PIGALILEO TOTAL HIP REPLACEMENT (THR) SYSTEM

K070731 · Plus Orthopedics AG · HAW · Jul 31, 2007 · Neurology

Device Facts

Record IDK070731
Device NamePIGALILEO TOTAL HIP REPLACEMENT (THR) SYSTEM
ApplicantPlus Orthopedics AG
Product CodeHAW · Neurology
Decision DateJul 31, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4560
Device ClassClass 2

Intended Use

The PiGalileo™ Total Hip Replacement (THR) System is intended to be used in computer assisted orthopedic surgery to aid the surgeon with bone cuts and implant positioning during joint replacement. It provides information to the surgeon that is utilized to place surgical instruments during surgery utilizing anatomical landmarks and other data specifically obtained intra-operatively (e.g. hip center, pelvic plane etc.). Examples of some surgical procedures include but are not limited to: Primary total hip replacement Revision hip surgery Minimally invasive hip arthroplasty

Device Story

PiGalileo™ THR System is a software-controlled electromechanical stereotaxic device for computer-aided orthopedic surgery. System uses infrared (IF) camera and passive markers on surgical instruments to track position in real-time. Intra-operative anatomical landmarks (e.g., spinae, symphysis, femur) are scanned via probe to calculate hip cup and stem alignment. System cart houses CPU, monitor, user interface, and IF camera. Surgeon operates system in OR to receive real-time guidance on bone cuts and implant positioning. Output provides visual data on screen to assist surgeon in placing instruments; surgeon maintains control and can revert to conventional techniques. Benefits include improved precision in implant alignment during hip arthroplasty.

Clinical Evidence

Bench testing only. Design verification and validation were performed in accordance with 21 CFR 820.30. No clinical data was provided. Compliance with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), and ISO 14971 (risk management) was demonstrated.

Technological Characteristics

Stereotaxic navigation system using infrared (IF) camera and passive markers. System components: CPU, monitor, keyboard/touchpad, IF camera, and surgical instruments. Software-controlled electromechanical device. Connectivity: Standalone system cart. Sterilization: Cleaning/disinfection/sterilization per AAMI TIR 12-1994, ISO 17664, and ISO 11134. Biocompatibility per ASTM/ISO standards for invasive devices.

Indications for Use

Indicated for patients undergoing orthopedic joint replacement surgery, including primary total hip replacement, revision hip surgery, and minimally invasive hip arthroplasty, requiring computer-assisted guidance for bone cuts and implant positioning.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # |070731 JUL 3 1 2007 # Attachment B - Revised Premarket Notification 510(k) Summary : Attachment B July 26, 2007 Company Confidential . {1}------------------------------------------------ # K070731 # Premarket Notification 510(k) Summary #### 1. Submitter Information Company Name and Address: PLUS ORTHOPEDICS AG Erlenstraße 4a CH-6343 Rotkreuz Switzerland Contact Name: Pamela J. Weagraff, Principal Consultant Quintiles Consulting 18 Bridie Lane Norfolk, MA 02056 Telephone:508-528-1745 Facsimile: 978-752-1225 E-mail: pamela.weagraff@quintiles.com Date Prepared: July 26, 2007 #### 2. Name of Device - Trade Name: PiGalileo™ Total Hip Replacement (THR) System (a) - (b) Common Name: Navigation System - (c) Classification Name and Reference: Title 21 Code of Federal Regulation (CFR), Part 882.4560 Stereotaxic instrument, Product Code: HAW # 3. Substantial Equivalence Claimed to Predicate Device VectorVision Hip, K060468, manufactured by BrainLAB AG Orthopilot® 2 THA V 2.0, K050752, manufactured by Aescular®, Inc. Stryker Navigation System, K022365, manufactured by Stryker Corporation PiGalileo™ Total Knee Replacement (TKR) System, K061362, manufactured by Plus Orthopedics AG {2}------------------------------------------------ ### 4. Device Description The PiGalileo™ Total Hip Replacement (THR) software application when provided with the PiGalileo™ system cart results in the PiGalileo™ THR System, a softwarecontrolled electromechanical stereotaxic device for computer-aided navigation of PiGalileo™ surgical instruments with the purpose of assisting the surgeon in optimally positioning hip prostheses. Refer to Section 11.0 for an in depth description of the PiGalileo™ THR System. The PiGalileo™ THR System is intended to be used in computer-assisted orthopedic surgery to aid the surgeon with bone cuts and implant positioning during joint replacement. It provides information to the surgeon that is utilized to place surgical instruments during surgery utilizing anatomical landmarks and other data specifically obtained intra-operatively. The PiGalileo™ THR System is based on common stereotaxic technology in which Infrared (IF) LED (light emitting diodes) or passive markers on the surgical instruments allow the instruments to be tracked in real time in the surgical field. In the case of PiGalileo™ THR System, patient data that is required to navigate the surgical instruments is collected during the procedure. The system utilizes this data to establish a connection between passive locaters, i.e., Infrared (IF) light, and the system's IF camera as previously described tracks the surgical instruments in real time in the surgical field. The precision of navigation-assisted surgery depends on accurate scanning of skeletal landmarks. Scanned morphological data represents the basis for calculating the position of the hip cup and hip stem. Scanned points must represent unique anatomic landmarks so that they can be located in a safe and reproducible manner throughout surgery. For cup navigation, the spinae, left/ right are scanned as well as the symphysis. Each point is scanned three times to enhance precision and used for inclination and anteversion alignment. For stem navigation, the sagital plane is determined from the tibialis anterior and the intersection point on the femur. The position and alignment of the proximal stem axis are determined with a probe which provides the basis of the varus/valgus alignment. The surgeon maintains control of the operation and any decisions required with regard to the surgery at all times. Risk mitigations implemented under Design Controls ensure that sufficient fail safe mechanisms allow the surgeon to convert to non-navigated conventional surgical techniques at any time. The navigation platform for the PiGalileo™ THR System is the same as the navigation platform cleared under PiGalileo™ TKR System, K061362, and includes the following elements: {3}------------------------------------------------ - (a) System Cart housing the following items; there are no changes to the system cart as compared to K061362. - 1 System electronics such as CPU and connection box - 트 Monitor - I User Interface consisting of keyboard, touch pad and hard controls, e.g., "ON / OFF", footswitch - Infrared (IF) Camera and Camera Stand/Tripod - l Optional printer - . System cabling - (b) PiGalileo™ THR Software refers to the PiGalileo™ System Software and the THR software application: - . PiGalileo™ System Software: this software is the same as that of the PiGalileo™ TKR System; there are no changes to either the system software or the calculations or the landmark technique as compared to K061362. - 지 PiGalileo™ THR software application consists of two applets, PiGalileo™ Hip Cup and PiGalileo™ Hip Stem. Each applet consists of a collection of software modules that support the surgeon in hip cup and hip stem replacement. Each module is designed specifically for an implant in which the software calculates the values provided on screen differently for each implant, the specific implant geometry and the implant specific instruments. - (c) Surgical Instruments: include universal instruments such as hemispheres, and various locators; navigated instruments are equipped with markers that are tracked by the stereotaxic camera. The POLARCUP® Dual Mobility System consists of two components: a thin press fit shell and a liner component. ## 5. Intended Use / Indication for Use The PiGalileo™ Total Hip Replacement (THR) System is intended to be used in computer assisted orthopedic surgery to aid the surgeon with bone cuts and implant positioning during joint replacement. It provides information to the surgeon that is utilized to place surgical instruments during surgery utilizing anatomical landmarks and other data specifically obtained intra-operatively (e.g. hip center, pelvic plane etc.). Examples of some surgical procedures include but are not limited to: Primary total hip replacement Revision hip surgery Minimally invasive hip arthroplasty {4}------------------------------------------------ # 6. Predicate Device Comparison of Indications for Use / Intended Use and Technical Characteristics The comparison of the PiGalileo™ Total Hip Replacement (THR) System to the predicate devices was based on a review of the Design Control documentation, relevant aspects of which are included in the company's 510(k) Premarket Notification, and information concerning the predicate device, PiGalileo™ TKR System, K061362, that was available to the company internally, information concerning the predicate device that was obtained from the predicate device manufacturer's web site. The comparison considered technical characteristics and the indications for use / intended use. # 7. Performance Data - (a) Performance Standards (Section 514 Compliance): No performance standards applicable to this device, 882.4580, stereotaxic instrument, have been adopted under Section 514 of the Food Drug and Cosmetic Act. - (b) PiGalileo™ Total Hip Replacement (THR) System does conform to the following FDA recognized standards: - Thermal, Electrical and Mechanical Safety i) IEC 60601-1: Medical Electrical Equipment Part 1: General Requirements for Safety - ii) Electromagnetic Compatibility: IEC 60601-1-2:01: Electromagnetic Compatibility, Requirements and Tests, Emissions - Class A Limit, and Immunity IEC 61000-3-2:01: Electromagnetic Compatibility - Part 3-2 Limits for Harmonic Current Emissions IEC 61000-3-3:94+A1:01: Electromagnetic Compatibility -- Part 3-3 Limitation of Voltage Changes, Voltage Fluctuations - iii) Software IEC 60601-1-4:1996, Medical electrical equipment - Part 1-4: General requirements for safety - Collateral standard: Programmable electrical medical systems FDA's "Reviewer Guidance for the Content of Premarket Submission for Software Contained in Medical Devices", May 11, 2005 FDA's "Guidance for Off-The-Shelf Software Use in Medical Devices" - iv) Sterilization Cleaning, disinfection and sterilization testing was performed according to the following test protocols: "Microbiological Efficiency Control of the Automated Cleaning and Disinfection Method", AAMI TIR 12-1994, ISO 17664, LM SOP 2-11-01 {5}------------------------------------------------ "Microbiological Efficiency Control of Steam Sterilization (Fractionated Vacuum Procedure)", AAMI TIR 12-1994, ISO 17664, (half-cycle method), AINSI/AAMI/ISO 11134, EN 556-1, AINSI/AAMI ST 67, LM P 2-11-04, LM SOP 2-11-04 - v) Risk Analysis: ISO 14971:2000, Application of risk management to medical devices. - vi) Biocompatibility (Note: applies only to surgical tools, i.e., materials for system and THR software application do not come into contact with the patient.) ASTM/ISO standards, i.e., suitable for surgically invasive devices for transient to short term use. - vii) General: FDA's "Deciding When to Submit a 510(k) for a Change to an Existing Device", January 10, 1997 FDA's "Premarket Notification 510(k): Regulatory Requirements for Medical Devices", August, 1995 FDA's Updated 510(k) Sterility Review Guidance K90-1, August 30, 2002 - viii) Performance Testing: Design verification and design validation, e.g., bench testing was performed according to FDA's Design Control Requirements, Title 21 Code of Federal Regulations, Part 820.30. #### 8. Conclusion: The information and data provided in this 510(k) Premarket Notification establish that the PiGalileo™ Total Hip Replacement (THR) System is substantially equivalent to the afore-mentioned predicate devices with respect to indications for use/intended use, and technical characteristics. {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of human figures. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # JUL 3 1 2007 Plus Orthopedics AG % Ouintiles Consulting Pamela J. Weagraff, MBA, RAC Principal Consultant 18 Bridie Lane Norfolk, Massachusetts 02056 Re: K070731 Trade/Device Name: PiGalileo™ Total Hip Replacement (THR) System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: II Product Code: HAW Dated: July 11, 2007 Received: July 16, 2007 Dear Ms. Weagraff: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {7}------------------------------------------------ Page 2 - Pamela J. Weagraff, MBA, RAC forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark A. Millerson Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ ## ODE Indications Statement # 510(k) Number (if known): Unknown ### Device Name: PiGalileo™ Total Hip Replacement (THR) System #### Indications for Use: The PiGalileo™ Total Hip Replacement (THR) System is intended to be used in computer assisted orthopedic surgery to aid the surgeon with bone cuts and implant positioning during joint replacement. It provides information to the surgeon that is utilized to place surgical instruments during surgery utilizing anatomical landmarks and other data specifically obtained intra-operatively (e.g. hip center, pelvic plane etc.). Examples of some surgical procedures include but are not limited to: Primary total hip replacement Revision hip surgery Minimally invasive hip arthroplasty Prescription Use: X (21 CFR 801 Subpart D) AND/OR Over-the-Counter Use: (21 CFR 807 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Mark A. Milliken (Division Sign-Off) Division of General, Restorative, and Neurological Devices 510(k) Number K070731
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...