MIDAS REX III MOTOR
Device Facts
| Record ID | K962483 |
|---|---|
| Device Name | MIDAS REX III MOTOR |
| Applicant | Midas Rex Pneumatic Tools, Inc. |
| Product Code | HBB · Neurology |
| Decision Date | Nov 19, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 882.4370 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Midas Rex Motor is designed for skull base and other microsurgical applications, including otolaryngology, ear, nose and throat, craniofacial and maxillofacial specifications in which a small, lightweight, high speed bone dissection system is desired.
Device Story
Pneumatic cranial drill motor; lightweight, ergonomic design; variable speed operation. Input: compressed air (20-150 psi). Output: high-speed mechanical bone dissection. Used in OR by surgeons for microsurgical applications. Integrates with existing Midas Rex GS dissecting tools, tube attachments, fixed/rotating footed attachments, telescoping microsystems, and micro mini drill guides. Facilitates precise bone removal in skull base and ENT procedures.
Clinical Evidence
Bench testing only; no clinical data provided. Substantial equivalence based on design and performance similarity to predicate devices.
Technological Characteristics
Pneumatic motor; operating pressure 20-150 psi (2-10 bar), recommended 100 psi (6.9 bar). Lightweight, small diameter housing. Quick disconnect interface for GS dissecting tools and attachments. Mechanical energy source.
Indications for Use
Indicated for patients requiring skull base and microsurgical bone dissection procedures, including otolaryngology, ENT, craniofacial, and maxillofacial surgeries.
Regulatory Classification
Identification
A pneumatic cranial drill motor is a pneumatically operated power source used with removable rotating surgical cutting tools or drill bits on a patient's skull.
Predicate Devices
- Midas Rex Motors, Midas (I, II, & Convertible) (K953434)
- Midas Rex Motors, Midas (I, II, & Convertible) (K952300)
- Whirl Wind TM (Pre-1976)
- Midas GS and GS20 Surgical Tool System (K952463)
- Midas Drill Guide Attachment (K954076)
- Midas Telescoping System (K953433)
Related Devices
- K970965 — MIDAS REX III MOTOR AND ATTACHMENTS (MIDAS III) · Midas Rex Pneumatic Tools, Inc. · Apr 9, 1997
- K972289 — MIDAS REX MOTOR · Midas Rex Pneumatic Tools, Inc. · Sep 12, 1997
- K020069 — MEDTRONIC MIDA REX LEGEND SYSTEM, MIDAS REX LEGEND PNEUMATIC HIGH SPEED SYSTEM OR LEGEND SYSTEM · Medtronic Midas Rex · Mar 18, 2002
- K013091 — ZMM 2000 ZEPPELIN MILLENNIUM MOTORDRILL SYSTEM · Von Zeppelin GmbH · Jan 25, 2002
- K030172 — SMALL HIGH SPEED MOTOR DRILL SYSTEM · Brasseler USA I, L.P. · Apr 3, 2003