MAYFIELD DISPOSABLE AND REUSABLE TITANIUM SKULL PINS

K072208 · Integra LifeSciences Corporation · HBL · Sep 7, 2007 · Neurology

Device Facts

Record IDK072208
Device NameMAYFIELD DISPOSABLE AND REUSABLE TITANIUM SKULL PINS
ApplicantIntegra LifeSciences Corporation
Product CodeHBL · Neurology
Decision DateSep 7, 2007
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 882.4460
Device ClassClass 2
AttributesPediatric

Intended Use

The MAYFIELD® Disposable and Reusable Titanium Skull Pins are intended for use with a MAYFIELD skull clamp that is placed on the patient's skull to hold their head and neck in a particular position during surgical procedures when rigid skeletal fixation is desired and Intra-Operative MR imaging is used. The MAYFIELD® Disposable and Reusable Titanium Skull Pins are indicated for use in open and percutaneous craniotomies and spinal surgeries when rigid skeletal fixation is necessary and when Intra-Operative MR imaging of the patient is used.

Device Story

Titanium skull pins designed for use with MAYFIELD skull clamps; provide rigid skeletal fixation for patient head and neck positioning during neurosurgical and spinal procedures. Pins are installed in rocker arm and opposite side of clamp; facilitate intra-operative MR imaging compatibility. Used by surgeons in operating room settings. Benefit includes secure patient stabilization during surgery while allowing for MR imaging without artifact interference.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Titanium construction; two sizes (Adult and Child); mechanical fixation pins for use with MAYFIELD skull clamps; MR compatible.

Indications for Use

Indicated for patients undergoing open or percutaneous craniotomies and spinal surgeries requiring rigid skeletal fixation and intra-operative MR imaging.

Regulatory Classification

Identification

A neurosurgical head holder (skull clamp) is a device used to clamp the patient's skull to hold head and neck in a particular position during surgical procedures.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## MAYFIELD® Disposable and Reusable Titanium Skull Pins 510(k) Summary 1072208 #### Submitter's name and address: Integra LifeSciences Corporation 4900 Charlemar Drive, Building A Cincinnati, Ohio 45227 USA ### Contact person and telephone number: Helder A. Sousa Integra LifeSciences Corporation 311 Enterprise Drive Plainsboro, NJ 08536 (609) 936- 6850 #### Date prepared: August 7, 2007 #### Name of device: | Trade Name: | MAYFIELD® Disposable and Reusable Titanium Skull Pins | |----------------------|-------------------------------------------------------| | Common Name: | Skull Pins | | Classification Name: | Neurological head holder (Skull Clamp) | | Regulation Number: | 21 CFR 882.4460 | | Product Code: | HBL | #### Substantial Equivalence: The MAYFIELD® Disposable and Reusable Titanium Skull Pins are substantially equivalent in function and intended use to the MAYFIELD® Disposable Steel Skull Pins (K932860), MAYFIELD® Reusable Steel Skull Pins (pre-amendment), MAYFIELD® Radiolucent Disposable Skull Pins (K071458) and DORO Titanium Pins (K063494). The MAYFIELD® Disposable and Reusable Titanium Skull Pins like the predicate devices are used in a skull clamp that is placed on the patient's head to hold the patient's head and neck in a particular position during surgical procedures when rigid skeletal fixation is desired. The MAYFIELD® Disposable and Reusable Titanium Skull Pins like the predicate devices may be used in open and percutaneous craniatories and spinal surgeries when rigid skeletal fixation is necessary. The MAYFIELD® Disposable and Reusable Titanium Skull Pins as the MAYFIELD® Radiolucent Disposable Skull Pins (K071458) and the DORO Titanium Pins (K063494) may be used in surgical procedures when Intra-Operative MR imaging of the patient is used. SEP - 7 2007 {1}------------------------------------------------ #### Indications Use: The MAYFIELD® Disposable and Reusable Titanium Skull Pins are intended for use with a MAYFIELD skull clamp that is placed on the patient's skull to hold their head and neck in a particular position during surgical procedures when rigid skeletal fixation is desired and Intra-Operative MR imaging is used. The MAYFIELD® Disposable and Reusable Titanium Skull Pins are indicated for use in open and percutaneous craniotomies and spinal surgeries when rigid skeletal fixation is necessary and when Intra-Operative MR imaging of the patient is used. #### Device Description: The MAYFIELD® Disposable and Reusable Titanium Skull Pins are available in two sizes, Adult and Child which are used with MAYFIELD® Skull Clamps that are applied to the patient's skull to hold their head and neck in a particular position during surgical procedures where rigid skeletal fixation is desired and intra-operative MR imaging is used. In preparation for surgery three Titanium Skull Pins are installed in a MAYFIELD® Skull Clamp. Two Titanium Skull Pins are inserted in the Rocker Arm side of the Clamp and a single Titanium Skull Pin is inserted on the opposite side. The MAYFIELD® Disposable and Reusable Titanium Skull Pins will be used during open and percutaneous craniotomies and spinal surgeries when rigid skeletal fixation is necessary and when Intra-Operative MR imaging of the patient is used. #### Conclusion: The MAYFIELD® Disposable and Reusable Titanium Skull Pins are substantially equivalent to the MAYFIELD® Disposable Steel Skull Pins (K932860), MAYFIELD® Reusable Steel Skull Pins (pre-amendment), MAYFIELD® Radiolucent Disposable Skull Pins (K071458) and DORO® Skull Pins (K063494). The MAYFIELD® Disposable and Reusable Titanium Skull Pins are similar to the predicate devices in the intended use, the fundamental scientific technology of the device, and does not raise new issues of safety and effectiveness. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three heads, representing the department's commitment to health, human services, and science. The eagle is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. ood and Drug Administration 200 Corporate Boulevard Rockville MD 20850 Integra LifeSciences corpration % Helder A. Sousa Regulatory Affairs Specialist 311 Enterprise Drive Plainsboro, New Jersey 08536 - 2007 Re: K072208 Trade/Device Name: MAYFIELD® Disposable and Reusable Titanium Skull Pins Regulation Number: 21 CFR 882.4460 Regulation Name: Neurosurgical head holder (skull clamp) Regulatory Class: II Product Code: HBL Dated: August 7, 2007 Received: August 8, 2007 Dear Helder A. Sousa: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Helder A. Sousa This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permitte your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 633-241 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours for Pete Mark N. Melkerson Director DEP D.RS gof 9/7/07 Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): K072208 MAYFIELD® Disposable and Reusable Titanium Skull Pins Device Name: Indications For Use: The MAYFIELD® Disposable and Reusable Titanium Skull Pins are intended for use with a MAYFIELD skull clamp that is placed on the patient's skull to held their bead and neck in a particular position during surgical procedures when rigid skeletal fixation is desired and Intra-Operative MR imaging is used. The MAYFIELD® Disposable and Reusable Titanium Skull Pins are indicated for use in open and percutaneous craniotomies and spinal surgeries when rigid skouled fir ation is necessary and when Intra-Operative MR imaging of the patient is used. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (Please do not WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General, Restarativest 1 and Neurological Devices **510(k) Number** K072207 C - 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...