TECA (TM) ELITE DISPOSABLE MONOPOLAR NEEDLES

K111131 · Carefusion 209, Inc. · IKT · Jun 17, 2011 · Physical Medicine

Device Facts

Record IDK111131
Device NameTECA (TM) ELITE DISPOSABLE MONOPOLAR NEEDLES
ApplicantCarefusion 209, Inc.
Product CodeIKT · Physical Medicine
Decision DateJun 17, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 890.1385
Device ClassClass 2

Indications for Use

TECA™ elite Disposable Monopolar Needles are intended for use with recording, monitoring and stimulation/recording equipment for the stimulation/recording of biopotential signals including electromyograph (EMG) and nerve potential signals.

Device Story

TECA™ elite Disposable Monopolar Needles are sterile, single-use passive electrodes; stainless steel, platinum alloy, or gold-plated stainless steel construction; PTFE-coated with low-friction lubricant. Needles insert into patient muscle to sense subcutaneous bioelectrical activity or stimulate nerve/muscular response. Device connects to external electromyographic (EMG) equipment via cable. EMG device detects, processes, and displays biopotential information to clinician. Used in clinical settings by healthcare professionals. Modifications include reformulated PTFE coating, added lubricant, and updated tip geometry. No electrical hardware or software components.

Clinical Evidence

Bench testing only. No clinical data. Performance validated via verification of coating, tip geometry, insertion force, and recording characteristics (noise/impedance) per ISO 7864. Biocompatibility verified per ISO 10993-1.

Technological Characteristics

Materials: stainless steel, platinum alloy, gold-plated stainless steel; PTFE insulator; low-friction lubricant. Passive device; no energy source. Dimensions/form factor: needle electrode. Connectivity: wired connection to external EMG equipment. Sterilization: sterile, single-use. No software or electrical hardware.

Indications for Use

Indicated for patients requiring stimulation or recording of biopotential signals, including EMG and nerve potential signals, via subcutaneous needle insertion.

Regulatory Classification

Identification

A diagnostic electromyograph needle electrode is a monopolar or bipolar needle intended to be inserted into muscle or nerve tissue to sense bioelectrical signals. The device is intended for medical purposes for use in connection with electromyography (recording the intrinsic electrical properties of skeletal muscle).

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # k 111131 JUN 1 7 2011 # Special 510(k) Premarket Notification CareFusion 209, Inc. TECA™ ëlitë Disposable Monopolar Needles 510(k) Summary | Submission Date: | 18 April 2011 | | | |--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|----------------------------------------------------------------------------------| | Submitter: | CareFusion 209, Inc.<br>1850 Deming Way<br>Middleton, WI 53562 | | | | Submitter Contact: | Mr. Glen Hermanson<br>Regulatory Affairs Manager - Manufacturing<br>CareFusion 209, Inc.<br>1850 Deming Way<br>Middleton, WI 53562<br>+1 (608) 829 8608<br>+1 (608) 829 8737 (fax)<br>Glen.Hermanson@CareFusion.com | | | | Manufacturing Site: | CareFusion Manufacturing Ireland 241, Ltd.<br>IDA Business Park<br>Gort, Co. Galway<br>Ireland | | | | Trade Name: | TECA™ elite Disposable Monopolar Needles | | | | Classification<br>Name: | Diagnostic electromyograph needle electrode | | | | Classification<br>Regulation: | 21 CFR §890.1385; 21 CFR §882.1350 | | | | Product Code: | IKT; GXZ | | | | Substantially<br>Equivalent Devices: | New CareFusion<br>Model | 510(k)<br>Number | Predicate<br>Manufacturer / Model | | | TECA™ elite<br>Disposable Monopolar<br>Needles | K011818 | Nicolet Biomedical, Inc.(now<br>owned by CareFusion) / Sterile<br>EMG Electrodes | {1}------------------------------------------------ # kulli 31 #### Special 510(k) Premarket Notification CareFusion 209, Inc. TECA™ elite Disposable Monopolar Needles 510(k) Summary . | Device Description: | CareFusion TECAT™ elite Disposable Monopolar Needles (TECATM elite Needles) are disposable, single use, sterile products intended to sense subcutaneous bioelectrical activity, or to stimulate nerve or muscular response by insertion into the patients muscle(s) through the skin where the recording area can differentiate between individual motor units. TECATM elite Needles are connected to an electromyographic (EMG) device through a cable. The EMG device detects and displays the biopotential information to the user.<br><br>TECAT™ elite Needles are stainless steel, platinum alloy, and/or gold plated stainless steel, and coated with a polytetrafluoroethylene (PTFE) insulator. A low-friction lubricant is applied to the exterior surface of the needle. TECAT™ elite Needles are passive devices, and do not contain electrical hardware components or software.<br><br>The following device modifications were made to the TECAT™ elite Needles: Reformulation of the PTFE exterior needle coating; Addition of a low friction lubricant to the exterior surface of the needle; and Modification of the needle tip geometry. | |--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use: | TECAT™ elite Disposable Monopolar Needles are intended for use with recording, monitoring and stimulation/recording equipment for the stimulation/recording of biopotential signals including electromyograph (EMG) and nerve potential signals. | | Technology Comparison: | TECAT™ elite Needles employ the same technological characteristics as the predicate device. | | Biocompatibility | Low friction lubricant used on the exterior surface of the TECAT™ elite<br>Needles is qualified by predicate use of an equivalent low friction<br>lubricant used on the following devices:<br>K092973, Bionen s.a.s Disposable Concentric Needle Electrode;<br>Disposable Monopolar/Subdermal Needle Electrode; and<br>Disposable Monopolar Needle Electrode K093109, Novo Nordisk Inc. NovaTwist® needle. | | | Biocompatibility verification was performed on direct patient contact<br>materials comprising the TECAT™ elite Needles that were affected by<br>the device modifications in accordance with and the applicable portions<br>of the following Standards:<br>ISO 10993-1: 2009, Biological evaluation of medical devices –<br>Part 1: Evaluation and testing within a risk management process. | | | Verification results and device material comparisons indicated that the<br>TECAT™ elite Needles comply with their predetermined specifications,<br>and with the applicable portions of the Standard. | | Performance Testing<br>- Bench | Performance testing was performed on device characteristics of the<br>TECAT™ elite Needles that were affected by the device modifications in<br>accordance with internal requirements and the applicable portions of<br>the following Standard:<br>ISO 7864: 1993, Sterile hypodermic needles for single use | | | Performance testing related to internal requirements consisted of:<br>Verification of coating performance; Verification of tip geometry; Validation of insertion force; and Validation of recording characteristics including noise and<br>impedance testing. | | | Verification and validation results indicated that the TECAT™ elite<br>Needles comply with their predetermined specification and with the | : " : ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴ {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows a series of dots and dashes, which could represent Morse code or some other form of encoded information. The dots and dashes are arranged in a horizontal line across the image. Below the line of dots and dashes, there is a number 219. The dots and dashes are spaced evenly apart. llll l 3 l # Special 510(k) Premarket Notification CareFusion 209, Inc. TECA™ elite Disposable Monopolar Needles 510(k) Summary #### Summary of Performance Testing: applicable Standard. : . ・・ Page | 3 {3}------------------------------------------------ # k 11113 l Page | 4 ### Special 510(k) Premarket Notification CareFusion 209, Inc. TECATM elite Disposable Monopolar Needles 510(k) Summary --- Conclusion Verification and validation activities were conducted to establish the performance and safety characteristics of the device modifications made to the TECA™ elite Needles. The results of these activities demonstrate that the TECA™ elite Needles are safe and effective when used in accordance with the intended use and labeling. Therefore, the TECA™ elite Needles are considered substantially equivalent to the predicate device. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular fashion around the symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 JUN 1 7 2011 CareFusion 209, Inc. c/o Glen Hermanson Regulatory Affairs Manager 1850 Deming Way Middleton, WI 53562 Re: K111131 Trade/Device Name: TECA ™ Elite Disposable Monopolar Needles Regulation Number: 21 CFR 890.1385 Regulation Name: Diagnostic electromyograph needle electrode Regulatory Class: Class II Product Code: IKT: GXZ Dated: May 25, 2011 Received: May 26, 2011 Dear Mr. Hermanson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {5}------------------------------------------------ Page 2 – Mr. Hermanson CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Debra Falls Malvina B. Eydelman. M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health {6}------------------------------------------------ Indications for Use 510(k) Number (if known): # K | | | | 3 | Device Name: Indications for Use: TECA™ elite Disposable Monopolar Needles TECA™ elite Disposable Monopolar Needles are intended for use with recording, monitoring and stimulation/recording equipment for the stimulation/recording of biopotential signals including electromyograph (EMG) and nerve potential signals. and the comments of the comments of the comments of Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use AND/OR (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) # ALEx BAILEY (Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices 510(k) Number_KULLI 3 K Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...