The OSTEOTRANS™-MX Bioabsorbable Bone Fixation System is intended for use in trauma and reconstructive procedures of the craniofacial skeleton, including fracture of the cranium, infant craniofacial surgery (i.e. craniosynostosis, congenital malformations), pediatric reconstructive procedures, reconstructive procedures of the cranium, craniotomy flap fixation.
Device Story
OSTEOTRANS-MX is a sterile, single-use bioabsorbable bone fixation system consisting of plates, meshes, and screws. Fabricated from hydroxyapatite and poly-L-lactide (HA/PLLA) composites. Used by surgeons in clinical settings to maintain alignment of bone fractures and osteotomies during craniofacial trauma or reconstructive procedures. The device provides mechanical stabilization of bone segments; the material is designed to degrade over time while supporting bone healing. It is a passive implantable device; no software or electronic components involved.
Clinical Evidence
Bench testing only. Performance data demonstrated the device possesses the requisite mechanical strength and a favorable degradation profile to provide sustained bone fixation for the indicated procedures.
Technological Characteristics
Materials: Hydroxyapatite and poly-L-lactide (HA/PLLA) composite. Form factor: Sterile, single-use bone plates, meshes, and screws of various sizes. Principle: Bioabsorbable mechanical fixation. Sterilization: Not specified.
Indications for Use
Indicated for trauma and reconstructive procedures of the craniofacial skeleton in pediatric and adult populations, including cranial fractures, craniosynostosis, congenital malformations, and craniotomy flap fixation.
Regulatory Classification
Identification
A preformed alterable cranioplasty plate is a device that is implanted into a patient to repair a skull defect. It is constructed of a material, e.g., tantalum, that can be altered or reshaped at the time of surgery without changing the chemical behavior of the material.
Predicate Devices
Inion CPS 1.5/2.0/2.5 Bioabsorbable Fixation System (K010352)
OSTEOTRANS™-MX Bioabsorbable Bone Fixation System (K061881)
Submission Summary (Full Text)
{0}------------------------------------------------
K073006 page 14
# 510(k) Summary OSTEOTRANS™-MX Bioabsorbable Bone Fixation System
MAY 2 9 2008
Submitter's name : Submitter's address:
Contact Person :
Date prepared:
Trade or proprietary name:
Common or usual name:
Classification name:
Device product code:
Takiron Co., Ltd. 3-13 Azuchi-machi 2-chome, Chuo-ku, Osaka 541-0052, Japan
Kunihiro Hata Regulatory Affairs Specialist 7-1-19, Minatojimaminamimachi, Chuo-ku, Kobe, Hyogo, 650-0047, Japan Phone: +81 78 306 2601 +81 78 306 2606 Fax: October 23, 2007
OSTEOTRANS™-MX Bioabsorbable Bone Fixation System Bioabsorbable bone fixation plate and screw Cranioplasty plate and screw, Bone plate and screw, Class II HRS - 21 CFR 888.3030 Plate, Fixation, Bone HWC - 21 CFR 888.3040 Screw, Fixation, Bone GWO - 21 CFR 882.5320 Plate, Cranioplasty, Preformed, Alterable HBW - 21 CFR 882.5360 Fastener, Plate, Cranioplasty
#### Establishment Registration Number:
Takiron Co., Ltd. has not yet obtained an Establishment Registration Number.
### Legally Marketed Predicate Devices:
- 1. Inion Ltd.; Inion CPS 1.5/2.0/2.5 Bioabsorbable Fixation System (K010352)
- 2. Biomet, Inc .; LactoSorb® Trauma Plating System (K992355, K971870)
- Codman & Shurtleff, Inc.; Codman® Craniosorb™ Absorbable Fixation System 3. (K992905, K003549)
- 4. Bioplate Inc .; Modified plate design for use in conjunction with the Bioplate Rigid Bone Plating System for Craniomaxillofacial Surgery (K062819)
- Takiron Co., Ltd.; OSTEOTRANSTM-MX Bioabsorbable Bone Fixation System 5. (K061881)
5-1
{1}------------------------------------------------
#### Intended Use:
The OSTEOTRANS™-MX Bioabsorbable Bone Fixation System is intended for use in trauma and reconstructive procedures of the craniofacial skeleton, including fracture of the cranium, infant craniofacial surgery (i.e. craniosynostosis, congenital malformations), pediatric reconstructive procedures, reconstructive procedures of the cranium, craniotomy flap fixation.
#### Device Description:
The OSTEOTRANS™-MX Bioabsorbable Bone Fixation System devices are the sterile, single-use bone plates, meshes and screws manufactured from composites of hydroxyapatite and poly-L-lactide (HA/PLLA). Plates, meshes and screws are provided with various shapes and sizes typical of other marketed fixation devices.
Used properly, in the presence of adequate immobilization, the OSTEOTRANS" MX Bioabsorbable Bone Fixation System devices maintain accurate alignment of bone fractures and osteotomies.
# Summary of Technology:
The OSTEOTRANS "" - MX Bioabsorbable Bone Fixation System has the same technological characteristics (i.e., design and material) when compared to the predicate devices.
Performance data demonstrate that the OSTEOTRANS ""-MX Bioabsorbable Bone Fixation System has the requisite strength and favorable degradation profile to provide sufficient and sustained bone fixation for intended uses.
### Substantial equivalence
The OSTEOTRANS The MX Bioabsorbable Bone Fixation System is indicated for the same uses and anatomical regions as the predicate devices.
The OSTEOTRANS 2011 MX Bioabsorbable Bone Fixation System is identical with the implants in the previously 510(lx) cleared OSTEOTRANS-MX Bioabsorbable Bone Fixation System (K061881) in terms of composition, manufacturing method, sterilization method and packaging solution.
The OSTEOTRANS™-MX Bioabsorbable Bone Fixation System has very similar physical design features and functional characteristics as the other predicate devices. Therefore the OSTEOTRANS MX Bioabsorbable Bone Fixation System is substantially equivalent in design, materials and intended use and principles of operation to the predicate devices.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a circular arrangement of text around an emblem. The emblem is a stylized depiction of an eagle with its wings spread, clutching a staff or caduceus.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# MAY 2 9 2008
Takiron Co., Ltd. % Mr. Kunihiro Hata Regulatory Affairs Specialist 7-1-19, Minatojimaminamimachi, Chuo-ku, Kobe, Hyogo, 650-0047 Japan
Re: K073006
Trade/Device Name: OSTEOTRANS™ -MX Bioabsorbable Bone Fixation System Regulation Number: 21 CFR 882.5320 Regulation Name: Preformed alterable cranioplasty plate. Regulatory Class: Class II Product Code: GWO, HBW Dated: April 7, 2008 Received: April 7, 2008
Dear Mr. Hata:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 – Mr. Kunihiro Hata
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N Millerson
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### INDICATIONS FOR USE
Applicant: _ Takiron Co., Ltd. _ 510(k) Number (if known): K073006 Device Name: OSTEOTRANS™-MX Bioabsorbable Bone Fixation System
## Indications For Use:
The OSTEOTRANS™-MX Bioabsorbable Bone Fixation System is intended for use in trauma and reconstructive procedures of the craniofacial skeleton, including fracture of the cranium, infant craniofacial surgery (i.e. craniosynostosis, congenital malformations), pediatric reconstructive procedures, reconstructive procedures of the cranium, craniotomy flap fixation.
Prescription Use X (Per 21 CFR 801.109)
OR Over-The-Counter Use
er Use
N/D
(Please do not write below this line-continue on another page if needed) Concurrence of CDRH, Office of Device Evaluation (ODE)
Nuke bul for axa
(Division Sign-Off)
Division of General, Restorative, and Neurological Devices
510(k) Number K073006
4-1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.