CELSIUS CONTROL SYSTEM

K033623 · Innercool Therapies, Inc. · NCX · May 21, 2004 · Cardiovascular

Device Facts

Record IDK033623
Device NameCELSIUS CONTROL SYSTEM
ApplicantInnercool Therapies, Inc.
Product CodeNCX · Cardiovascular
Decision DateMay 21, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5900
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Celsius Control™ System is a thermal regulating system intended to induce, maintain and reverse mild hypothermia in neurosurgical patients in surgery and in recovery/intensive care, to achieve and/or maintain normothermia in cardiac surgery patients in surgery and in recovery/intensive care, and for use in fever reduction, as an adjunct to other antipyretic therapy, in patients with cerebral infarction and intracerebral hemorrhage who required access to the central venous circulation and who are intubated and sedated.

Device Story

Celsius Control System is a thermal regulating system for clinical use in surgery and intensive care. System components: endovascular catheter with distal heat exchange element; console with refrigeration/heating elements, heat exchanger, pump, and control software; sterile tubing set. Operation: thermal transfer fluid circulates through catheter to cool or warm blood. Temperature monitoring via conventional YSI-400 esophageal probes or catheter-integrated thermistor. Physician/clinician operates console to manage patient temperature. Output: controlled patient thermal state. Benefit: precise temperature management for neurosurgical/cardiac patients and fever reduction in specific neurological conditions.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on design, technological characteristics, and performance comparisons to predicate devices. A warning is included regarding increased mortality in subarachnoid hemorrhage and primary traumatic brain injury patients based on a randomized controlled study of endovascular cooling.

Technological Characteristics

System comprises an endovascular catheter (9-14 French), console with refrigeration/heating unit, and sterile tubing. Materials are biocompatible per ISO 10993-1. Sensing via YSI-400 thermistor probes or catheter-integrated thermistor. Energy source: electrical console. Connectivity: standalone. Software controls fluid circulation and thermal regulation.

Indications for Use

Indicated for neurosurgical and cardiac surgery patients requiring induction, maintenance, or reversal of mild hypothermia or maintenance of normothermia; also indicated for fever reduction in intubated, sedated patients with cerebral infarction or intracerebral hemorrhage requiring central venous access. Contraindicated for fever reduction in patients with subarachnoid hemorrhage or primary traumatic brain injury due to increased mortality risk.

Regulatory Classification

Identification

A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 033623 1/2 # MAY 2 1 2004 # 510(k) SUMMARY STO(K) SOMMEREE If accerdance with and the end 807 (21 CFR §807), and in particular §807.92, the following summary of safety and effectiveness information is provided: | Submitted by | | |--------------------------------|------------------------------------------------------------------------| | INNERCOOL therapies, Inc. | | | 3931 Sorrento Valley Boulevard | | | San Diego, California 92121 | | | Telephone: (858) 713-9507 | | | Contact: | Steve Reitzler, Vice President, Regulatory Affairs & Quality Assurance | | Date Prepared: | September 26, 2003 | | | | | Device Name | | | Trade or Proprietary Name: | Celsius Control™ System | | Common or Usual Name: | Thermal Regulating System | | Classification Name: | Thermal Regulating System | ### Predicate Devices Classification Name: The subject device is substantially equivalent, in whole or in part, to predicate devices manufactured by Radiant Medical (K012512), and Alsius (K014241, K030421). ### Device Description The subject device is a thermal regulating system consisting of three (3) parts: - an endovascular catheter having a heat exchange element at the distal end, . through which a thermal transfer fluid is circulated to cool or warm the blood, and which is available in various diameters from 9 french to 14 french; - a console containing refrigeration/lieating elements, a heat exchanger to cool . and warm the circulating fluid, a pump to circulate that fluid, and controls and software necessary to operate the system; and {1}------------------------------------------------ - a sterile tubing set to connect the console to the catheter, and through which the � a sterite cabing over 10033623 4/2 Two (2) models of the System are available: One which uses conventional, off-theshelf thermistor probes such as YSI-400 esophageal probes, to monitor patient temperature and control System operation, and one that uses a thermistor integral to the catheter. #### Intended Use The Celsius Control ™ System is a thermal regulating system intended to induce, maintain and reverse mild hypothermia in neurosurgical patients in surgery and in recovery/intensive care, to achieve and/or patients in surgery rmia in cardiac surgery patients in surgery and in maintain normosive care, and for use in fever reduction, as an adjunct to other antipyretic therapy, in patients with cerebral infarction and intracerebral hemorrhage who required access to the central venous circulation and who are intubated and sedated. #### Warning - Fever Reduction The safety of this device has not been demonstrated for fever reduction in patients presenting with subarachnoid hemorrhage or primary traumatic brain injury. A randomized controlled study of endovascular cooling in patients with subarachnoid hemorrhage and primary traumatic brain injury has shown increased mortality as compared to patients receiving standard of cure. ## Comparison to Predicate Devices The subject device has the same, or equivalent, indications for use as do other thermal regulating systems cleared for commercial distribution in the U.S .; The subject device has the same or equivalent design characteristics as other thermal regulating systems cleared for commercial distribution in the U.S .; The subject device is composed of biocompatible materials meeting the requirements of ISO 10993-1, as are other devices cleared for commercial distribution in the U.S .; The subject device has equivalent performance in inducing and reversing hypothermia, and in maintaining normothermia, as do other thermal regulating systems commercially available in the U.S. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract image of an eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAY 21 2004 Mr. Steve Reitzler Vice President, Regulatory Affairs and Quality Assurance INNERCOOL Therapies, Inc. 3931 Sorrento Valley Boulevard San Diego, California 92121 Re: K033623 Trade/Device Namc: Celsius Control System Regulation Number: 21 CFR 870.5900 Regulation Name: Thermal Regulating System Regulatory Class: Class II Product Code: NCX Dated: February 19, 2004 Received: February 23, 2004 Dear Mr. Reitzler: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the cnclosure) to legally marketed predicate devices marketed in interstate for associated in the May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls and Ocentral the Act and the limitations described below. The general controls provisions of the provisions of are recessarion, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. The Office of Device Evaluation has determined that there is a reasonable likelihood that this I no Orivel be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following evalu vaaler nust appear in the Warnings section of the device's labeling as a box warning immediately following the indications for use statement: ### Warning - Fever Reduction The safety of this device has not been demonstrated for fever reduction in patients r ne saling with subarrachnoid hemorrhage or primary traumatic brain injury. A procomized controlled study of endovascular cooling in patients with subarrachnoid hemorrhage and primary traumatic brain injury has shown increased mortality as compared to patients receiving standard of care. {3}------------------------------------------------ ## Page 2 - Mr. Steve Reitzler Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling. The FDA finding of substantial equivalence of your device to a legally marketed predicate I no a Dreams in a classification for your device and permits your device to procced to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it your be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination docs not mean r toase be determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific information about the application of other labeling requirements to your device (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, at note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International, and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, VoeleTu Donna-Bea Tillman, Ph.D. Acting Director Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### DRAFT LABELING 5.0 #### Indications for Use న. 1 510(k) Number (if known): Device Name: INNERCOOL therapies, Inc., Celsius Control ™ System Indications for Use: "The Celsius Control™ System is a thermal regulating system intended to induce, maintain and reverse mild hypothermia in neurosurgical patients in thatee, maintain and work of to achieve and/or maintain surgery and in recordiac surgery patients in surgery and in recovery/ intensive normont final as our were enduction, as an adjunct to other antipyretic therapy, cure, and for use in Jore. Fearction and intracerebral hemorrhage who required in patchis while venous circulation and who are intubated and sedated." V.P.C.A.T.M. Division of General, Restorative. and Neurological Devices **510(k) Number** K033623 Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Pcr 21 CFR 801.109) OR Over-The-Counter Use
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...