MIDAS REX MOTORS, (MIDAS I, MIDAS II & CONVERTIBLE MODELS)

K953434 · Midas Rex Pneumatic Tools, Inc. · HBB · May 16, 1996 · Neurology

Device Facts

Record IDK953434
Device NameMIDAS REX MOTORS, (MIDAS I, MIDAS II & CONVERTIBLE MODELS)
ApplicantMidas Rex Pneumatic Tools, Inc.
Product CodeHBB · Neurology
Decision DateMay 16, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4370
Device ClassClass 2
AttributesTherapeutic

Indications for Use

To provide power to operate removable rotating surgical cutting tools or drill bits on a patient's skull.

Device Story

Midas Rex Motors (Midas I, Midas II, Convertible) are air-powered surgical motors; designed to drive rotating cutting tools or drill bits for cranial surgery. System components include stainless steel/aluminum/brass motor housing, air supply hose, and joint connectors. Operates at 20-120 psi or 200 psi. Features include optional rotating cuffs at joints to allow independent swivel of motor and hose, preventing kinking. Used in clinical surgical settings by physicians. Device provides mechanical power to surgical attachments; enables bone/tissue cutting. Supplied non-sterile; requires hospital-performed sterilization (steam or ethylene oxide) prior to use.

Clinical Evidence

Bench testing only. Performance characteristics (RPM, stall torque) compared to predicate device demonstrate equivalent functional output.

Technological Characteristics

Pneumatic motor; stainless steel, aluminum, and brass construction. Operating pressure: 20-120 psi or 200 psi. RPM: 30,000-75,000. Stall torque: 1-5.3 in-oz. Form factor: round or hexagonal casing (0.375, 0.5, or 0.75 inch diameter). Connectivity: air hose connection with optional swivel joints. Sterilization: compatible with ethylene oxide, high vacuum steam (270° F), and gravity displacement steam (250° F or 270° F).

Indications for Use

Indicated for use in surgical procedures requiring the operation of rotating cutting tools or drill bits on a patient's skull.

Regulatory Classification

Identification

A pneumatic cranial drill motor is a pneumatically operated power source used with removable rotating surgical cutting tools or drill bits on a patient's skull.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1} K953434 # 510 (k) SUMMARY MAY 16 1995 ## I. ADMINISTRATIVE Submitter: Midas Rex Pneumatic Tools, Inc. 3001 Race Street Fort Worth, Texas 76111 (817) 831-2604 Contact Person: Forest Barber, Jr. Date of Preparation: July 20, 1995 ## II. DEVICE NAME Proprietary Name: Midas Rex Motors (Midas I, Midas II and Convertible Models) Common Name: Surgical Instrument Motor Classification Name: Pneumatic Cranial Drill Motor (84HBB) ## III. PREDICATE DEVICE Midas Rex Whirlwind™ Motor, legally marketed prior to May 28, 1976. ## IV. DEVICE DESCRIPTION The Midas Rex Motors (Midas I, Midas II and Convertible models) are air-powered, rotating motors designed to operate at 20-120 psi or at up to 200 psi. The motor is connected to an air-supply source by means of a stainless steel joint and a plastic air hose with inner rubber pressure hose. The motors are constructed of stainless steel, aluminum and brass and are designed to operate at 20-120 psi or at up to 200 psi with the same performance characteristics. Depending on user preference, the 200 psi motor is available in either a round or hexagonal casing, approximately 0.75 inch in diameter. Similarly, the 120 psi motor can be supplied in either a round or hexagonal casing in diameters of approximately 0.375, 0.5 or 0.75 inch. Again depending on user preference, the joint connecting the motor to the air hose can be equipped with a rotating cuff at one or both ends, allowing the motor and/or air hose to swivel independently with respect to the joint, thus providing greater freedom of movement during surgical procedures and avoiding possible "kinking" of the air hose. Thus, a variety of device configurations are possible based on the surgeon's personal preference, as shown below: Premarket Notification: Midas Rex Motors {2} | | Motor | Joint | Hose | | --- | --- | --- | --- | | Midas I | Fixed | Straight (180°) | Fixed | | | Swivels | Straight (180°) | Swivels | | Midas II | Fixed | Angled (45°) | Fixed | | | Fixed | Angled (45°) | Swivels | | | Swivels | Angled (45°) | Fixed | | | Swivels | Angled (45°) | Swivels | | Convertible | Fixed | 180° or 45° | Fixed | | | Fixed | 180° or 45° | Swivels | | | Swivels | 180° or 45° | Fixed | | | Swivels | 180° or 45° | Swivels | Midas Rex Motors are supplied non-sterile for sterilization by hospital staff immediately prior to each use. The materials of construction of the motor and air hose are compatible with both ethylene oxide and steam sterilization (flash or standard autoclave procedures). A 4-minute cycle is recommended for high vacuum steam sterilization at 270° F. Gravity displacement steam sterilization may be performed at 250° F for 35 minutes or at 270° F for 15 minutes. A minimum of 18 hours mechanical aeration is recommended following ethylene oxide sterilization. ## V. INTENDED USE To provide power to operate removable rotating surgical cutting tools or drill bits on a patient's skull. ## VI. COMPARISON TO PREDICATE DEVICES See Table. Premarket Notification: Midas Rex Motors {3} Table of Comparison to Legally Marketed Device | | Midas I | Midas II | Convertible | Midas I, II and Convertible (K952300) | Whirlwind™ (pre-1976) | | --- | --- | --- | --- | --- | --- | | Construction | Stainless steel, brass and aluminum | Stainless steel, brass and aluminum | Stainless steel, brass and aluminum | Stainless steel, brass and aluminum | Stainless steel, brass and aluminum | | Case Shape/Diameter | Round or hexagonal: 0.375, 0.5 or 0.75 inch | Round or hexagonal: 0.375, 0.5 or 0.75 inch | Round or hexagonal: 0.375, 0.5 or 0.75 inch | Round or hexagonal: 0.375, 0.5 or 0.75 inch | Round/0.75 inch | | Motor/Air hose Joint | Straight | Angled (45°) | Straight or Angled (swivel joint) | Straight, Angled and Swivel Joint | Straight | | Motor Alignment (relative to joint) | Fixed or Swivel | Fixed or Swivel | Fixed or Swivel | Fixed or Swivel | Fixed | | Hose Alignment (relative to joint) | Fixed or Swivel | Fixed or Swivel | Fixed or Swivel | Fixed or Swivel | Fixed | | Air Operating Pressure | 20-120 psi or 200 psi | 20-120 psi or 200 psi | 20-120 psi or 200 psi | 20-120 psi or 200 psi | 200 psi | | RPM (000's) | 40-75 (20-120 psi) 30-75 (200 psi) | 40-75 (20-120 psi) 30-75 (200 psi) | 40-75 (20-120 psi) 30-75 (200 psi) | 40-75 (20-120 psi) 30-75 (200 psi) | 30-75 | | Stall Torque | 1-5.3 (20-120 psi) 1.3-5.3 (200 psi) | 1-5.3 (20-120 psi) 1.3-5.3 (200 psi) | 1-5.3 (20-120 psi) 1.3-5.3 (200 psi) | 1-5.3 (20-120 psi) 1.3-5.3 (200 psi) | 1.3-5.3 | | Intended Use | To provide power to operate removable rotating surgical cutting tools or drill bits on a patient's skull. | To provide power to operate removable rotating surgical cutting tools or drill bits on a patient's skull. | To provide power to operate removable rotating surgical cutting tools or drill bits on a patient's skull. | To provide power to operate various accessories or attachments to cut hard tissue or bone and soft tissue. | To provide power to operate various accessories or attachments to cut hard tissue or bone and soft tissue. | Premarket Notification: Midas Rex Motors
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...