K171306 · Electrocore, LLC · PKR · May 30, 2017 · Neurology
Device Facts
Record ID
K171306
Device Name
gammaCore-S
Applicant
Electrocore, LLC
Product Code
PKR · Neurology
Decision Date
May 30, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.5892
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The gammaCore-S Non-invasive Vagus Nerve Stimulator is intended to provide noninvasive vagus nerve stimulation (nVNS) on the side of the neck. The gammaCore-S device is indicated for the acute treatment of pain associated with episodic cluster headache in adult patients.
Device Story
Non-invasive vagus nerve stimulator (nVNS) applied to side of neck; delivers mild electrical stimulation to vagus nerve. Patient-operated; controls stimulation intensity via two-button interface. Device features LCD screen and LED for status indication. Each stimulation session lasts two minutes. Intended for acute pain relief in episodic cluster headache patients. Device output (signal waveforms) remains identical to predicate. Modifications limited to user interface (buttons/LCD replacing thumbwheel/single LED) and associated firmware updates. No clinical data required for modifications; verification/validation testing confirmed safety and performance.
Clinical Evidence
No clinical data required; substantial equivalence supported by bench testing, verification, and validation activities confirming that modified device meets predetermined acceptance criteria.
Technological Characteristics
Non-invasive electrical stimulator. Patient-contact materials identical to predicate. Two-button interface for power and intensity control; LCD screen and LED for status. Power source identical to predicate. Software updated to support new UI components.
Indications for Use
Indicated for acute treatment of pain associated with episodic cluster headache in adult patients.
Regulatory Classification
Identification
An external vagal nerve stimulator for headache is a prescription device used to apply an electrical current to a patient's vagus nerve through electrodes placed on the skin for the treatment of headache.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The technical parameters of the device, including waveform, output modes, maximum output voltage and current (with 500, 2,000, and 10,000 ohm loads), pulse duration, frequency, net charge (µC) per pulse, maximum phase charge at 500 ohms, maximum current density (mA/cm
2 , r.m.s.), maximum average current (mA), maximum average power density (W/cm2 ), and the type of impedance monitoring system shall be fully characterized through non-clinical performance testing.(2) Software verification, validation, and hazard analysis shall be performed.
(3) Biocompatibility evaluation of the patient-contacting components of the device shall be performed.
(4) The device shall be tested for electrical, thermal, and mechanical safety, and for electromagnetic compatibility (EMC).
(5) The labeling must include:
(i) Instructions for proper use of the device, including placement of the device on the patient; and
(ii) Instructions on care and cleaning of the device.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 30, 2017
electroCore, LLC Mike Romaniw VP, Quality Assurance & Regulatory Affairs 150 Allen Road. Suite 201 Basking Ridge, New Jersey 07920
Re: K171306
Trade/Device Name: gammaCore-S Regulation Number: 21 CFR 882.8592 Regulation Name: External Vagal Nerve Stimulator For Headache Regulatory Class: Class II Product Code: PKR Dated: May 2, 2017 Received: May 3, 2017
Dear Mike Romaniw:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
William J. Heetderks -S 2017.05.30 16:12:38 -04'00'
- for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
Device Name gammaCore-S
Indications for Use (Describe)
The gammaCore-S Non-invasive Vagus Nerve Stimulator is intended to provide noninvasive vagus nerve stimulation (nVNS) on the side of the neck. The gammaCore-S device is indicated for the acute treatment of pain associated with episodic cluster headache in adult patients.
| Type of Use ( <i>Select one or both, as applicable</i> ) | |
|----------------------------------------------------------|--|
|----------------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 6: 510(k) Summary
The following information is provided as required by 21 CFR § 807.87 for the gammaCore-S 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the information upon which the substantial equivalence determination is based.
| Applicant: | electroCore® LLC |
|------------------------------------|---------------------------------|
| | 150 Allen Road, Suite 201 |
| | Basking Ridge, New Jersey 07920 |
| | Ph: 973-290-0097 |
| | Fax: 973-290-9171 |
| Establishment Registration Number: | 3000060063 |
Establishment Registration Number: 3009060963
| Contact: | Mike Romaniw |
|--------------------------------------------|--------------|
| VP, Quality Assurance & Regulatory Affairs | |
| Office: 973-355-6702 | |
| Fax: 973-290-9171 | |
| mike.romaniw@electrocorellc.com | |
- Alternate Contact: Marie Marlow Chief Executive Officer M Squared Associates, Inc. Office: 855-776-0638 x201 Fax: 703-562-9797 MMarlow@msquaredassociates.com Date submitted: 02 May 2017 gammaCore-S® Proprietary Name: Common Name: External vagal nerve stimulator for headache Classification Status: Class II
- Product Codes: PKR
{4}------------------------------------------------
#### Predicate Device: gammaCore-DEN150048
Device Description: gammaCore-S is a device that provides non-invasive Vagus Nerve Stimulation (nVNS) when applied to the side of the neck. This is a mild electrical stimulation of the vagus nerve, which runs through the neck and carries information to the central nervous system. Each stimulation with gammaCore-S lasts two minutes. The patient controls the stimulation strength.
Indication for Use: The gammaCore-S Non-invasive Vagus Nerve Stimulator is intended to provide noninvasive vagus nerve stimulation (nVNS) on the side of the neck. The gammaCore -S device is indicated for the acute treatment of pain associated with episodic cluster headache in adult patients.
Summary of Technological Characteristics: The gammaCore-S modifications include a change from a thumbwheel control to a two button control for power on/off and control of the stimulation intensity as well as the addition of a small LCD screen which indicates the device status. Additionally, minor software updates were required as a result of these changes.
Summary of Nonclinical Testing: The verification and validation activities, as identified by the risk analysis to ensure that the modified device is as safe and effective as the predicate device, have been completed and demonstrate that the predetermined acceptance criteria have been met.
Additional risks are the same as those submitted in the original submission and have been mitigated in by the same methods. Declarations of Conformity with design controls are provided.
#### Substantial Equivalence Discussion:
#### Similarities
The GammaCore-S device technology is identical to the device technology used in the gammaCore. The similarities include:
- Intended use and indication for use: .
- Signal Outputs and waveforms .
- Materials used for patient contact surfaces .
- Power Source .
{5}------------------------------------------------
#### Differences
The differences between the gammaCore-S and the gammaCore can be summarized as "a change in the user interface". The thumb-wheel potentiometer used to turn the gammaCore device on and off, as well as to increase or decrease the treatment signal amplitude, and the single LED that served as a status indicator to the user have been replaced by a pair of buttons, an LCD panel, and an LED. The buttons serve to turn the gammaCore-S on and off, as well as to increase / decrease the treatment signal amplitude. The LCD panel and LED provide device status information to the user.
Summary: The gammaCore-S has the same intended use as the predicate gammaCore device. The gammaCore-S does not alter the fundamental scientific technology of the device because it does not change the operating principle or device output. The modification to the user interface does not impact the device for its intended use in the acute treatment of pain associated with episodic cluster headaches in adult patients.
Clinical Data: Clinical studies were not required to validate the modifications in the gammaCore-S.
Conclusion: The gammaCore-S described in this submission based on the information provided is substantially equivalent to the predicate.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.