The COOLGARD™ 3000/Alsius Catheter Thermal Regulation System, using either the IcyTM or Fortius™ model catheter, is indicated for use: - In cardiac surgery patients to achieve and or maintain normothermia during surgery and recovery/intensive care, and - To induce, maintain and reverse mild hypothermia in neurosurgery patients in surgery and recovery/intensive care. The Alsius CoolGard® 3000 and Cool Line™ Catheter Thermal Regulation System is indicated for use in fever reduction, as an adjunct to other antipyretic therapy, in patients with cerebral infarction and intracerebral hemorrhage who require access to the central venous circulation and who are intubated and sedated.
Device Story
CoolGard 3000 is an integrated electro-mechanical thermal regulation system. It uses a temperature probe in the patient to monitor core temperature and a controller to adjust the temperature of a sterile saline bath. This temperature-controlled saline is circulated through an indwelling intravascular catheter (Icy, Fortius, or Cool Line models) placed percutaneously. The system is operated by clinicians in surgical or intensive care settings to manage patient body temperature. By circulating saline through the heat-exchanging catheter, the device adds or removes heat from the patient's blood to achieve a physician-set target temperature. This helps maintain normothermia or induce therapeutic hypothermia, potentially benefiting patients in neuro-critical or cardiac recovery states.
Clinical Evidence
Clinical evidence includes a randomized controlled trial of 296 patients evaluating fever reduction safety across four cohorts: cerebral infarction (CI), intracerebral hemorrhage (ICH), primary traumatic brain injury (PTBI), and subarachnoid hemorrhage (SAH). Mortality results were reported for each cohort. Due to safety concerns in specific populations, a boxed warning was mandated for PTBI and SAH patients.
Technological Characteristics
Integrated electro-mechanical heater/cooler system. Components: temperature monitor, temperature controller, heat exchanger, and roller pump. Uses disposable start-up kits and intravascular catheters for percutaneous placement. Operates by circulating temperature-controlled sterile saline. Software-controlled thermal regulation.
Indications for Use
Indicated for cardiac surgery patients (normothermia maintenance) and neurosurgery patients (mild hypothermia induction/maintenance/reversal). Also indicated for fever reduction in intubated/sedated patients with cerebral infarction or intracerebral hemorrhage. Contraindicated for fever reduction in patients with subarachnoid hemorrhage or primary traumatic brain injury.
Regulatory Classification
Identification
A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.
Predicate Devices
Alsius CoolGard 3000 and Catheter Thermal Regulation System (acts as its own predicate)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the text "APR of Coming". The text is in a simple, sans-serif font. The letters are all capitalized, and the text is left-aligned. The image is a close-up of the text, and the background is blurred.
K0600308
Image /page/0/Picture/2 description: The image shows the logo for Alsius Corporation. The logo features the word "Alsius" in a stylized font, with the word "Corporation" in smaller letters underneath. The logo is simple and professional, and it is likely used to represent the company in its marketing and branding materials.
Company Confidential
CoolGard 3000 Thermal Regulation System
.
#### 510(k) Summary or 510(k) Statement 5
## 510(k) SUMMARY FOR ALSIUS CORPORTATION'S COOLGARD AND CATHETER THERMAL REGULATION SYSTEM
## Submitter's Name, Address, Telephone Number, and Contact Person
ALSIUS CORPORATION 15770 Laguna Canyon Road, Suite 150 Irvine, CA 92618
| Contact: | Ken Collins |
|----------|---------------------|
| Phone: | (949)-453-0150 |
| Fax: | (949)-453-0250 |
| Email: | kcollins@alsius.com |
#### Name of Device
The Alsius CoolGard And Catheter Thermal Regulation System.
#### Common or Usual Name
Central Venous Catheter (short term) and Thermal Regulating System.
#### Classification Name
FDA has classified the Alsius CoolGard 3000/Alsius heat exchange catheters as Class Il devices under 21 C.F.R. §§ 870.5900 - System, Hypothermia, Intravenous, Cooling. Classification Product Code: NCX.
## Predicate Devices
The product acts as its own predicate. This is a modification to the software of the device.
## Indications for Use
The following Indications for Use have clearance within the USA. The COOLGARD™ 3000 can be used with any of the Alsius Catheters. The indications for use are specific to the catheter. Please refer to the Indications for Use statement in the catheter specific Instructions for Use.
## Indications for Use (K030421, K051912, K052443)
The COOLGARD™ 3000/Alsius Catheter Thermal Regulation System, using either the IcyTM or Fortius™ model catheter, is indicated for use:
- . In cardiac surgery patients to achieve and or maintain normothermia during surgery and recovery/intensive care, and
Page 17
{1}------------------------------------------------
Company Confidential
- . To induce, maintain and reverse mild hypothermia in neurosurgery patients in surgery and recovery/intensive care.
## Indications for Use (K014241, K051912)
The Alsius CoolGard® 3000 and Cool Line™ Catheter Thermal Regulation System is indicated for use in fever reduction, as an adjunct to other antipyretic therapy, in patients with cerebral infarction and intracerebral hemorrhage who require access to the central venous circulation and who are intubated and sedated.
## Warning - Fever Reduction
The safety of this device has not been demonstrated for fever reduction in patients presenting with subarachnoid hemorrhage or primary traumatic brain injury. The safety and effectiveness of this device was examined in a randomized controlled trial of 296 patients. The mortality results reported in this trial, for the four patient cohorts enrolled, are presented in the table below (CI - cerebral infarction, ICH - intracerebral hemorrhage, PTBI -- primary traumatic brain injury, SAH -- subarachnoid hemorrhage).
| | Cool Line | | | Control | | | |
|------|-----------|----|------|---------|----|------|----------|
| | n | N | % | n | N | % | p-value* |
| CI | 3 | 16 | 18.8 | 3 | 14 | 21.4 | 0.74 |
| ICH | 8 | 33 | 24.2 | 7 | 27 | 25.9 | 1.00 |
| PTBI | 10 | 44 | 22.7 | 4 | 38 | 10.5 | 0.24 |
| SAH | 13 | 61 | 21.3 | 7 | 63 | 11.1 | 0.15 |
| Table 1-1. Mortality by Diagnosis (ITT) | | | | | | | |
|------------------------------------------|--|--|--|--|--|--|--|
|------------------------------------------|--|--|--|--|--|--|--|
*Fischer's exact test
For more details on the results of this study please refer to Physician's Manual -- "Normothermia for the Neuro-critically III Stroke Patient" #101416-001.
## Technical Characteristics
The CoolGard and Catheter Thermal Regulation System consists of the CoolGard™ 3000, a disposable Start Up Kit used in the CoolGard™ for interface with the cooling bath and patient catheter and the Intravascular Catheter. The Alsius CoolGard™ 3000 is an integrated electro-mechanical heater/cooler that consists of a temperature monitor, a temperature controller unit, a heat exchanger unit, and roller pump. It supplies temperature controlled sterile saline to the indwelling Catheter that is placed percutaneously in the patient.
The technical characteristics of the Catheter are identical to the predicate device with the exception of revisions to the software that controls the device.
The following models of Intravascular catheters are available for use with the CoolGard and Catheter Thermal Regulation System:
- 1. ICY® Catheter Kit Model IC-3585A(CO)
| 15770 Laguna Canyon Road, Suite 150 | Page 18 |
|-------------------------------------|---------|
| Irvine, California 92618 | |
Tel: 949.453.0150 FAX: 949,453.0250
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Alsius Corporation. The logo is in black and white and features the word "Alsius" in a stylized font. Below the word "Alsius" is the word "Corporation" in a smaller, sans-serif font. The logo is simple and professional.
- 2. ICY® Catheter Kit Model IC-3893A(CO)
- 3. Fortius® Catheter Kit Model FR-5093B(CO)
- 4. Cool Line® Catheter Kit Model CL 2085B(CO)
- 5. Cool Line® Catheter Kit Model CL 2295A(CO)
#### Principles of Operation
The CoolGard™ 3000 system automatically adjusts the temperature of the heater/chiller saline bath to achieve the patient target temperature that has previously been set by the attending physician. This is done via data from a temperature probe in the patient that interfaces with the temperature controller. This principle of operation is identical to the predicate device,
#### Summary of the Basis for Finding of Substantial Equivalence
The System in this premarket notification acts as its own predicate. There are no changes to the indications for use of the device. Revisions have been made to the software of the device for which supportive evidence is provided to establish that there are no new questions of safety or efficacy.
The System is substantially equivalent to the predicate device, the system itself.
#### Conclusion
In summary, descriptive information and performance data demonstrate that the Alsius CoolGard and Catheter Thermal Regulation System characteristics do not raise new questions of safety and effectiveness. Where appropriate, performance data demonstrate equivalence. The system in this revision acts as its own predicate device. The CoolGard and Catheter Thermal Regulation System is safe and effective when used in accordance with the Directions For Use and substantially equivalent to the predicate device.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" are arranged in a circular pattern around the eagle. The logo is black and white.
Public Health Service
APR 2 6 2006
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Alsius Corporation c/o Kenneth A. Collins, M.D. Executive Vice President 15770 Laguna Canyon Road, Suite 150 Irvine, California 92618
Re: K060308
Trade/Device Name: Alsius CoolGard 3000 and Catheter Thermal Regulation System Regulation Number: 21 CFR 870.5900 Regulation Name: Thermal regulating system Regulatory Class: II Product Code: NCX Dated: February 2, 2006 Received: February 9, 2006
Dear Dr. Collins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
The Office of Device Evaluation has determined that there is a reasonable likelihood that the Cool Line Catheter will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with section 513(i)(1)(E) of the Act, the following limitation must appear in the Warnings section of the device's labeling as a box warning immediately following the indications for use statement: The safety of this device has not been demonstrated for fever reduction in patients presenting with subarachnoid hemorrhage or primary traumatic brain injury. The safety and effectiveness of this device was examined in a randomized controlled trial of 296 patients. The mortality results reported in this trial, for the four patient cohorts enrolled, are presented in the table below (CI, cerebral infarction, ICH -- intracerebral hemorrhage, PTBI - primary traumatic brain injury, SAH - subarachnoid hemorrhage).
{4}------------------------------------------------
#### Page 2 - Kenneth A. Collins, M.D.
| | Cool Line | | | Control | | | |
|------|-----------|----|------|---------|----|------|----------|
| | n | N | % | n | N | % | p-value* |
| CI | 3 | 16 | 18.8 | 3 | 14 | 21.4 | 0.74 |
| ICH | 8 | 33 | 24.2 | 7 | 27 | 25.9 | 1.00 |
| PTBI | 10 | 44 | 22.7 | 4 | 38 | 10.5 | 0.24 |
| SAH | 13 | 61 | 21.3 | 7 | 63 | 11.1 | 0.15 |
#### Mortality by Diagnosis (ITT)
#### *Fischer's exact test
For more details on the results of this study please refer to Physician's Manual - "Normothermia for the Neuro-critically ill stroke patient" #101416-001.
Please note that the above labeling limitation is required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before this limitation is modified in any way or removed from the device's labeling. This limitation does not apply to the Icy & Fortius Catheters.
The FDA finding of substantial equivalence of your device to a legally marked predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 3 - Kenneth A. Collins. M.D.
If you desire specific information about the application of other labeling requirements to your device (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification: (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International, and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
V.C.M.
Donna-Bea Tillman, Ph.D., M.P.A. Director Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K060308
Alsius CoolGard 3000 and Catheter Thermal Regulation System Device Name:
Indications For Use:
Indications For Use:
#### Cool Line Cathers - Indications for Use:
The Alsius CoolGard® 3000 and Cool Line ™ Catheter Thermal Regulation System is indicated for use in fever reduction, as an adjunct to other antipyretic therapy, in nationts with cerebral infarction and intracerebral hemorrhage who require access to the central venous circulation and who are intubated and sedated.
Warning - Fever Reduction
The safety of this device has not been demonstrated for fever reduction in patients presenting with subarrachnoid hemorrhage or primary traumatic brain injury. The safety and effectiveness of this device was examined in a randomized controlled trial of 296 patients. The mortality results reported in this trial, for the four patient cohorts enrolled, are presented in the table below (C1 - cerebral infarction, ICH - intracerebral hemormage, PTBI - primary traumatic brain injury, SAH - subarrachnoid hemorrhage).
| Mortality by Diagnosis (ITT analysis) | | | | | | | | |
|---------------------------------------|-----------|----|------|--|---------|----|------|------|
| | Cool Line | | | | Control | | | |
| | n | N | % | | n | N | % | p* |
| CI | 3 | 16 | 18.8 | | 3 | 14 | 21.4 | 0.74 |
| ICH | 8 | 33 | 24.2 | | 7 | 27 | 25.9 | 1.00 |
| PTBI | 10 | 44 | 22.7 | | 4 | 38 | 10.5 | 0.24 |
| SAH | 13 | 61 | 21.3 | | 7 | 63 | 11.1 | 0.15 |
For more details on the clinical trial results please refer to Physician's Manual -"Normothermia for the Neuro-critically III stroke patient" #101416-001.
Prescription Use X
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Deyise Evalua
{7}------------------------------------------------
# Indications for Use
## 510(k) Number (if known): K060308
Alsius CoolGard 3000 and Catheter Thermal Regulation System Device Name:
Indications For Use:
## Fortius and Icy Catheters - Indications for Use:
The COOLGARD™ 3000/Alsius Catheter Thermal Regulation System, using either the Icy™ or Fortius™ model catheter, is indicated for use:
- in cardiac surgery patients to achieve and or maintain normothermia during . surgery and recovery/intensive care, and
- to induce, maintain and reverse mild hypothermia in neurosurgery patients in . surgery and recovery/intensive care.
AND/OR -Over-The-Counter Use Prescription Use _X (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.