Evolution Spine Navigation Instruments

K250167 · Evolution Spine · OLO · Aug 18, 2025 · Neurology

Device Facts

Record IDK250167
Device NameEvolution Spine Navigation Instruments
ApplicantEvolution Spine
Product CodeOLO · Neurology
Decision DateAug 18, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4560
Device ClassClass 2

Indications for Use

The Evolution Spine navigation instruments are indicated for use during the preparation and placement of Whistler Modular Pedicle Screw System polyaxial screws or Stowe Pedicle Screw System polyaxial screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The navigation instruments are intended to be used with the Medtronic® StealthStation® S8 Navigation System (Software Version 1.3.0), which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where reference to a rigid anatomical structure such as a vertebra can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.

Device Story

Evolution Spine Navigation Instruments are reusable surgical tools (awls, probes, taps, drivers) used in spinal surgery. Instruments interface with Medtronic StealthStation S8 Navigation System to assist surgeons in precise localization of anatomical structures and placement of specific pedicle screw systems. System inputs include CT/MR models, fluoroscopy images, or digitized anatomical landmarks. Navigation system tracks instrument position relative to rigid anatomical structures (vertebrae). Output is real-time visual guidance for the surgeon during screw placement. Used in OR by surgeons to improve accuracy of pedicle screw insertion, potentially reducing risk of malposition and improving patient outcomes. Instruments are stainless steel; designed exclusively for use with specified Medtronic hardware/software and Evolution Spine implants.

Clinical Evidence

Bench testing only. Performance evaluated per ASTM F2554-18 (Measurement of Positional Accuracy of Computer Assisted Surgical Systems). Worst-case instrument constructs met or exceeded performance criteria established in ASTM F2554-18 and Kamimura et al. No clinical data required.

Technological Characteristics

Reusable surgical instruments (awls, probes, taps, drivers) constructed of stainless steel per ASTM F899. Designed for use with Medtronic StealthStation S8 Navigation System. Connectivity is via integration with the navigation system hardware. Sterilization method not specified. No active electronic components in the instruments themselves; passive tracking markers used for navigation.

Indications for Use

Indicated for patients undergoing spinal surgery requiring preparation and placement of Whistler Modular or Stowe Pedicle Screw System polyaxial screws; assists surgeons in locating anatomical structures during open or minimally invasive procedures using stereotactic navigation.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION August 18, 2025 Evolution Spine James Mirda Principal Engineer 2300 N. Haskell Ave Dallas, Texas 75204 Re: K250167 Trade/Device Name: Evolution Spine Navigation Instruments Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: July 21, 2025 Received: July 21, 2025 Dear James Mirda: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K250167 - James Mirda Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE {2} K250167 - James Mirda Page 3 by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Shumaya Ali -S Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K250167 Device Name Evolution Spine Navigation Instruments Indications for Use (Describe) The Evolution Spine navigation instruments are indicated for use during the preparation and placement of Whistler Modular Pedicle Screw System polyaxial screws or Stowe Pedicle Screw System polyaxial screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The navigation instruments are intended to be used with the Medtronic® StealthStation® S8 Navigation System (Software Version 1.3.0), which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where reference to a rigid anatomical structure such as a vertebra can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4} 510(k) #: K250167 510(k) Summary Prepared on: 2025-08-13 # Contact Details 21 CFR 807.92(a)(1) | Applicant Name | Evolution Spine | | --- | --- | | Applicant Address | 2300 N. Haskell Ave Dallas TX 75204 United States | | Applicant Contact Telephone | 901-550-3377 | | Applicant Contact | Mr. James Mirda | | Applicant Contact Email | jmirda@evolutionspine.com | | Correspondent Name | Evolution Spine | | Correspondent Address | 2300 N. Haskell Ave Dallas TX 75204 United States | | Correspondent Contact Telephone | 9015503377 | | Correspondent Contact | Mr. James Mirda | | Correspondent Contact Email | jmirda@evolutionspine.com | # Device Name 21 CFR 807.92(a)(2) | Device Trade Name | Evolution Spine Navigation Instruments | | --- | --- | | Common Name | Stereotaxic instrument | | Classification Name | Orthopedic Stereotaxic Instrument | | Regulation Number | 882.4560 | | Product Code(s) | OLO | # Legally Marketed Predicate Devices 21 CFR 807.92(a)(3) | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | | --- | --- | --- | | K223649 | VERTICALE® Navigation Instruments | OLO | | K16120 | Medtronic Navigated Instruments | HWE | | K140454 | Navigated CD HORIZON® SOLERA® Screwdrivers and Taps | OLO | # Device Description Summary 21 CFR 807.92(a)(4) The Evolution Spine Navigation Instruments are reusable surgical instruments for use with the Medtronic® StealthStation® S8 Navigation System. The system is designed to assist surgeons in the precise localization of anatomical structures, preparation and placement of pedicle screw implants during spinal procedures. The Evolution Spine navigation Instruments include awls, probes, taps and drivers. The navigated instruments are to be used with Whistler Modular Pedicle Screw System (K182478) and Stowe Pedicle Screw System (K181554). All instruments are made of stainless steel per ASTM F899. The Evolution Spine navigation Instruments are not compatible with implants Page 1 of 2 {5} Page 2 of 2 from other manufacturers and are designed for use only with Medtronic® StealthStation® S8 Navigation System hardware and software. ## Intended Use/Indications for Use **21 CFR 807.92(a)(5)** The Evolution Spine navigation instruments are indicated for use during the preparation and placement of Whistler Modular Pedicle Screw System polyaxial screws or Stowe Pedicle Screw System polyaxial screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The navigation instruments are intended to be used with the Medtronic® StealthStation® S8 Navigation System (Software Version 1.3.0), which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where reference to a rigid anatomical structure such as a vertebra can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy. ## Indications for Use Comparison **21 CFR 807.92(a)(5)** The subject "Indications for Use" statement is nearly identical to those of the predicate devices with only minor changes that do not impact the indicated use of the devices. ## Technological Comparison **21 CFR 807.92(a)(6)** Evolution Spine Navigation Instruments is substantially equivalent to the predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to have substantially equivalent technological characteristics to the predicate devices through comparison of design, intended use, material composition, function, and range of sizes. ## Non-Clinical and/or Clinical Tests Summary & Conclusions **21 CFR 807.92(b)** The worst-case instruments constructs were tested per ASTM F2554-18, "Standard Practice for Measurement of Positional Accuracy of Computer Assisted Surgical Systems". The worst-case construct meets and exceeds all the performance criteria identified in Kamimura et al: Accurate pedicle screw insertion under the control of a computer-assisted guiding system: Laboratory test and clinical study and ASTM F2554-18. Based upon the test data presented and comparison of features, Evolution Spine has determined that the proposed System is substantially equivalent to previously cleared systems. Clinical testing was not necessary to demonstrate the substantial equivalence of the subject devices. A comparison of critical dimensions was conducted to support the substantial equivalence of the subject devices. Biocompatibility was addressed through use of identical manufacturing materials and processes to the devices cleared in K182478 and K181554. Based on the indications for use, technological characteristics, performance testing, and comparison to predicate devices, the subject Evolution Spine Navigation Instruments has been shown to be substantially equivalent to legally marketed predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...