K043602 · Medtronic Sofamor Danek · GAD · Feb 23, 2005 · General, Plastic Surgery
Device Facts
Record ID
K043602
Device Name
QUADRANT RETRACTOR SYSTEM
Applicant
Medtronic Sofamor Danek
Product Code
GAD · General, Plastic Surgery
Decision Date
Feb 23, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4800
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The MAST QUADRANT™ Retractor System is intended to provide surgeons with instruments such as dilators, retractors, light sources and pedicle access needles used to perform a variety of spinal fixation procedures utilizing a minimally invasive approach. The MAST QUADRANT™ Retractor System is indicated for visualization of the surgical field in any area of the body cut open during a surgical procedure. When used in the cervical, thoracic, or lumbar spine either from an anterior or posterior direction, for example, the MAST QUADRANT™ Retractors and accessories are intended to aid the surgeon's visualization of the surgical area and allow him/her to perform any type of surgical spinal procedure such as herniated disc repair, visualization of the circumferential decompression of the nerve roots, aiding in the search and removal of nucleus material, spinal fusion, or insertion of spinal implants.
Device Story
MAST QUADRANT™ Retractor System is a tubular-based surgical instrument set for minimally invasive spinal access. System components include dilators, articulating/distracting retractor blades, light sources, and pedicle access needles. Operated by surgeons in an OR setting to dilate overlying tissues and maintain surgical field access. Compatible with external visualization aids like microscopes, cameras, and light sources. By providing a stable, adjustable portal to the spine, the system facilitates procedures such as decompression, discectomy, and fusion. Benefits include improved visualization and access to the surgical site through a minimally invasive approach, potentially reducing tissue trauma compared to traditional open surgery.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Tubular-based manual surgical retractor system. Components include dilators, articulating blades, light sources, and pedicle access needles. Designed for use with external visualization aids (microscopes, cameras). Non-powered, manual instrumentation.
Indications for Use
Indicated for visualization of the surgical field in any body area during surgery. Specifically for cervical, thoracic, or lumbar spine procedures (anterior or posterior approach) including herniated disc repair, nerve root decompression, nucleus material removal, spinal fusion, or spinal implant insertion.
Regulatory Classification
Identification
A manual surgical instrument for general use is a nonpowered, hand-held, or hand manipulated device, either reusable or disposable, intended to be used in various general surgical procedures. The device includes the applicator, clip applier, biopsy brush, manual dermabrasion brush, scrub brush, cannula, ligature carrier, chisel, clamp, contractor, curette, cutter, dissector, elevator, skin graft expander, file, forceps, gouge, instrument guide, needle guide, hammer, hemostat, amputation hook, ligature passing and knot-tying instrument, knife, mallet, disposable or reusable aspiration and injection needle, disposable or reusable suturing needle, osteotome, pliers, rasp, retainer, retractor, saw, scalpel blade, scalpel handle, one-piece scalpel, snare, spatula, stapler, disposable or reusable stripper, stylet, suturing apparatus for the stomach and intestine, measuring tape, and calipers. A surgical instrument that has specialized uses in a specific medical specialty is classified in separate regulations in parts 868 through 892 of this chapter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a date stamp that reads "FEB 9 2003". The month is abbreviated to "FEB", followed by the day "9", and the year "2003". The stamp appears to be somewhat faded or of low resolution, but the date is still legible.
## MAST QUADRANT™ Retractor System 510(k) Summary February 2005 K043602
- Medtronic Sofamor Danek Company: I. 1800 Pyramid Place Memphis, TN 38132 (901) 396-3133
Richard W. Treharne, PhD Contact: Senior Vice President Regulatory Affairs
#### Proprietary Trade Name: MAST QUADRANT™ Retractor System II.
- Classification Name: Retractors, Manual Surgical Instrument III.
- Regulation Number: Class I Exempt as described in 21 CFR 878.4800 and 888.1100. IV. Product Code GAD, NBH
#### V. Product Description
The MAST QUADRANT™ Retractor System is a tubular-based retraction system, designed to provide surgeons with the freedom to retract tissue through any combination of distracting or articulating the blades. The MAST QUADRANT™ System includes instruments used to access the spine by dilating the overlying tissues, as well serving as a retracting device to maintain the access. The system can be used in conjunction with microscopes, light sources, cameras, or other visualization aids.
#### VII Indications
The MAST QUADRANT™ Retractor System is intended to provide surgeons with instruments such as dilators, retractors, light sources and pedicle access needles used to perform a variety of spinal fixation procedures utilizing a minimally invasive approach.
The MAST QUADRANT™ Retractor System is indicated for visualization of the surgical field in any area of the body cut open during a surgical procedure. When used in the cervical, thoracic, or lumbar spine either from an anterior or posterior direction, for example, the MAST QUADRANT™ Retractors and accessories are intended to aid the surgeon's visualization of the surgical area and allow him/her to perform any type of surgical spinal procedure such as herniated disc repair, visualization of the circumferential decompression of the nerve roots, aiding in the search and removal of nucleus material, spinal fusion, or insertion of spinal implants.
{1}------------------------------------------------
K043602
of 2
P...
#### Substantial Equivalence VIII
Documentation was provided which demonstrated the subject MAST QUADRANT™ Retractors to be substantially equivalent to the METRx, MED, Inclusive, and/or INCL Microscopes described in Medtronic Sofamor Danek 510(k) No. K002931 (SE 11/24/00), while the pedicle access needles included in the system are identical to those cleared in the EQUESTRA™ Fluid Delivery System (K040483, SE 07/23/04).
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
FEB 2 3 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Richard W. Treharne, Ph.D. Senior Vice President, Regulatory Affairs Medtronic Sofamor Danek 1800 Pyramid Place Memphis, Tennessee 38132
Re: K043602
Trade/Device Name: MAST QUADRANTTM Retractor System Regulation Number: 21 CFR 878.4580 Regulation Name: Surgical Lamp Regulatory Class: II Product Code: FSZ, GAD, NBH Dated: February 2, 2005 Received: February 4, 2005
Dear Dr. Treharne:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Dr. Treharne
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
Miriam C Provost
Celia M. Witten, PhD., MD Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# KOA 3607
Page 1 of 1 December 2004
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
MAST QUADRANTTM Retractor System _____________________________________________________________________________________________________________________________________________ Device Name: _________________________________________________________________________________________________________________________________________________________________
### Indications for Use:
The MAST OUADRANT™ Retractor System is intended to provide surgeons with instruments such as dilators, retractors, light sources and pedicle access needles used to perform a variety of spinal fixation procedures utilizing a minimally invasive approach.
The MAST QUADRANT™ Retractor System is indicated for visualization of the surgical field in any area of the body cut open during a surgical procedure. When used in the cervical, thoracic, or lumbar spine either from an anterior or posterior direction, for example, the MAST QUADRANT™ Retractors and accessories are intended to aid the surgeon's visualization of the surgical area and allow him/her to perform any type of surgical spinal procedure such as herniated disc repair, visualization of the circumferential decompression of the nerve roots, aiding in the search and removal of nucleus material, spinal fusion, or insertion of spinal implants.
Prescription Use AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Muriam C. Provost
(Division Sign-Off) Division of General, Restorative. and Neurological Devices
K043602 510(k) Number ________________________________________________________________________________________________________________________________________________________________
0000000
Pano. | at |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.