K182025 · Tivic Health Systems, Inc. · GZJ · Jan 2, 2019 · Neurology
Device Facts
Record ID
K182025
Device Name
ClearUP Sinus Pain Relief
Applicant
Tivic Health Systems, Inc.
Product Code
GZJ · Neurology
Decision Date
Jan 2, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Tivic Health ClearUP(TM) Sinus Pain Relief device is intended to be used for the temporary relief of sinus pain associated with Allergic Rhinitis.
Device Story
Handheld microcurrent stimulation device for temporary relief of sinus pain associated with allergic rhinitis. Device applies low-level electrical current to facial sinus passages. Features patent-pending algorithm to locate optimal treatment points; system of lights and vibrations guides user. Used in home environment by patients (OTC). User operates device by moving it over facial skin; device provides 7-second treatment period. Output intensity adjustable in 3 steps. Benefits include non-pharmacological pain relief and ease of use. Healthcare providers do not operate; patients self-administer based on device feedback.
Clinical Evidence
Double-blinded randomized controlled clinical study at Stanford Sinus Center (n=81: 27 chronic rhinosinusitis, 49 allergic rhinitis, 5 other). Compared active device to sham. 74% of active users experienced sinus pain reduction; 24% experienced reduction of 3+ points on visual analog scale (0-10). 82% preferred device over current treatments. Statistically significant greater mean pain reduction in active group vs. sham after single treatment. Safety profile favorable; only one minor skin reddening event reported.
Indicated for temporary relief of sinus pain associated with Allergic Rhinitis in adults.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
Rejuvatone MD, TSTONEMD by Trophy Skin, Inc. (K152199)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 2, 2019
Tivic Health Systems Inc. % Nina Peled Consultant Medical Device Regulatory Consultant 550 Davis St. Unit 27 San Francisco, California 94111
Re: K182025
Trade/Device Name: ClearUP™ Sinus Pain Relief Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: GZJ Dated: November 25, 2018 Received: December 3, 2018
Dear Nina Peled:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Pamela D. Scott -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182025
Device Name ClearUP(TM) Sinus Pain Relief
Indications for Use (Describe)
The Tivic Health ClearUP(TM) Sinus Pain Relief device is intended to be used for the temporary relief of sinus pain associated with Allergic Rhinitis.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 1. Submitter Information
- Company: Tivic Health Systems Inc. 750 Menlo Ave, #200 Menlo Park, CA 94025
- Contact Person: Nina Peled Regulatory Consultant 550 Davis St. Unit 27 San Francisco, CA 94111 npeled@yahoo.com 650 454 0322
- 2. Date Prepared: December 26.2018
- ClearUP™ Sinus Pain Relief 3. Device Name:
#### 4. Product Code, Class, Regulation Number and Panel
| Product Code | Class | Regulation Number | Panel |
|--------------------------------------------------------------------|----------|-------------------|-----------|
| GZJ, Transcutaneous electrical<br>nerve stimulator for pain relief | Class II | 21.CFR 882.5890 | Neurology |
- 5. Predicate Devices: -Pointer excel by Lhasa Oms, Inc., K060517 -Rejuvatone MD, TSTONEMD by Trophy Skin, Inc. K152199
- 6. Device Description and Principle of Operation:
Page 1 of 10 Traditional 510(k) Notification Tivic Health ClearUP™ Sinus Pain Relief
{4}------------------------------------------------
Clear UP™ Sinus Pain Relief is a handheld microcurent stimulation device used for the temporary relief of sinus pain associated with allergic rhinitis. The unit applies a low level electrical current to the facial sinus passages.
The ClearUP Sinus Pain Relief unit provides a customized treatment with the following features:
- Microcurrent waveforms to relieve sinus pain associated with allergic rhinitis ●
- Patent-pending algorithm to locate the optimal treatment points ●
- Simple-to-use system of lights and vibrations to guide the user ●
- Small handheld shape for easy holding and comfort ●
### 7. Indications for Use:
The Tivic Health ClearUP Sinus Pain Relief device is intended to be used for the temporary relief of sinus pain associated with Allergic Rhinitis.
# Over-the Counter (OTC) use:
The labeling, instructions, and user operations (21 CFR § 801.60 and 61), are designed for layman understanding and use. Proposed labeling for the device passed a Flesch-Kincaid reading below 80 grade. Clinical trial users were able to operate the device without any further training or help from the clinical staff.
### 8. Technological Characteristics and Predicate Device Comparison
The subject device and the predicate devices use the same fundamental technology. A comparison of the subject device technology and any differences to the predicate devices is provided in the table below.
### Substantial Equivalence Comparison Table
| | New Device | Predicate | Predicate | Remarks | |
|----|----------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | 510k # | K182025 | K060517 | K152199 | |
| | | New Device | Predicate | Predicate | Remarks |
| 2 | Regulation<br>number | 882.5890 | 882.5890 | 882.5890 | |
| 3 | OTC/prescriptio<br>n? | OTC | Prescription | OTC | |
| 4 | Product code | GZJ | GZJ | NFO | |
| 5 | Device name<br>and model | ClearUP Sinus Pain Relief | Pointer Excel | Rejuvatone MD, Model<br>TSTONEMD | |
| 6 | Manufacturer | Tivic Health Systems Inc. | Lhasa OMS, Inc. | Trophy Skin, Inc. | |
| 7 | Intended use | The Tivic ClearUP Sinus<br>Pain Relief is intended to<br>be used for the temporary<br>relief of sinus pain<br>associated with Allergic<br>Rhinitis. | The Pointer-Excel is<br>intended for use in the<br>symptomatic relief of<br>chronic intractable pain,<br>postoperative pain, and<br>acute pain. | The device is intended<br>for facial stimulation<br>and is indicated for over<br>the counter cosmetic<br>use. | |
| 8 | Power source | 3.7V rechargeable battery<br>supply | 9V battery supply | 9V battery (Type:<br>6LR61) | |
| | Method of line<br>current<br>isolation | N/A | N/A | N/A | |
| | Patient leakage<br>current (μΑ) | 0.6uA (normal condition)<br>0.7uA (single fault<br>condition: humidity) | N/A | 60 μΑ<br>360 μΑ | The new device presents<br>reduced risk to electrical<br>shock due to patient<br>leakage current as it is<br>battery-operated and is<br>disabled when charging |
| | | New Device | Predicate | Predicate | Remarks |
| 9 | Average DC<br>current through<br>electrodes<br>when device is<br>on but no<br>pulses are being<br>applied (μA) | 0 μA | 0 | 0 | Device is transformer<br>output hence no DC<br>component to waveform |
| 10 | Applied part<br>Number of<br>output modes | Type BF<br>1 | Type BF<br>1 | Type BF<br>1 | |
| 11 | Number of<br>output channels | 1 | 1 | 1 | |
| 12 | Output<br>intensity level | 3 steps, 85%, 90%, 100% | Continuously zero to full<br>power | 5 steps | |
| 13 | Regulated<br>current or<br>regulated<br>voltage? | Current controlled | Current controlled | Both | No impact on risk as long<br>as either voltage or<br>current aspect of<br>stimulation energy is<br>controlled |
| 14 | Software/firmw<br>are/micro-<br>processor<br>control? | Yes | Yes | Yes | |
| 15 | Automatic<br>overload trip | Not required due to circuit<br>design | Unknown | Not required due to<br>circuit design | |
| | | New Device | Predicate | Predicate | Remarks |
| 16 | Automatic no<br>load contact<br>trip | Yes | Yes | No | The new device presents<br>reduced impact on risk of<br>delayed treatment. User<br>is prompted when no<br>contact is detected with<br>treatment site (open<br>load) |
| 17 | Automatic shut<br>off | Yes | No | Yes | |
| 18 | User override<br>control? | Yes<br>Power on/off on the device | Yes, Normally Open<br>stimulation activation | Yes | |
| 19 | Indicator<br>display | | | | |
| | -On/off status | No | No | No | No risk impact on new |
| | -Low battery? | Yes | Yes | Yes | device as constant |
| | -Voltage/<br>current level? | Yes | Yes | No | current output set point is |
| | | No | Yes | Yes | fixed |
| 20 | Timer range<br>(minutes) | 7-second treatment period | Unknown | 20 minutes | |
| 21 | Weight (grams) | < 90 grams | 90 grams (main device) | 248 grams | |
| 22 | Dimensions<br>(cm.) HxWxL | 76x38x17 (mm) | 225x50x38 mm (main<br>device) | 177.2x35x55mm | |
| 23 | Housing<br>material and<br>construction | Housing - ABS,<br>Output contacts - Stainless<br>steel -active electrode,<br>Chrome return electrode | Housing - ABS,<br>Output contacts - ACD12<br>Chrome | Housing - ABS<br>Output contacts –<br>ACD12 Chrome | |
| | | New Device | Predicate | Predicate | Remarks re new device |
| 1 | 510k # | K182025 | K060517 | K152199 | |
| 2 | Device name<br>and model | ClearUP Sinus Pain Relief | Pointer Excel | Rejuvatone MD, Model<br>TSTONEMD | |
| 3 | Waveform | Biphasic with low duty<br>cycle | Biphasic | Pulsed monophasic | The new device presents<br>reduced risk of tissue<br>irritation due to net zero<br>transfer of charge with<br>balanced biphasic output |
| 4 | Shape | AC-coupled square | Square | Modulated square | Waveform design for<br>optimal transfer of<br>current across capacitive<br>feature of skin. |
| 6 | LOAD | Stimulation<br>Peak Volt / Peak mA | Stimulation<br>Peak Volt / Peak mA | Sensing<br>Peak Volt / Peak mA | |
| | 500Ω | +/- 3V / 6mA | 20V / 40mA | 256mV / 0.5mA | |
| | 2kΩ | +/- 10V / 5mA | 46V / 23mA | 806mV/ 0.4mA | |
| | 10kΩ | +/- 20V / 2mA | 85V / 8.5mA | 4.02V / 0.4mA | No impact on risk |
| 9 | Pulse<br>duration<br>(µsec) | 250 µsec | 260 µsec ±20% | On phase: 60 ms<br>Off phase: 60ms<br>Pulse width:120 ms | No impact on risk |
| 10 | Frequency<br>(Hz) | 15Hz | 1-16 Hz | 8.33 Hz | No impact on risk |
| 11 | Maximum<br>phase charge<br>(μC) 500Ω | 1.5 µC | 30 | 24.18 | The new device presents<br>reduced risk of tissue<br>damage |
| | | New Device | Predicate | Predicate | Remarks re new device |
| 1 | Maximum<br>current<br>density 500 Ω | $6 mA / 0.0625 cm² = 96 mA/cm²$ | $40 mA / 0.12 cm² = 330 mA/cm²$ | $0.498 mA/cm²$ (Minimum<br>electrode contact area<br>0.81 cm²) | The new device presents<br>reduced risk of tissue<br>damage |
| 1 | Maximum<br>current | $2.4-2.52 mA @ 500 Ω$<br>$1.5-1.75 mA @ 2k Ω$<br>$0.6-0.7 mA @ 10k Ω$ | Blue zone 0 - 2 mA<br>Black zone 2 - 45 mA<br>± 20% at 500 ohms<br>loading | $512 μΑ @ 500 Ω$<br>$403 μA @ 2k Ω$<br>$403 μA @ 10k Ω$ | The new device presents<br>reduced risk of tissue<br>damage |
| 1 | Maximum<br>average<br>power<br>density @<br>500 Ω | $42uA * 42uA * 500 /0.0625 cm² = 14uW/ cm²$ | $160uA * 160uA * 500 / 0.12 cm² = 106uW/ cm²$ | 0.03277 (Minimum<br>electrode contact area<br>0.81 cm²) | The new device presents<br>reduced risk of tissue<br>damage |
| 1 | Maximum<br>power<br>density @<br>500 ohms | $6mA * 6mA * 500 /0.0625 cm² = 0.29W/ cm²$ | $40mA * 40mA *500 / 0.12 cm² = 6.6W/ cm²$ | $2.005 mW/cm² @ 10k Ω$<br>(Minimum electrode<br>contact area 0.81 cm²) | Less than the predicate<br>device. |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
# Output specifications:
{9}------------------------------------------------
{10}------------------------------------------------
## 9. Summary of Verification and Validation activities
Bench testing passed predefined acceptance criteria. Software verification testing was carried out to ensure all device functions perform as intended. Testing was conducted and passed criteria as specified in the following standards:
- 트 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process [Including: Technical Corrigendum 1 (2010)], ISO, 10993-1 Edition 4: 2009-10, FDA Recognition Number: 2-220
- 트 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity, ANSI AAMI ISO, 10993-5 2009/(R)2014, FDA Recognition Number: 2-245
- . <87> Biological Reactivity Test, In Vitro -- Elution Test, USP, 40-NF35: 2017, FDA Recognition Number: 2-252
- 트 Medical device software - Software life cycle processes, IEC, 62304 Edition 1.1: 2015-06, FDA Recognition Number: 13-79
- . Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability, IEC, 60601-1-6 Edition 3.1: 2013-10, FDA Recognition Number: 5-89
- Medical electrical equipment - Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators, IEC, 60601-2-10 Edition 2.1: 2016-04, FDA Recognition Number: 17-16
- . Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment, IEC, 60601-1-11 Edition 2: 2015-01, FDA Recognition Number: 19-14
- . Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests, IEC, 60601-1-2 Edition 4: 2014-02, FDA Recognition Number: 19-8
- . Medical Devices - Part 1: Application of usability engineering to medical devices, IEC, 62366-1, Edition 1: 2015-02, FDA Recognition Number: 5-114
- C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD), ES60601-1, FDA Recognition Number 19-4
- Secondary cells and batteries containing alkaline or other non-acid electrolytes -Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications |Including: Corrigendum 1 (2013)], IEC 62133 Edition 2.0 2012-12, FDA Recognition Number 19-13
- Standard Specification for Wrought Stainless Steels for Surgical Instruments, ASTM F899-12b, FDA Recognition Number 8-343
Page 8 of 10 Traditional 510(k) Notification Tivic Health ClearUP™ Sinus Pain Relief
{11}------------------------------------------------
#### 10. Labeling Validation
A Usability/Label Comprehension Validation Study was conducted with 71 participants in order to validate the OTC appropriateness of the device labeling. The study group found that:
- 93% agreed that the device is easy to understand.
- 트 94% agreed that the instructions are easy to follow.
- 97% agreed that performing treatment was easy.
- 트 88% agreed that performing treatment was fast
# 11. Clinical Validation
A double-blinded randomized controlled clinical validation study was conducted at the Stanford Sinus Center to demonstrate the ability of subjects suffering from sinus pain to self-treat with the ClearUP Sinus Pain Relief device and to obtain pain relief compared to a sham/placebo device. Included were 27 sinus pain subjects with chronic rhinosinusitis, 49 patients suffering from allergic rhinitis and 5 subjects with other sinus conditions. The mean age for sham subjects was 44 and for active subjects 45 both groups had approximately 2:1 ratio of females to males.
Results from the clinical study were:
- -74% who used the microcurrent device experienced a reduction in sinus pain
- -24% who used the microcurrent device experienced sinus pain reduction of 3 points or more on the visual analog pain scale (0 no pain to 10 severe pain)
- -82% preferred ClearUP Sinus Pain Relief device to their current sinus treatment(s)
- -Only one subject experienced minor reddening of the skin which disappeared within minutes.
The statistically significant results demonstrated that the ClearUp device is a safe and effective treatment for sinus pain associated with allergic rhinitis. After a single treatment, active-treated subjects exhibited a statistically significant greater mean reduction in pain than subjects using the sham device.
## 12. Conclusion
The ClearUP Sinus Pain Relief device has the same fundamental scientific technology, intended use, and similar output parameters I comparison to predicate devices. . The differences between the proposed device and the predicate devices do not raise new
> Page 9 of 10 Traditional 510(k) Notification Tivic Health ClearUP™ Sinus Pain Relief
{12}------------------------------------------------
issues of safety and effectiveness and the proposed indication for sinus pain associated with allergic rhinitis was supported by clinical performance data as described above. In conclusion, the Tivic Health ClearUP Sinus Pain Relief device is substantially equivalent to the identified predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.