PKR · Non-Invasive Vagus Nerve Stimulator - Headache

Neurology · 21 CFR 882.5892 · Class 2

Overview

Product CodePKR
Device NameNon-Invasive Vagus Nerve Stimulator - Headache
Regulation21 CFR 882.5892
Device ClassClass 2
Review PanelNeurology

Identification

An external vagal nerve stimulator for headache is a prescription device used to apply an electrical current to a patient's vagus nerve through electrodes placed on the skin for the treatment of headache.

Classification Rationale

Class II (special controls). The special controls for this device are:

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) The technical parameters of the device, including waveform, output modes, maximum output voltage and current (with 500, 2,000, and 10,000 ohm loads), pulse duration, frequency, net charge (µC) per pulse, maximum phase charge at 500 ohms, maximum current density (mA/cm 2 , r.m.s.), maximum average current (mA), maximum average power density (W/cm2 ), and the type of impedance monitoring system shall be fully characterized through non-clinical performance testing.(2) Software verification, validation, and hazard analysis shall be performed. (3) Biocompatibility evaluation of the patient-contacting components of the device shall be performed. (4) The device shall be tested for electrical, thermal, and mechanical safety, and for electromagnetic compatibility (EMC). (5) The labeling must include: (i) Instructions for proper use of the device, including placement of the device on the patient; and (ii) Instructions on care and cleaning of the device.

Recent Cleared Devices (9 of 9)

RecordDevice NameApplicantDecision DateDecision
K211856gammaCore SapphireElectrocore, Inc.Sep 10, 2021SESE
K203546gammaCore SapphireElectrocore, Inc.Feb 12, 2021SESE
K191830gammaCore SapphireElectrocore, Inc.Mar 26, 2020SESE
K182369gammaCore SapphireElectrocore, Inc.Nov 27, 2018SESE
K180538gammaCore SapphireElectrocore, LLCMar 30, 2018SESE
K173442gammaCore-SElectrocore, LLCJan 23, 2018SESE
K172270gammaCore-2Eletrocore, LLCDec 7, 2017SESE
K171306gammaCore-SElectrocore, LLCMay 30, 2017SESE
DEN150048gammaCore Non-invasive Vagus Nerve StimulatorElectrocore, LLCApr 14, 2017DENG

Top Applicants

Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...