The FHC microTargeting™ XL STar™ Drive System is intended to be used with commercially available stereotactic systems for neurosurgical procedures which require the accurate positioning of microelectrodes, stimulating electrodes, or other instruments in the brain or nervous system
Device Story
The microTargeting™ XL STar™ Drive System is a stereotaxic instrument used by neurosurgeons to precisely position intracranial electrodes and instruments during functional neurosurgery. It interfaces with commercially available stereotactic frames via specialized mounting hardware. The system consists of a drive unit, insertion tubes, a verification probe, and cleaning/sterilization accessories. It utilizes a manual or optional motor-driven mechanism to advance instruments into the brain or nervous system. The device provides a position indicator (mechanical or digital) to assist the surgeon in achieving accurate placement. By enabling precise electrode positioning, the system facilitates functional neurosurgical interventions. It is designed for use in clinical settings where stereotactic procedures are performed.
Clinical Evidence
Bench testing only. Performance testing documented in design control verification phases demonstrated equivalence or improvement over the predicate system in mechanical and electrical quietness, ease of use, repeatability, accuracy, rigidity, and adaptability.
Technological Characteristics
Materials: 304 stainless steel (insertion tubes), hardcoated aluminum, stainless steel. Drive mechanism: Manual or optional motor. Travel: 125mm maximum. Connectivity: Interfaces with stereotactic frames (Radionics, Leksell). Sterilization: Steam. Position indicator: Manual, mechanical, or digital readout.
Indications for Use
Indicated for neurosurgical procedures requiring accurate positioning of microelectrodes, stimulating electrodes, or other instruments in the brain or nervous system.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
Predicate Devices
FHC, Inc. microTargeting™ STar Drive™ System (K092562)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the letters "FHC" in a bold, textured font. The letters are arranged horizontally, with "F" on the left, "H" in the middle, and "C" on the right. The texture of the letters gives them a slightly rough or patterned appearance.
Microelectrodes and Instrumentation for Neuroscience
Research and Clinical microTargeting®
FHC Corporate & Manufacturing
1201 Main Street
Bowdoin, ME 04287-7302 USA
Tel: 207-666-8190
US & Can: 800-326-2905
Fax: 207-666-8292
E-mail: fhciric@ih-co.com
www.fh-co.com
# SPECIAL 510(k) SUMMARY
FHC, Inc., 1201 Main Street, Bowdoin, Maine 04287 Submitter: Tel: 207-666-5651; Fax: 207-666-8539
Contact Person: Keri Seitz
Date of Summary Preparation: June 28, 2012
Trade Name: microTargeting™ XL STar™ Drive System
Common Name: Stereotaxic instrument
Classification Name: Stereotaxic instrument (21 CFR 882.4560, Product Code HAW)
Substantially Equivalent To: FHC, Inc. microTargeting™ STar Drive™ System K092562, September 18, 2009
#### Description:
1.121950
When used in conjunction with commonly available stereotactic systems, the microTargeting "M XL STar™ Drive System allows a neurosurgeon to precisely position intracranial microelectrodes, stimulating electrodes, lesion electrodes and other instruments during functional neurosurgical procedures.
# microTargeting™ XL STar™ Drive System Components
- microTargeting™ XL STarTM Drive .
- insertion tubes
- verification probe ●
- sterilization case ●
- cleaning brushes .
# Device Mounting Hardware and other Components
Additional components for device mounting include hardware specifically designed to interface between the microTargeting™ XL STar™ Drive system and other stereotactic frames or instruments, and also include optional components to increase utility.
- Adapters to fit Radionics and Leksell stereotactic systems .
- microTargeting controller firmware based on user specifications of drive travel.
{2}------------------------------------------------
#### Intended Use:
The FHC microTargeting™ XL STar™ Drive System is intended to be used with commercially available stereotactic systems for neurosurgical procedures which require the accurate positioning of microelectrodes, stimulating electrodes, or other instruments in the brain or nervous system
# Technological Characteristics:
# Comparison Table
| Parameter | microTargeting™ Star<br>Drive™ | microTargeting™ XL Star™<br>Drive |
|--------------------------------|----------------------------------------------------------------------------------------------------------|-----------------------------------|
| Indications for Use | Accurate positioning of probes in<br>the brain or nervous system | Same |
| Drive mechanism | Manual and/or optional motor<br>drive | Same |
| Biocompatibility | | |
| Drive system &<br>Accessories | No contact with tissue | Same |
| Insertion Tubes | 304 stainless steel | Same |
| Travel | 50 mm | Maximum of 125mm |
| Sterilization | Steam | Steam |
| Position Indicator | Manual, mechanical and/or digital<br>readout capable | Same |
| Stereotactic frame<br>adapters | Radionics, Leksell, Leibinger<br>RM, Leibinger ZD, M-IGN<br>NeXframe and FHC<br>microTargeting™ Platform | Radionics and Leksell |
| Materials | Hardcoated Aluminum, Stainless<br>Steel | Same |
### Performance testing
Performance testing of the microTargeting™ XL STar™ Drive System documented in the verification phases of design control (See Section 10 and APPENDIX C of this document) show the system to be equivalent or improved over the predicate system in ease of manufacturing, mechanical and electrical quietness, ease of use, repeatability, accuracy, rigidity, and adaptability.
# Substantial Equivalence statement:
The microTargeting™ XL STar™ Drive System is substantially equivalent in design, construction, materials, intended use and performance characteristics to its predicate
{3}------------------------------------------------
4121950
P. 3 of 3
-
.
device, the FHC microTargeting™ STar Drive™ System, which was cleared under 510(k) K092562, September 18, 2009.
.
.
. . . .
く
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUL 26 2012
FHC. Incorporated % Ms. Keri Seitz President and CEO 1201 Main Street Bowdoin, Maine 04287
Re: K121950
Trade/Device Name: microTargeting™ XL STar Drive™ System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: II Product Code: HAW Dated: June 28, 2012 Received: July 3, 2012
Dear Ms. Seitz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Coometer for (110) in 10. (110) in the general controls provisions of the Act. The r ou may, alere, mains of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and advileerstion. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Keri Seitz
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Eula A. Rumm, MD, of
Malvina B. Evdelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Page 1 of 1
KI21950 510(k) Number (if known):
Device Name: microTargeting™ XL STar™ Drive System
Indications for Use:
The FHC microTargeting™ XL STar™ Drive System is intended to be used with commercially available stereotactic systems for neurosurgical procedures which require the accurate positioning of microelectrodes, stimulating electrodes, or other instruments in the brain or nervous system.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K121950
(Optional Format 3-10-98)
**Prescription Use**
(Per 21 CFR 801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.