K232140 · Ortoma AB · OLO · Mar 11, 2024 · Neurology
Device Facts
Record ID
K232140
Device Name
OTS Hip
Applicant
Ortoma AB
Product Code
OLO · Neurology
Decision Date
Mar 11, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Attributes
AI/ML
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Segmentation and landmark identification
Fixed/static machine learning algorithms
—
Met the acceptance criteria
90 datasets from US (45.6%), Japan (33.3%), and the European Union (21.1%).
>1 (blinded annotators) + 1 (third reviewer)
—
—
Indications for Use
OTS Hip is indicated to enable planning of orthopedic surgical procedures based on CT medical imaging data of the patient anatomy. It is an intraoperative image-guided localization system that enables navigated surgery. It links a freehand probe, tracked by a passive marker sensor system, to virtual computer image space on a patient's preoperative image data being processed by the OTS platform. The system is indicated for orthopedic hip surgical procedures where a reference to a rigid anatomical structure, such as the pelvis, can be identified relative to a CT-based model of the anatomy. The system aids the surgeon to accurately navigate a compatible prosthesis to the preoperatively planned position. The system is designed for orthopedic surgical procedures including: - Pre-operative planning of Total Hip Arthroplasty (THA) - Intraoperative navigated surgery for THA using a posterior approach
Device Story
OTS Hip is a stereotaxic, image-guided surgical navigation system for Total Hip Arthroplasty (THA). It processes preoperative CT medical imaging data to create 3D anatomical models for surgical planning (OTS Hip Plan) and provides real-time intraoperative navigation (OTS Hip Guide). The system uses a camera, computer, and passive marker sensor system (Tracers) to track surgical instruments and prosthetic components in 3D space relative to the patient's anatomy. Operated by orthopedic surgeons in the OR, the system displays real-time positional data to guide tool and implant placement toward preoperatively planned goals. This navigation aids in achieving accurate implant positioning, potentially improving surgical outcomes. The system includes hardware (camera, PC, footswitch, adapters) and software for segmentation and landmark identification.
Clinical Evidence
No clinical data. Performance supported by quantitative system-level validation using real-life surgical data, bench testing (ASTM F2554-18), and ML algorithm validation. ML models for segmentation and landmark identification were validated against expert-annotated ground truth using 90 datasets (US/OUS) representative of diverse populations and CT equipment. Electrical safety and EMC testing conform to IEC 60601-1 and IEC 60601-1-2.
Technological Characteristics
Stereotaxic system using infrared optical tracking of passive markers (Tracers). Components: camera, medical PC, footswitch, adapters, and surgical instruments. Software: 3D planning (OHP) and real-time navigation (OHG). ML algorithms used for segmentation and landmark identification. Biocompatibility per ISO 10993. Sterilization: steam for reusable, gamma for single-use. Connectivity: DICOM-compliant. Standards: ASTM F2554-18 (tracking), IEC 60601-1 (safety), IEC 60601-1-2 (EMC), ISO 14971 (risk).
Indications for Use
Indicated for orthopedic hip surgical procedures (e.g., Total Hip Arthroplasty) requiring navigation relative to rigid anatomical structures (e.g., pelvis) using CT-based models. Contraindicated for patients with implants in the treatment side.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
March 11, 2024
Ortoma AB % John Smith Partner Hogan Lovells, LLC Columbia Square, 555 Thirteenth Street, NW Washington, District of Columbia 20004
Re: K232140
Trade/Device Name: OTS Hip Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: February 9, 2024 Received: February 9, 2024
# Dear John Smith:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
Tejen D. Soni -S
For Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair
{2}------------------------------------------------
and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232140
Device Name OTS Hip
#### Indications for Use (Describe)
OTS Hip is indicated to enable planning of orthopedic surgical procedures based on CT medical imaging data of the patient anatomy. It is an intraoperative image-guided localization system that enables navigated surgery. It links a freehand probe, tracked by a passive marker sensor system, to virtual computer image space on a patient's preoperative image data being processed by the OTS platform.
The system is indicated for orthopedic hip surgical procedures where a reference to a rigid anatomical structure, such as the pelvis, can be identified relative to a CT-based model of the system aids the surgeon to accurately navigate a compatible prosthesis to the preoperatively planned position.
The system is designed for orthopedic surgical procedures including:
- Pre-operative planning of Total Hip Arthroplasty (THA)
- Intraoperative navigated surgery for THA using a posterior approach
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for ORTOMA. The logo consists of a blue geometric shape on the left and the word "ORTOMA" in black on the right. The geometric shape is made up of three triangles, two larger triangles on top and a smaller triangle on the bottom.
# 510(k) SUMMARY - K232140 Ortoma AB OTS Hip
| Submitter | Ortoma AB |
|-----------------|--------------------------------------------|
| | Falkenbergsgatan 3 |
| | 412 85 Göteborg |
| | Sweden |
| Phone: | +46 73 81 03 660 |
| Contact Person: | John Smith |
| | Hogan Lovells, LLC |
| | Columbia Square, 555 Thirteenth Street, NW |
| | D.C., 20004 |
| | +1 202 367 3638 |
| | john.smith@hoganlovells.com |
Date Prepared: March 11, 2024
Name of Device: OTS Hip Common or Usual Name: Stereotaxic Instrument
Classification Name: 21 CFR Section 882.4560, Stereotaxic Instrument Regulatory Class: II Product Code: OLO Predicate Devices Ortoma AB, Ortoma Treatment Solution system (K181449)
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the word "ORTOMA" in bold, black letters. To the left of the word is a blue geometric shape that resembles a stylized letter A. The shape is made up of several triangles that are arranged to form the letter.
### Device Description
OTS Hip is a system to support a surgeon with preoperative planning and intraoperative guidance during orthopedic hip joint replacement surgery.
OTS Hip is comprised of software systems and hardware components that work together to form a stereotaxic system. The system uses medical imaging data in DICOM format that is loaded into the system for access in the software that are part of the system.
OTS Hip software consists of OTS Hip Plan (OHP), which is a 3D preoperative planning software, and OTS Hip Guide (OHG) that provides intraoperative real-time navigation for the quidance of surgical tools and prosthetic components in relation to the preoperatively determined goal positions.
OHP is a software for preoperative planning prior to a THA (Total Hip Arthroplasty) surgery. OHP enables the orthopedic surgeon to prepare surgery by analyzing the patient anatomy in a 3D environment based on medical imaging data.
OHG imports the result from the preceding planning stage, a released plan, with the 3D model and planned data, from the database of the OTS system. In addition, OHG monitors the real-time information of the position of instruments and prosthetic components in a 3D environment by means of medical imaqinq data.
The components of the OHG device include a camera and computer stand with an electrical system to which a camera and a medical panel PC are attached, a footswitch, a keyboard, Tracers (passive markers), adapters that hold the Tracers and can be mounted to compatible surgical instruments and that are used for calibration, and tools and instruments that are used during surgery.
The OTS is compatible with the following Depuy Synthes components:
- -PINN GB OFFSET GRATER HANDLE. DePuv Synthes 255000100
- Emphasys offset reamer, DePuy Synthes 4811-00-510 -
- Greatbatch Offset Cup Impactor, DePuy Synthes 255000115 -
- Pinnacle straight impactor, DePuy Synthes 221750041 l
- -Emphasys straight impactor, DePuy Synthes 4812-00-150
### Intended Use / Indications for Use
OTS Hip is indicated to enable planning of orthopedic surgical procedures based on CT medical imaging data of the patient anatomy. It is an intraoperative image-quided localization system that enables navigated surgery. It links a freehand probe, tracked by a passive marker sensor system, to virtual computer image space on a patient's preoperative image data being processed by the OTS platform.
The system is indicated for orthopedic hip surgical procedures where a reference to a rigid anatomical structure, such as the pelvis, can be identified relative to a CT-based model of the anatomy. The system aids the surgeon to accurately navigate a compatible prosthesis to the preoperatively planned position.
The system is designed for orthopedic surgical procedures including:
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for ORTOMA. The logo consists of a blue geometric shape on the left and the word "ORTOMA" in black, bold letters on the right. The geometric shape is made up of three triangles that are arranged to form a larger triangle.
- . Pre-operative planning of Total Hip Arthroplasty (THA)
- Intraoperative navigated surgery for THA using a posterior approach.
## Additional Considerations for Use
The device should not be used for patients with implants in the treatment side.
# Summary of Technological Characteristics
Both the predicate device and the subject device enable pre-operative planning and navigation of prosthetic components. OTS Hip is comprised of software systems and hardware components that work together to form a stereotaxic system.
Like the predicate device, the subject device includes software for pre-operative planning of orthopedic prosthetic components, OTS Hip Plan (OHP). The predicate device and the subject device have the same workflow for anatomical landmarks, and planning of implant size and positions. In the subject device functionality is changed to detect landmarks and perform segmentation using fixed/static machine learning (ML) algorithms, and to generate a 3D model based on the segmentation.
Like the predicate device, the subject device enables intraoperative image-guided navigated surgery using the OHG software and hardware components. The subject device and the predicate device include hardware components and a software for real-time navigation of surgical instruments and implants relative to the patient. A disposable tracking array called Tracer in the subject device replaces the combination of Marker and Marker Reflector Disc in the predicate device. The software included in OHG of the subject device and the predicate device is workflow based, where the user is guided to perform various steps in the workflow.
The subject device includes a new version of the computer and with electrical components. A computer has been added to the subject device the camera of the system has been changed to a camera from a new supplier. The underlying technology with infra-red tracking of passive markers remains the same as in the predicate device. In the subject device, a new version of the passive markers, called Tracers, has replaced the passive markers in the predicate device.
Components that are used in surgery include adapters, which are used to hold the passive markers. The adapters have the same function, to hold the passive marker relative to a compatible instrument, in the predicate device and the subject device. In the subject device, new adapters have been added with a revised design to hold the Tracer and to fit to compatible instruments that are supported by the subject device. Components, OTS Instruments that includes screws and drivers, have been added in the subject device, replacing components from external suppliers.
The OTS Hip has the equivalent indications for use and similar design features as compared with the predicate system.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for ORTOMA. The logo consists of a blue geometric shape on the left and the word "ORTOMA" in black, bold letters on the right. The geometric shape is made up of three triangles arranged to form a larger triangle.
The performance testing demonstrates that the performance characteristics of the OTS Hip are equivalent to those of the predicate device, and therefore supports a determination of Substantial Equivalence for the proposed indications for use.
Any differences between the subject and predicate device would not render the device NSE, affect the safety or effectiveness, or raise different questions of safety and effectiveness.
## Performance Testing
The following performance testing has been completed for the subject device, in support of the substantial equivalence decision:
## Quantitative System Level Validation
Quantitative system level validation testing was performed using real-life data from surgeries under clinical conditions enabling validation of system level accuracy and specific functionality within the system.
The results demonstrated that the inclination and anteversion, and the mean deviation for position error were acceptable for stereotactic systems.
Leg length measurements were evaluated for residual Leg Length Inequality (LLI) and was found to be within clinically acceptable values and would not affect the safety or effectiveness, or raise different questions of safety and effectiveness.
## Electrical Safety and Electromagnetic Compatibility (EMC) Testing
Electrical safety and EMC testing demonstrates conformance to IEC 60601-1:2005 (3rd Ed) and IEC 60601-1-2:2014 (4th Ed).
### Machine Learning Algorithm Validation
Validation testing demonstrated the accuracy of Machine Learning (ML) algorithms for segmentation and landmark identification.
The results of segmentation and landmark ML algorithms were compared with the manually annotated "ground truth" segmentations and landmarks of the test dataset. Appropriately qualified experts established the ground truth. Using objective criteria, cases were evaluated by blinded annotators. Cases were then separated into training and testing datasets in an unbiased fashion. Cases assigned to the test dataset were then validated by a third reviewer who evaluated the initial annotation.
For the seqmentation validation, two clinically complex cases that would be difficult for clinicians to manually interpret did not pass the acceptance criteria. However, the ML-models overall met the acceptance criteria.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image contains the logo for ORTOMA. The logo consists of a blue geometric shape on the left and the word "ORTOMA" in black, bold letters on the right. The geometric shape is made up of three triangles that are arranged to form a larger triangle.
The test datasets consisted of 90 datasets from both US and OUS data which were representative of the US population in terms of gender, age, and ethnicity and included images from multiple CT equipment manufacturers.
The dataset consisted of patients from the US (45.6%), Japan (33.3%), and the European Union (21.1%). Notably, the data from Japan included a high percentage of dysplastic hips with accompanying marked degenerative change.
The OUS dataset was unblinded.
For the OUS datasets, independent datasets were used between training and testing, though the two datasets were collected from the same site.
Subgroup analyses were conducted based on data variables such as scanner manufacture, slice thickens and imaging parameters. The test datasets were independent from the training dataset, where none of the datasets used for training was used for testing.
# Design Verification
The following design verification activities have been performed to ensure the correct functionality of the system as it has been specified. Tests were successfully completed.
- . Verifying the accuracy performance of the localization and tracking technology using the standardized test procedure according to ASTM Standard F2554-18.
- . Functional testing to ensure that all functional requirements are fulfilled.
- Safety testing verifying the effectiveness of all risk controls determined in the device risk analysis. .
- Risk assessment was performed per ISO 14971:2019 Medical devices Application or Risk ● Management to medical devices.
- . A detailed verification was performed covering the detailed functionality of the software (e.g., calculations of measurements from CT scans).
Non-clinical tests were performed to confirm the system targets. Specific OR setups and surgical procedures were simulated in laboratory environments and cadaver labs.
## Substantial Equivalence Comparison
| Characteristic | OTS Hip – Subject Device | Ortoma Treatment<br>Solution – Predicate<br>Device | Equivalence<br>Assessment |
|----------------|--------------------------|----------------------------------------------------|---------------------------|
| 510(k) Number | K232140 | K181449 | N/A |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains the word "ORTOMA" in bold, black letters. To the left of the word is a blue graphic consisting of three triangles. The triangles are arranged in a way that they form a larger triangle shape.
| Manufacturer | Ortoma AB | Same | N/A |
|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation | 21 CFR 882.4560 | 21 CFR 882.4560 | Identical |
| Product Code | OLO | OLO | Identical |
| Intended Use | To enable planning of<br>orthopedic surgical<br>procedures and enable<br>intraoperative image-<br>guided surgery. | To enable planning of<br>orthopedic surgical<br>procedures and enable<br>intraoperative image-<br>guided surgery. | Identical |
| Indications for<br>Use | OTS Hip is indicated to<br>enable planning of<br>orthopedic surgical<br>procedures based on CT<br>medical imaging data of the<br>patient anatomy. It is an<br>intraoperative image-<br>guided localization system<br>that enables navigated<br>surgery. It links a freehand<br>probe, tracked by a passive<br>marker sensor system, to<br>virtual computer image<br>space on a patient's<br>preoperative image data<br>being processed by the<br>OTS platform.<br>The system is indicated for<br>orthopedic hip surgical<br>procedures where a<br>reference to a rigid<br>anatomical structure, such<br>as the pelvis, can be<br>identified relative to a CT-<br>based model of the<br>anatomy. The system aids<br>the surgeon to accurately<br>navigate a compatible<br>prosthesis to the<br>preoperatively planned<br>position.<br><br>The system is designed for<br>orthopedic surgical<br>procedures including: | The Ortoma Treatment<br>Solution system is intended<br>to be an intraoperative<br>image-guided localization<br>system to enable navigated<br>surgery. It links a freehand<br>probe, tracked by a passive<br>marker sensor system, to<br>virtual computer image<br>space either on a patient's<br>preoperative image data<br>being processed by OTS<br>platform, or on an individual<br>3D-model of the patient's<br>bone, which is generated<br>through acquiring multiple<br>landmarks on the bone<br>surface.<br>The system is indicated for<br>hip surgical procedures, in<br>which the use of navigated<br>surgery is considered to be<br>safe and effective, and<br>where a reference to a rigid<br>anatomical structure, such<br>as the skull, a long bone, or<br>vertebra, can be identified<br>relative to a CT-based<br>model of the anatomy. The<br>system aids the surgeon to<br>accurately navigate a hip<br>prosthesis to the<br>preoperatively planned<br>position.<br><br>Example orthopedic<br>surgical procedures include<br>but are not limited to: | Equivalent<br>Revised indications<br>do not affect the<br>device's diagnostic<br>or therapeutic<br>effects. |
| | Pre-operative planning<br>and of Total Hip<br>Arthroplasty (THA) Intraoperative<br>navigated surgery for<br>THA using a posterior<br>approach. | Total Hip Arthroplasty<br>(THA) using posterior<br>approach. Preoperative planning<br>and intraoperative<br>navigated surgery for<br>joint replacement with<br>Stryker Exeter X3 Rimfit<br>cups. | |
| User Population | Orthopedic surgeon | Orthopedic surgeon | Identical |
| Anatomical Site | Hip | Hip | Identical |
| Where Used | Office of user and<br>Operating room | Operating room | Substantially<br>Equivalent.<br>Support for pre-<br>operative planning<br>in the user office to<br>support user<br>standard working<br>environment.<br>Preoperative<br>planning is not<br>dependent on<br>environment in<br>operating room<br>since input data is<br>generated before<br>entering the<br>operating room. |
| Technological<br>Principle of<br>Operation | Intraoperative image-<br>guided localization system<br>allowing user to plan<br>surgery using<br>premeasurements of<br>patient anatomy. Software<br>tracks anatomy, implants<br>and surgical tools in real-<br>time. | Intraoperative image-<br>guided localization system<br>allowing user to plan<br>surgery using<br>premeasurements of<br>patient anatomy. Software<br>tracks anatomy, implants<br>and surgical tools in real-<br>time. | Identical |
| Main<br>Technology of<br>MIS | The technology of<br>Minimally Invasive Surgery<br>(MIS) is based on Image<br>Guided Surgery (IGS)<br>devices. | The technology of<br>Minimally Invasive Surgery<br>(MIS) is based on Image<br>Guided Surgery (IGS)<br>devices. | Identical |
| Principle of<br>Operation Flow | Preoperative image ><br>surgical planning > surgical<br>guiding > recording | Preoperative image ><br>surgical planning > surgical<br>guiding > recording | Identical |
| Major<br>Components | Software for planning and<br>guiding | Software for planning and<br>guiding | Substantially<br>Equivalent. |
| | Calibration Adapter Unit<br>OTD | Calibration Adapter Unit | Substantially<br>Equivalent. |
| | | | Updated version to<br>support compatible<br>implant system and<br>new tracking array<br>(Tracers). |
| | Calibration Adapter Unit<br>OTD Emphasys | Calibration Adapter Unit | Substantially<br>Equivalent. |
| | | | New version to<br>support compatible<br>implant system and<br>new tracking array<br>(Tracers). |
| | Inserter Adapter OTD | Inserter Adapter | Substantially<br>Equivalent. |
| | | | Updated version to<br>support compatible<br>implant system and<br>new tracking array<br>(Tracers). |
| | Reamer Adapter OTD | Reamer Adapter | Substantially<br>Equivalent. |
| | | | Updated version to<br>support compatible<br>implant system and<br>new tracking array<br>(Tracers). |
| | Inserter Adapter OTD<br>Straight | Inserter Adapter | Substantially<br>Equivalent. |
| | | | New version to<br>support compatible<br>implant system and<br>new tracking array<br>(Tracers). |
| | Inserter Adapter OTD<br>Pinnacle Straight | Inserter Adapter | Substantially<br>Equivalent.<br>New variant to<br>support compatible<br>implant system and<br>new tracking array<br>(Tracers). |
| | Attachment Adapter Fix | Attachment Adapter for<br>Stryker OrthoLock EX-Pin<br>4x150…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.