K994387 · Woodside Biomedical, Inc. · GZJ · Mar 16, 2000 · Neurology
Device Facts
Record ID
K994387
Device Name
RELIEFBAND NST DEVICE, MODEL WB-2L, WB-6L, WB-RL
Applicant
Woodside Biomedical, Inc.
Product Code
GZJ · Neurology
Decision Date
Mar 16, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ReliefBand® NST™ Device is indicated for use in the treatment of nausea and vomiting (NV) due to chemotherapy, motion sickness, and pregnancy. The ReliefBand® NST™ Device is indicated as an adjunct to antiemetics in reducing postoperative nausea (PON).
Device Story
Non-invasive nerve stimulation therapy device worn on wrist; provides electrical stimulation to nerves in wrist to treat nausea and vomiting. Device consists of wristband with user display; five blinking LEDs indicate stimulation intensity; pushbutton allows patient to select intensity level; sixth LED indicates low battery. Powered by two 3V lithium batteries; disposable models (WB-2L, WB-6L) have non-replaceable batteries; reusable model (WB-RL) has user-replaceable batteries. Patient operates device by selecting intensity level via pushbutton; stimulation affects nerves to provide relief from nausea/vomiting. Used in clinical or home settings.
Clinical Evidence
No clinical data provided; substantial equivalence based on performance data comparison to predicate devices.
Technological Characteristics
Non-invasive nerve stimulation device; wrist-worn form factor; electrical stimulation principle; powered by two 3V lithium batteries; user interface includes 5-level intensity LED display and pushbutton control; disposable (2-day, 6-day) and reusable models.
Indications for Use
Indicated for treatment of nausea and vomiting due to chemotherapy, motion sickness, and pregnancy; also indicated as an adjunct to antiemetics for reducing postoperative nausea in patients.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
MAR 1 6 2000
K99438 WOODSIDE BIOMEDICAL, INC.
ReliefBand™ NSTT™ Model WB-xL Family
Original Premarket 510(k) Notification
10F3
## SUMMARY OF SAFETY AND EFFECTIVENESS Woodside Biomedical Inc. ReliefBand® NST™ Device Models WB-2L, WB-6L, and WB-RL
### SUBMITTER INFORMATION
| A | Company Name: | Woodside Biomedical, Inc. |
|----|--------------------------------|-----------------------------------------------------------------------------------|
| B. | Company Address: | 1915 Aston Avenue, Suite 102<br>Carlsbad, CA 92008 |
| C. | Company Phone:<br>Company Fax: | (760) 804-6900<br>(760) 804-6925 |
| D. | Contact Person: | Tom Grey<br>Vice-President of Product<br>Development<br>Woodside Biomedical, Inc. |
| E. | Date Summary Prepared: | December 19, 1999 |
# DEVICE IDENTIFICATION
| A. | Generic Device Name: | Nerve Stimulation Therapy Device |
|----|-------------------------|---------------------------------------------------------|
| B. | Trade/Proprietary Name: | ReliefBand® NSTTM (Nerve Stimulation<br>Therapy) Device |
| C. | Classification: | Class II |
| D. | Product Code: | GZJ and BWK |
### IDENTIFICATION OF PREDICATE DEVICE
| Device | Manufacturer | 510(k) No. | Date Cleared |
|----------------|---------------------------|------------|------------------|
| ReliefBand, | Woodside Biomedical, Inc. | K982436 | October 8, 1998 |
| Models WB-2, | | K983907 | December 9, 1999 |
| WB-6, and WB-R | | | |
{1}------------------------------------------------
WOODSIDE BIOMEDICAL, INC. RelietBand™ NST™ Model WB-xL Family Original Premarket 510{k) Notification
20F3
#### DEVICE DESCRIPTION
The ReliefBand® NST™ Device Models WB-2L, WB-6L, and WB-RL are non-invasive nerve stimulation therapy devices, and are indicated for use in the treatment of nausea and vomiting (NV) due to chemotherapy, motion sickness, and pregnancy. The ReliefBand ® NST™ Device is indicated as an adjunct to antiemetics in reducing postoperative nausea (PON). The devices are contained within a wristband, and provide relief of NV and PON through electrical stimulation of the nerves in the patient's wrist.
The devices can be worn on the ventral or palmar (i.e., inside) surface of either or both wrists, approximately 2 fingers breadth above the distal skin crease of the wrist joint, between the tendons of the palmaris longus and flexor carpi radialis muscles.
The ReliefBand™ NST™ device Model WB-xL family has a user display that incorporates five blinking LEDs which are used to identify the intensity level (5 discrete LEDs, one for each intensity level), so that the patient can easily select the desired stimulation. Selection of the intensity level is performed via a pushbutton located on the user display, which controls the peak pulse amplitude of the electrical impulse and thereby determines the intensity of the stimulation. A sixth blinking LED is used to display the low battery indicator.
The ReliefBand™ NST™ device Model WB-xL family is powered by two commercially available 3V lithium batteries. These batteries are not user replaceable in the disposable 2 day (WB-2L) and 6 day (WB-6L) devices, but are user replaceable in the Reusable device (WB-RL). The battery life for the 2 day device is specified to be 50 hours when used at setting 3. The battery life for the 6 day and Reusable devices is specified to be 150 hours when used at setting 3.
{2}------------------------------------------------
WOODSIDE BIOMEDICAL, INC ReliefBand™ NSTTM Model WB-xL Family Original Premarket 510(k) Notification
3 of 5
#### SUBSTANTIAL EQUIVALENCE
The Woodside Biomedical, Inc. ReliefBand® NST™ device Models WB-2L, WB-6L, and WB-RL are of comparable type and are substantially equivalent to the predicate ReliefBand® NST™ device Models WB-2, WB-6, and WB-R (K982436 and K983907).
#### INDICATION FOR USE
The ReliefBand® NST™ Device is indicated for use in the treatment of nausea and vomiting (NV) due to chemotherapy, motion sickness, and pregnancy. The ReliefBand® NST™ Device is indicated as an adjunct to antiemetics in reducing postoperative nausea (PON).
#### TECHNOLOGICAL CHARACTERISTICS
A comparison of the technological characteristics of the ReliefBand® NST™ Device and the predicate device has been performed. The results of this comparison demonstrate that the ReliefBand® NST™ Device is equivalent to the marketed predicate device.
#### PERFORMANCE DATA
The performance data indicate that the ReliefBand® NST™ Device Models WB-2L, WB-6L, and WB-RL are substantially equivalent to the ReliefBand® Devices distributed under K982436 and K983907.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a serpent entwined around it, overlaid on a circular seal. Encircling the caduceus are the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The text is in all caps and appears to be in a sans-serif font.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 6 2000
Woodside Biomedical, Inc. Ms. Carol L. Patterson c/o Patterson Consulting Group 21911 Erie Lane Lake Forest, California 92630
Re: K994387
Trade Name: ReliefBand® NST™ Device Models WB-2L, WB-6L, WB-RL Regulatory Class: II Product Code: GZJ and BWK Dated: December 24, 1999 Received: December 28, 1999
Dear Ms. Patterson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general control provisions of the Act. The general control provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{4}------------------------------------------------
Page 2 - Ms. Carol L. Patterson
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
i. W. R.P. Ogden
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE
| 510(k) Number: | K994387 (To Be Assigned By FDA) |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | ReliefBand® NST™ Device Models WB-2L, WB-6L, and<br>WB-RL |
| Indications For Use: | The ReliefBand® NST™ Device is indicated for use in the<br>treatment of nausea and vomiting (NV) due to<br>chemotherapy, motion sickness, and pregnancy. The<br>ReliefBand® NST™ Device is indicated as an adjunct to<br>antiemetics in reducing postoperative nausea (PON). |
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
NRO for cmo
(Division Sign-Off)
Division of General Restorative Devices
510(k) Number K994387
OR
× Prescription Use
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(Per 21 CFR 801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.