PZR · Percutaneous Nerve Stimulator For Opioid Withdrawal
Neurology · 21 CFR 882.5896 · Class 2
Overview
| Product Code | PZR |
|---|---|
| Device Name | Percutaneous Nerve Stimulator For Opioid Withdrawal |
| Regulation | 21 CFR 882.5896 |
| Device Class | Class 2 |
| Review Panel | Neurology |
Identification
A percutaneous nerve stimulator for substance use disorders is a device that stimulates nerves percutaneously to aid in the reduction of withdrawal symptoms associated with substance use disorders.
Classification Rationale
Class II (special controls). The special controls for this device are:
Special Controls
In combination with the general controls of the FD&C Act, the percutaneous nerve stimulator for substance use disorders is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible. (2) Electromagnetic compatibility and electrical, mechanical, and thermal safety testing must be performed. (3) Electrical performance testing of the device and electrodes must be conducted to validate the specified electrical output and duration of stimulation of the device. (4) Software verification, validation, and hazard analysis must be performed. (5) Sterility testing of the percutaneous components of the device must be performed. (6) Shelf life testing must be performed to demonstrate continued sterility, package integrity, and device functionality over the specified shelf life. (7) Labeling must include the following: (i) A detailed summary of the device technical parameters; (ii) A warning stating that the device is only for use on clean, intact skin; (iii) Instructions for use, including placement of the device on the patient; and (iv) A shelf life.
Recent Cleared Devices (8 of 8)
| Record | Device Name | Applicant | Decision Date | Decision |
|---|---|---|---|---|
| K251246 | Sparrow Ascent | Spark Biomedical, Inc. | Aug 25, 2025 | SESE |
| K233166 | NET Recovery Corp/NET Device | Net Recovery | May 29, 2024 | SESE |
| K230796 | Sparrow Ascent | Spark Biomedical, Inc. | Jun 20, 2023 | SESE |
| K221231 | Drug Relief v1 | Dyansys, Inc. | Jun 6, 2022 | SESE |
| K211971 | Drug Relief v1 | Dyansys, Inc. | Nov 5, 2021 | SESE |
| K201873 | Sparrow Therapy System | Spark Biomedical, Inc. | Jan 2, 2021 | SESE |
| K173861 | Drug Relief | Dyansys, Inc. | May 2, 2018 | SESE |
| DEN170018 | NSS-2 System | Innovative Health Solutions (Ihs), Inc. | Nov 15, 2017 | DENG |
Top Applicants
- Dyansys, Inc. — 3 clearances
- Spark Biomedical, Inc. — 3 clearances
- Innovative Health Solutions (Ihs), Inc. — 1 clearance
- Net Recovery — 1 clearance