Valence Robotic Navigation System; Valence Robotic Navigation Instruments (For Use with StealthStation)
K252597 · Alphatec Spine, Inc. · OLO · Feb 19, 2026 · Neurology
Device Facts
Record ID
K252597
Device Name
Valence Robotic Navigation System; Valence Robotic Navigation Instruments (For Use with StealthStation)
Applicant
Alphatec Spine, Inc.
Product Code
OLO · Neurology
Decision Date
Feb 19, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
Valence Robotic Navigation System: Valence Robotic Navigation System is indicated for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of a tool holder or guide tube to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous spinal procedures in reference to rigid patient anatomy and fiducials that can be identified on a 3D imaging scan. The Valence Robotic Navigation System is indicated for assisting surgeons in spinal procedures, such as: • Pedicle screw placement • Interbody device placement Valence Robotic Navigation Instruments (for use with the StealthStation™): The Valence Posterior Disc Prep Instruments are indicated to facilitate discectomy, bony resection, and implant selection during spinal surgery. The Valence Posterior Fixation Instruments are intended to be used during the preparation and placement of Alphatec screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open, or minimally invasive, procedures. Navigated instruments are specifically designed for use with the Medtronic® StealthStation™ System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where reference to a rigid anatomical structure, such as vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.
Device Story
Image-guided system for spinal surgery; comprises computer workstation, software, trajectory system, targeting platform, camera, and navigated instruments. Inputs: intraoperative 3D scan data; optical tracking of instruments and patient anatomy. Operation: system registers instruments to patient anatomy; provides robotic-assisted or freehand navigation for pedicle screw placement and disc preparation. Used in OR by surgeons; provides real-time visual feedback on workstation display. Output: spatial positioning and orientation of tool holders/guide tubes. Clinical benefit: assists surgeon in precisely locating anatomical structures; facilitates accurate implant placement and bony resection; reduces reliance on manual landmarking.
Clinical Evidence
No clinical data provided. Evidence consists of nonclinical performance testing: software verification, hardware verification, accuracy verification testing, dimensional analysis of instruments, and simulated use testing using cadaveric specimens.
Technological Characteristics
Image-guided stereotaxic system; optical instrument tracking; computer workstation with integrated software; robotic trajectory system; compatible with 3D imaging scans. Instruments designed for spinal fixation and disc preparation. Connectivity: workstation-based navigation. Sterilization: reusable instruments.
Indications for Use
Indicated for patients undergoing open, percutaneous, or minimally invasive spinal procedures requiring pedicle screw placement, interbody device placement, discectomy, or bony resection. Applicable where stereotactic surgery is appropriate and rigid anatomical structures (vertebrae) can be referenced via 3D imaging, CT, MR, fluoroscopy, or digitized landmarks.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 19, 2026
Alphatec Spine Inc.
Garima Shrivastava
Senior Regulatory Affairs Specialist
1950 Camino Vida Roble
Carlsbad, California 92008
Re: K252597
Trade/Device Name: Valence Robotic Navigation System; Valence Robotic Navigation Instruments (For Use with StealthStation)
Regulation Number: 21 CFR 882.4560
Regulation Name: Stereotaxic Instrument
Regulatory Class: Class II
Product Code: OLO
Dated: January 20, 2026
Received: January 20, 2026
Dear Garima Shrivastava:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252597 - Garima Shrivastava
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE
{2}
K252597 - Garima Shrivastava
Page 3
by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali -S
Shumaya Ali, M.P.H.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252597 | |
| Device Name Valence Robotic Navigation System; Valence Robotic Navigation Instruments (For Use with StealthStation) | |
| Indications for Use (Describe) Valence Robotic Navigation System: Valence Robotic Navigation System is indicated for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of a tool holder or guide tube to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous spinal procedures in reference to rigid patient anatomy and fiducials that can be identified on a 3D imaging scan. The Valence Robotic Navigation System is indicated for assisting surgeons in spinal procedures, such as: • Pedicle screw placement • Interbody device placement | |
| Valence Robotic Navigation Instruments (for use with the StealthStation™): The Valence Posterior Disc Prep Instruments are indicated to facilitate discectomy, bony resection, and implant selection during spinal surgery. The Valence Posterior Fixation Instruments are intended to be used during the preparation and placement of Alphatec screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open, or minimally invasive, procedures. Navigated instruments are specifically designed for use with the Medtronic® StealthStation™ System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where reference to a rigid anatomical structure, such as vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
atec
Traditional 510(k) Premarket Notification
Valence™ Robotic Navigation System
This 510(k) summary of safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92.
A. SUBMITTER:
Alphatec Spine, Inc.
1950 Camino Vida Roble
Carlsbad, CA 92008
Phone : (760) 356-6783
Fax : (760) 431-0289
Contact Person:
Garima Shrivastava
Sr. Regulatory Affairs Specialist
Email: gshrivastava@atecspine.com
Date Summary Prepared:
August 15, 2025
B. DEVICE
Trade Name: Valence Robotic Navigation System
Common or Usual Name: Stereotaxic Instrument
Classification Name: Orthopedic stereotaxic instrument
Regulation Number: 21 CFR 882.4560
Regulatory Class: Class II
Product Code: OLO
LEGALLY MARKETED PREDICATE DEVICES
Primary Predicate:
| 510(k) | Product Name | Product Code | Clearance Date |
| --- | --- | --- | --- |
| K232413 | Fusion Robotic Navigation System | OLO | 09/08/2023 |
Page 1 of 4
{5}
atec
Traditional 510(k) Premarket Notification
Valence™ Robotic Navigation System
Additional Predicates:
| 510(k) | Product Name | Product Code | Clearance Date |
| --- | --- | --- | --- |
| K161210 | Medtronic Navigated Manual Reusable Instruments for use with the StealthStation™ | OLO, OUR, HWE | 08/12/2016 |
| K232345 | ATEC Posterior Navigated Disc Prep Instruments | OLO | 11/02/2023 |
| K240951 | Invictus Robotic Navigation Instruments | OLO | 06/06/2024 |
| K200936 | Invictus™ CT Spinal Fixation System | NKG | 05/22/2020 |
| K232275 | Invictus® Spinal Fixation System | NKB, KWP, KWQ, OUR, PML | 09/27/2023 |
| K234092 | SafeOp 3: Neural Informatix System | GWF, GXY, ETN, GXZ, PDQ, IKN | 04/19/2024 |
Reference Devices:
| 510(k) | Product Name | Product Code | Clearance Date |
| --- | --- | --- | --- |
| K211616 | ExcelsiusHub | OLO | 08/25/2021 |
| K192938 | Invictus™ Spinal Fixation System | NKB, KWP | 12/12/2019 |
| K161363 | Arsenal Spinal Fixation System | NKB, KWP, MNH, MNI, OSH | 06/10/2016 |
## DEVICE DESCRIPTION
The Valence Robotic Navigation System is an image guided system primarily comprised of a computer workstation, software, a trajectory system, including a targeting platform, a camera, and various image guided instruments intended for assisting the surgeon in placing pedicle screws and interbody implants in the thoracolumbar and sacral spine.
The system uses optical instrument tracking and registration to an intraoperative 3D scan to navigate patient anatomy across three independent workflows:
- Robotic assisted screw navigation
- Freehand screw navigation
- Freehand disc prep and interbody implant navigation
In addition to compatibility with the Valence Robotic Navigation System, select instruments referred to as Valence Robotic Navigation Instruments (for use with StealthStation™) are designed to be compatible with the Medtronic® StealthStation™ Surgical Navigation System. The navigated
Page 2 of 4
{6}
atec
Traditional 510(k) Premarket Notification
Valence™ Robotic Navigation System
disc preparation instruments are intended to facilitate discectomy, and bony resection while the posterior fixation instruments are intended to assist surgeons in precisely locating anatomical structures in either open, minimally invasive, or percutaneous procedures for preparation and placement of pedicle screw system implants.
## INDICATIONS FOR USE
### Valence Robotic Navigation System
The Valence Robotic Navigation System is indicated for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of a tool holder or guide tube to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous spinal procedures in reference to rigid patient anatomy and fiducials that can be identified on a 3D imaging scan.
The Valence Robotic Navigation System is indicated for assisting surgeons in spinal procedures, such as:
- Pedicle screw placement
- Interbody device placement.
### Valence Robotic Navigation Instruments (for Use with StealthStation™)
The Valence Posterior Disc Prep Instruments are indicated to facilitate discectomy, bony resection, and implant selection during spinal surgery.
The Valence Posterior Fixation Instruments are intended to be used during the preparation and placement of Alphatec screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open, or minimally invasive, procedures.
Navigation instruments are specifically designed for use with the Medtronic® StealthStation™ System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where reference to a rigid anatomical structure, such as vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.
## TECHNOLOGICAL COMPARISON TO PREDICATE
The subject device Valence Robotic Navigation System is substantially equivalent to the primary predicate, Fusion Robotic Navigation System (K232413). Additionally, the subject Valence Robotic Navigation Instruments (for use with Stealth Station) are substantially equivalent to Medtronic Navigated Manual Reusable Instruments for use with the StealthStation™ (K161210) and ATEC Posterior Navigated Disc Prep Instruments (K232345). The technological design features of the subject device were compared to the predicate devices in intended use, indications for use, design, function and technology and it was demonstrated that they are substantially equivalent. Any technological differences within this 510(k), between the subject device and the predicate device, does not impact substantial equivalence, or safety and effectiveness.
Page 3 of 4
{7}
atec
Traditional 510(k) Premarket Notification
Valence™ Robotic Navigation System
# PERFORMANCE DATA
The following nonclinical performance tests were performed to support the substantial equivalence of the subject Valence Robotic Navigation System to its predicates:
- Software Verification
- Hardware Verification
- Accuracy Verification Testing
- Dimensional analysis of the Valence Instruments for Use with StealthStation compared to Medtronic instruments
- Simulated use testing using cadaveric specimens
The results of these tests demonstrate that the subject Valence Robotic Navigation System is substantially equivalent to the primary predicate Fusion Robotic Navigation System (K232413) and additional predicates Medtronic Navigated Manual Reusable Instruments for use with the StealthStation™ (K161210) and ATEC Posterior Navigated Disc Prep Instruments (K232345).
# CONCLUSION
The information provided in this submission, and in the discussion in Section 3, Substantial Equivalence, demonstrates that the subject device does not pose additional risk to safety and effectiveness when compared to the predicate device. The subject device Valence Robotic Navigation System is substantially equivalent to the primary predicate, Fusion Robotic Navigation System (K232413) and additional predicates, ATEC Posterior Navigated Disc Prep Instruments (K232345) and Medtronic Navigated Manual Reusable Instruments for use with the StealthStation™ (K216120).
Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.