K140886 · Innovative Neurotronics, Inc. · GZI · Aug 8, 2014 · Neurology
Device Facts
Record ID
K140886
Device Name
WALKAIDE SYSTEM
Applicant
Innovative Neurotronics, Inc.
Product Code
GZI · Neurology
Decision Date
Aug 8, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Innovative Neurotronics WalkAide External Functional Neuromuscular Stimulator (WalkAide System) is intended to address the lack of ankle dorsiflexion in patients who have sustained damage to upper motor neurons or pathways to the spinal cord. During the swing phase of gait, the WalkAide System electrically stimulates the appropriate muscles that cause ankle dorsiflexion and may thus improve the gait in patients with chronic stroke. Medical benefits of Functional Electrical Stimulation (FES) may include prevention/ retardation of disuse atrophy, increased local blood flow, muscle re-education, and maintained or increased joint range of motion.
Device Story
External functional neuromuscular stimulator; attaches to leg below knee near fibula head. Stimulates common peroneal nerve to trigger tibialis anterior muscle contraction; produces ankle dorsiflexion during gait swing phase. Input: tilt sensor (internal) or optional foot/heel sensor (force-sensitive resistor) detects gait cycle phase. Output: biphasic electrical stimulation pulses. Clinician uses WalkAnalyst software on PC/tablet via Bluetooth-linked WalkLink to configure stimulation parameters, manage patient data, and generate reports. Patient uses device independently at home; powered by single AA battery. Benefits: improved gait speed, muscle re-education, increased blood flow, prevention of disuse atrophy, maintained joint range of motion.
Clinical Evidence
Randomized, unblinded, parallel-group clinical trial comparing WalkAide (WA) to Ankle-Foot Orthosis (AFO) in chronic stroke patients (at least 6 months post-stroke). Study population averaged 6.9 years post-stroke. Primary endpoints included gait velocity and Stroke Impact Scale (SIS) composite scores. Results demonstrated WA is equivalent to AFO for gait velocity, SIS scores, and safety, supporting clinical relevance in functional ambulation for chronic stroke patients.
Technological Characteristics
Microprocessor-controlled external stimulator; ABS housing. Stimulation: constant voltage (0-110V), biphasic asymmetrical waveform, 16.7-33.3 pps, 25-300 µs pulse duration. Electrodes: SS 316 and hydrogel. Connectivity: Bluetooth (via WalkLink for setup only). Power: 1.5V AA battery. Standards: IEC 60601-2-10, IEC 60601-1-2, AAMI/ANSI ES60601-1.
Indications for Use
Indicated for patients with upper motor neuron damage or spinal cord pathways damage causing lack of ankle dorsiflexion, including chronic stroke patients, to improve gait during the swing phase.
Regulatory Classification
Identification
An external functional neuromuscular stimulator is an electrical stimulator that uses external electrodes for stimulating muscles in the leg and ankle of partially paralyzed patients (e.g., after stroke) to provide flexion of the foot and thus improve the patient's gait.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the seal is a stylized image of three human profiles facing right, with flowing lines representing hair or fabric. To the right of the seal, the letters "DE" are visible, suggesting the full word "DEPARTMENT" was intended.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 8, 2014
Innovative Neurotronics, Inc. c/o Mr. Glen Neally 3600 North Capital of Texas Highway Building B. Suite 150 Austin, Texas 78746
Re: K140886
Trade Name: WalkAide System Regulation Number: 21 CFR 882.5810 Regulation Name: External functional neuromuscular stimulator Regulatory Class: Class II Product Code: GZI Dated: May 5, 2014 Received: May 13, 2014
Dear Mr. Neally:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Felipe Aguel -S
for
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K140886
Device Name WalkAide System
#### Indications for Use (Describe)
The Innovative Neurotronics WalkAide External Neuromuscular Stimulator (WalkAide System) is intended to address the lack of ankle dorsiflexion in patients who have sustained damage to upper motor neurons or pathways to the spinal cord. During the swing phase of gait, the WalkAide System electrically stimulates the appropriate muscles that cause ankle dorsiflexion and may thus improve the gait in patients with chronic stroke. Medical benefits of Functional Electrical Stimulation (FES) may include prevention of disuse atrophy, increased local blood flow, muscle reeducation, and maintained or increased joint range of motion.
#### Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary for the Innovative Neurotronics, Inc. WalkAide System
#### 1. SPONSOR
Innovative Neurotronics, Inc. 3600 N. Capital of Texas Highway Bldg. B. Suite 150 Austin, Texas 87846
| | Contact Person: Glen Neally – Director of Quality and Regulatory |
|------------|------------------------------------------------------------------|
| Telephone: | 1-512-721-1903 |
| Fax: | 1-512-721-1939 |
May 5, 2014 Date Prepared:
#### 2. DEVICE NAME
| Proprietary Name: | WalkAide |
|--------------------------------------------|----------------------------------------------|
| Common/Usual Name: | External Neuromuscular Functional Stimulator |
| Classification Names: | External Neuromuscular Functional Stimulator |
| Classification Number: 21 CFR 882.5810 | |
| Product Code: | GZI |
- 3. PREDICATE DEVICES WalkAide System- K123972
#### 4. INTENDED USE
The Innovative Neurotronics WalkAide External Functional Neuromuscular Stimulator (WalkAide System) is intended to address the lack of ankle dorsiflexion in patients who have sustained damage to upper motor neurons or pathways to the spinal cord. During the swing phase of gait, the WalkAide System electrically stimulates the appropriate muscles that cause ankle dorsiflexion and may thus improve the gait in patients with chronic stroke. Medical benefits of Functional Electrical Stimulation (FES) may include prevention/ retardation of disuse atrophy, increased local blood flow, muscle re-education, and maintained or increased joint range of motion.
{4}------------------------------------------------
#### 5. DEVICE DESCRIPTION
The WalkAide is an external functional electrical stimulator. It is a small device that attaches to the leg just below the knee, near the head of the fibula. During a gait cycle, the WalkAide stimulates the common peroneal nerve, which innervates the tibialis anterior and other muscles that cause dorsiflexion of the ankle. The WalkAide System consists of the WalkAide Patient Kit and the WalkAide Clinician Kit . The WalkAide Patient Kit comprises of all the components and accessories that the patient will take home and use. The Clinician System comprises the accessories that a clinician (i.e., orthotic specialist, physiotherapist, occupational therapist, etc.) will use to set up a patient's WalkAide.
### WalkAide Patient Kit
The WalkAide Patient Kit consists of the components and accessories that the patient takes home. The WalkAide Patient Kit includes (See Fig 4.1):
- WalkAide Control Module (Stimulator Unit) ●
- WalkAide Leg Cuff ●
- WalkAide Electrode Cable ●
- Electrodes (ordered seperately) ●
- Foot Sensor (Optional) .
Image /page/4/Picture/9 description: The image shows a medical device with a gray and black design. The device has a control panel with a blue knob and two white buttons. A gray cable is connected to the device, and a tan-colored sensor with a black wire is also visible. The device appears to be used for medical monitoring or treatment.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a collection of medical devices and accessories. There is a foot sensor template with cutting guidelines and size markings. Additionally, there are two circular electrode pads with attached wires, a beige-colored adjustable strap, a device with a blue knob, and various connecting cables with red and black connectors.
WalkAide Patient Kit
## WalkAide Control Module (WalkAide)
As with the parent K123972 WalkAide control module, the Innovative Neurotronics WalkAide Control Module houses the electronics and controls of the WalkAide system, and delivers electrical stimulation to the end-user's peroneal nerve via surface electrodes. The WalkAide attaches to the Leg Cuff and provides the stimulation pulse to the electrodes. The WalkAide requires a single AA battery to operate.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a person's leg with a medical device attached to the upper part of the leg. The device is attached with a strap. The leg is shown from the thigh to the foot. The person's toes are visible.
### WalkAide attached to WalkAide Cuff on a Patient's Leg
As with the K123972 WalkAide, the Innovative Neurotronics WalkAide is a microprocessor controlled device, and uses an EEPROM to store device, user, and configuration information. The intensity is controlled by the potentiometer on the top face of the WalkAide control module. The position of the potentiometer controls the appropriate stimulation intensity.
Stimulation could be triggered by three mechanisms: the Tilt Sensor inside the control module, a Foot Sensor or Heel Sensor, or through the Hand Switch (the WalkLink.) Three LEDs (green, amber, and red) provide indications such as start up, stimulation, low battery, and fault. Additional information about these indicators is provided in the User Manual in Appendix A-2.
### WalkAide Bi-Flex Cuff
The WalkAide is designed for single-handed application and removal. It may take a bit of practice to develop a routine that works best for each person. The WalkAide is applied directly to the leg and can be easily worn under most clothing. The clinician will find the optimal placement of the electrodes on the initial visit. The electrode placement will be marked on
{7}------------------------------------------------
the inside of the cuff via Red and Black Electrode Locators and the position should not be moved by the patient.
Image /page/7/Picture/1 description: The image shows two different views of a medical device. The device appears to be a head-mounted apparatus with a gray frame and black straps. The top image shows the device from a side angle, revealing a padded interior with circular openings and a control panel with a blue knob. The bottom image shows the underside of the device, highlighting the frame's structure and the attachment points for the straps.
WalkAide Cuff
## Electrodes
As with the K123972 Walk-Aide, the Innovative Neurotronics WalkAide System uses two surface electrodes. The WalkAide electrodes consist of electrodes that have been cleared under 510(k) (currently numbers K872976, K962332, K070807).
{8}------------------------------------------------
Image /page/8/Figure/0 description: This image shows a technical drawing of an object with various dimensions and labels. On the left, a circular object labeled "1" has a diameter of 1.25 inches (32mm) and a height of 0.75 inches (19mm). The total height of the object, including the handle labeled "2", is 3.50 +/- 0.25 inches (89mm). The right side of the image shows a side view of the object with labels 1A, 1B, 1C, 2A, 2B, 2C, and 2D.
WalkAide Electrodes
## Electrode Cable
This is a Y connector cable for connecting the electrodes with WalkAide. One end is connected to the connector slot on the back of the electrode, and the other two ends (black and red) are connected to respective electrodes.
Image /page/8/Figure/4 description: The image shows a Y-splitter cable with two female connectors on one end and a male connector on the other end. The two female connectors are identical and have a cylindrical shape with a hollow center. The male connector is right-angled and has a cylindrical shape with a pin in the center. The cable is white and has a smooth surface.
WalkAide Electrode Cable
{9}------------------------------------------------
### Foot Sensor (Optional Item)
The Foot Sensor is the same as in the parent WalkAide system. No changes have been made to the foot sensor. The Foot Sensor is designed for those individuals whose leg movement is limited such that the Tilt Sensor cannot be adequately used. The Foot Sensor is based on a force sensor resistor (FSR) as the detection element. The Foot Sensor plugs into the receptacle on the side of the WalkAide control module as it did in the parent device.
Image /page/9/Picture/2 description: The image shows a line drawing of a heated insole with a cord and plug. The insole is shaped like a foot and has several layers. The cord is attached to the insole and has a plug at the end. The plug is a standard 3.5mm audio plug. The insole is likely used to keep feet warm in cold weather.
Foot Sensor
### WalkAide Clinician Kit
The WalkAide Clinician Kit, which is similar to the Clinician System used in the parent WalkAide, is used by trained professionals (clinicians) in a clinical setting to configure WalkAide for an individual user. The WalkAide Clinician Kit is only used by the clinician and it is not sent with the patient.
Innovative Neurotronics uses the WalkAide Clinician System Software called the WalkAnalyst, with Notebook/Tablet PC/Desktop platform for the ease of use and better display. The Notebook/Tablet PC/Desktop is not provided by Innovative Neurotronics. The WalkAnalyst has been developed and validated according to regulatory guidance documents.
The WalkAide consists of the following components:
- WalkAnalyst Software ●
- WalkLink (Hand Switch)
- Heel Sensor
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows a white box with a black lanyard attached to the top. There is a black cord coming out of the bottom of the box. To the left of the box is a gray cord with a plug on the end and a tan-colored object. To the right of the box is a black cord.
Clinician Kit without WalkAnalyst Software
## WalkAnalyst Software
The WalkAnalyst software has been developed exclusively for the proposed Innovative Neurotronics WalkAide. This software can be installed on a Tablet PC, notebook, or desktop PC hardware platform. The WalkAnalyst performs the following functions:
- Set up configuration of WalkAide Control Module, which includes management of data input from tilt sensor, and heel/foot sensor
- Management of patient information ●
- . Graphical display of stimulation information and gait recordings
- Analysis of data and optimization of stimulation parameters
- Report generation .
The WalkAnalyst software communicates with WalkAide via a Bluetooth wireless communication device in the WalkLink.
## WalkLink (Hand Switch)
The WalkLink is only used during the patient set-up and control module configuration process. The WalkLink serves two purposes (1) the
{11}------------------------------------------------
Clinician can stimulate the patient at the appropriate times during the setup process (2) It communicate gait recording from WalkAide to the PC (Tablet/Notebook/Desktop) using Bluetooth (BT) wireless communication protocol. It should be noted that WalkAide itself does not have Bluetooth functionalities. WalkAide communicates with WalkLink using the WalkLink cable. WalkLink transmits the information received from WalkAide to the PC using the Bluetooth protocol.
The WalkLink contains a stimulation button and three LED indicators. The green blinking indicator shows adequate battery voltage, the red indicator indicates low battery or error, and the blue blinking indicator indicates that BT link is established.
Image /page/11/Picture/2 description: The image shows two devices, one is a slim device with a pairing button, and the other is a WalkLink device. The WalkLink device has a logo at the top and three lights, one green, one red, and one blue. The device also has an FC label and the text "Tested to comply with FCC standards".
WalkLink
## WalkLink Cable
The WalkLink cable connects WalkLink with WalkAide. This cable contains standard RS232 communication lines. This is a 5 conductor stranded cable approximately 40 inches long cable.
Image /page/11/Figure/6 description: The image shows a drawing of a flexible endoscope. The endoscope is a long, thin tube with a camera and light at the end. It is used to view the inside of the body. The endoscope in the image has a curved shape. The endoscope has a connector at each end.
{12}------------------------------------------------
#### WalkAide to WalkLink Cable
### Heel Sensor
The heel sensor is the same patient heel sensor used in the parent WalkAide system. No changes have been made to the Heel Sensor. The design is similar to the Foot Sensor, as it incorporates a force sensitive sensor (FSR) as the detection element. The Heel Sensor is intended for use in the clinic by the Clinician during the device set-up stage.
Fig 4.10: Heel Sensor
#### 4.5 Operations and Set-up of the WalkAide System
The operation and set-up of the WalkAide system is detailed in the Clinician Manual. The following is a summary:
The patient places the back electrode slightly posterior and distal to the head of the fibula and the front electrode on the upper 1/3 of the tibialis anterior muscle over the muscle belly. The back electrode is connected to the black lead and the front electrode to the red lead. The WalkAide is turned on by rotating the intensity control on the Control Module. The round intensity knob is moved counter-clockwise through "0" or in the direction of increasing numbers. The STIM button labeled (J) is depressed for 1 to 2 seconds to provide a test stimulus. An amber light on the top of the
{13}------------------------------------------------
WalkAide Control Module indicates that the unit is stimulating. If the ankle movement produced by the stimulus is too small, the intensity control is turned up or the black electrode is adjusted slightly to the rear, then application of the test stimulus is attempted again. Stimulation consists of delivering a train of biphasic current pulses to the common peroneal nerve via two surface electrodes. The goal is to produce a relatively pure dorsiflexion of the ankle with as low stimulus intensity as possible. Once the optimal electrode positions have been found, the cuff is placed over the electrodes.
The cuff is then removed with the electrodes that are affixed by Velcro to the inside of the cuff. The Clinician then positions the electrodes on the leg cuff for patient reference during future electrode replacement. The leads from the electrodes are threaded through the holes on the cuff. The black lead of the Y-connector cable is connected to the back electrode and the red to the anterior electrode. The electrodes, leads and connectors are placed in the groove or pocket of the cuff and covered with a Velcro strap. A heel sensor is placed in the patient's shoe on the affected side and is connected to the WalkAide Control Module. As with the parent Walk-Aide, the WalkAide Control Module houses the electronics and controls of the WalkAide, and delivers electrical stimulation to the patient via two surface electrodes.
The Clinician collects and analysis gait recording by using WalkAnalyst Software and according to set-up shown in Fig 4.11. After optimal parameters are selected, WalkAide is configured using the selected parameters.
{14}------------------------------------------------
Image /page/14/Figure/0 description: The image shows a diagram of a clinician using a WalkLink device to apply manual stimulation to a patient. The patient is wearing a heel sensor in their shoe, and they are walking in a figure-eight pattern with a WalkAide device. The WalkAide is connected to the WalkLink, and a tablet PC, notebook, or desktop computer is collecting heel sensor and stimulation data for the clinician using WalkAnalyst. The diagram is labeled with the numbers 1-4, which correspond to the different steps in the process.
Fig 4.11: WalkAide Set-up Process
Once set up and configuration has been completed by the Clinician, WalkLink is disconnected from the WalkAide. WalkAide, including the accessories such as the WalkAide Cuff, Electrodes, Electrode Cable, and Foot Sensor (optional item) is provided to the patient.
WalkAide in normal operation mode will use the Tilt Sensor or optional Foot Sensor to sense the appropriate point in the patient's gait (walk) cycle to trigger stimulation. The WalkAide provides visual and audible indications for various conditions such as operational status and low battery condition.
{15}------------------------------------------------
#### 6. BASIS FOR SUBSTANTIAL EQUIVALENCE
The purpose of this 510(k) is to update the submission of the Walk-Aide System (K123972) to include in the Indications For Use Summary Statement the specific population of chronic stroke patients for the specific indication of improved gait. This is the only change in this submission.
Clinical data is submitted to support the difference in indication for use. The randomized trial utilized an unblinded, parallel group design. The purpose of this study was to compare of the WalkAide Functional Electrical Stimulation device (WA) versus an Ankle-Foot Orthosis (AFO) in subjects with chronic stroke. Subjects were required to be at least 6 months post stroke. The study data was analyzed using an intention to treat analysis with missing data points calculated using multiple imputations. The clinical trial data outcomes support chronic stroke population for gait improvement.
Comparison Conclusion Table:
{16}------------------------------------------------
| FEATURE | INNOVATIVE<br>NEUROTRONICS, INC.<br>WALKAIDE<br>Current Submission | INNOVATIVE<br>NEUROTRONICS, INC.<br>WALKAIDE<br>K123972 | COMPARISON |
|-----------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| Intended Use | SUMMARY STATEMENT<br>The Innovative Neurotronics<br>WalkAide External Functional<br>Neuromuscular Stimulator<br>(WalkAide System) is intended<br>to address the lack of ankle<br>dorsiflexion in patients who have<br>sustained damage to upper motor<br>neurons or pathways to the spinal<br>cord. During the swing phase of<br>gait, the WalkAide System<br>electrically stimulates the<br>appropriate muscles that cause<br>ankle dorsiflexion and may thus<br>improve the gait in patients with<br>chronic stroke. Medical benefits<br>of Functional Electrical<br>Stimulation (FES) may include<br>prevention/ retardation of disuse<br>atrophy, increased local blood<br>flow, muscle re-education, and<br>maintained or increased joint<br>range of motion. | SUMMARY STATEMENT<br>The Innovative Neurotronics<br>WalkAide External Functional<br>Neuromuscular Stimulator<br>(WalkAide System) is intended<br>to address the lack of ankle<br>dorsiflexion in patients who<br>have sustained damage to<br>upper motor neurons or<br>pathways to the spinal cord.<br>During the swing phase of gait,<br>the WalkAide System<br>electrically stimulates the<br>appropriate muscles that cause<br>ankle dorsiflexion and may<br>thus improve the patient's gait.<br>Medical benefits of Functional<br>Electrical Stimulation (FES)<br>may include prevention/<br>retardation of disuse atrophy,<br>increased local blood flow,<br>muscle re-education, and<br>maintained or increased joint<br>range of motion. | This current<br>submission has<br>included the<br>indication of gait<br>improvement for<br>chronic stroke<br>patients. |
| FEATURE | INNOVATIVE<br>NEUROTRONICS, INC.<br>WALKAIDE<br>Current Submission | INNOVATIVE<br>NEUROTRONICS, INC.<br>WALKAIDE<br>K123972 | COMPARISON |
| Power Source | 1.5 V AA Battery<br>(not rechargeable) | 1.5<br>V<br>AA<br>Battery<br>(not rechargeable) | Identical |
| Microprocessor Controlled | Yes | Yes | Identical |
| Indicators<br>Unit functioning<br>Low Battery | Yes<br>Yes | Yes<br>Yes | Identical |
| Patient Device Set-up and<br>Training | Done by a clinician using<br>a Notebook/Tablet PC | Done by a clinician<br>using a Notebook/Tablet<br>PC | Identical |
| Number of Output Modes | 1 | 1 | Identical |
| Channels | 1 | 1 | Identical |
| Output Stage<br>Type<br>Range/Accuracy<br>Load | Constant Voltage<br>(Adjustable)<br>0 - 110 V (±10%)<br>1000 Ohm load | Constant Voltage<br>(Adjustable)<br>0 - 110 V (±10%)<br>1000 Ohm load | Identical |
| Max Output Current | <208 mA peak @<br>500 Ohm load<br><121 mA peak @<br>1000 Ohm load | <208 mA peak @<br>500 Ohm load<br><121 mA peak @<br>1000 Ohm load | Identical |
| Max Output Voltage<br>(baseline to peak; load) | 121 V<br>1000 Ohm load<br><150 V<br>1M Ohm load | 121 V<br>1000 Ohm load<br><150 V<br>1M Ohm load | Identical |
| FEATURE | INNOVATIVE<br>NEUROTRONICS, INC.<br>WALKAIDE<br>Current Submission | INNOVATIVE<br>NEUROTRONICS, INC.<br>WALKAIDE<br>K123972 | COMPARISON |
| Ramp Modulations (for gait)<br>Ramp Up<br>Ramp Down | Yes<br>Yes | Yes<br>Yes | Identical |
| Waveform<br>Monophasic or Biphasic<br>Symmetrical or<br>Asymmetrical Shape | Biphasic<br>Asymmetrical | Biphasic<br>Asymmetrical | Identical |
| Pulse Duration | 25-300 microseconds.<br>Accuracy ±5% or ±7<br>microseconds, whichever<br>is greater. | 25-300 microseconds.<br>Accuracy ±5% or ±7<br>microseconds, whichever<br>is greater. | Identical |
| Frequency Range<br>(Pulses per second) | 16.7 to 33.3 (Adjustable)<br>pps | 16.7 to 33.3 (Adjustable)<br>pps | Identical |
| Stimulation Trigger Source<br>(When used for gait) | Tilt Sensor or Foot Sensor | Tilt Sensor or Foot<br>Sensor | Identical |
| Burst Duration<br>Stimulation when used for<br>gait<br>Max. Burst Duration<br>(Seconds) | Dependent on length and<br>speed of stride<br>5 | Dependent on length and<br>speed of stride<br>5 | Identical |
| Max Phase Charge:<br>500 Ohms:<br>1K Ohms: | 50 microCoulombs<br>30 microCoulombs | 50 microCoulombs<br>30 microCoulombs | Identical |
| Max. Current Density | 25.3 mA/cm², peak | 25.3 mA/cm², peak | Identical |
| FEATURE | INNOVATIVE<br>NEUROTRONICS, INC.<br>WALKAIDE<br>Current Submission | INNOVATIVE<br>NEUROTRONICS, INC.<br>WALKAIDE<br>K123972 | COMPARISON |
| Maximum Average Current<br>Density | 0.25 mA/cm², avg. | 0.25 mA/cm², avg. | Identical |
| Max. Average Power Density | 27.6 mW/cm²<br>(Using 500 Ohm) | 27.6 mW/cm²<br>(Using 500 Ohm) | Identical |
| Electrode Size and Shape<br>(Smallest recommended) | 3.175 cm (1.25 inches)<br>diameter<br>Round | 3.175 cm (1.25 inches)<br>diameter<br>Round | Identical |
| Electrode material<br>(Basic mechanical element) | Hydrogel and Stainless<br>Steel (SS) 316 | Hydrogel and Stainless<br>Steel (SS) 316 | Identical |
| Material contacting skin | | | |
| Housing Material | ABS | ABS | Identical |
{17}------------------------------------------------
{18}------------------------------------------------
{19}------------------------------------------------
Non-clinical performance testing summary:
{20}------------------------------------------------
17-11 IEC 60601-2-10 Edition 2.0 2012-06, medical electrical equipment -- part 2-10: particular requirements for the basic safety and essential performance of nerve and muscle stimulators. (Neurology)
19-1
IEC 60601-1-2 Edition 3: 2007-03, medical electrical equipment - part 1-2: general requirements for basic safety and essential performance - collateral standard: electromagnetic compatibility - requirements and tests. (General II (ES/EMC))
19-4
AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012,, c1:2009/(r)2012 and a2:2010/(r)2012 (consolidated text) medical electrical equipment - part 1: general requirements for basic safety and essential performance (iec 60601-1:2005, mod). (General II (ES/EMC))
7. CONCLUSIONS
{21}------------------------------------------------
As indicated in the comparison table above the device and software are identical to the predicate K123972. The only change is in the indications for use, and clinical data has been presented for this.
The results of this study show the WA to be equivalent to the AFO for improvements in gait velocity, SIS composite score and safety. The WA produces physiological dorsiflexion, with all the motor and sensory benefits inherent in active muscle contraction, including improved gait speed and improved overall quality of gait for individuals poststroke. Results from this study support the fact that use of the WA produces clinically relevant in functional ambulation and thus should be considered a viable alternative to conventional bracing for individuals with foot drop secondary to chronic stroke. Because these results were reported in a population of individuals averaging 6.9 years from onset of stroke, this study demonstrates that functional improvements can be obtained in the chronic phase of stroke with appropriate intervention, and that continued recovery of function should remain a goal throughout the lifetime of a person with stroke.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.