K223649 · Silony Medical GmbH · OLO · Dec 22, 2022 · Neurology
Device Facts
Record ID
K223649
Device Name
VERTICALE® Navigation Instruments
Applicant
Silony Medical GmbH
Product Code
OLO · Neurology
Decision Date
Dec 22, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
Silony Medical Navigation Instruments are intended to be used during the preparation and placement of VERTICALE® pedicle screws during spinal surgery to assist the surgeon in precise locating anatomical structures in spinal procedures. These instruments are designed for use with stereotactic navigation system Medtronic StealthStation®, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.
Device Story
VERTICALE® Navigation Instruments are reusable surgical tools (awls, probes, taps, drivers, inserts) used in spinal surgery; designed for use with Medtronic StealthStation® navigation system. Instruments interface with navigation hardware to assist surgeons in precise localization of anatomical structures and placement of VERTICALE® pedicle screws. Surgeon operates instruments manually while navigation system tracks position relative to patient anatomy (CT/MR/fluoroscopy/digitized landmarks). Output is visual guidance on navigation display; assists surgeon in accurate implant placement; benefits patient through increased surgical precision. Used in OR setting.
Clinical Evidence
Bench testing only. Navigational accuracy verified per ASTM F2554-18. Technical evaluation demonstrated compatibility and functionality of new instruments by comparing against previously cleared VERTICALE® Navigation Instruments (K212007) and Medtronic Navigated Instruments (K161210). Results from K212007 verification are transferable to new components.
Technological Characteristics
Reusable surgical instruments (awls, probes, taps, drivers, inserts) made of stainless steel per ASTM F899. Designed for use with Medtronic StealthStation® stereotactic navigation system. Connectivity is physical/mechanical interface with navigation hardware. Sterilization method not specified, but instruments are reusable. Accuracy testing performed per ASTM F2554-18.
Indications for Use
Indicated for use during spinal surgery for preparation and placement of VERTICALE® pedicle screws to assist surgeons in precise localization of anatomical structures. Designed for use with Medtronic StealthStation® stereotactic navigation systems in patients where stereotactic surgery is appropriate and anatomical structures can be referenced via CT, MR, fluoroscopy, or digitized landmarks.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
VERTICALE® Posterior Spinal Fixation System (K171421)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
December 22, 2022
Silony Medical GmbH Lisa Schwarz Regulatory Affairs Manager Leinfelder Straße 60 Leinfelden-Echterdingen, Baden-Württemberg D-70771 Germany
Re: K223649
Trade/Device Name: VERTICALE® Navigation Instruments Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: December 6, 2022 Received: December 6, 2022
Dear Lisa Schwarz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jesse Muir -S
For: Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223649
Device Name VERTICALE® Navigation Instruments
#### Indications for Use (Describe)
Silony Medical Navigation Instruments are intended to be used during the preparation and placement of VERTICALE® pedicle screws during spinal surgery to assist the surgeon in precise locating anatomical structures. These instruments are designed for use with stereotactic navigation system Medtronic StealthStation®, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5.510(K) SUMMARY
| Submitter's Name: | Silony Medical GmbH |
|----------------------------|---------------------------------------------------------------------------------------|
| Submitter's Address: | Leinfelder Straße 60<br>D-70771 Leinfelden-Echterdingen, Germany |
| Submitter's Telephone: | +49 (0) 711-782 525 0 |
| Contact Person: | Lisa Schwarz<br>Silony Medical GmbH<br>+49711 78252515<br>LSchwarz@Silony-Medical.com |
| Date Summary was Prepared: | December 6th, 2022 |
| Trade or Proprietary Name: | VERTICALE® Navigation Instruments |
| Common or Usual Name: | Orthopedic Stereotaxic Instrument |
| Classification Name: | Stereotaxic instrument |
| Classification: | Class II per 21 CFR 882.4560 |
| Product Code: | OLO |
| Classification Panel: | Division of Orthopedic Devices |
## DEVICE DESCRIPTION:
Silony Medical's VERTICALE® navigation instruments are reusable surgical instruments for use with the Medtronic® StealthStation® Navigation System is designed to assist surgeons in the precise localization of anatomical structures, preparation and placement of pedicle screw implants during spinal procedures.
The VERTICALE® Navigation Instruments include awls, probes, taps, drivers and inserts. The VERTICALE® Navigation Instruments are to be used with the VERTICALE® Posterior Spinal Fixation System cleared under K171421.
All instruments are made of stainless steel per ASTM F899. The VERTICALE Navigation Instruments are not compatible with implants from other manufacturers and are designed for use only with Medtronic StealthStation Navigation System hardware and software.
## INTENDED USE / INDICATIONS FOR USE
Silony Medical Navigation Instruments are intended to be used during the preparation and placement of VERTICALE® pedicle screws during spinal surgery to assist the surgeon in precise locating anatomical structures in spinal procedures. These instruments are designed for use with stereotactic navigation system Medtronic StealthStation®, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.
{4}------------------------------------------------
# SUBSTANTIAL EQUIVALENCE SUMMARY
The technological characteristics of the subject VERTICALE Navigation Instruments components remain the same as, or similar to, the predicate VERTICALE Navigation Instruments (K212007) in regards to intended use, indications for use, design, manufacturing methods and fundamental technology. The purpose of this submission is to seek clearance for four additional VERTICALE Navigation instruments for spinal surgery.
| | | Table 5 Predicate Devices | |
|--|--|---------------------------|--|
|--|--|---------------------------|--|
| 510k Number | Trade or Proprietary or Model Name | Manufacturer | Predicate Type |
|----------------------------------------------------|---------------------------------------------|------------------------------------|---------------------|
| K212007 | VERTICALE® Navigation Instruments | Silony Medical GmbH | Primary |
| Predicate Device from the previous 510(k); K212007 | | | |
| K161210<br>(PRIMARY)/K143628/<br>K140454/K143375 | Medtronic Navigated Instruments | Medtronic Sofamore Danek USA, Inc. | Reference (K212007) |
| K171421 | VERTICALE® Posterior Spinal Fixation System | Silony Medical GmbH | Reference (K212007) |
## PERFORMANCE DATA
The VERTICALE Navigation Instruments have been tested per ASTM F2554-18. "Standard Practice for Measurement of Positional Accuracy of Computer Assisted Surgical Systems".
These performance data are also applicable for the new instruments.
The additional instruments have been developed to complement Silony Medical's already approved navigable instruments (VERTICALE® Navigation Instruments; K212007) and to fully cover the intended use of the VERTICALE® Posterior Spinal Fixation System ((K171421); Section 11).
By means of a technical evaluation, the compatibility and functionality of the additional navigable instrument for spine surgery was demonstrated. Data from the already cleared navigable instruments (VERTICALE® Navigation Instruments; K212007 and Medtronic Navieated Instruments; K161210) served as the basis for the technical evaluation. The technical evaluation relates to the navigational accuracy of the instruments in terms of position verification.
As result of the verification, relevant interfaces and functionalities between the existing and the new instruments are (essentially) equivalent to the existing instruments. Therefore, the results of the verification tests (K212007) are directly transferable to the new additional instruments and repetition of these tests is not necessary.
{5}------------------------------------------------
### RISK MANAGEMENT
Risk Assessment of the VERTICALE Navigation Instruments was performed in accordance with ISO 14971:2019, Medical Devices - Application of Risk Management to Medical Devices. The purpose of this analysis was to identify potential new hazards and failure modes introduced by the new devices. This risk analysis resulted in the inclusion of additional references in the labeling (surgical technique).
The previous verifications and risk controls for the predicate VERTICALE Navigation Instruments (K212007) remain valid and acceptable as the controls used are unchanged by the introduction of the new instrument components presented in this submission.
The overall residual risk is acceptable and the medical benefits continue to outweigh the overall residual risk.
The Special 510(k) Program issued by the FDA on September 13, 2019 is appropriate for this submission as the proposed change is being submitted by the legal manufacturer and a risk analysis format supports substantial equivalence
### SUBSTANTIAL EQUIVALENCE CONCLUSION
The new VERTICALE Navigation Instruments are substantially equivalent to the predicate system as a spinal fixation device in regards to intended use, indications for use, fundamental technology including design, materials, manufacturing methods, sterility, and general operational principles. Furthermore, accuracy testing and other supporting information sufficiently demonstrate the substantial equivalence of the subject components to the navigated instrument system which has been cleared for stereotactic guidance during orthopedic surgery procedures.
Based on this information, the subject device does not raise any new issues regarding the safety or efficacy when compared to its predicates (K212007).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.