MICROTARGETING STAR DRIVE SYSTEM, COMPONENTS

K092562 · FHC, Inc. · HAW · Sep 18, 2009 · Neurology

Device Facts

Record IDK092562
Device NameMICROTARGETING STAR DRIVE SYSTEM, COMPONENTS
ApplicantFHC, Inc.
Product CodeHAW · Neurology
Decision DateSep 18, 2009
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 882.4560
Device ClassClass 2

Intended Use

The FHC microTargeting™ STar Drive™ System is intended to be used with commercially available stereotactic systems for neurosurgical procedures which require the accurate positioning of microelectrodes, stimulating electrodes, or other instruments in the brain or nervous system

Device Story

The microTargeting™ STar Drive™ System is a stereotactic instrument used by neurosurgeons during functional neurosurgical procedures. It functions as a mechanical positioning device that mounts to commercially available stereotactic frames (e.g., Radionics, Leksell, Leibinger, Medtronic-IGN NeXframe). The system allows for the precise manual or motor-driven placement of intracranial microelectrodes, stimulating electrodes, and lesion electrodes. Components include the drive unit, optional lower guides, lead holders, depth stops, and verification probes. The device provides mechanical or digital position indication to assist the surgeon in navigating to specific intracranial targets. By facilitating accurate probe placement, the system supports clinical decision-making during procedures such as deep brain stimulation or lesioning. It is intended for use in a clinical/OR setting.

Clinical Evidence

Bench testing only. Verification and validation phases of design control demonstrated the system's equivalence or improvement over the predicate in mechanical and electrical quietness, repeatability, accuracy, rigidity, and ease of use.

Technological Characteristics

Materials: Hardcoated aluminum, 304 stainless steel. Drive mechanism: Manual or optional motor drive. Travel: 50 mm. Position indicator: Mechanical and/or digital readout. Connectivity: Mechanical interface to various stereotactic frames (Radionics, Leksell, Leibinger, M-IGN NeXframe). Sterilization: Steam.

Indications for Use

Indicated for use in neurosurgical procedures requiring accurate positioning of microelectrodes, stimulating electrodes, or other instruments within the brain or nervous system. Used in conjunction with commercially available stereotactic positioning systems.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K092562 SEP 1 8 2009 Microelectrodes and Instrumentation for Neuroscience Research and Clinical microTargeting FHC Corporate & Manufacturing 1201 Main Stree! Bowdom, ME 04287-7302 USA Tel. 207-666-8190 US & Can 800-326-2905 Fax 207-656-8292 E-mail fhcinc@fh.co.com www.th-co.com # SPECIAL 510(k) SUMMARY Submitter: FHC, Inc., 1201 Main Street, Bowdoin, Maine 04287 Tel: 207-666-5651; Fax: 207-666-8539 Contact Person: Lee D. Margolin, MS, PhD. Date of Summary Preparation: August 19, 2009 Trade Name: microTargeting™ STar Drive™ System Common Name: Stereotaxic instrument Classification Name: Stereotaxic instrument (21 CFR 882.4560, Product Code HAW) Substantially Equivalent To: FHC, Inc. microTargeting™ Drive System K003776, February 23, 2001 #### Description: When used in conjunction with commonly available stereotactic systems, the microTargeting™ STar Drive™ System allows a neurosurgeon to precisely position intracranial microelectrodes, stimulating electrodes, lesion electrodes and other instruments during functional neurosurgical procedures. microTargeting™ STar Drive™ System Components - microTargeting™ STar Drive™ . - optional lower guide . - lead holder with lead /lesion stop . - . verification probe - sterilization case . - cleaning brushes � #### Device Mounting Hardware and other Components Additional components for device mounting include hardware specifically designed to interface between the microTargeting TM STar Drive™ system and other stereotactic frames or instruments, and also include optional components to increase utility. {1}------------------------------------------------ - Adapters to fit Radionics, Leksell, Leibinger RM, Leibinger ZD. . FHC MicroTargeting™ Platform and Medtronic-IGN NeXframe stereotactic systems - ♥ single electrode insertion tube set - . array electrode insertion tube set - . lesion insertion tube kit and depth stops - . custom microelectrode depth stops ### Intended Use: The FHC microTargeting™ STar Drive™ System is intended to be used with commercially available stereotactic systems for neurosurgical procedures which require the accurate positioning of microelectrodes, stimulating electrodes, or other instruments in the brain or nervous system ### Technological Characteristics: Comparison Table | Parameter | microTargeting™ Drive | microTargeting™ STar Drive™ | |------------------------------------------------|---------------------------------------------------------------------------------------------------|-----------------------------| | Indications for Use | Accurate positioning of probes<br>in the brain or nervous system | Same | | Drive mechanism | Manual and/or optional motor drive | Same | | Biocompatibility<br>Drive system & Accessories | No contact with tissue | Same | | Insertion Tubes | 304 stainless steel | Same | | Travel | 50 mm | Same | | Sterilization | Steam; ethylene oxide | Steam | | Position Indicator | Mechanical and/or<br>digital readout capable | Same | | Stereotactic frame adapters | Radionics, Leksell, Leibinger RM,<br>Leibinger ZD, M-IGN NeXframe<br>FHC microTargeting™ Platform | Same | | Materials | Hardcoated Aluminum, Stainless Steel | Same | #### Performance testing Performance testing of the microTargeting™ STar Drive™ System documented in the verification and validation phases of design control, (See Section 10 and APPENDIX C of this document), show the system to be the equivalent or improved over the predicate system in ease of manufacturing, mechanical and electrical quietness, ease of use, repeatability, accuracy, rigidity, and adaptability. {2}------------------------------------------------ ## Substantial Equivalence statement: . . The microTargeting™ STar Drive™ System is substantially equivalent in design, construction, materials, intended use and performance characteristics to its predicate device, the FHC microTargeting™ Drive System, which was cleared under 510(k) K003776, February 23, 2001. : : : {3}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract bird or a series of flowing lines. SEP 1 8 2009 Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002 FHC, Inc c/o Lee Margolin, MS, PhD Senior Associate Director of Research & Development Quality System Officer 1201 Main Street Bowdoin, Maine 04287 Re: K092562 Trade/Device Name: microTargeting™ STar Drive™ System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: II Product Code: HAW Dated: August 19, 2009 Received: August 20, 2009 Dear Dr. Margolin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ ### Page 2 - Dr. Lee Margolin If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Kesia Alexander for Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): K092562 Device Name: microTargeting™ STar Drive™ System Indications for Use: The FHC microTargeting™ STar Drive™ System is intended to be used with commercially available positioning systems for neurosurgical procedures which require the accurate positioning of microelectrodes, stimulating electrodes, or other instruments in the brain or nervous system. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) ` Page of (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Nath Kohm (Division Sign Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices 510(k) Number k092562
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...