknown | Yes | |
| | Visual<br>inspection | Yes | Unknown | Unknown | |
| | Bending<br>stiffness | Yes | Yes | Unknown | |
| | Endotoxin | Yes | Unknown | Yes | |
| | Packaging | Foil pouch within<br>protective outer<br>pouch | Double peel<br>packages | Double peel<br>packages | |
| | Shelf Life | 3 years | 5 years | 24 months | |
{5}------------------------------------------------
{6}------------------------------------------------
The Nerbridge™ and the predicate devices have the same intended use: to act as a conduit to assist the nerve in repairing itself following damage.
The Nerbridge™ and the predicate devices are sutured to the nerve and nerve tissue and surround the damaged nerve to protect the damaged area, allowing it to heal. At a high level, the subject and predicate devices are based on the following same technological elements:
- Provided or can be made into a cylindrical shape
- Flexible ●
- Smooth ●
- Wettable ●
- Resorbable material .
Certain technological differences exist between the subject and predicate devices as detailed in the above. In particular, the subject device is a PGA tube with an outer coating and internal filling of a porous Type I and III collagen blend; the Neurotube is a PGA tube that does not contain collagen; the Cova Ortho-Nerve is a Type I collagen membrane. While having some technological differences with the predicates, the Clinical and Non-Clinical tests performed on the Nerbridge™ show that it is safe and effective for its intended use and substantially equivalent to the predicate devices.
### 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence:
The following functional performance verification and product characterization testing was conducted on finished sterile Nerbridge™ devices using known standards and/or clinically relevant criteria:
- Suture retention strength ●
- . Mechanical compression and rebound
- Porositv ●
- Permeability .
- In vitro degradation
- Tensile strength ●
- pH ●
- Swelling rate
- Visual inspection ●
- Bending stifffness ●
- Endotoxin
The testing supports and further characterizes Nerbridge™ that was studied in the Nerbridge™ Clinical Study, and shown to be clinically safe and effective for its intended use as a nerve conduit, and supporting substantial equivalence to other nerve conduits with the same intended use.
| Test | Test Method Summary | Results |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| Test | Test Method Summary | Results |
| Suture retention<br>strength | Testing according to ISO 7198:1998<br>Acceptance criteria: Suture must fail before the<br>Nerbridge | Acceptance<br>criteria met |
| Mechanical<br>compression and<br>rebound | Testing according to JIS T0401:2013<br>Acceptance criteria: Nerbridge must maintain<br>mechanical strength after continuous<br>compression and rebound loading | Acceptance<br>criteria met |
| Porosity | Testing according to ISO 845:2006 and JIS<br>Z8807:2012<br>Acceptance criteria: Porosity of Nerbridge must<br>be greater than or equal to porosity of PGA tube<br>alone | Acceptance<br>criteria met |
| Permeability | Devices were filled with protein solution and then<br>immersed in saline. At periods of time up to 96<br>hours, saline was measured for protein particles<br>Acceptance criteria: Protein must permeate the<br>wall of the Nerbridge | Acceptance<br>criteria met |
| In vitro<br>degradation | Testing according to ISO 15814:1999 and ISO<br>527-1:2012<br>Acceptance criteria: Nerbridge must maintain a<br>tensile strength greater than that of suture for 8<br>weeks | Acceptance<br>criteria met |
| Tensile strength | Testing according to ISO 527-1:2012<br>Acceptance criteria: Nerbridge tensile strength is<br>at least as much as the predicate | Acceptance<br>criteria met |
| pH | Testing according to JIS T 3211:2011<br>Acceptance criteria: Nerbridge satisfies the<br>acceptance criteria as specified in JIS T<br>3211:2011 | Acceptance<br>criteria met |
| Swelling rate | Testing according to ISO 10545-3:1995<br>Acceptance criteria: Swelling rate of Nerbridge<br>must be greater than or equal to swelling rate of<br>PGA tube alone | Acceptance<br>criteria met |
| Visual inspection | Visual inspection with a magnifying glass<br>Acceptance criteria: No observation of significant<br>streaks, wrinkles, uneven, scratches and cracks,<br>bubbles or other abnormalities | Acceptance<br>criteria met |
| Test | Test Method Summary | Results |
| Bending stiffness | Testing according to JIS T0401:2013<br>Acceptance criteria: Bending stiffness of<br>Nerbridge must be greater than or equal to<br>bending stiffness of PGA tube alone | Acceptance<br>criteria met |
| Endotoxin | Testing according to Japanese Pharmacopoeia<br>16th edition and FDA Guidance for Industry:<br>Pyrogen and Endotoxin Testing: Questions and<br>Answers (June 2012)<br>Acceptance criteria: Endotoxin level is less than<br>or equal to 20 EU/device | Acceptance<br>criteria met |
{7}------------------------------------------------
{8}------------------------------------------------
## Biocompatibility Testing
The biocompatibility evaluation for the Toyobo Nerbridge device was performed according to ISO 10993 to demonstrate that the device was both safe for implantation and establish substantial equivalence among the predicate device.
| Test | Test Method Summary | Results |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------|---------------------------------------------------|
| Cytotoxicity | ISO Direct contact Cytotoxicity<br>Assay | Non-cytotoxic |
| Sensitization | ISO Guinea pig Maximization test<br>with device extracts (saline and<br>sesame oil extracts) | No evidence of sensitization |
| Acute<br>intracutaneous<br>Reactivity | ISO Acute intracutaneous<br>Reactivity Test in rabbits with<br>device extracts (saline and<br>sesame oil extracts) | No evidence of irritation |
| Acute Systemic<br>Toxicity | ISO Acute System Toxicity in Mice<br>with device extracts (saline and<br>sesame oil extracts) | No mortality or evidence of<br>systemic toxicity |
| Rabbit Pyrogen<br>Study | USP Material-mediated Rabbit<br>pyrogen test with saline extract of<br>the device | No evidence of material-<br>mediated pyrogenicity |
| Hemolysis | Hemolysis test by direct contact<br>with human red blood cells | No hemolytic activity |
| Genotoxicity | ISO Ames Mutagenicity Assay<br>with device extracts (saline<br>and ethanol extracts) | No evidence of<br>mutagenicity |
{9}------------------------------------------------
| Test | Test Method Summary | Results |
|---------------------------------------|-------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| Genotoxicity | ISO Mouse bone marrow micronucleus with device extracts (saline and sesame oil extracts) | No evidence of clastogenicity |
| Genotoxicity | CHL/IU cells with device extracts (MEM & 10%CS/MEM) | No evidence of inducing chromosomal aberrations or polyploid cells |
| Implantation Absorption | Subcutaneous implantation in rats | Absorption of material by 13 weeks. No inflammation observed |
| Implantation (safety and performance) | In vivo safety and performance study in rats after 3, 30 and 90 days | Protection during nerve repair. No fibrous peri-nervous tissue was observed after 3, 30 or 90 days. |
| Subchronic / Chronic toxicity | 13-week systemic toxicity and local tolerance study in rats following subcutaneous implantation | No adverse tissue reaction to the implant up to 13 weeks of implantation<br>No systemic toxicity |
### Viral Inactivation Study
Viral Inactivation steps were validated for collagen extracted from tissues used in the manufacture of the device. For the validation, different viruses were selected according to their physio-chemical resistance and representativeness. The viral inactivation steps are routinely performed in process using appropriate treatment. It was demonstrated that the inactivation steps reduced down the final viral load to the limit of detection.
#### 8. Clinical Tests Performed
A multi-center joint randomized evaluator-blinded comparative clinical study was conducted on the Nerbridge™ device where the Nerbridge™ device was used in 60 subjects at 20 trial sites with hand injuries. The primary efficacy endpoint was evaluated on the Full Analysis Set (58 subjects) and the Per Protocol Set (54 subjects). The control group consisted of 6 subjects implanted with an autogenous free nerve graft. The mean percentage of improvement in primary evaluation in the sensory function test by the Semmes-Weinstein method for the Nerbridge™ device was higher than that in the autogenous free nerve grafting group. The incidence of adverse events were lower for the Nerbridge™ device than the control. Therefore, the Nerbridge "" device is safe
{10}------------------------------------------------
and effective for its intended use and substantially equivalent to the predicate devices.
#### 9. Animal Tests Performed
An animal study was conducted in rabbit model to evaluate the performance and safety (local tissue effects) of the Nerbridge™ after nerve section and direct suture in the rabbit sciatic nerve. Nerbridge™ resorbed by day 72, there was evidence of nerve regeneration, and the histological reactions at the site of introduction of the device were slight.
#### 10. Conclusions
The subject device, the Nerbridge™, has the same intended use as the predicate devices, Biom'Up SA. Cova ™ ORTHO-NERVE (K103081) and Neuroregen LLC, Neurotube™ (K983007). The Clinical and Non-Clinical testing supplied within our submission demonstrates that there are not any significant differences in their technological characteristics thereby not raising any new questions of safety and effectiveness. Therefore, the Nerbridge™ is substantially equivalent to the predicate devices, Biom'Up SA. CovaT™ ORTHO-NERVE (K103081) and Neuroregen LLC, Neurotube™ (K983007).
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.