K012985 · Boston Scientific, Target · HCG · Jan 31, 2002 · Neurology
Device Facts
Record ID
K012985
Device Name
MATRIX DETACHABLE COILS
Applicant
Boston Scientific, Target
Product Code
HCG · Neurology
Decision Date
Jan 31, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5950
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Matrix™ Detachable Coils are intended for embolization of those intracranial aneurysms that because of their morphology, their location, or the patient's general medical condition - are considered by the treating neurosurgical team to be: a) very high risk for management by traditional operative techniques, or, b) inoperable, and for embolization of other neuro vascular abnormalities such as arteriovenous malformations and arteriovenous fistulae. The Matrix™ Detachable Coils are also intended for arterial and venous embolizations in the peripheral vasculature.
Device Story
Matrix™ Detachable Coils are platinum wire coils coated with a biocompatible absorbable polymer; used for embolization of intracranial aneurysms, vascular malformations, and peripheral vasculature. System includes coils, delivery wire, and battery-operated SynerG™ Power Supply. Physician (neurosurgeon/interventionalist) navigates coil to target site; power supply applies constant current to delivery wire; electrolytic dissolution detaches coil from wire. Power supply senses resistance to detect detachment, signals completion, and terminates current. Device provides mechanical occlusion of vascular structures; polymer coating intended to enhance healing/occlusion. Used in clinical/OR settings.
Clinical Evidence
Bench testing only. No clinical data provided. Verification testing included friction, catheter compatibility, tensile/junction strength, coil stiffness, stability in aneurysm, polymer mass loss, deployment/retraction, bond integrity, saline compatibility, detachment time, heating effects (electrolysis/MRI), and electromagnetic compatibility.
Technological Characteristics
Platinum wire coil with biocompatible absorbable polymer coating and UV-curing adhesive. Delivery wire: stainless-steel core with Teflon® jacket. Energy source: battery-operated (two 9V) power supply providing constant current (0.5-2.0 mA). Biocompatibility per ISO 10993-1. Sterilization not specified. Connectivity: wired connection between power supply and delivery wire.
Indications for Use
Indicated for patients with intracranial aneurysms deemed high-risk or inoperable by neurosurgical teams, and for patients with neurovascular abnormalities (arteriovenous malformations, arteriovenous fistulae) or peripheral vascular arterial/venous embolization needs.
Regulatory Classification
Identification
A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).
Predicate Devices
Guglielmi Detachable Coil (GDC™) and GDC™ Power Supply (K001083)
Submission Summary (Full Text)
{1}------------------------------------------------
JAN 3 1 2002
### IX. 510(k) Summary of Safety and Effectiveness
SUBMITTER:
Boston Scientific Target 47900 Bayside Parkway Fremont, CA 94538
CONTACT PERSON:
Seth A. Schulman TEL: 510-440-7753 FAX: 510-440-7752 E-MAIL: schulmas@bsci.com
DATE PREPARED: August 28, 2001
CLASSIFICATION NAME: Artificial Embolization Device (21 CFR Section 882.5950)
- COMMON NAME: Occlusion Coil
PROPRIETARY NAME: Matrix™ Detachable Coils
#### Guglielmi Detachable Coil (GDC™) and GDC™ Power PREDICATE DEVICE: Supply (K001083)
DEVICE DESCRIPTION: The Matrix™ Detachable Coil system consists of the following, each of which is sold seperately:
- · GDCTM SynerGTM Power Supply
- · Matrix™ Detachable Coil attached to a delivery wire
- · Set of GDCTM SynerGTM connecting cables
- Patient return electrode
- · Two 9-volt batteries
The coil is detached by electrolytically dissolving a small portion of the delivery wire upon desired placement of the coil in the anatomy.
Matrix™ Detachable Coils are manufactured from platinum wire, which is first wound into a primary coil and then formed into a secondary helical shape. Coating the platinum coil is a proprietary biocompatible absorbable polymer mixture.
Coils are attached to a delivery wire, which consists of a ground stainless-steel core wire with a stainless-steel coil welded at the distal end and a Teflon® outer jacket. The delivery wire is identical to that employed for the predicate GDCTM Coils cleared under K001083.
Image /page/0/Picture/21 description: The image contains the logo for Boston Scientific Target. The words "Boston Scientific" are stacked on top of each other in a large, bold font. Below that, the word "TARGET" is written in a large, bold font with a line above it. The logo is black and white.
510(k) Premarket Notification -Matrix™ Detachable Coil
Page 39 of 98
116
{2}------------------------------------------------
The GDCTM SynerGTM power supply is a battery-operated, selfcontained unit designed to initiate and control the electrolytic detachment of a GDC™ Coil or Matrix™ Detachable Coil inside an aneurysm.
Kolz985
Each time the power supply is turned on, the unit defaults to the 1.0 mA current setting. Pressing the "Current" switch one time changes the setting to the 2.0 mA current setting; pressing a second time changes the setting to 0.5 mA; pressing a third time returns the unit to the default 1.0 mA setting. Each time the switch is pressed, the current display flashes the new current setting.
The GDC™ SynerG™ Power Supply is designed to apply a constant current through the GDC™ System and to detect when coil detachment has occurred. It maintains a constant current by:
1) sensing the amount of resistance to current flow through the Matrix™ Detachable Coil, and
2) adjusting the voltage needed to maintain the desired current setting. It is also designed to identify subtle changes in the way current is flowing through the Matrix™ Detachable Coil and to recognize those changes, which indicate detachment.
Once those patterns are identified, the GDCTM SynerGTM power supply signals detachment and stops the flow of current through the Matrix™ Detachable Coil.
Matrix™ Detachable Coils are intended for embolizing certain intracranial aneurysms that - because of their morphology, their location, or the patient's general medical condition - are considered by the treating neurosurgical team to be:
a) very high risk for management by traditional operative techniques, or,
b) inoperable,
and for embolizing other vascular malformations such as arteriovenous malformations and arteriovenous fistulae of the neurovasculature.
MatrixTM Detachable Coils are also intended for arterial and venous embolizations in the peripheral vasculature.
All component materials of the Matrix™ Detachable Coils are biocompatible in accordance with ISO 10993-1.
### MATERIALS:
INTENDED USE:
Image /page/1/Picture/14 description: The image shows the logos for Boston Scientific and Target. The Boston Scientific logo is in a serif font, with the words stacked on top of each other. The Target logo is in a sans-serif font, with the word "TARGET" in all caps. The Target logo is underlined.
{3}------------------------------------------------
# TECHNOLOGICAL CHARACTERISTICS COMPARISON:
| Coil Dimensional Attributes | Matrix Detachable Coils |
|------------------------------------------------------|---------------------------|
| Coil Primary Wind OD<br>(Platinum alloy and Polymer) | Same as predicate device. |
| Secondary Coil OD | Same as predicate device. |
| Coil Wire OD | Same as predicate device. |
| Delivery Wire Length | Same as predicate device. |
| Delivery Wire Proximal OD | Same as predicate device. |
| Delivery Wire Distal OD | Same as predicate device. |
# Boston
## Scientific
### TARGET
510(k) Premarket Notification -Matrix™ Detachable Coil
K012985
{4}------------------------------------------------
| Materials | Matrix™ Detachable Coils |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Main Coil | |
| Main Coil Wire | Same as predicate device<br>(The Matrix™ detachable Coils<br>additionally incorporate a biocompatible<br>absorbable polymer and ultraviolet curing<br>adhesive in the Main Coil.) |
| Main Coil / delivery wire junction tubing | Same as predicate device |
| Delivery Wire | |
| Core wire w/coating | Same as predicate device |
| Proximal Coil | Same as predicate device |
| Proximal Marker Coil | Same as predicate device |
| Sheath, Delivery Wire (heat shrink tubing) | Same as predicate device |
| Proximal Tubing | Same as predicate device |
| Bushing | Same as predicate device |
| Inner Coil | Same as predicate device |
Technological Characteristics Comparison (cont.):
Image /page/3/Picture/4 description: The image shows the logo for Boston Scientific Target. The words "Boston Scientific" are stacked on top of each other in a large, bold font. There is a line underneath the words "Boston Scientific". The word "TARGET" is in a large, bold font and is located underneath the line.
510(k) Premarket Notification
{5}------------------------------------------------
| Power Supply | Matching Detachable Coils |
|-----------------------|-------------------------------|
| Power | Same as for predicate device. |
| Batteries | Same as for predicate device. |
| Expected Battery Life | Same as for predicate device. |
| Red Cable | Same as for predicate device. |
| Black Cable | Same as for predicate device. |
| Current Settings | Same as for predicate device. |
| Current | Same as for predicate device. |
| Voltage | Same as for predicate device. |
| Operating Temp. | Same as for predicate device. |
| Storage Temp. | Same as for predicate device. |
| Relative Humidity | Same as for predicate device. |
| Unit Size | Same as for predicate device. |
| Unit Weight | Same as for predicate device. |
Technological Characteristics Comparison (cont.):
Image /page/4/Picture/4 description: The image shows the logo for Boston Scientific Target. The words "Boston Scientific" are stacked on top of each other, with "Boston" on top and "Scientific" below it. There is a line underneath the word "Scientific". Below the line is the word "TARGET" in a larger, bolder font.
{6}------------------------------------------------
### Verification Test Summary Table: Predicate GDC™ Coils vs Matrix™ Detachable Coils
| Test or Point of Comparison | Matrixy Detachable Coils |
|--------------------------------------------------------------|-------------------------------------------------------------|
| Friction | Meets acceptance criteria established for predicate device. |
| Catheter / Coil Compatibility | Meets acceptance criteria established for predicate device. |
| Main Coil Tensile Strength | No change made which would affect this test. |
| Main Coil Junction Strength | Meets acceptance criteria established for predicate device. |
| Coil Stiffness Test | Meets acceptance criteria established for predicate device. |
| Coil Stability in Aneurysm Test | Meets acceptance criteria established for predicate device. |
| Polymer Mass Loss Tests | Meets acceptance criteria established for device. |
| Deployment / Retraction Test | Meets acceptance criteria established for device. |
| Polymer to Platinum Bond Integrity<br>Testing | Meets acceptance criteria established for device. |
| Polymer Integrity Tensile Strength<br>Test | Meets acceptance criteria established for device. |
| Compatibility with Saline Test | Meets acceptance criteria established for device. |
| Detachment Time | No change made which would affect this test. |
| Heating Effect of Electrolysis | No change made which would affect this test. |
| Heating Effect of MRI | No change made which would affect this test. |
| Electrostatic Discharge | No change made which would affect this test. |
| Electromagnetic Compatibility-<br>Radiated Susceptibility | No change made which would affect this test. |
| Electromagnetic Compatibility-<br>Radiated Emissions Class B | No change made which would affect this test. |
| Electromagnetic Compatibility-<br>Magnetic Immunity | No change made which would affect this test. |
| Operating System Test<br>(Assembly Source Code) | Meets acceptance criteria established for predicate device. |
Image /page/5/Picture/4 description: The image shows the logo for Boston Scientific Target. The words "Boston Scientific" are stacked on top of each other in a large, bold font. Below that, the word "TARGET" is written in a larger, bold font and is underlined. The logo is simple and professional.
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three overlapping profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## JAN 3 1 2002
Mr. Seth A. Schulman Senior Regulatory Affairs Specialist Boston Scientific, Target 47900 Bayside Parkway Fremont, California 94538
Re: K012985
Trade/Device Name: Matrix™ Detachable Coil Regulation Number: 882.5950 Regulation Name: Artificial embolization device Regulatory Class: III Product Code: HCG Dated: December 12, 2001 Received: December 14, 2001
Dear Mr. Schulman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{8}------------------------------------------------
Page 2 - Mr. Seth A. Schulman
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis iction with and w you'ls cognifinding of substantial equivalence of your device to a legally premaired predicated device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and If you desire specific art 809.10 for in vitro diagnostic devices), please contact the Office of additionally 21 OF IC For at 659. Additionally, for questions on the promotion and advertising of Compliance at (301) 591-1097 Tice of Compliance at (301) 594-4639. Also, please note the your do rest, titled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). regulation chities, "Thiordians on your responsibilities under the Act may be obtained from the Oiner general mistmational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
Go Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
### IX. 510(k) Summary of Safety and Effectiveness
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Name: Matrix™ Detachable Coil
Indications For Use:
The Matrix™ Detachable Coils have the same indications for use as the predicate GDC" coils.
Matrix™ Detachable Coils are intended for embolization of those intracranial aneurysms that because of their morphology, their location, or the patient's general medical condition - are considered by the treating neurosurgical team to be:
a) very high risk for management by traditional operative techniques, or,
b) inoperable,
and for embolization of other neuro vascular abnormalities such as arteriovenous malformations and arteriovenous fistulae.
The Matrix™ Detachable Coils are also intended for arterial and venous embolizations in the peripheral vasculature.
(Please do not write below this line - continue on another page if needed)
Concurrence of CDRH, Office of Evaluation (ODE)
OR Over-The-Counter Use:______________________________________________________________________________________________________________________________________________________ Prescription Use: (Per 21 CFR §801.109)
Miriam C. Provost
(Division Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number .
# Boston
Scientific
TARGET
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.