The Invictus Robotic Navigation Instruments are indicated for use during the placement of Invictus Spinal Fixation System screws in either open or minimally invasive procedures. The Invictus Robotic Navigation Instruments are specifically designed for use with Globus Medical's Excelsius GPS® Robotic Navigation Platform, which is intended for use as an aid for precisely locating anatomical structures and for the spationing and orientation of an instrument holder or guide tube to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous procedures, provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or fluoroscopy.
Device Story
Invictus Robotic Navigation Instruments are reusable surgical instruments designed for use with Globus Medical's Excelsius GPS Robotic Navigation Platform. Instruments facilitate screw placement during open or minimally invasive spinal surgery. System uses CT or fluoroscopy inputs to identify fiducial markers and rigid patient anatomy; platform provides spatial positioning and orientation for instrument holder or guide tube. Surgeon operates system to navigate and guide instruments. Output is physical guidance for precise screw placement. Benefits include improved accuracy in locating anatomical structures during spinal fixation. Instruments are provided non-sterile and require steam sterilization by end user.
Clinical Evidence
No clinical data. Substantial equivalence determined via bench testing, including dimensional analysis, geometric comparison, and cleaning/reprocessing validation.
Technological Characteristics
Reusable surgical instruments; non-sterile, steam-sterilizable. Designed for compatibility with Excelsius GPS Robotic Navigation Platform. Materials and design dimensions equivalent to predicate devices.
Indications for Use
Indicated for use during placement of Invictus Spinal Fixation System screws in open or minimally invasive spinal procedures. Designed for use with Excelsius GPS Robotic Navigation Platform for locating anatomical structures and guiding surgical instruments via CT or fluoroscopy.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 6, 2024
Alphatec Spine, Inc. Alanna Joshi Regulatory Affairs Specialist 1950 Camino Vida Roble Carlsbad, California 92008
Re: K240951
Trade/Device Name: Invictus Robotic Navigation Instruments Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: April 5, 2024 Received: April 8, 2024
Dear Alanna Joshi:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Shumava Ali -S
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K240951
Device Name Invictus Robotic Navigation Instruments
### Indications for Use (Describe)
The Invictus Robotic Navigation Instruments are indicated for use during the placement of Invictus Spinal Fixation System screws in either open or minimally invasive procedures. The Invictus Robotic Navigation Instruments are specifically designed for use with Globus Medical's Excelsius GPS® Robotic Navigation Platform, which is intended for use as an aid for precisely locating anatomical structures and for the spationing and orientation of an instrument holder or guide tube to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous procedures, provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or fluoroscopy.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------|--|
| <div> <span> <span style="font-size:16px">✓</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | |
| <div> <span> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for "atec". The "a" is green, while the "tec" is dark blue. There is a trademark symbol to the right of the "c". There is a black line underneath the logo.
This 510(k) summary of safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92.
| I. | SUBMITTER: | Alphatec Spine, Inc.<br>1950 Camino Vida Roble<br>Carlsbad, CA 92008<br>Phone: (760) 431-9286<br>Fax: (760) 431-0289 |
|----|------------------------|----------------------------------------------------------------------------------------------------------------------|
| | Contact Person: | Alanna Joshi<br>Regulatory Affairs Specialist |
| | Date Summary Prepared: | April 5, 2024 |
#### II. DEVICE
| Trade or Proprietary Name: | Invictus Robotic Navigation Instruments |
|----------------------------|-----------------------------------------|
| Common Name: | Navigation Instruments |
| Classification Name: | Orthopedic Stereotaxic Instrument |
| Regulation Number: | 21 CFR 882.4560 |
| Classification: | Class II |
| Product Code: | OLO |
#### III. LEGALLY MARKETED PREDICATE DEVICES
## Primary Predicate Device:
| 510(k) | Product Name | Clearance Date |
|---------|-----------------------------------------|-----------------|
| K232044 | Invictus Robotic Navigation Instruments | October 6, 2023 |
Additional Predicate Devices:
| 510(k) | Product Name | Clearance Date |
|---------|---------------------------------|--------------------|
| K232275 | Invictus Spinal Fixation System | September 27, 2023 |
| K171651 | EXCELSIUS GPS® | August 16, 2017 |
### Reference Devices:
| 510(k) | Product Name | Clearance Date |
|---------|-------------------------------------------|-----------------|
| K161363 | Arsenal Spinal Fixation System | June 10, 2016 |
| K203742 | IdentiTi ALIF Standalone Interbody System | March 18, 2021 |
| K083416 | Revolve Stabilization System | January 9, 2009 |
| K111449 | Revolve Stabilization System | August 17, 2011 |
| K124058 | CREO Stabilization System | May 13, 2023 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for "atec". The "a" is green, while the "tec" is dark blue. There is a horizontal line underneath the logo.
### IV. DEVICE DESCRIPTION
The Invictus Robotic Navigation Instruments are surgical instruments that is designed to be compatible with Globus Medical's Excelsius GPS® Robotic Navigation Platform. The subject device is intended to facilitate the placement of screws during spinal surgery. The subject device are reusable instruments and offered non-sterile to be cleaned and steam sterilized by the end user.
#### V. INDICATIONS FOR USE
The Invictus Robotic Navigation Instruments are indicated for use during the placement of Invictus Spinal Fixation System screws in either open or minimally invasive procedures. The Invictus Robotic Navigation Instruments are specifically designed for use with Globus Medical's Excelsius GPS® Robotic Navigation Platform, which is intended for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of an instrument holder or guide tube to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous procedures, provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or fluoroscopy.
#### VI. TECHNOLOGICAL COMPARISON TO PREDICATES
The subject Invictus Robotic Navigation Instruments have equivalent technological characteristics to that of the predicate devices and the minor differences do not raise any new issues of safety and effectiveness. Specifically, the following characteristics are similar between the subject and predicates: device design and dimensions, indications for use, materials of manufacture, and principles of operation. Identical to the predicates, the subject Invictus Robotic Navigation Instruments are also non-sterile, reusable instruments to be steam sterilized by the end user.
The technological design features of the subject instrument were compared to the predicates in intended use, indications for use, design, function and technology and it was demonstrated that they are substantially equivalent.
### Clinical Information
Not applicable; determination of substantial equivalence is not based on an assessment of clinical performance data.
#### VII. PERFORMANCE DATA
The Invictus Robotic Navigation Instruments have been evaluated through a dimensional analysis and geometric comparison to the predicate devices to establish the safety and effectiveness for accuracy performance.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for a company called "atec". The "a" is green and stylized, while the "tec" is dark blue. There is a trademark symbol next to the "c". A black line is underneath the logo.
The Invictus Robotic Navigation Instruments were compared to the predicate devices and found to be substantially equivalent in terms of cleaning and reprocessing.
The results of this engineering analysis show that the subject is substantially equivalent to the cleared predicate.
# VIII. CONCLUSION
Based upon the information provided in this 510(k) submission, it has been determined that the subject device is substantially equivalent to legally marketed devices in regard to indications for use, intended use, design, technology, and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.