VYCOR SURGICAL ACCESS SYSTEMS

K060973 · Vycor Medical, LLC · GZT · Jul 26, 2006 · Neurology

Device Facts

Record IDK060973
Device NameVYCOR SURGICAL ACCESS SYSTEMS
ApplicantVycor Medical, LLC
Product CodeGZT · Neurology
Decision DateJul 26, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4800
Device ClassClass 2
AttributesTherapeutic, 3rd-Party Reviewed

Indications for Use

To provide for access and allow for visualization of the surgical field during brain and spinal surgery.

Device Story

Vycor ViewSite Surgical Access System comprises family of retractor devices for brain and spinal surgery. System includes introducer and transparent port; introducer spreads delicate tissue; port remains in place to provide hollow working channel for direct visualization of underlying anatomical structures. Devices are lightweight, non-conductive, and compatible with Lelya positioning clamps for fixation. Spinal access variant includes fixation screws for temporary attachment to surrounding bone. Used in OR by neurosurgeons; provides stable, clear access to target tissues; facilitates surgical procedures; reduces accidental movement via fixation.

Clinical Evidence

No clinical data. Device relies on biocompatibility testing and historical usage of materials.

Technological Characteristics

Transparent, non-conductive retractor system. Components: introducer, port, and fixation screws (spinal). Materials have history of medical use or passed biocompatibility testing. Sterilization via gamma radiation. Compatible with Lelya positioning clamps.

Indications for Use

Indicated for patients undergoing brain or spinal surgery requiring surgical access and visualization of the surgical field.

Regulatory Classification

Identification

A self-retaining retractor for neurosurgery is a self-locking device used to hold the edges of a wound open during neurosurgery.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ bo973 # 510 (k) Summary ## Date Prepared [21 CFR 807.92(a)(1)] Revised July 20, 2006 ## Submitter's Information [21 CFR 807.92(a)(1)] This 510(k) is being submitted by Joseph Azary on behalf of Vycor Medical LLC Joseph Azary can be contacted by telephone at (203) 944-9320 or fax at (203) 944-9317. Mailing address; 543 Long Hill Avenue, Shelton, CT 06484. Vycor Medical LLC is located at 40 Chardonnay Drive, Coram, NY 11727. The establishment registration for Vycor Medical is pending. ## Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)] Possible device trade names are: - . Vycor ViewSite Access System - � Vycor ViewSite Brain Access System (VBAS) - Vycor ViewSite Spinal Access System (VSAS) . ## Device Common, Usual, or Classification Names Retractor, Retractor for Neurosurgery, Brain Retractor, Self Retaining Retractor #### Classification Panel Classification of this device would fall under the responsibility of the Division of General, Restorative and Neurological Devices. #### Class Classification: Class 2 Product Code: GZT, 21 CFR 882.4800 ## Description of the Device [21 CFR 807.92(a)(4)] The Vycor Medical Surgical Access Systems include a family of retractor devices of varying shapes and sizes designed for providing diagnostic and surgical access to various portions of the brain and spinal region. The variety of devices include: - Brain Access System including three models of varying shapes and sizes. . - . TC model (elliptical diameters 2.1cm or 1.7cm, lengths 3,5,7, or 9 cm) - EC model (3.4 cm diameter, 3,5, 7 cm lengths) . - . SF model (channel 2.2cm or 2.5cm, length 3,5, or 7 cm) - . Spinal Access System - (aperture widths are 2.6 to 5.4 cm, and 3 or 5cm access channel height) . Each device consists of an introducer and port. The port and introducer are packaged assembled and ready for use. Upon insertion of the device, the introducer is removed and the port is left in place. The introducer has a length greater than the port. The smooth and soft tapered introducer works to spread JUL 26 2006 {2}------------------------------------------------ apart the brain or other portions of delicate tissue. Upon removal of the introducer, the port provides a hollow working channel allowing the surgeon access to the target tissues. The port is transparent to allow direct visualization of the underlying retracted anatomical structures. The retractors are light weight, non-conductive, and are compatible with Lelya positioning clamps to provide fixation and reduce accidental movement. The spinal access device can access the spine for surgical procedures. The spinal access device has three components including the introducer, port, and fixation screws. The fixation screws are used to hold the device in place during the surgery. The fixation screws are temporarily screwed into surrounding bone tissue during the surgical procedure. #### Packaging The devices will be packaged in a heat sealed off the shelf low density polyethylene pouch. The types of pouches have been utilized for medical devices for many years and are compatible with gamma radiation sterilization. The pouches will be placed in different sized labeled boxes depending on the model. #### Intended Use [21 CFR 807.92(a)(5)] To provide for access and allow for visualization of the surgical field during brain and spinal surgery. ## Technological Characteristics [21 CFR 807.92(a)(6)] Vycor Medical LLC believes that the subject device is substantially equivalent to the predicate device. ## Performance Data [21 CFR 807.92(b)(1)] The device complies is composed of materials that either have a long history of usage in the medical community or have passed biocompatibility testing requirements. ## Predicate Device [21 CFR 807.92(a)(3)] The predicate devices are listed as follows: For Spinal Indications: - � Medtronic - Mast Quadrant Retractor System - K043602 For Brain Indications: - V. Mueller -- Brain Retractor -- K895395 . - . Budde Halo Retractor - K830332 - Compass Stereotactic System -- Retractor K896156 ● #### Conclusion [21 CFR 807.92(b)(3)] We believe the differences between the subject device and predicate device are minor and conclude that the subject devices are as safe and effective as the predicate devices. {3}------------------------------------------------ What about the cohesive gel? The question about 18 years for saline and 22 years for silicone is a problem. I can think of no The question about To years for saillio and 22 your for cliforne is a predictions of silent rupture..." . . : {4}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and wings. The eagle is facing to the left. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUL 26 2006 Vycor Medical LLC % TUV Rheinland of North America. Inc. Mr. Tamas Borsai 12 Commerce Road Newton, Connecticut 06470 Re: K060973 Trade/Device Name: Vycor ViewSite Surgical Access System Regulation Number: 21 CFR 882.4800 Regulation Name: Self-retaining retractor for neurosurgery Regulatory Class: II Product Code: GZT Dated: July 7, 2006 Received: July 11, 2006 Dear Mr. Borsai We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Codc of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements. including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {5}------------------------------------------------ Page 2 - Mr. Tamas Borsai This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Hubert Leinweber Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ## Indications for Use 大060973 510(k) Number (if known): Vycor ViewSite Surgical Access System Device Name: Indications For Use: To provide for access and allow for visualization of the surgical field during brain and spinal surgery. Prescription Use × (Part 21 CFR 801 Subpart D) 4 AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Page 1 of 1 Division of General, Restorative, and Neurological Devices | 510(k) Number | K060973 | |---------------|---------| |---------------|---------|
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...