K173861 · Dyansys, Inc. · PZR · May 2, 2018 · Neurology
Device Facts
Record ID
K173861
Device Name
Drug Relief
Applicant
Dyansys, Inc.
Product Code
PZR · Neurology
Decision Date
May 2, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5896
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Drug Relief is a percutaneous nerve field stimulator (PNFS) system, that can be used as an aid to reduce the symptoms of opioid withdrawal, through application to branches of Cranial Nerves V, VII, IX and X, and the occipital nerves identified by transillumination.
Device Story
Drug Relief is a wearable, battery-operated percutaneous nerve field stimulator (PNFS) used in clinics, hospitals, or home environments to aid opioid withdrawal symptom reduction. Device delivers low-level electrical pulses to auricular cranial nerve branches (V, VII, IX, X) and occipital nerves via titanium stimulation needles. System comprises a pulse generator, wire assembly, and needles; powered by three 1.4V zinc-air batteries. Operation lasts 120 hours (5 days) with a 2-hour ON/1-minute OFF cycle. Constant current source ensures consistent stimulation energy regardless of skin impedance. Healthcare providers or patients apply the device; needles are inserted at specific auricular points. Output is a rectangular pulse waveform with 1-10 Hz modulating frequency. A 3-pin connector allows voltage output verification before application. Device provides non-pharmacological symptom relief.
Clinical Evidence
Bench testing only. Performance testing verified pulse width, duration, amplitude, and current values over a 120-hour continuous operation. Biocompatibility testing conducted per ISO 10993. Electrical safety and EMC testing performed per IEC 60601-1 and 60601-1-2. Sterilization validated per ISO 11135.
Technological Characteristics
Wearable PNFS system; titanium stimulation needles; zinc-air battery powered (3x1.4V); rectangular pulse waveform (1-10 Hz); 120-hour operating life. EtO sterilization. Complies with IEC 60601-1, IEC 60601-1-2, ISO 10993 (1, 5, 6, 7, 10, 11), and ISO 11135. Dimensions: 50x23x7 mm. Weight: 6g.
Indications for Use
Indicated for adults as an aid to reduce symptoms of opioid withdrawal via percutaneous nerve field stimulation of cranial nerves V, VII, IX, X and occipital nerves.
Regulatory Classification
Identification
A percutaneous nerve stimulator for substance use disorders is a device that stimulates nerves percutaneously to aid in the reduction of withdrawal symptoms associated with substance use disorders.
Special Controls
In combination with the general controls of the FD&C Act, the percutaneous nerve stimulator for substance use disorders is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible.
(2) Electromagnetic compatibility and electrical, mechanical, and thermal safety testing must be performed.
(3) Electrical performance testing of the device and electrodes must be conducted to validate the specified electrical output and duration of stimulation of the device.
(4) Software verification, validation, and hazard analysis must be performed.
(5) Sterility testing of the percutaneous components of the device must be performed.
(6) Shelf life testing must be performed to demonstrate continued sterility, package integrity, and device functionality over the specified shelf life.
(7) Labeling must include the following:
(i) A detailed summary of the device technical parameters;
(ii) A warning stating that the device is only for use on clean, intact skin;
(iii) Instructions for use, including placement of the device on the patient; and
(iv) A shelf life.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services USA and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized eagle emblem encircled by the department's name. To the right is the FDA logo, with the acronym "FDA" in a blue square, followed by "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 2, 2018
DyAnsys, Inc. Srini Nageshwar CEO 300 North Bavshore Boulevard San Mateo, California 94401
Re: K173861
Trade/Device Name: Drug Relief Regulation Number: 21 CFR 882.5896 Regulation Name: Percutaneous Nerve Stimulator For Opiod Withdrawal Regulatory Class: Class II Product Code: PZR Dated: April 18, 2018 Received: April 20, 2018
Dear Srini Nageshwar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
William J. Heetderks -S 2018.05.02 15:36:03 -04'00'
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173861
Device Name Drug Relief
Indications for Use (Describe)
The Drug Relief is a percutaneous nerve field stimulatory (PNFS) system, that can be used as an aid to reduce the symptoms of opioid withdrawal, through application to branches of Cranial Nerves V, VII, IX, and X, and the occipital nerves identified by transillumination.
Type of Use (*Select one or both, as applicable*)
| <div> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|----------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for DyAnsys. The logo consists of a stylized symbol to the left of the company name. The symbol is a combination of a greater-than sign and a curved line, with the curved line colored in red. The company name, "DyAnsys", is written in a simple, sans-serif font and is colored in black.
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.87 and 21 CFR 807.92.
510(k) Number: K173861
## Applicant Information:
| Date Prepared: | |
|----------------|--|
|----------------|--|
| Name: | DyAnsys, Inc |
|-------------------|---------------------------------------------------------|
| Address: | 300, North Bayshore Boulevard, San Mateo, CA 94401, USA |
| Contact Person: | Srini Nageshwar |
| Phone Number: | 408.480.4700 |
| Facsimile Number: | (650)556-1621 |
#### Device Information
| Classification | : Class II |
|---------------------|------------------------------------------------------------|
| Trade Name | : Drug Relief |
| Classification Name | : Percutaneous nerve stimulator for substance use disorder |
| Product Code | : PZR |
#### Predicate Device:
DEN Number: DEN170018 Model Name: NSS-2 BRIDGE Manufacturer: Innovative Health Solutions, Inc.
#### Device Description:
The Drug Relief™ is designed to aid in the treatment of opiate withdrawal symptoms by the method of electrical stimulation at the auricular stimulation points. The Drug Relief is a wearable, battery-operated device that is designed to administer periodical low-level electrical pulses to the ear over five days / 120 hours (2 hours ON/1 minute OFF) from the time of activation of the device.
The electrical pulse from the device will be delivered to the branches of Cranial Nerves on the ear through a set of wire assembly and Stimulation needles. Three zinc air batteries with 1.4 V each
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for DyAnsys. The logo consists of a stylized symbol to the left, resembling a combination of a greater-than sign and a curved line, with the right portion of the symbol colored in red. To the right of the symbol is the text "DyAnsys" in a simple, sans-serif font, with the letters in black.
provides the required stimulation energy for a maximum of 120 hours. There are three Stimulation electrodes and one ground electrode - which constitute of a needle and lead/ wire with the snap-fit ring. The stimulation needles are inserted at three specific points, which have the ability to stimulate the cranial nerves. The ground electrode is inserted at one specific point (constant in all treatments) which forms the functional earthing to the device.
This constant current source guarantees equivalent stimulation energy regardless of the individual impedance of the skin.
The stimulation pattern consists of rectangular pulses with differing inter-pulse intervals.
A 3-pin connector is provided, which is used to check the output voltage of the device once it is activated and before applying to the patient with any one of the voltage measuring devices available in the market with the appropriate regulatory compliance.
#### Intended Use:
The Drug Relief is a percutaneous nerve field stimulator (PNFS) system, that can be used as an aid to reduce the symptoms of opioid withdrawal, through application to branches of Cranial Nerves V, VII, IX and X, and the occipital nerves identified by transillumination.
#### Comparison to Predicate Device:
The DyAnsys, Inc Drug Relief device is substantially equivalent to it's own legally marketed predicate device NSS-2 BRIDGE (DEN 170018). It was evaluated through Performance and Non-Clinical testing.
| MODEL NAME | Drug Relief<br>(Subject) | NSS-2 BRIDGE<br>(DEN 170018) |
|-------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| MANUFACTURER | DyAnsys Inc | Innovative Health Solutions, Inc. |
| INDICATIONS FOR USE | The Drug Relief is a percutaneous nerve field<br>stimulator (PNFS) system, that can be used as<br>an aid to reduce the symptoms of opioid<br>withdrawal, through application to branches<br>of Cranial Nerves V, VII, IX and X, and the<br>occipital nerves identified by<br>transillumination. | The NSS-2 BRIDGE is a percutaneous nerve<br>field stimulator (PNFS) system, that can be<br>used as an aid to reduce the symptoms of<br>opioid withdrawal, through application to<br>branches of Cranial Nerves V, VII, IX and X,<br>and the occipital nerves identified by<br>transillumination. |
| PATIENT POPULATION | Adults | Adults |
| SHAPE | Rectangle | Rectangle |
| DIMENSION, mm | 50*23*7 mm | 36*16*7 mm |
| WEIGHT, Kg | 6 gm (including battery) | 4 gm |
| DEVICE ADHESIVE | The device's foam pad is connected to the<br>adhesive – gel pad | Tegaderm is used to fasten the device. |
| NEEDLE DIMENSIONS, | 0.4*2 mm (width*length) | 0.5mm width x 2mm length |
| mm | | |
| WIRE ASSEMBLY | 4 units of wire with snap-fit ring.<br>Where 3 nos constitute a single assembly for<br>Stimulation and the other one separate wire<br>will act as a ground electrode. | One stainless steel wire with four<br>stimulation needles and two stainless steel<br>wires with one stimulation needle each.<br>There is one ground wire with one needle. |
| WIRE ASSEMBLY TYPE | Wire assembly, without stimulation needles<br>soldered with the pulse generator. | Wire assembly is connected with the<br>stimulation needles as one end and has the<br>facility to connect with the pulse generator. It<br>is sterilized and packed separately. |
| VOLTAGE<br>MEASUREMENT<br>FEASIVLITY | A 3-pin connector provided to measure the<br>output voltage of the device once it is<br>activated. | Not feasible.<br>There is no provision to measure the voltage.<br>Hence not feasible. |
| POWER: | | |
| FREQUENCY | Pulses with modulating frequency (1 Hz - 10<br>Hz) | Pulses with modulating frequency (1 Hz - 10<br>Hz) |
| WAVEFORM | Rectangular pulse | Rectangular pulse |
| (ENERGY SOURCE)<br>BATTERY OPERATION | Yes | Yes |
| BATTERY TYPE | Zinc Air batteries, P10 | Lithium ion battery, CR1220 |
| BATTERY CAPACITY | 100 mAh | 40 mAh |
| NO. / VOLTAGE | 3*1.4 V | 1*3 V |
| PULSE WIDTH | 0.980mSec | 0.980mSec |
| OPERATING TIME,<br>HOURS | 120 hrs (5 days), 2 Hours ON (Periodically) | 120 hrs (5 days), 2 Hours ON (Periodically) |
| ENVIRONMENTAL | | |
| OPERATING<br>TEMPERATURE | 5°C to 45°C | 5°C to 30°C |
| OPERATING<br>HUMIDITY (NON<br>CONDENSING) | 40% to 80% | 40% to 60% |
| ENVIRONMENT OF<br>USE | Clinics, Hospital and Home environments | Clinics, Hospital and Home environments |
| STERILIZATION | EtO Sterilization | Irradiation (Gamma) |
| RE-USE | Single use Device | Single use Device |
| SHELF LIFE | 6 Months | 12 months |
| PLANNING &<br>PURCHASE | | |
| WARRANTY | NA | NA |
| Battery Type and Voltage (V) | Zinc Air batteries, P10 | Lithium ion battery, CR1220 |
| Voltage Controlled? | Yes | Yes |
| Software Controlled? | Yes | Yes |
| Weight (grams) | 6 g | 4 g |
| Dimensions (mm) | 50*23*7 mm | 36*16*7 mm |
| Electrode Needle Arrays | | |
| Number of leads | 4 | 4 |
| Lead type | 1 Single needle in all the 4 leads | One stainless steel wire with four<br>stimulation needles and two stainless<br>steel wires with one stimulation needle<br>each. There is one ground wire with<br>one needle. |
| Dimensions of each needle (width<br>and length in mm) | 0.4 mm width x 2 mm length | 0.5 mm width x 2 mm length |
| Surface Area of needle (cm2) | 0.0201 cm² | 0.0276 cm² |
| Needle Material | Titanium | Titanium |
| System Characteristics (Output Specs) | | |
| Max Charge Density<br>(microcoulomb/cm2) Per needle | 65.67 @ 1K ohm<br>7.96 @ 10K ohm | 48.91 @ 1K ohm<br>5.79 @ 10K ohm |
| Max Average Power Density (W/cm2) | 0.346 @ 1K ohm<br>0.0509 @ 10K ohm | 0.2645 @ 1K ohm<br>0.0371 @ 10K ohm |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for DyAnsys. The logo consists of a stylized red symbol to the left of the company name, which is written in black. The symbol appears to be a combination of a greater-than sign and a stylized letter A.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for DyAnsys. The logo consists of a stylized red symbol resembling a greater-than sign combined with a curved line, followed by the company name "DyAnsys" in black sans-serif font. The symbol is positioned to the left of the company name.
# Pulse Generator
These differences do not affect the safety and effectiveness of the device.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for DyAnsys. The logo consists of a stylized red and black symbol to the left of the company name, "DyAnsys," which is written in a simple, sans-serif font. The symbol appears to be an abstract representation of a waveform or signal, with a portion of it colored red. The overall design is clean and modern, conveying a sense of technology and innovation.
#### Performance Testing Summary
The Drug Relief device and its components are subjected to performance testing to validate the effectiveness of each unit. The final product testing is performed to verify and compare the effectual output along with that of the predicate device. The functional test is performed for 120 hours to monitor the continuous performance. The pulse width, pulse duration, amplitude and current values are captured for the Drug Relief device. The Drug Relief has equivalent Performance specifications when compared to the predicate device.
All the hardware components, form factors, material for sterilization and patient contacting materials of the Drug Relief are similar to the 510(K) cleared device ANSiStim-PP (K170391).
#### Compliance with Standards
The ANSiStim complies with the following standards
- 1. IEC 60601-1 2. IEC 60601-1-2 3. ISO 10993-1 4. ISO 10993-5 5. ISO 10993-6 6. ISO 10993-7 7. ISO 10993-10 8. ISO 10993-11 9. ISO 11135
#### Sterilization Testing Summary
The needle package was subjected to Bio-burden test. The Needle packs are exposed to EtO Sterilization to curtail the presence of microorganisms and to achieve the defined sterility assurance level (SAL). During the sterilization process the biological indicators are used to ensure the desired sterility assurance level. These BIs were placed at the appropriate location, where the sterilizing conditions are the most difficult to achieve. These needle packages carry a chemical indicator on the rear side which indicates the exposure to EtO. The sterility test performed on the needles indicates that there is no turbidity. The residual risk report carried out on the sterilized needle packs evidenced that the results are in-line with standards requirement. All Sterilization testing was performed in accordance with ISO 11135:2014-Sterilization of healthcare productsethylene oxide, ISO 11140-1:2005/(R) 2010 Sterilization of healthcare products- chemical indicators, ISO 10993-7:2008/(R) 2012 Biological evaluation of medical devicesethylene oxide sterilization residuals, ISO 11737-1:2006/(R) 2011 Sterilization of medical devices- Microbiological methods- Part 1: Determination of a population of microorganisms on products, ISO 11737-2: 2009 Sterilization of medical devices-Microbiological methods- Part2: Tests of Sterility performed in the definition, validation and maintenance of a Sterilization process and ISO 11138-2:2006/(R) 2010- Sterilization
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for DyAnsys. The logo consists of a stylized red symbol resembling an arrow or a stylized letter 'A' on the left, followed by the word "DyAnsys" in black, sans-serif font. The logo appears to be clean and modern, with a focus on simplicity and readability.
Of Healthcare Products-Biological Indicators- Part2: Biological Indicators for ethylene oxide Sterilization Processes.
## 9. Non-Clinical Testing Summary
The bench test has been performed and found that the Drug Relief met all the requirements specifications and standards requirements. The testings includes the following:
- 1. MEE testing as per IEC 60601-1
- 2. EMI/EMC testing as per IEC 60601-1-2
- 3. Biocompatibility testing as per ISO 10993
- 4. Performance testing
#### 10. Conclusion
Hence it is concluded that by demonstrating the Performance testing and with the Indications For Use, Environment for Use, Biocompatibility and compliance with the same harmonized standards, the product Drug Relief is substantially equivalent to the predicate device NSS-2 BRIDGE.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.