K143531 · Magstim Company Limited · OBP · May 8, 2015 · Neurology
Device Facts
Record ID
K143531
Device Name
Rapid Therapy System
Applicant
Magstim Company Limited
Product Code
OBP · Neurology
Decision Date
May 8, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5805
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Rapid2 Therapy System is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode.
Device Story
Rapid2 Therapy System is a computerized, electromechanical rTMS device; delivers non-invasive, pulsed magnetic fields to induce electrical currents in cerebral cortex. System components: user interface (LCD touchscreen), mainframe, power supply, air-film figure-8 coil, coil stand, footswitch, cables. Used in clinics/hospitals by physicians; operator controls stimulation via interface, charging/discharging high-voltage capacitor. Coil positioned over patient's frontal cortex via stand. Treatment involves repetitive pulse trains at predetermined frequencies to stimulate cortical neurons. Output parameters (intensity, frequency, train duration) are set by clinician based on patient motor threshold. Device provides non-pharmacological intervention for depression; clinical benefit derived from cortical stimulation.
Clinical Evidence
Bench testing only. Evidence includes electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and biocompatibility (ISO 10993-1). Performance testing included magnetic field plots, acoustic output measurements, electric field spatial distribution modeling (finite element method on homogeneous sphere), and cooling system effectiveness testing. Software verification and validation confirmed performance per specifications.
Indicated for treatment of Major Depressive Disorder in adults who failed to achieve satisfactory improvement from prior antidepressant medication in the current episode.
Regulatory Classification
Identification
A repetitive transcranial magnetic stimulation system is an external device that delivers transcranial repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex to treat the symptoms of major depressive disorder without inducing seizure in patients who have failed at least one antidepressant medication and are currently not on any antidepressant therapy.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's “Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation System.” See § 882.1(e) for the availability of this guidance document.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol consists of three overlapping profiles facing right, with a design element resembling a wing or ribbon extending from the first profile.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 8, 2015
Magstim Company Limited C/O Janice Hogan Partner, Hogan Lovells US LLP 1835 Market Street, 29th Floor Philadelphia, PA 19103
Re: K143531
Trade/Device Name: Rapid2 Therapy System Regulation Number: 21 CFR 882.5805 Regulation Name: Repetitive transcranial magnetic stimulation system Regulatory Class: Class II Product Code: OPB Dated: April 10, 2015 Received: April 10, 2015
Dear Ms. Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Felipe Aquel - S
Carlos L. Peña, Ph.D., M.S. for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510(k) Number (if known)
Device Name
Rapid2 Therapy System
Indications for Use (Describe)
The Rapid2 Therapy System is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode.
Type of Use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
#### This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{4}------------------------------------------------
## 510(k) SUMMARY Magstim's Rapid2 Therapy System
## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Maqstim Company Limited Spring Gardens, Whitland, Carmathenshire SA34 OHR, United Kingdom
Phone: +44 (0) 1994 240798 Facsimile: +44 (0) 1994 240061
Contact Person: Charles Hounsell
Date Prepared: December 11, 2014
### Name of Device
Magstim Rapid- Therapy System
### Common or Usual Name/
Repetitive Transcranial Magnetic Stimulation (rTMS) System
### Classification
Repetitive Transcranial Magnetic Stimulation (rTMS) System
21 C.F.R. § 882.5805, Class II, product code OBP
### Predicate Devices
NeuroStar TMS Therapy System, Neuronetics, Inc. (K133408)
### Device Description
The Rapid Therapy System is a computerized, electromechanical medical device that produces and delivers non-invasive, magnetic stimulation using brief duration rapidly alternating, or pulsed, magnetic fields to induce electrical currents directed at spatially discrete regions of the cerebral cortex. This method of cortical stimulation by application of brief magnetic pulses to the head is known as Transcranial Magnetic Stimulation.
The Rapid" Therapy System is a non-invasive tool for the stimulation of cortical neurons for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from antidepressant medication in the current episode. The Rapid-Therapy System is used for patient treatment by prescription only under the supervision of a licensed physician. It can be used in both inpatient settings, including physicians' offices, clinics, and hospitals.
{5}------------------------------------------------
The Rapid Therapy System is an integrated system consisting of a combination of hardware, software, and accessories. Its technological characteristics are described in further detail below.
## Intended Use / Indications for Use
The Rapid Therapy System is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from prior antidepressant medication in the current episode.
## Technological Characteristics
The Rapid2 Therapy System is comprised of five principal components. These include:
- 1) User Interface;
- 2) Mainframe;
- 3) Power supply;
- 4) Air Film Coil;
- 5) Coil Stand
- 6) Accessory footswitch and
- 7) Accessory cables
The operator controls the Rapid2 Therapy System via the User Interface, using a graphic LCD panel with touchscreen technology. The operator instructions, given through the User Interface, direct the Rapid Stimulator mainframe in charging and discharging the device's high voltage discharge capacitor. The discharge is delivered to the treatment area via the Air Film Coil, which is positioned above the treatment area. Positioning, and fixation, of the coil over the treatment area is accomplished using the Coil Stand. The Rapid power supply provides power to charge the high voltage capacitor in the Rapid2 Stimulator.
Software documentation for a "moderate" level of concern has been provided.
### Non-Clinical Testing
Electrical safety and electromagnetic compatibility ("EMC") testing was conducted to demonstrate that the device is compliant with IEC 60601-1 (2ªª & 3ª ed.) and IEC 60601-1-2 Environmental testing also demonstrated compliance with IEC 60601-1. The (2007). biocompatibility evaluation demonstrated that the coil meets ISO 10993-1 (2009) standards. In addition, magnetic field plots, and acoustic output measurements were also conducted to demonstrate safety and performance.
Software verification and validation testing further demonstrated that the software performs as intended and in accordance with specifications. The potential risks of Rapid2 have been identified and evaluated in compliance with ISO14971, and the risks were determined to be acceptable, or have been addressed with risk control measures.
The non-clinical testing with the Rapid2 Therapy System included testing of the magnetic field characteristics of the system, as required by FDA's quidance document "Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation (rTMS) Systems", as follows:
{6}------------------------------------------------
#### Output Waveform .
The output waveform produced by the biphasic figure 8 coil was measured using a calibrated search coil connected to an oscilloscope. This way the waveform could be quantified in three directions as well as in time.
#### Electric Field Spatial Distribution .
The electric field distribution was modeled using a finite element method on a homogeneous sphere with a radius of 8.5cm. This allows a quantification of the depth (d1/2) and spread (S1/2) of the field.
#### Magnetic Field Strength Gradient .
The magnetic field strength was measured in a relevant volume above the stimulating coils using a calibrated search coil connected to an oscilloscope. This way the magnetic field strength could be quantified in three directions as well as its magnitude.
#### . Cooling System Test
The temperature on the patient surface was measured over successive treatment sessions in order to evaluate the effectiveness of the cooling system at a higher than normal ambient temperature to ensure safe operation.
The performance tests demonstrate that the Rapid2 Therapy System may be safely and effectively used for treatment of Major Depressive Disorder.
## Substantial Equivalence
The Rapid Therapy System has the same intended use and indications for use as the predicate device, as well as similar technological characteristics and principles of operation. The minor technological differences between the Rapid2 Therapy System and its predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that the Rapid- Therapy System is as safe and effective as the predicate.
The design of the Rapid- Therapy System is similar to the design of the NeuroStar TMS Therapy System, as both systems are based on applying transcranial magnetic stimulation by means of repetitive pulse trains at a predetermined frequency. Both systems use the same mechanism of action, i.e., an electromechanical instrument that produces and delivers brief duration, rapidly alternating (pulsed) magnetic fields to induce electrical currents in localized regions of the prefrontal cortex.
Transcranial magnetic stimulation is enabled in the Rapid2 Therapy System and in the NeuroStar TMS Therapy System, as both devices have the same system components, consisting of a mobile cart or console, an electromagnetic coil, a positioning arm, a TMS stimulator, and software. The basic operational procedure is the same in both the Rapid Therapy System and in the NeuroStar TMS Therapy System consisting of system setup, patient preparation, determination of patient's motor threshold, coil position, and administration of treatment at predefined treatment stimulation parameters. The similarities and minor differences between the Rapid² Therapy System and the NeuroStar TMS Therapy System are described in the table below:
{7}------------------------------------------------
# Table 1: Substantial Equivalence Summary
| Criteria | Rapid² Therapy System | NeuroStar TMS Therapy System |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Magnetic Field<br>Intensity | 120% of the MT | 120% of the MT |
| Frequency | 10 Hz | 10 Hz |
| Train duration | 4 sec | 4 sec |
| Inter-train interval | 26 sec | 26 sec |
| Number of trains | 75 | 75 |
| Magnetic Pulses per<br>Session | 3000 | 3000 |
| Treatment Session<br>Duration | 37.5 min | 37.5 min |
| Sessions/wk | 5 | 5 |
| Treatment Schedule | 5 daily sessions for 6 weeks | 5 daily sessions for 6 weeks |
| Area of brain to be<br>stimulated | Frontal Cortex | Frontal Cortex |
| Applicator<br>Configuration<br>Core Material | Biphasic Figure 8 Coil<br>Air | Biphasic Figure 8 Coil<br>Ferromagnetic |
| Coil Parameters | Area = 33000mm²<br>Flat spiral winding<br>WC = 1.75 mm × 6 mm<br>N = 3x19 turns/wing x 2 wings | Area = 13000mm²<br>Stranded rectangular windings<br>Wire wrapped around cross-section of the core<br>N = 5 turns/wing (stranded) |
| Machine Output<br>Parameters: Amplitude<br>in Standard Motor<br>Threshold (SMT) units<br>Pulse width (usec)<br>Frequency range (Hz)<br>Pulse train duration<br>range (sec)<br>Inter-train interval<br>range (sec) | 0.28 - 1.9<br>300<br>0.1 - 30<br>1 - 20<br>10 - 60 | 0.22 - 1.6<br>185<br>0.1 - 30<br>1 - 20<br>10 - 60 |
| Criteria | Rapid2 Therapy System | NeuroStar TMS Therapy System |
| Maximum trains per<br>session | ~ 140 | ~ 140 |
| Maximum # of pulses<br>per session<br>(cumulative exposure) | 5000 | 5000 |
{8}------------------------------------------------
The basic software capabilities related to treatment administration are the same in the Rapid2 Therapy System and in the NeuroStar TMS Therapy System.
Both the Rapid Therapy System and in the NeuroStar TMS Therapy System meet the same electrical and mechanical safety standards (IEC 60601-1) and the same EMC standards (IEC 60601-1-2).
## Conclusions
In summary, the intended use and indications for use for the Rapid2 Therapy System and in the NeuroStar TMS Therapy System are identical. Furthermore, the key technological characteristics of the Rapid2 Therapy System and in the NeuroStar TMS Therapy System, including basic design, mechanism of action, components, specifications, and treatment procedure, are similar. Performance testing demonstrates that the minor differences do not impact safety or effectiveness.
Thus, the Rapid2 Therapy System is substantially equivalent to the NeuroStar TMS Therapy System predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.