ODFS PACE

K102115 · Odstock Medical, Ltd. · GZI · Mar 30, 2011 · Neurology

Device Facts

Record IDK102115
Device NameODFS PACE
ApplicantOdstock Medical, Ltd.
Product CodeGZI · Neurology
Decision DateMar 30, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ODFS® Pace is intended to provide ankle dorsiflexion in individuals who have a dropped foot as a consequence of upper motor neuron injury. By detecting the swing phase of gait through a foot switch signal, appropriate electrical stimulation of the leg and ankle muscles may improve gait by flexing the foot of persons who have lost or impaired function. There may be additional benefits from FES such as muscle re-education, prevention/retardation of disuse atrophy, increased or maintained range of joint motion and increase in local blood flow. The ODFS® Pace is a medical device and should only be used under medical supervision for the treatment of dropped foot following an upper motor neuron injury.

Device Story

ODFS Pace is a body-worn, single-channel external functional neuromuscular stimulator (FES). Input: gait signal from a wired footswitch worn in the shoe. Operation: device detects swing phase of gait; triggers electrical pulses to leg/ankle muscles via self-adhesive skin surface electrodes (typically common peroneal nerve). Output: muscle contraction causing ankle dorsiflexion. Optional mode: powered muscle stimulator for exercise (footswitch disabled). Used by patients with drop foot from upper motor neuron injury; clinician-supervised setup. Benefits: improved gait, muscle re-education, reduced disuse atrophy, maintained joint range of motion, increased local blood flow.

Clinical Evidence

Bench testing only. Verification of functional and output specifications (stimulator controls, settings, indicators, stimulus output/timing, footswitch operation, power consumption). Compliance with ISO 60601-1:2006 (safety), ISO 60601-1-2:2007 (EMC), and ISO 60601-2-10:2001 (nerve/muscle stimulators).

Technological Characteristics

Battery-powered external FES/NMES. Waveform: Biphasic (symmetrical/asymmetrical). Pulsewidth: 0-360μs. Frequency: 20-60Hz. Dimensions: small, body-worn. Connectivity: wired footswitch. Electrodes: 5cm x 5cm self-adhesive. Standards: ISO 60601-1, ISO 60601-1-2, ISO 60601-2-10.

Indications for Use

Indicated for individuals with dropped foot due to upper motor neuron injury. Used to provide ankle dorsiflexion to improve gait via FES. Optional use as a powered muscle stimulator for muscle re-education, prevention of disuse atrophy, maintenance of joint range of motion, and increased local blood flow. Contraindicated for use without medical supervision.

Regulatory Classification

Identification

An external functional neuromuscular stimulator is an electrical stimulator that uses external electrodes for stimulating muscles in the leg and ankle of partially paralyzed patients (e.g., after stroke) to provide flexion of the foot and thus improve the patient's gait.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K10211.5 Image /page/0/Picture/1 description: The image shows a logo with the letters "OML" in a bold, sans-serif font. The letters are white and are set against a dark, textured background. The logo is contained within a rounded rectangle with a thick border. The overall appearance of the logo is somewhat distressed or vintage due to the texture and shading. LEADING REHABILITATION THROUGH TECHNOLOGY # Odstock Medical Limited National Clinical FES Centre MAR 3 0 2011 Salisbury District Hospital Salisbury Wilitshire SP2 8BJ വന Tel +44 (0) 1722 439541 Fax +44 (0) 1722 425263 steven.crook@odstockmedical.com www.odstockmedical.com # 510(k) Summary For the # ODFS® Pace # [Section 5 of the Submission] 5.1 Sponsor/Owner Odstock Medical Limited Laing Building Salisbury District Hospital Salisbury Wiltshire SP2 8BJ United Kingdom Odstock Medical Ltd is a Registered Establishment, number: 3005344585 | Contact person: | Steven Crook, Quality and Regulatory Manager | |-----------------|----------------------------------------------| | Telephone: | 0044 1722 439541 | | Fax: | 0044 1722 425263 | | Email: | steven.crook@odstockmedical.com | 21 July 2010, with revisions 11&14 March 2011 Date prepared #### US Agent | <b>Boston Brace International</b><br>20 Ledin Drive<br>Avon, MA 02322 USA | Phone: 800-262-2235<br>Fax: 800-634-5048 | |---------------------------------------------------------------------------|------------------------------------------| |---------------------------------------------------------------------------|------------------------------------------| Barry Bull (Chairman) • Malcom Cassells (Director) • Philip Casson (CEO) • Ian Swain (Director) Odstock Medical Limited (Company No: 5532620) is part of Salisbury NHS Foundation Trust {1}------------------------------------------------ #### 510(k) Summary for the ODFS® PACE Dropped Foot Stimulator ## 5.2 DEVICE NAME | Trade/Proprietary Name: | ODFS®-Pace | |-------------------------|----------------------------------------------------------------------------------------| | Common/Usual Name: | External Functional Neuromuscular Stimulator (FES) | | Classification Name: | External Functional Neuromuscular Stimulator (FES)<br>Powered Muscle Stimulator (NMES) | | Classification No. | 21 CFR 882.5810 | | Classification Code: | Class II | | Product Codes: | GZI and IPF | #### 5.3 Predicate Devices | Predicate Device | K number | Comment | | |----------------------|----------|---------------------------------------|--| | EMPI Focus Model 795 | K951951 | EMPI | | | ODFS III V6.2 | K020991 | Odstock Medical Ltd (transferred from | | | | | Salisbury District Hospital) | | #### 5.4. Device Description The ODFS® Pace is an external functional neuromuscular stimulator. It consists of a small, body worn, single channel stimulator which is triggered by a wired footswitch worn in the shoe. Electrical pulses are generated by the device in order to stimulate muscles in the lower limb. Triggering with a footswitch permits the stimulation to be varied according to the gait of the user. Stimulation produces contraction of the appropriate muscles to cause dorsiflexion of the ankle in individuals who have impaired or absent ability to pick up their foot during walking as a result of a neurological injury. This partial paralysis affecting the foot is commonly known as 'drop foot' or 'dropped foot'. The electrical stimulation is delivered via self adhesive skin surface electrodes. The electrodes may be flexibly placed according to the optimum response of the patient but typically are located over the common peroneal nerve as it passes near the head of the fibula. For individuals presenting with weakness and/or spasticity a muscle stimulation exercise program can be delivered by the ODFS® Pace. In this case the footswitch trigger is not used and the ODFS® Pace functions as a powered muscle stimulator. A clinician would determine the need for this application which can enabled from a setup option in the ODFS® Pace. (Default is to have the exercise disabled) #### 5.5 Intended Use The ODFS® Pace is intended to provide ankle dorsiflexion in individuals who have a dropped foot as a consequence of upper motor neuron injury. By detecting the swing phase of gait through a foot switch signal, appropriate electrical stimulation of the leg and ankle muscles may improve gait by flexing the foot of persons who have lost or impaired function. There may be additional benefits from FES such as muscle re-education, Odstock Medical Ltd 510(k) Summary, v4, ODFS® PACE Page 2 of 4, Section 5 {2}------------------------------------------------ prevention/retardation of disuse atrophy, increased or maintained range of joint motion and increase in local blood flow. The ODFS® Pace is a medical device and should only be used under medical supervision for the treatment of dropped foot following an upper motor neuron injury. #### 5.6 Technological Characteristics and Substantial Equivalence The technological characteristics of the ODFS® Pace and predicate devices are tabulated below. All the devices are battery powered and deliver electrical stimulation pulses of similar waveform, pulsewidth and frequency. The ODFS® Pace is a development of the ODFS III v6.2 and therefore has identical indications for use as an FES device. These indications correspond to the FES functions of the EMPI device. The exercise function of the ODFS® Pace corresponds to and is substantially equivalent to the NMS/NMES indications of the EMPI device. | Feature | ODFS® Pace | ODFS III v6.2 | EMPI Focus 795 | |---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Waveform | Biphasic (symmetrical<br>and balanced<br>asymmetrical) | Biphasic (symmetrical<br>and balanced<br>asymmetrical) | Biphasic (symmetrical<br>and balanced<br>asymmetrical) | | Pulsewidth | $0μs - 360μs (±10%)$<br>according to patient<br>selection (3.6 μs steps) | $7μs - 365μs (±10%)$<br>according to patient<br>selection | $300μs$ | | Frequency | Default is 40Hz.<br>20 - 60 Hz in 5Hz steps<br>available | 40Hz ± 10% | 25, 30, 35, 45,50,80Hz | | Indications for use | The ODFS® Pace is<br>intended to provide<br>ankle dorsiflexion in<br>individuals who have a<br>dropped foot as a<br>consequence of upper<br>motor neuron injury. By<br>detecting the swing<br>phase of gait through a<br>foot switch signal,<br>appropriate electrical<br>stimulation of the leg<br>and ankle muscles may<br>improve gait by flexing<br>the foot of persons who<br>have lost or impaired<br>function. There may be<br>additional benefits from<br>FES such as muscle re-<br>education,<br>prevention/retardation of<br>disuse atrophy,<br>increased or maintained<br>range of joint motion<br>and increase in local | To provide ankle<br>dorsiflexion in<br>individuals who have a<br>dropped foot as a<br>consequence of upper<br>motor neuron injury. By<br>detecting the swing<br>phase of gait through a<br>foot switch signal,<br>appropriate electrical<br>stimulation of the leg<br>and ankle muscles may<br>improve gait by flexing<br>the foot of persons who<br>have lost or impaired<br>function. There may be<br>additional benefits from<br>FES such as muscle re-<br>education,<br>prevention/retardation of<br>disuse atrophy,<br>increased or maintained<br>range of joint motion<br>and increase in local<br>blood flow. | TENS device<br>Symptomatic relief and<br>management of chronic<br>intractable pain.<br>Adjunctive treatment for<br>post-surgical and post<br>trauma acute pain<br>As a NMS/NMES<br>device: relaxation of<br>muscle spasm;<br>prevention or retardation<br>of disuse atrophy;<br>increasing local blood<br>circulation; muscle<br>reduction; immediate<br>post surgical stimulation<br>of the calf muscles to<br>prevent venous<br>thrombosis; maintaining<br>or increasing range of<br>motion.<br>As a FES device:<br>Stimulating muscles in | # 5.6 Technological Characteristics and Substantial Equivalence - Table Odstock Medical Ltd {3}------------------------------------------------ # 510(k) Summary for the ODFS® PACE Dropped Foot Stimulator | | blood flow.<br>As a NMS/NMES<br>device: relaxation of<br>muscle spasm;<br>prevention or retardation<br>of disuse atrophy;<br>increasing local blood<br>circulation; muscle re-<br>education; maintaining<br>or increasing range of<br>motion. | | the leg and ankle of<br>partially paralysed<br>patients to provide<br>flexion of the foot and<br>thus improve the<br>patients gait. | |------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|-------------------------------------------------------------------------------------------------------------------------------------| | Accessories | Heel switch | Heel switch | Heel switch, hand<br>switch | | Electrode size & shape | 5cm x 5cm | Scm x 5cm | 5.1cm4 | # 5.7 Performance testing The ODFS® Pace system has been subjected to testing to verify that the device meets its functional and output specifications. Production units pass functional testing of stimulator controls, settings and indicators, stimulus output and timing, footswitch operation, and power consumption. Parameters measured under the conditions specified in FDA Guidance Document for Powered Muscle Stimulator 510(k)s are included in the ODFS® Pace 510(k) submission. In addition to functional and operational performance testing the ODFS® Pace passed the applicable requirements of the following standards: - ISO 60601-1-1:2006 Medical Electrical Equipment General requirements for . basic safety and essential performance. - I ISO 60601-1-2:2007 Medical Electrical Equipment – General requirements for basic safety and essential performance. Collateral standard: Electromagnetic compatibility - Requirements and tests - . ISO 60601-2-10:2001 Medical electrical equipment. Particular requirements for the safety of nerve and muscle stimulators Based on the information provided above the ODFS® Pace is substantially equivalent to legally marketed predicate devices. | | --X-- | | |--|-------|--| |--|-------|--| | Version | Comment | Author | | |---------|------------------------------------------------------------------------------|-------------------|--| | 3->4 | Amend indications for use and add regulation number for classification name. | 1 S.Crook 14/3/11 | | | | | | | | | | | | Odstock Medical Ltd 510(k) Summary, v4, ODFS® PACE {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized caduceus, a symbol often associated with medicine and healthcare. The symbol consists of three curved lines that converge at the bottom. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Odstock Medical, Ltd. c/o Dr. Steven Crook Quality and Regulatory Manager Salisbury District Hopital Salisbury, Wilitshire United Kingdom, SP2 8BJ MAR 3 0 2011 Re: K102115 Trade/Device Name: ODFS® Pace Regulation Number: 21 CFR 882.5810 Regulation Name: External Functional Neuromuscular Stimulator Regulatory Class: Class II Product Code: GZI Dated: January 21, 2011 Received: January 26, 2011 Dear Dr. Crook: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, signature Malvina B. Eydelman, M.D. Director Division of Ophthalmic. Neurological. and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indications for Use K102115 510(k) Number (if known) Device Name: ODFS® Pace Indications for use: The ODFS® Pace is intended to provide ankle dorsiflexion in individuals who have a dropped foot as a consequence of upper motor neuron injury. By detecting the swing phase of gait through a foot switch signal, appropriate electrical stimulation of the leg and ankle muscles may improve gait by flexing the foot of persons who have lost or impaired function. There may be additional benefits from FES such as muscle re-education, prevention/retardation of disuse atrophy, increased or maintained range of joint motion and increase in local blood flow. The ODFS® Pace is a medical device and should only be used under medical supervision for the treatment of dropped foot following an upper motor neuron injury. Prescription use X (Part CFR 801 Subpart D) and/or Over-The-Counter Use (CFR 801 Subpart C) # (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY) of CDRH, Office of Device Evaluation, (ODE) (Division Sign-O Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices 510(k) Number K102115 Odstock Medical Ltd Indications statement v4, ODFS® PACE [section 4]
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...