K250879 · Proprio, Inc. · OLO · Jun 4, 2025 · Neurology
Device Facts
Record ID
K250879
Device Name
Paradigm System
Applicant
Proprio, Inc.
Product Code
OLO · Neurology
Decision Date
Jun 4, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The Paradigm System is a stereotaxic image guidance system intended for the spatial positioning and orientation of spinal surgical instruments used by surgeons during open surgical procedures with appropriate bone preparation. The device is indicated for posterior approach spine surgery where reference to a rigid anatomical structure that can be identified on CT derived patient images for pedicle screw cannulation of the thoracic to sacrum vertebrae.
Device Story
Stereotaxic image guidance system for spinal surgery; inputs include preoperative CT scans and real-time intraoperative camera imaging of surgical site. System software segments CT data to generate 3D model of patient vertebrae; intraoperative Prism Sensor Array matches live anatomy topography to 3D model for registration. Creates common coordinate space between preoperative data and patient anatomy; overlays virtual information on live anatomy for guidance. Used in OR by surgeons; provides visual navigation for instrument placement (e.g., pedicle screws). Benefits include precise instrument trajectory and placement relative to patient anatomy.
Clinical Evidence
Bench testing only. Verification and validation activities conducted per standards including IEC 62366-1, ISO 10993-1, ISO 14971, ISO 17665-1, ANSI/AAMI ST79, ASTM F2554-22, ANSI/AAMI ST98, and IEC 62304.
Technological Characteristics
Stereotaxic image-guided navigation system; components include cart, arm, sensor array, monitor, and instruments. Uses reflective markers (Northern Digital K033621). Sterilization: steam for instruments (SAL 1x10^-6), ethylene oxide for drapes (SAL 1x10^-6). Software lifecycle per IEC 62304. Positional accuracy testing per ASTM F2554-22.
Indications for Use
Indicated for patients undergoing posterior approach spine surgery requiring pedicle screw cannulation of thoracic to sacrum vertebrae where rigid anatomical structures are identifiable on CT images.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
K252950 — Paradigm System · Proprio, Inc. · Dec 15, 2025
K243258 — Paradigm System · Proprio, Inc. · Feb 21, 2025
K252597 — Valence Robotic Navigation System; Valence Robotic Navigation Instruments (For Use with StealthStation) · Alphatec Spine, Inc. · Feb 19, 2026
K201189 — Stealthstation S8 Spine Software v1.3.0 · Medtronic Navigation, Inc. · May 29, 2020
K242271 — Caduceus S · Taiwan Main Orthopaedic Biotechnology Co., Ltd. · Oct 30, 2024
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 4, 2025
Proprio, Inc.
Shannon Eubanks
Vice President, Systems Engineering & Regulatory
111 W. John St., Suite 308
Seattle, Washington 98119
Re: K250879
Trade/Device Name: Paradigm System
Regulation Number: 21 CFR 882.4560
Regulation Name: Stereotaxic Instrument
Regulatory Class: Class II
Product Code: OLO
Dated: April 4, 2025
Received: April 7, 2025
Dear Shannon Eubanks:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250879 - Shannon Eubanks
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE
{2}
K250879 - Shannon Eubanks
Page 3
by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali -S
Shumaya Ali, M.P.H.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250879 | |
| Device Name Paradigm System | |
| Indications for Use (Describe) The Paradigm System is a stereotaxic image guidance system intended for the spatial positioning and orientation of spinal surgical instruments used by surgeons during open surgical procedures with appropriate bone preparation. The device is indicated for posterior approach spine surgery where reference to a rigid anatomical structure that can be identified on CT derived patient images for pedicle screw cannulation of the thoracic to sacrum vertebrae. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K250879 Page 1 of 4
# 510(k) Summary Statement
| Submitter: | Proprio, Inc. |
| --- | --- |
| Contact Person: | Shannon Eubanks
VP of Systems Engineering and Regulatory
Phone: (425) 802-6063
Email: seubanks@propriovision.com |
| Trade Name: | Paradigm System |
| Common Name: | Orthopedic stereotaxic instrument |
| Classification: | Class II |
| Product Code: | OLO |
| Regulation: | 21 CFR 882.4560 |
| Predicate Device(s): | The subject device is equivalent to the following device: Paradigm System (K243258) |
| Device Description: | The Paradigm System is a stereotaxic image guidance system intended for the spatial positioning and orientation of spinal surgical instruments used by surgeons. The Paradigm System fuses high-resolution, real-time camera imaging of the surgical site (such as the spine) with medical imagery (such as CT) and surgical navigation data, such as predetermined instrument trajectories for precise instrument placement.
The Paradigm System encompasses non-sterile operating room equipment components as well as sterilizable, reusable instruments, and off-the-shelf sterile-packaged, single-use accessories.
The Paradigm System is used with an external monitor. |
| Indications For Use: | The Paradigm System is a stereotaxic image guidance system intended for the spatial positioning and orientation of spinal surgical instruments used by surgeons during open surgical procedures with appropriate bone preparation. The device is indicated for posterior approach spine surgery where reference to a rigid anatomical structure that can be identified on CT derived patient images for pedicle screw cannulation of the thoracic to sacrum vertebrae. |
Proprio, Inc.
Premarket Notification for the Paradigm System
{5}
K250879 Page 2 of 4
| | Paradigm System
(Subject Device) | Paradigm System
(Predicate Device) | Comment |
| --- | --- | --- | --- |
| Device Overview | | | |
| 510(k) Number
Decision Date | K250879 | K243258
February 21, 2025 | |
| Manufacturer | Proprio, Inc. | Proprio, Inc. | |
| Classification | Class II | Class II | Same |
| Product Code | OLO | OLO | Same |
| Regulation | 21 CFR 882.4560 | 21 CFR 882.4560 | Same |
| Medical Specialty | Neurology | Neurology | Same |
| Indications For Use | The Paradigm System is a stereotaxic image guidance system intended for the spatial positioning and orientation of spinal surgical instruments used by surgeons during open surgical procedures with appropriate bone preparation. The device is indicated for posterior approach spine surgery where reference to a rigid anatomical structure that can be identified on CT derived patient images for pedicle screw cannulation of the thoracic to sacrum vertebrae. | The Paradigm System is a stereotaxic image guidance system intended for the spatial positioning and orientation of spinal surgical instruments used by surgeons during open surgical procedures with appropriate bone preparation. The device is indicated for posterior approach spine surgery where reference to a rigid anatomical structure that can be identified on CT derived patient images for pedicle screw cannulation of the thoracic to sacrum vertebrae. | Same |
| Principle of operation | Paradigm System fuses high-resolution, real-time camera imaging of the surgical site (such as the spine) with medical imagery (such as CT) and surgical navigation data, such as predetermined instrument trajectories for precise instrument placement.
The Paradigm System enables surgical navigation of instruments relative to the patient spinal anatomy by combining preoperative imaging segmentation and intraoperative registration of the patient’s live anatomy. Preoperative CT data is uploaded to the console and the system software generates a 3D model of the patient’s operative vertebrae. The 3D model is used for preoperative planning such as implant placements or instrument trajectories. During the procedure, once the patient’s anatomy is exposed, the Prism Sensor Array acquires imaging to match the topography of the live anatomy to the 3D model. This intraoperative registration of live anatomy to the 3D model creates a common coordinate space between the | Paradigm System fuses high-resolution, real-time camera imaging of the surgical site (such as the spine) with medical imagery (such as CT) and surgical navigation data, such as predetermined instrument trajectories for precise instrument placement.
The Paradigm System enables surgical navigation of instruments relative to the patient spinal anatomy by combining preoperative imaging segmentation and intraoperative registration of the patient’s live anatomy. Preoperative CT data is uploaded to the console and the system software generates a 3D model of the patient’s operative vertebrae. The 3D model is used for preoperative planning such as implant placements or instrument trajectories. During the procedure, once the patient’s anatomy is exposed, the Prism Sensor Array acquires imaging to match the topography of the live anatomy to the 3D model. This intraoperative registration of live anatomy to the 3D model creates a common coordinate space between the | Same |
Proprio, Inc.
Premarket Notification for the Paradigm System
Page 2 of 4
{6}
K250879 Page 3 of 4
| | Paradigm System
(Subject Device) | Paradigm System
(Predicate Device) | Comment |
| --- | --- | --- | --- |
| | preoperative CT data and the patient anatomy. This enables virtual information to be overlaid on the live anatomy during the procedure for intraoperative guidance. | preoperative CT data and the patient anatomy. This enables virtual information to be overlaid on the live anatomy during the procedure for intraoperative guidance. | |
| **Technical Comparison** | Stereotaxic image guided surgical navigation system during spine surgery. Preoperative CT data is segmented and a 3D model is generated. Preoperative planning can be performed. Intraoperative image capture using reflective markers on spinal instruments and registers the patient’s anatomy to the preoperative 3D model for intraoperative guidance during the procedure. | Stereotaxic image guided surgical navigation system during spine surgery. Preoperative CT data is segmented and a 3D model is generated. Preoperative planning can be performed. Intraoperative image capture using reflective markers on spinal instruments and registers the patient’s anatomy to the preoperative 3D model for intraoperative guidance during the procedure. | Same |
| **Components** | Cart, Arm, Sensor Array, External Monitor, Surgical Instruments, Software | Cart, Arm, Sensor Array, External Monitor, Surgical Instruments, Software | Same |
| **Instruments** | Dynamic Reference Frames, Awl, Pedicle Probes
Drill Guide, Drill Bits, Depth Cannulas, Trocar | Dynamic Reference Frames, Awl, Pedicle Probes | Identical intended use and identical materials |
| **Energy Source** | 120V | 120V | Same |
| **Sterilization Method – Surgical Instruments** | Steam | Steam | Same |
| **Minimum SAL** | 1 x 10^{-6} | 1 x 10^{-6} | Same |
| **Required Accessories** | Reflective Markers (Northern Digital K033621) | Reflective Markers (Northern Digital K033621) | Same |
| **Optional Accessories** | Paradigm Drape | Paradigm Drape | Same |
| **Sterilization Method – drape** | Ethylene Oxide | Ethylene Oxide | Same |
| **Minimum SAL** | 1 x 10^{-6} | 1 x 10^{-6} | Same |
Proprio, Inc.
Premarket Notification for the Paradigm System
Page 3 of 4
{7}
K250879 Page 4 of 4
| **Functional and Safety Testing:** | Verification and Validation activities have been conducted to provide assurance that the device meets the performance requirements under the indications for use conditions and to establish substantial equivalence with the predicate. Proprio performed the following testing to ensure the safety and effectiveness of the Paradigm System:
• Non-Clinical Verification Tests
• Non-Clinical Design Validation
• Compliance Conformity Assessments
The following standards were used in testing:
• IEC 62366-1: 2015/AMD1:2020 Medical Devices – Part 1: Application of Usability Engineering to Medical Devices, including Amendment 1
• ISO 10993-1:2018 Biological evaluation of medical devices: Part 1: Evaluation and testing within a risk management process
• ISO 14971:2019 Medical devices – Application of risk management to medical devices
• ISO 17665-1:2006 Sterilization of health care products – Moist heat – Part:1 Requirements for the development, validation and routine control of a sterilization process for medical device
• ANSI/AAMI ST79:2017 Comprehensive guide to steam sterilization and sterility assurance in health care facilities
• ASTM F2554-22 Standard Practice For Measurement Of Positional Accuracy Of Computer Assisted Surgical Systems
• ANSI/AAMI ST98:2022 Cleaning validation of healthcare products – Requirements for development and validation of a cleaning process for medical devices
• IEC 62304 Edition 1.1 2015-06 Medical device software – Software life cycle processes |
| --- | --- |
| **Conclusion:** | The Paradigm System intended use, indications for use, and fundamental scientific technology is the same as the predicate device. Performance, safety and usability testing demonstrate that the differences between the subject device and the predicate device do not raise new risks of safety and effectiveness. |
Proprio, Inc.
Premarket Notification for the Paradigm System
Page 4 of 4
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.