VASCULAR OCCLUSION SYSTEM (TRUFILL PUSHABLE COILS), AND TRUFILL DCS ORBIT DETACHABLE COIL SYSEM
K123560 · Codman & Shurtleff, Inc. · HCG · Jan 25, 2013 · Neurology
Device Facts
Record ID
K123560
Device Name
VASCULAR OCCLUSION SYSTEM (TRUFILL PUSHABLE COILS), AND TRUFILL DCS ORBIT DETACHABLE COIL SYSEM
Applicant
Codman & Shurtleff, Inc.
Product Code
HCG · Neurology
Decision Date
Jan 25, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.5950
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Pushable Coils may be used to reduce or block the rate of blood flow in vessels of the peripheral and neurovasculature. They are intended for use in the interventional radiologic management of arteriovenous malformations, arteriovenous fistulas, and other vascular lesions of the brain, spinal cord and spine. The TRUFILL DCS ORBIT Detachable Coil System is indicated for embolizing intracranial aneurysms and other vascular malformations such as arteriovenous malformations and arteriovenous fistulae of the neurovasculature. The TRUFILL DCS ORBIT Detachable Coil is also intended for arterial and venous embolization in the peripheral vasculature.
Device Story
Device consists of TRUFILL Pushable Coils and TRUFILL DCS ORBIT Detachable Coil System; used for endovascular embolization of aneurysms, arteriovenous malformations, and fistulas. Pushable coils: platinum/tungsten and synthetic fibers; advanced via microcatheter; deployed using TRUPUSH pusher. Detachable coils: platinum alloy wire; secondary helical/complex shape; detached via hydraulic release mechanism using TRUFILL DCS Syringe II. Used under fluoroscopy by interventional radiologists/physicians. Mechanism: coils pack aneurysm/vessel lumen; disrupt blood flow; induce thrombosis; isolate weakened vessel walls from arterial pressure. Benefits: exclusion of blood flow from target lesions; involution of aberrant vessels.
Clinical Evidence
No clinical data provided. Substantial equivalence established via bench testing of packaging modifications.
Technological Characteristics
Materials: platinum/tungsten, synthetic fibers, platinum alloy wire. Form factor: straight/shaped pushable coils; helical/complex detachable coils. Delivery: microcatheter-based, fluoroscopy-guided. Detachment: hydraulic release mechanism. Sterilization: not specified (standard industry methods implied).
Indications for Use
Indicated for patients requiring embolization of intracranial aneurysms, arteriovenous malformations, arteriovenous fistulae, and other vascular lesions of the brain, spinal cord, and spine, as well as arterial and venous embolization in the peripheral vasculature.
Regulatory Classification
Identification
A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).
TRUFILL DCS ORBIT Detachable Coil System (K030963)
TRUFILL DCS ORBIT Detachable Coil System (K032553)
TRUFILL DCS ORBIT Detachable Coil System (K053197)
TRUFILL DCS ORBIT Detachable Coil System (K080967)
Submission Summary (Full Text)
{0}------------------------------------------------
K123560
# 510(k) SUMMARY
:
・
.
JAN 2 5 2013
# A. Company Information
.
·
| Company Name: | Codman & Shurtleff, Inc. |
|---------------------|--------------------------|
| Address: | 325 Paramount Drive |
| | Raynham, MA 02767 |
| Telephone: | 408-433-1400 |
| Fax: | 408-433-1585 |
| Contact Person: | Richard Kimura |
| Date of Submission: | November 16, 2012 |
B. Trade/Device Name: TRUFILL<sup>®</sup> Pushable Coils and TRUFILL<sup>®</sup> DCS ORBIT<sup>TM</sup> Detachable Coil System
| Common Name: | Artificial Embolization Device |
|----------------------|-----------------------------------|
| Classification Name: | Neurovascular Embolization Device |
| Regulation Number: | 21 CFR 882.5950 |
| Product Code: | HCG |
# C. Predicate Device Information:
| Device | Company | 510(k) Number/<br>Concurrence Date | Product<br>Code | Predicate For: |
|---------------------------------|----------------------------|------------------------------------|-----------------|-----------------------------------------------------------------------|
| Vascular<br>Occlusion<br>System | Codman &<br>Shurtleff, Inc | K964367<br>January 30, 1997 | HCG | Intended Use<br>Design<br>Materials<br>Manufacturing<br>Sterilization |
| Vascular<br>Occlusion<br>System | Codman &<br>Shurtleff, Inc | K972881<br>June 4, 1998 | HCG | Intended Use<br>Design<br>Materials<br>Manufacturing<br>Sterilization |
| Vascular<br>Occlusion<br>System | Codman &<br>Shurtleff, Inc | K983483<br>March 24, 1999 | HCG | Intended Use Design<br>Materials<br>Manufacturing<br>Sterilization |
l
·
.
{1}------------------------------------------------
| Device | Company | 510(k) Number/<br>Concurrence Date | Product<br>Code | Predicate For: |
|---------------------------------------------------|----------------------------|------------------------------------|-----------------|-----------------------------------------------------------------------|
| TRUFILL DCS<br>ORBIT<br>Detachable Coil<br>System | Codman &<br>Shurtleff, Inc | K030963<br>June 20, 2003 | HCG | Intended Use<br>Design<br>Materials<br>Manufacturing<br>Sterilization |
| TRUFILL DCS<br>ORBIT<br>Detachable Coil<br>System | Codman &<br>Shurtleff, Inc | K032553<br>September 23, 2003 | HCG | Intended Use<br>Design<br>Materials<br>Manufacturing<br>Sterilization |
| TRUFILL DCS<br>ORBIT<br>Detachable Coil<br>System | Codman &<br>Shurtleff, Inc | K053197<br>December 15, 2005 | HCG | Intended Use<br>Design<br>Materials<br>Manufacturing<br>Sterilization |
| TRUFILL DCS<br>ORBIT<br>Detachable Coil<br>System | Codman &<br>Shurtleff, Inc | K080967<br>May 2, 2008 | HCG | Intended Use<br>Design<br>Materials<br>Manufacturing<br>Sterilization |
#### D. Device Description:
The intent of endovascular treatment using embolic microcoils is to pack the lumen of the aneurysm with the microcoil mass, thereby eliminating the influx of blood into the aneurysm. As blood flow is disrupted within the aneurysm, intraluminal thrombosis leads to exclusion of blood flow from the parent artery into the aneurysm. Blood flow in the parent artery is then secluded from the aneurysm and the weakened wall of the aneurysm is isolated from arterial pressures. Coil embolization may also be used to treat arteriovenous malformations and fistulas by packing coils into the lumen of the parent vessel feeding into the unwanted arteriovenous structure, resulting in cessation of blood
{2}------------------------------------------------
flow. Stagnated blood within the arteriovenous malformation or fistula then thromboses, leading to involution and absorption of the aberrant vessels.
#### The TRUFILL Vascular Occlusion System:
The Vascular Occlusion System consists of the TRUFILL Pushable Coils and TRUPUSH Coil Pusher. The TRUFILL Pushable Coil is loaded into the proximal end of a compatible microcatheter and advanced through the catheter to the desired vessel location. The coil is deployed using the TRUPUSH Coil Pusher which is sold separately. The TRUFILL Pushable Coils are made from platinum/tungsten and synthetic fibers, and are designed for use under fluoroscopy with the TRUPUSH Coil Pusher and microcatheters having a minimum ,021" I.D. (0.5 mm). The TRUFILL Pushable coils are available in straight and shaped configurations.
### The TRUFILL DCS ORBIT Detachable Coil System:
The TRUFILL DCS ORBIT Detachable Coil System consists of a delivery system (delivery tube and coil introducer) and an embolic coil. The embolic coil is the implantable segment of the device, and is comprised of a vasoocclusion coil wound from a platinum alloy wire into a primary coil and then formed into a secondary helical or complex shape. The embolic coil is detached from the delivery tube via a proprietary hydraulic release mechanism using the TRUFILL DCS Syringe II which is sold separately. The TRUFILL DCS ORBIT Detachable Coil System is designed for use under fluoroscopy with compatible infusion catheters, and is available in Standard and Fill configurations.
#### E. Intended Use:
Pushable Coils may be used to reduce or block the rate of blood flow in vessels of the peripheral and neurovasculature. They are intended for use in the interventional radiologic management of arteriovenous malformations, arteriovenous fistulas, and other vascular lesions of the brain, spinal cord and spine.
The TRUFILL DCS ORBIT Detachable Coil System is indicated for embolizing intracranial aneurysms and other vascular malformations such as arteriovenous malformations and arteriovenous fistulae of the neurovasculature. The TRUFILL DCS
3
{3}------------------------------------------------
ORBIT Detachable Coil is also intended for arterial and venous embolization in the peripheral vasculature.
## F. Summary of technological characteristics of the proposed to the predicate device:
The proposed TRUFILL Pushable Coils and the TRUFILL DCS ORBIT Detachable Coils are substantially equivalent to the currently cleared TRUFILL Pushable Coils and TRUFILL DCS ORBIT Detachable Coils. No new technological characteristics are being introduced with the proposed device.
The proposed TRUFILL Pushable Coils and the TRUFILL DCS ORBIT Detachable Coils have the same intended use, same operating principle, same design, manufacturing and sterilization process. All materials are the same as the currently marketed devices with the exception of the change to the coating material on the Tyvek pouch (provided by a different vendor) and one minor change to one pouch dimension (for the TRUFILL DCS ORBIT Detachable Coil only). The following Table provides an overview of the similarities and differences for both products.
| Table 1: Similarities and Differences compared to Current Product | | |
|-------------------------------------------------------------------|-------------------------------------------------------|-------------------------------------------------------------------------------|
| Characteristic | TRUFILL Pushable Coils | TRUFILL DCS ORBIT<br>Detachable Coils |
| Intended Use | No change | No change |
| Operating Principle | No change | No change |
| Design | No change | No change |
| Materials | No change (except coating<br>material on Tyvek pouch) | No change (except coating<br>material on Tyvek pouch) |
| Dimensions | No change | No change (except<br>packaging pouch) |
| Manufacturing Process &<br>Locations | No change | No change |
| Packaging Components | No change (except coating<br>material on Tyvek pouch) | No change (except coating<br>material on Tyvek pouch,<br>and pouch dimension) |
| Sterilization Process &<br>Location | No change | No change |
| Shelf Life | No change. | No change |
4
{4}------------------------------------------------
## G. Summary of Nonclinical Testing:
Codman performed non-clinical testing necessary to demonstrate substantial equivalence to the predicate devices. Bench testing demonstrated that the TRUFILL Pushable Coils and the TRUFILL DCS ORBIT Detachable Coils perform according to their description and intended use, and established the performance characteristics of the packaging modifications.
Results of verification and validation testing conducted on the TRUFILL Pushable Coils and TRUFILL DCS ORBIT Detachable Coils demonstrated that the proposed device is substantially equivalent to the predicate device and that the packaging modifications do not impact the design, safety, operation, or performance characteristics of the device.
The following tests were conducted to verify the packaging modifications for both the TRUFILL Pushable Coils and TRUFILL DCS Detachable Coils:
- . Package Integrity Testing
- O Dye Leak
.
- Visual Inspection o
- Seal Strength o
- Biocompatibility Testing .
- Cytotoxicity Testing
- Sterilization Verification Testing . o EO/ECH Residuals Testing
- Packaging Shelf Life Verification .
- O Dye Leak
- Visual Inspection 0
- Seal Strength 0
Based upon the design, materials, function, intended use, comparison with currently marketed devices and the non-clinical testing performed by Codman & Shurtleff, Inc., it
5
{5}------------------------------------------------
is concluded that the TRUFILL Pushable Coils and the TRUFILL DCS ORBIT Detachable Coils are substantially equivalent to the currently cleared TRUFILL Pushable Coils and the TRUFILL DCS ORBIT Detachable Coils and, therefore, do not raise any new questions of safety or effectiveness.
## H. Summary of Clinical Testing:
Since there are no changes proposed to the intended use, design, performance characteristics, manufacturing process, sterilization, or principles of operation of the devices, clinical testing was not required to establish substantial equivalence.
{6}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three overlapping wings, representing service to the nation. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular fashion around the eagle.
#### Public Health Service
January 25,2013
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Codman & Shurtleff, Inc. c/o Richard K. Kimura Regulatory Affairs Specialist 325 Paramount Drive Raynham, MA 02767-0350
Re: K123560
Trade/Device Name: TRUFILL Pushable Coils TRUFILL DCS ORBIT Detachable Coil System Regulation Number: 21 CFR 882.5950 Regulation Name: Neurovascular Embolization Device Regulatory Class: Class II Product Code: HCG Dated: December 21, 2012 Received: December 26, 2012
Dear Mr. Kimura:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above, and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device
{7}------------------------------------------------
Page 2 - Richard K. Kimura
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Victor Krauthamer -S
Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
{8}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known): K123560
### Device Name: TRUFILL Pushable Coils TRUFILL DCS ORBIT Detachable Coil System
Indications For Use:
Pushable Coils may be used to reduce or block the rate of blood flow in vessels of the peripheral and neurovasculature. They are intended for use in the interventional radiologic management of arteriovenous malformations, arteriovenous fistulas, and other vascular lesions of the brain, spinal cord and spine.
The TRUFILL DCS ORBIT Detachable Coil System is indicated for embolizing intracranial aneurysms and other vascular malformations such as arteriovenous malformations and arteriovenous fistulae of the neurovasculature. The TRUFILL DCS ORBIT Detachable Coil is also intended for arterial and venous embolization in the peripheral vasculature.
Prescription Use _____________________________________________________________________________________________________________________________________________________________
. AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Joyce M. Whang
(Division Sign Off) Division of Neurological and Physical Medicine Devices (DNPMD)
510(k) Number K123560
> Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.