K110068 · Omron Healthcare, Inc. · NUH · Dec 8, 2011 · Neurology
Device Facts
Record ID
K110068
Device Name
MODEL PM3030
Applicant
Omron Healthcare, Inc.
Product Code
NUH · Neurology
Decision Date
Dec 8, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This device is intended for the relief of pain associated with sore or aching muscles of the lower back, arms, or legs due to strain from exercise or normal household and work activities.
Device Story
PM3030 is a battery-operated, single-channel TENS device for OTC pain relief. It delivers biphasic, rectangular electrical pulses to the skin via electrode pads. The device features three output modes for application to the shoulder/arm, lower back, and leg/foot. Operated by the patient in home, clinic, or hospital settings, the device uses a microprocessor to control stimulation parameters. The user adjusts output levels via a control interface. By providing electrical stimulation, the device aims to alleviate muscle pain. Safety features include automatic overload, no-load, and shut-off functions. The device is powered by two AAA batteries.
Clinical Evidence
No clinical data. Substantial equivalence is based on bench testing, including electrical output characterization and compliance with IEC 60601-1 and IEC 60601-1-2 safety standards. Patient-contacting materials were tested per ISO 10993-1.
Technological Characteristics
Battery-powered (2x AAA) TENS device; single-channel; biphasic rectangular waveform; frequency < 100 Hz. Housing material: ABS. Microprocessor-controlled with automatic overload, no-load, and shut-off safety features. Electrode compliance with 21 CFR 898. Patient-contacting materials tested per ISO 10993-1. Standards: IEC 60601-1, IEC 60601-1-2.
Indications for Use
Indicated for adult patients for temporary relief of pain in lower back, arms, or legs caused by exercise or household/work strain. Contraindicated for patients with cardiac pacemakers, implanted defibrillators, or other implanted metallic/electronic devices; do not use with life-supporting medical electronic devices (e.g., artificial heart/lung) or body-worn medical devices (e.g., ECG).
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Omron OTC TENS
#### Premarket Notification 510(k) Section 5 - 510(k) Summary
# 510(k) Summary Page 1 of 6 7-Jan-11
| Omron Healthcare, Inc.<br>1200 Lakeside Dr.<br>Bannockburn, IL 60015 | Tel - 847-247-5626 |
|----------------------------------------------------------------------|------------------------------------------------------------------------------|
| | Fax - 847-680-6911 |
| Official Contact: | Mirna DiPano - Director of Quality & Regulatory |
| Proprietary or Trade Name: | PM3030 |
| Common/Usual Name: | Stimulator, nerve, transcutaneous, over-the-counter |
| Classification Name/Code: | NUH - stimulator, nerve, transcutaneous, over-the-counter<br>21 CFR 882.5890 |
| Device: | PM3030 |
| Predicate Devices: | Endurance Therapeutics, Model T1040, 510(k) K060846 |
#### Device Description:
The PM3030 is a small, battery operated, three (3) output mode TENS device for pain relief intended for OTC use.
The output modes are intended for application to the following areas: Shoulder/Arm, Lower Back and Ley/Foot. The specifications of each mode will be discussed in greater detail later in this section.
The accessories include an electrode cord / cable and electrodes pads (Long Life) which are placed on the specific body part.
As above the device is battery powered there is no connection to AC mains supply.
The device has been tested to and meets the requirements of the following recognized consensus standards:
- IEC 60601-1 Medical Electrical Equipment Part 1: General Requirements for . Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995
{1}------------------------------------------------
#### Omron OTC TENS
#### 510(k) Summary Page 2 of 6 7-Jan-11
- IEC 60601-1-2, Medical Electrical Equipment Part 1-2: General Requirements for Safety - Collateral standard: Electromagnetic:Compatibility - Requirements and Tests (Edition 2:2001 with Amendment 1:2004; Edition 2.1 (Edition 2:2001 consolidated with Amendment 1:2004)).
#### Indications for Use:
This device is intended for the relief of pain associated with sore or aching muscles of the lower back, arms, or legs due to strain from exercise or normal household and work activities.
Environment of Use: Clinics, hospital and home environments
#### Contraindications:
- Do not use this device if you have a cardiac pacemaker, implanted defibrillator, or other implanted metallic or electronic device. Such use could cause electric shock, burns, electrical interference, or death.
- · Do not use this device together with a life-supporting medical electronic device such as an artificial heart or lung.
- · Do not use this device together with a body-worn medical electronic device such as an ECG.
{2}------------------------------------------------
Premarket Notification 510(k) Section 5 – 510(k) Summary
# Omron OTC TENS
# 510(k) Summary Page 3 of 6 7-Jan-11
# Summary of substantial equivalence
| Attribute | Model PM3030 | Predicate |
|------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | Endurance Model T1040 |
| | | K060846 |
| Product Code | NUH | NUH, NGX, GZX |
| CFR | 882.5890 | 882.5890, 890.5850 |
| Indications for Use | This device is intended<br>for the relief of pain<br>associated with sore or<br>aching muscles of the<br>lower back, arms, or legs<br>due to strain from<br>exercise or normal<br>household and work<br>activities. | To be used for temporary relief of<br>pain associated with sore and aching<br>muscles in the<br>lower back due to strain from<br>exercise or normal household work<br>activities choose<br>Manual modes 1, 2, 3, 4, 5, 6 or<br>Auto 4<br>To be used for temporary relief of<br>pain associated with sore and aching<br>muscles in the<br>upper extremities (arm) due to strain<br>from exercise or normal household<br>work activities<br>choose Manual modes 1, 2, 3, 4, 5,<br>6 or Auto 1 or Auto 3<br>To be used for temporary relief of<br>pain associated with sore and aching<br>muscles in the<br>lower extremities (leg) due to strain<br>from exercise or normal household<br>work activities<br>choose Manual modes 1, 2, 6 or<br>Autol or Auto 4<br>Used to stimulate healthy muscles<br>in order to improve and facilitate<br>muscle<br>performance choose Manual Mode<br>1 or Auto Mode 2<br>Prescription |
| Patient Population | Adult | Not specified |
| Prescriptive or OTC | OTC | OTC |
| Environment of use | Clinics, hospital and home environments | Not specified |
| Contraindications - do not use if you have a cardiac pacemaker, implanted defibrillator or other implanted metallic or electronic device | Yes | Yes |
| Warnings standard | Yes | Yes |
| Precautions - standard | Yes | Yes |
| Adverse reactions - standard | Yes | Yes |
| Power source | 2 - AAA | 3 - AAA |
| Number of Output modes | 3 | 10 |
| Number of output channels | 1 | 1 |
| Waveform | Biphasic | |
| Shape | Rectangular | |
| Maximum Output Voltage (max) | | |
| 500 ohm | 35.4 V | 40.7 |
| 2k ohm | 46.7 V | 105.1 |
| 10k ohm | 50.8 V | 154.1 |
| Maximum Output Current (max) | | |
| 500 ohm | 4.4 mA | 81.4 |
| 2k ohm | 1.7 mA | 47.8 |
| 10k ohm | 0.4 mA | 15.4 |
| Maximum Phase charge (500 ohm) | 133 microC | 16.9 microC |
| Maximum Current Density (500 ohm) | 0.095 mA/cm² | 2.71 mA/cm² |
| Maximum Average Current (500 ohm) | 3.5 mA | Not specified |
| Maximum Average Power Density (500 ohm) | 89 mW/cm² | 10.2 mW/cm² |
| Frequency (Hz) | < 100 Hz | 245 Hz |
| Burst Mode | None | |
| Timer range (min) | 15 minutes | 30 minutes |
| Indication display | | |
| On/Off status | Yes | Yes |
| Low battery | No | Yes |
| Voltage / Current level | Yes | Yes |
| Output mode | Yes | Yes |
| Time to cut-off | No | Yes |
| Dimensions | 55 mm x 95 mm x 19 mm | 150 mm x 68 mm x 26 mm |
| Weight | 60 grams | 90 grams |
.
{3}------------------------------------------------
# Premarket Notification 510(k) Section 5 – 510(k) Summary
.
# Omron OTC TENS
... : : : : .. : ..
# 510(k) Summary
:: ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ ....
... ... ... ... ... ... ... ... ... ... ... .............................................................................................................................
{4}------------------------------------------------
#### Omron OTC TENS
#### 510(k) Summary Page 5 of 6 7-Jan-11
| Housing material | ABS | Not specified |
|--------------------------------------|---------------------|---------------|
| Microprocessor control | Yes | Yes |
| Automatic Overload trip | Yes | Not specified |
| Automatic no-load trip | Yes | Not specified |
| Automatic shut-off | Yes | Not specified |
| User override control | Power On/Off button | Not specified |
| Electrode compliance with 21 CFR 898 | Yes | Not specified |
| Electrode cable | Yes | Yes |
# Differences Between Other Legally Marketed Predicate Devices
The Omron PM3030 OTC TENSis viewed as substantially equivalent to the predicate device because: The electrical stimulation provided by the PM3030 is substantially equivalent to that commonly employed by TENS devices that have been cleared for marketing without prescription labeling; i.e., for OTC sale. The pulses in the waveform combinations are restricted in amplitude and duration to values consistent with other cleared devices.
Technological characteristics, features, specifications, materials and intended uses of the PM3030 are substantially equivalent to the quoted predicate devices.
The differences that exist between the devices are insignificant in the terms of safety or effectiveness.
#### Indications -
The PM3030 and the predicate (Endurance Model T1040 K060846) are for the temporary relief of pain associated with sore and aching muscles in the upper and lower extremities and lower back due to strain from exercise or normal household and work activities and we view them as equivalent. Performance testing has been performed that shows equivalent waveform characteristics.
# Technology -
- Identical both the PM3030 and the predicate provide electrically generated pulses . applied to the skin via electrodes.
# Operating specifications -
- Equivalent. Section 12 provides detailed descriptions of the characteristics of the . PM3030 and the predicate. They are equivalent.
{5}------------------------------------------------
Premarket Notification 510(k) Section 5 – 510(k) Summary
### Omron OTC TENS
# 510(k) Summary. Page 6 of 6 7-Jan-11
# Materials –
- The patient contacting materials of the electrode has been tested in accordance to . ISO 10993-1 and FDA Guidance.
Environment of Use -
- Both are OTC devices so assume same environment .
# Patient Population -
- . PM3030 is restricted to adults
{6}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird figure.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
DEC = 8 2011
Omron Healthcare, Inc. c/o Mr. Paul E. Dryden President, Regulatory Consultant ProMedic, Inc. 24301 Woodsage Drive Bonita Springs, FL 34134-2958
Re: K110068
Trade/Device Name: Model PM3030 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator for Pain Relief Regulatory Class: Class II Product Code: NUH Dated: October 25, 2011 Received: October 26, 2011
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 - Mr. Paul E. Dryden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
on
http://www.fda.gov/MedicalDevices/ResourceforYou/Industry/default.htm.
Sincerely yours,
Kesia Alexander
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use Statement
Page 1 of 1
510(k) Number:
(To be assigned)
Device Name:
Model PM3030
Indications for Use:
This device is intended for the relief of pain associated with sore or aching muscles of the lower back, arms, or legs due to strain from exercise or normal household and work activities.
Prescription Use (Part 21 CFR 801 Subpart D) or
Over-the-counter use XX (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) 1
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mussla L. Burke Nicholas
(Division Sign-Off)
Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K110068
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.