K970604 · Philips Medical Systems (Cleveland), Inc. · HAW · May 19, 1997 · Neurology
Device Facts
Record ID
K970604
Device Name
VIEWPOINT - 3.0 SOFTWARE
Applicant
Philips Medical Systems (Cleveland), Inc.
Product Code
HAW · Neurology
Decision Date
May 19, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The ViewPoint is intended for use as a device which uses diagnostic images of the patient acquired specifically to assist the physician with presurgical planning and to provide orientation and reference information during intra-operative procedures.
Device Story
ViewPoint 3.0 is an image-assisted surgery system for presurgical planning and intra-operative orientation. It processes diagnostic patient images (CT/MR) to provide spatial reference. The system uses hand-held tools equipped with infrared-emitting diodes (IREDs) tracked by a position sensor assembly (two optical detectors). The software supports registration via scanned or anatomical fiducials. New features include a detector positioning guide to assist users in locating the center of the active digitizer volume (1m x 1m silo). Used in OR settings by surgeons; the system provides visual feedback via a graphical user interface (GUI) to assist in navigation during intracranial and spinal procedures. It supports accessories including drill guides, tracking devices, and a CT spine phantom. The system aids clinical decision-making by providing real-time orientation relative to pre-acquired diagnostic images.
Clinical Evidence
No clinical data provided; bench testing only. Substantial equivalence is based on technical comparison of system parameters, including tool accuracy (2.0-5.0 mm) and tracking methodology.
Technological Characteristics
System utilizes infrared-based optical tracking with IRED-equipped hand-held tools. Detector assembly uses two optical sensors. Software operates in a UNIX environment with a GUI for image manipulation (MPR, surface rendering). Complies with ASTM 10th Draft 9/94 Neurological Standard for Image-Interactive Stereotactic and Localization Systems. Connectivity includes support for diagnostic image import. System is designed for intra-operative use.
Indications for Use
Indicated for intra-cranial surgical procedures involving space occupying lesions or malformations (soft tissue, vascular, osseous) and spinal surgical procedures involving spinal stabilization, neural decompression, or resection of spinal neoplasms.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
K990860 — VIEWPOINT MICROSCOPE OPTION · Philips Medical Systems (Cleveland), Inc. · Mar 30, 1999
K990868 — VIEWPOINT PASSIVE TOOL OPTION · Philips Medical Systems (Cleveland), Inc. · Mar 30, 1999
K963221 — OPTICAL DIGITIZER OPTION FOR VIEW POINT · Philips Medical Systems (Cleveland), Inc. · Jan 10, 1997
K212245 — Spine and Trauma Navigation System, Spine & Trauma 3D Navigation, Instrument Selection, Fluoro 3D, Registration Software Fluoro 3D, Registration Software Paired Point, Registration Software Spine Surface Matching, Spine Planning, Elements Screw Planning Spine, Elements Spine Screw Planning · Brainlab AG · Apr 13, 2022
K023975 — VOYAGER LINUX · Z-Kat, Inc. · Dec 20, 2002
Submission Summary (Full Text)
{0}
K970604
# SUMMARY OF SAFETY AND EFFECTIVENESS
(As required by 21 CFR 807.92)
MAY 19 1997
## 1. General Information
Classification: Class II
Image Assisted Surgery Device
Common/Usual Name: Image Assisted Surgery Device Option
Proprietary Name: ViewPoint - 3.0 Operating Software
Establishment Registration: Picker International, Inc.
World Headquarters
595 Miner Road
Highland Heights, Ohio 44143
FDA Owner Number: #1580240
FDA Registration Number: #1525965
Performance Standards: No applicable performance standards have been issued under section 514 of the Food, Drug and Cosmetic Act.
The device does comply with the draft voluntary industry standard titled “Neurological Standard for Image-Interactive Stereotactic and Localization Systems (ASTM 10th Draft 9/94).”
## 2. Intended Use and Indications for Use
The intended use of the ViewPoint is unchanged by the 3.0 software, but the indications for use have been expanded to include spinal surgical procedures. The intended use and indications for use are as follows:
The ViewPoint is intended for use as a device which uses diagnostic images of the patient acquired specifically to assist the physician with presurgical planning and to provide orientation and reference information during intra-operative procedures.
The ViewPoint is indicated for use in:
- Intra-cranial surgical procedures involving space occupying lesions or malformations (including soft tissue, vascular and osseous)
- Spinal surgical procedures involving spinal stabilization, neural decompression, or resection of spinal neoplasms.
PICKER INTERNATIONAL, INC.
(VIEW30)
2/11/97
I - 1
{1}
# 3. Device Description
The new features in the 3.0 Software include a detector positioning feature and support for the following optional hardware accessories: a tracking device, drill guides and a CT spine phantom. The indications for use of the ViewPoint have also been expanded to include use in spinal surgeries.
# 4. Safety and Effectiveness
The ViewPoint System operating with the 3.0 software is substantially equivalent to the ViewPoint System described in the 510(k) submissions K961168 and K963221. The following chart has been compiled to demonstrate the substantial equivalence of the ViewPoint operating with the 3.0 software to the predicate device.
SUBSTANTIAL EQUIVALENCE CHART
| Parameter | Predicate Device ViewPoint & Optical Digitizer Option (K961168, K963221) | ViewPoint with 3.0 Software |
| --- | --- | --- |
| Tools | A long and short tool with a minimum of four IREDs per tool. (See K963221) | Same. |
| Average Tool Accuracy | 2.0 - 5.0 mm (See K961168) | Same. |
| Type of Detector | Infrared signals emitted from diodes on a hand-held tool are detected by a Position Sensor Assembly with two optical detectors. The assembly is either on a mobile pedestal, mounted to the OR table or mounted to the ceiling. (See K963221) | Same. |
| Active Digitizer Volume | Silo shape, with 1 meter diameter and 1 meter length. (See K963221) | Same, Detector Positioning Feature added to guide user in finding center of active digitizer volume. |
| Accessories | MR/CT Head Phantoms. (See K961168) | MR/CT Head Phantoms, CT Spine Phantom, Tracking device, Drill Guide. |
PICKER INTERNATIONAL, INC.
(VIEW30)
2/11/97
I - 2
{2}
| Parameter | Predicate Device ViewPoint & Optical Digitizer Option (K961168, K963221) | ViewPoint with 3.0 Software |
| --- | --- | --- |
| Registration Technique | Scanned Fiducials. (See K961168) | Scanned Fiducials and Anatomical Fiducials. |
| Operating Software
Structure | UNIX environment with three major processes: Import, Surgery Application and Foot Switch. Uses a Graphical User Interface to facilitate interaction with user. | Same structure. Modified Graphical User Interface with similar functionality. |
| Image Manipulation | MPR and surface rendering. | Same. |
| Other Features | None. (See K961168) | Detector Positioning Feature. |
| Intended Use | The ViewPoint is intended for use as a device which uses diagnostic images of the patient acquired specifically to assist the physician with presurgical planning and to provide orientation and reference information during intra-operative procedures. (See K961168) | Same. |
| Indications for Use | The ViewPoint is indicated for use in intra-cranial surgical procedures involving space occupying lesions or malformations (including soft tissue, vascular and osseous). (See K961168) | The ViewPoint is indicated for use in:
• Intra-cranial surgical procedures involving space occupying lesions or malformations (including soft tissue, vascular and osseous)
• Spinal surgical procedures involving spinal stabilization, neural decompression, or resection of spinal neoplasms. |
PICKER INTERNATIONAL, INC.
(VIEW30)
2/11/97
I - 3
{3}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
MAY 19 1997
Elaine K. Keeler, Ph.D.
Manager, Clinical Science
Picker International, Inc.
5500 Avion Park Drive
Highland Heights, Ohio 44143
Re: K970604
Trade Name: ViewPoint - 3.0 Operating Software
Regulatory Class: II
Product Code: 84HAW
Dated: February 14, 1997
Received: February 18, 1997
Dear Dr. Keeler:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}
Page 2 - Elaine K. Keeler, Ph.D.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular, Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{5}
Page 1 of 1
510(k) Number (if known): K970604
Device Name: ViewPoint - 3.0 Operating Software
Indications for Use:
The intended use of the ViewPoint is unchanged by the 3.0 software, but the indications for use have been expanded to include spinal surgical procedures. The intended use and indications for use are as follows:
The ViewPoint is intended for use as a device which uses diagnostic images of the patient acquired specifically to assist the physician with presurgical planning and to provide orientation and reference information during intra-operative procedures.
The ViewPoint is indicated for use in:
- Intra-cranial surgical procedures involving space occupying lesions or malformations (including soft tissue, vascular and osseous)
- Spinal surgical procedures involving spinal stabilization, neural decompression, or resection of spinal neoplasms.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Thomas J. Callahan
(in-ON)
Fluovascutor, Respiratory,
Devices
Prescription Use ☑ (Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐ (Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.