K152199 · Trophy Skin, Inc. · NFO · Mar 23, 2016 · Neurology
Device Facts
Record ID
K152199
Device Name
RejuvatoneMD
Applicant
Trophy Skin, Inc.
Product Code
NFO · Neurology
Decision Date
Mar 23, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The RejuvatoneMD is intended for facial stimulation and is indicated for over-the-counter cosmetic use.
Device Story
RejuvatoneMD is a non-invasive, at-home facial stimulation device for cosmetic use. It delivers low-level electrical microcurrent impulses via dual contact spheres to facial areas. Users adjust output intensity levels for comfort. The device operates on a 9V battery and features a 20-minute timer with automatic shut-off. It is intended for self-use by laypeople. The healthcare provider/user applies the device to the skin to provide electrical stimulation; the output is intended for cosmetic purposes. The device uses a pulsed monophasic, modulated square waveform. It is designed to be substantially equivalent to the NuFace Mini.
Clinical Evidence
No clinical data. Bench testing only. Device performance validated through electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and nerve/muscle stimulator performance (IEC 60601-2-10) testing. Biocompatibility confirmed per ISO 10993-5 and ISO 10993-10. Usability/label comprehension study conducted with 30 participants; >95% success rate in self-selection, comprehension, and device use.
Technological Characteristics
ABS plastic housing; ACD12 Chrome output contacts. Pulsed monophasic, modulated square waveform. 9V battery powered. Type BF applied part. Single channel. 5-step intensity adjustment. 20-minute timer. Complies with IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2, IEC 60601-1-6, IEC 62366, ISO 14971, and IEC 62304.
Indications for Use
Indicated for over-the-counter cosmetic facial stimulation in adult users.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are connected and appear to be emerging from a single form. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 23, 2016
Trophy Skin, Inc. % Sue Anthoney-Dewet FDA Consultant Aegis Regulatory, Inc. 2424 Dempster Drive Coralville, IA 52241
Re: K152199
Trade/Device Name: RejuvatoneMD Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: NFO Dated: February 17, 2016 Received: February 23, 2016
Dear Ms. Anthoney-Dewet:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of
{1}------------------------------------------------
medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Michael J. Hoffmann -A
- for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K152199
Device Name RejuvatoneMD
Indications for Use (Describe)
The RejuvatoneMD is intended for facial stimulation and is indicated for over-the-counter cosmetic use.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
#### K152199
This 510(k) summary information is being submitted in accordance with the requirements of 21 CFR § 807.92
Date Prepared: March 23, 2016
- 1. Submitter Information: AEGIS Regulatory, Inc. – Susan Anthoney-DeWet 2424 Dempster Drive Coralville, IA 52241 Tel.: 865-982-5552 Email: bob@fdalistingconsultants.com
#### On Behalf of Sponsor: Trophy Skin, Inc.
Attn: Imran Karim, President 4372 Kenmare Trail Frisco, TX 75034 Phone: (469) 233-1768 Email: imran@trophyskin.com
#### 2. General Information
2.1 Classification Name: Transcutaneous Electrical Nerve Stimulator for Aesthetic Purposes
- 2.2 Common/Usual Name: TENS Device
- 2.3 Proprietary Names: RejuvatoneMD
- 2.4 Classification: Class II
- 2.5 Classification Number: 882.5890
- 2.6 Review Panel: Neurology
- 2.6 Product Code: NFO
{4}------------------------------------------------
#### 3. Device Description:
The RejuvatoneMD device is a non-invasive at-home facial stimulation device intended for cosmetic use. The device works by delivering low-level electrical microcurrent impulses through dual contact spheres to strategic locations on the face and allows users to adjust the output level for personalized comfort.
#### 4. Indications / Intended Use:
The RejuvatoneMD is intended for facial stimulation and is indicated for over-the-counter cosmetic use.
#### Rx or OTC:
The labeling, instructions, and User Operations (21 CFR § 801.60 and 61), are designed for layman understanding and use. The predicate devices are indicated for (OTC) use.
#### 5. Predicate Device:
The RejuvatoneMD device is substantially equivalent in design, function and intended use to the following predicate device:
1. K133823 – NuFace Mini (Carol Cole Company)
#### 6. Technological Characteristics and Substantial Equivalence
The technological characteristics of the RejuvatoneMD device and the predicate device, are substantially equivalent in that both devices are designed to provide electrical stimulation to the facial skin for cosmetic use. The RejuvatoneMD and the predicate device function using the same mode of action by producing microcurrent that is discharged through electrodes that are in contact with the patient's skin, and both have user adjustable settings . The technical characteristics and physical presented in this submission demonstrate that the RejuvatoneMD device is substantially equivalent in design, function and intended use to the predicate device.
Sections below compare the characteristics of the RejuvatoneMD and the predicate device:
{5}------------------------------------------------
| Elements of Comparison | Subject Device | Predicate Device | |
|-----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|-------------------------------------|
| Device Name and Model | RejuvatoneMD, Model: TSTONEMD | NuFACE® Mini Device | |
| 510(k) Number | Applying | K133823 | |
| Product Code | NFO | NFO | |
| Manufacturer | Trophy Skin, Inc. | Carol Cole Company | |
| Intended Use | This device is intended for facial stimulation and is indicated for over-the-counter cosmetic use. | The NuFACE® Mini Device is intended for facial stimulation and is indicated for over-the-counter cosmetic use. | |
| Power Source(s) | 9V battery (Type: 6LR61) | 2 rechargeable AA NiMH batteries | |
| Applied Part | Type BF | Type BF | |
| Patient Leakage Current | NC 60μA<br>Comply with IEC 60601-1 and IEC 60601-2-10 | 56μA<br>Comply with IEC 60601-1 and IEC 60601-2-10 | |
| | SFC 360μA<br>Comply with IEC 60601-1 and IEC 60601-2-10 | 374μA<br>Comply with IEC 60601-1 and IEC 60601-2-10 | |
| Average DC current through electrodes when device is on but no pulses are being applied | 0A | -- | |
| Number of Output Channels | One channel | One channel | |
| Number of Output Modes | 1 mode | 1 mode | |
| Output Intensity Level | 5 steps | 3 steps | |
| Synchronous or Alternating? | N/A - 1 Output Channel | N/A - 1 Output Channel | |
| Method of Channel Isolation | N/A - 1 Output Channel | N/A - 1 Output Channel | |
| Regulated Current or Regulated Voltage? | Both | Both | |
| Elements of Comparison | Subject Device | Predicate Device | |
| Software/Firmware/Microprocessor Control? | Yes | Yes | |
| Automatic Overload Trip | Not required due to circuit design | Not required due to circuit design | |
| Automatic No-Load Trip | No | Yes | |
| Automatic Shut Off | Yes | Yes | |
| Patient Override Control | Yes | Yes | |
| Indicator On/Off Status | Yes | Yes | |
| Display Low Battery | No | Yes | |
| Voltage/ Current Level | Yes | Yes | |
| Timer Range | Yes (20 minutes) | Yes (21 minutes) | |
| LCD Display | No | No | |
| Compliance with Voluntary Standards | Yes<br>Comply with IEC 60601-1 and IEC 60601-2-10, IEC 60601-1-2 | Yes<br>Comply with IEC 60601-1 and IEC 60601-1-2 | |
| Biocompatibility | All user directly contacting materials are compliance with ISO10993-5 and ISO10993-10 requirements. | All user directly contacting materials are compliance with ISO10993-5 and ISO10993-10 requirements. | |
| Compliance* with 21 CFR 898 | Yes | Yes | |
| Main Unit Weight | 248g | 816g | |
| Dimensions | 55.0mm x 35.0mm x 177.2mm (L x W x H) | 2.5" W x 4.2" L x 1.2" D | |
| Electrode Size | 0.81cm² | -- | |
| Housing Materials and Construction | Housing made from ABS plastic and output contacts made from ACD12 Chrome. | Thermo Plastic | |
| Environment for operating | Temperature: 5 ~ 45° | Temperature: 5 ~ 40° C | |
| Elements of Comparison | Subject Device | Predicate Device | |
| | Humidity: 20 ~ 65% RH | Humidity: 20 ~ 65% RH | |
| Environment for storage | Temperature: 0 ~ 45° | Temperature: 0 ~ 40° C | |
| | Humidity: 10 ~ 90% RH | Humidity: 10 ~ 90% RH | |
| Waveform | Pulsed Monophasic | Pulsed Monophasic | |
| Shape | Modulated Square | Modulated Square | |
| Maximum Output<br>Voltage(+/- 10%) | 256mV@500Ω | 222mV@ 500Ω | |
| | 806mV@2KΩ | 781mV@ 2KΩ | |
| | 4.02V@10KΩ | 3.90V@ 10KΩ | |
| Maximum Output<br>Current(+/- 10%) | 512µA @ 500Ω | 396µA @ 500Ω | |
| | 403µA @ 2KΩ | 395µA @ 2KΩ | |
| | 403µA @ 10KΩ | 391µA @10KΩ | |
| Pulse Duration | ON phase: 60 ms | ON phase: 60.4 ms | |
| | OFF phase: 60 ms | OFF phase: 60.4 ms | |
| | Total Pulse Width: 120 ms | Total Pulse Width: 120.8<br>ms | |
| Pulse Frequency | 8.33Hz | 8.28 Hz | |
| Net Charge (per pulse) | 0µC @ 500Ω | 1.43µC @ 500Ω | |
| Maximum Phase Charge | 24.18µC @ 500Ω | 23.7µC @ 500Ω | |
| Maximum Current Density | 0.498mA/cm²@500Ω<br>(Minimum Electrode Contact<br>Area:0.81cm²) | 0.514mA/cm²@500Ω<br>(Minimum Electrode<br>Contact Area:0.771cm²) | |
| Maximum Average Power<br>Density | 32.77µW/cm² @500Ω<br>(Minimum Electrode Contact<br>Area:0.81cm²) | 99.05µW/cm² @500Ω<br>(Minimum Electrode<br>Contact Area:0.771cm²) | |
| Maximum Power Density | 2005µW/<br>cm²@10kΩ(Minimum<br>Electrode Contact Area:<br>0.81cm²) | 1981µW/<br>cm²@10kΩ(Minimum<br>Electrode Contact Area:<br>0.771cm²) | |
| Elements of Comparison | Subject Device | Predicate Device | |
| Burst Mode (i.e., pulse trains) | a. Pulses per burst | 20 | 20 |
| | b. Pulses per second | 8.33 | 8.28 |
| | c. Burst duration (seconds) | 2.4 | 2.42 |
| | d. Duty Cycle[line (b)xline (c)] | 20 | 20 |
| Contraction and Relaxation Time | | Adjustable, due to different modes. | Adjustable, due to different modes. |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
The RejuvatoneMD has similar technological characteristics with the predicate device in the product design, material, program mode and output waveform.
There are minor differences between the RejuvatoneMD device and the predicate device in the electrode size and subsequently the output characteristics including "Maximum Output Voltage", "Maximum Output Current", "Pulse Duration", "Pulse Frequency", "Net Charge (per pulse)", "Maximum Average Current", "Maximum Phase Charge", "Maximum Current Density", "Maximum Average Power Density", and "Burst Mode (i.e., pulse trains)." The output specification of the RejuvatoneMD device is very close to the predicate device in these areas and within an acceptable range. Both the predicate device and the RejuvatoneMD device comply with IEC 60601-2-10 requirement. So the differences of function specification do not raise any safety or effectiveness issue and does not effect the determination of substantial equivalence.
#### 7. Performance Testing and Standards:
Safety & performance testing demonstrates that the RejuvatoneMD is at least as safe and effective as the legally marketed predicate device.
The RejuvatoneMD device has been tested and conforms to international consensus standards:
#### ELECTRICAL SAFETY:
#### Recognition Number 19-4:
- . IEC/EN 60601-1:2005 Edition 3/(R)2012 And A1:2012 Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (lec 60601-1:2005, Mod). (General II (ES/EMC))
PERFORMANCE: Recognition Number 17-11:
{9}------------------------------------------------
- IEC 60601-2-10 Edition 2.0 2012-06, Medical Electrical Equipment -- Part 2-10: Particular Requirements For The Basic Safety And Essential Performance Of Nerve And Muscle Stimulators. (Neurology)
EMC:
Recognition Number 19-1:
- IEC 60601-1-2 Edition 3: 2007-03, Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests. (General II (ES/EMC))
#### LABEL VALIDATION:
A Usability/Label Comprehension Study was conducted in order to validate the device labeling with 30 participants.
The results of the final study group found that over 95% of the participants were able to:
- Correctly Self-Select the device as the correct or incorrect device for themselves
- Correctly complete the Comprehension portion of the study
- Correctly perform the Device Use portion of the study
In addition to the testing above, the RejuvatoneMD was designed referencing the following standards and FDA Guidance documents:
#### USABILITY:
#### Recognition Number 5-85:
- . IEC 60601-1-6 Medical electrical equipment – Part 1-6: General requirements for safety – Collateral Standard: Usability Edition 3.0, 2010
Recognition Number 5-67:
- . IEC 62366:2007/(R)2013, Medical Devices - Application Of Usability Engineering To Medical Devices. (General I (QS/RM))
#### RISK MANAGEMENT:
#### Recognition Number 5-40:
- ISO 14971:2007, ed.2 + MDD 93/42/EEC Annex I, IEC 60601-1, ed. 3+A1, Medical Devices -. Application Of Risk Management To Medical Devices. (General I (QS/RM))
#### SOFTWARE DESIGN:
#### Recognition Number 13-8:
- . IEC 62304 First Edition 2006-05, Medical Device Software - Software Life Cycle Processes. (Software/Informatics)
- . Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 11, 2005; and
- General Principles of Software Validation: Final Guidance for Industry and FDA Staff, January 11, 2002
#### 8. Substantial Equivalence Conclusion
After an analysis of the safety, indications, intended uses, performance features, design materials,
biocompatibility, power output, technological properties, treatment regimes and
{10}------------------------------------------------
methods of operation, the Sponsor believes that no significant differences exist between the device and the predicate device. Minor differences in the technological characteristics of the RejuvatoneMD device and the predicate devices do not raise any issues of safety or effectiveness Therefore substantial equivalency is requested.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.