SudoC

K152216 · Ld Technology, LLC · GZO · Sep 24, 2015 · Neurology

Device Facts

Record IDK152216
Device NameSudoC
ApplicantLd Technology, LLC
Product CodeGZO · Neurology
Decision DateSep 24, 2015
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 882.1540
Device ClassClass 2

Indications for Use

SudoC is a Galvanic skin response measurement device related to the function of the sweat glands. The SudoC provides values. It is the physician's responsibility to make proper judgments based on these numbers. The device is indicated for use in general adult population. Prescription Use: Federal law restricts this device to sale by or on the order of a physician.

Device Story

SudoC is a programmable electro-medical system measuring galvanic skin response (conductance) via sweat glands. Input consists of electrical signals acquired from the soles of the feet using four conductive disposable cloth electrodes. The system includes a USB-connected electronic box and software installed on a computer. The device applies a 1.28V electrical output to the skin; measures conductance in the range of 1 to 120 micro Siemens over a 120-second acquisition duration. The software controls the measurement sequence between feet. The physician reviews the output values to make clinical judgments. The device is intended for use in a clinical setting under physician order. It benefits patients by providing quantitative data on sweat gland function.

Clinical Evidence

Bench testing only. Evidence includes software verification (SRS/SDS/STD/STR), design control activities, and a comparative study of conductance values between stainless steel electrodes and the new conductive disposable cloth electrodes. No clinical data was required as the fundamental technology and hardware remain unchanged.

Technological Characteristics

Programmable electro-medical system; USB-connected electronic box; 4 conductive disposable cloth electrodes; 1.28V electrical output; 1-120 micro Siemens conductance range; 120s acquisition duration. Standards: IEC 60601-1-1, IEC 60601-1-2. Software-controlled measurement sequence. Sterilization: Electrodes are single-use; cables/wires cleaned with 70-90% ethyl or isopropyl alcohol.

Indications for Use

Indicated for the general adult population for the measurement of galvanic skin response related to sweat gland function.

Regulatory Classification

Identification

A galvanic skin response measurement device is a device used to determine autonomic responses as psychological indicators by measuring the electrical resistance of the skin and the tissue path between two electrodes applied to the skin.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 882.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 24, 2015 LD Technology LLC Albert Maarek CEO 100 N. Biscayne Blvd. Suite 502 Miami. Florida 33132 Re: K152216 Trade/Device Name: SudoC Regulation Number: 21 CFR 882.1540 Regulation Name: Galvanic Skin Response Measurement Device Regulatory Class: Class II Product Code: GZO Dated: July 28, 2015 Received: August 7, 2015 Dear Albert Maarek: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Michael J. Hoffmann -S for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K152216 Device Name SudoC Indications for Use (Describe) SudoC device is a medical device for the measurement of galvanic skin response related to the sweat glands. The SudoC provides values. It is the physician's responsibility to make proper judgments based on these numbers. The device is indicated for use in general adult population. Prescription Use: Federal law restricts this device to sale by or on the order of a physician. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ L.D Technology LLC. Image /page/3/Picture/1 description: The image contains the logo for LD Technology. The logo consists of a green plant inside of a circle on the left, and the text "LD Technology" on the right. The text is in a light green color and the word "Technology" is italicized. 510(k) Premarket Notification Number: Special 510k number k152216 Preparation date: July 28, 2015 # Special 510(k) Summary SudoC #### This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.87 - 1. Device Trade Name of the device: SudoC Device Common name: Galvanic Skin Response Regulation number: 21 CFR 882.1540 Product Code: GZO Classification: Class II Classification Advisory: Neurology Submitter's Identification: 2. Manufacturer: L.D TECHNOLOGY LLC CEO of LD Technology: Albert MAAREK Address: L.D Technology 100 N. Biscayne Blvd, Suite 502 Miami, FL, 33132, USA Tel: 305-379-9900 Fax: 305 397 1115 Email: albert.ldteck@gmail.com #### 3. Predicate legally marketed (unmodified) device Trade name: SudoPath, Common name: Galvanic skin response 510K number K131568 Applicant and Manufacturer: LD TECHNOLOGY LLC (Same as new device) Product code GZO. #### 4. Intended use SudoC is a Galvanic skin response measurement device related to the function of the sweat glands. The SudoC provides values. It is the physician's responsibility to make proper judgments based on these numbers. The device is indicated for use in general adult population Prescription Use: Federal law restricts this device to sale by or on the order of a physician. #### 5. Device Description and Comparison Devices' comparison Table modified SudoPath / unmodified SudoPath 13 {4}------------------------------------------------ | Specifications | SudoC | SudoPath k 131568 | |--------------------------------------------------|-------------------------------------------------------------------------------------------|---------------------------------------------------------------------------| | Intended use | Galvanic Skin Response measurement device related to the function of the sweat glands | Galvanic Skin Response device measurement | | Prescription for use | Physician | Physician | | Hardware | Same Hardware with exactly the same specification | SudoPath Hardware | | Material used in contact with the patient | Conductive disposable cloth electrode | Stainless steel | | Fundamental Scientific Technology | Sympathetic skin potential response and sweat glands | Sympathetic skin potential response and sweat glands | | Number of electrodes in contact with the patient | 4 | 4 | | Sequence of measurement | Software control and sequence between feet. | Software control and sequence between feet and hands | | Time of measurement for each pathway | Software control , each pathway is measured during 30 seconds | Software control , each pathway is measured during 30 seconds | | Cleaning and disinfection | Ethyl or isopropyl alcohol (70-90%) for lead wires and cables. Electrodes are single use. | Ethyl or isopropyl alcohol (70-90%) for electrodes, lead wires and cables | | Standards met | IEC 60601-1 -1<br>IEC 60601-1-2 | IEC 60601-1-1<br>IEC60601-1-2 | | Range of conductance measurements | 1 to 120 micro Siemens | 1 to 120 micro Siemens | | Data acquisition duration | 120s | 120s | | electrical output to skin | 1.28V | 1.28V | | active surface area of electrode | 44 cm2 for each disposable electrode placed on the feet | 272 cm2 for the hand plates<br>330 cm2 for the foot plates | | current density at electrodes | < 0.01 UA/mm2 | < 0.01 UA/mm2 | # Type of device The SudoC is a programmable electro medical system including: - USB plug and play hardware device including an electronic box, 2 reusable cables to connect . the box to electrodes and 2 tactile electrodes placed on the sole of each foot. - Software installed on a computer. ● As a galvanic skin response measurement device, it measures the skin resistance (i.e., conductance). # Comparison with the legally marketed (unmodified) device: The submission is complying with the Items required under §807.87 Similarities: The SudoC has the following similarities to unmodified SudoPath which has previously received 510(k) clearance: {5}------------------------------------------------ - has the same intended use, - uses the same Hardware and not affect the hardware manufacture process - does not affect the Fundamental Scientific Technology ● - does not change the prescription use ● ### Modifications: Accessories: The hand electrodes are no longer used and the stainless steel foot electrodes has been replace by a conductive disposable cloth electrodes. Software: New design and change in time and sequence of measurement. Indication for use were updated the addition of the phrase "related to the function of the sweat glands" which does not constitute a new intended use. It clarifies the utility of the device. Labeling: Instructions for Use were modified according to the changes. # 6. Performances and Effectiveness The modifications perform to the proposed device do not change the performance of the device has shown with: - 1. New risk management - 2. Software verification (SRS/SDS/STD/STR) 3. Summary of Design Control Activities and Declaration of Design control conformity. 4. Comparison of the conductance values using stainless steel electrodes versus conductive disposable cloth electrode. 5. The removal of the hand electrodes does not affect the performance of the galvanic skin response measurement is performed in the area with the higher density of sweat glands (sole of the feet). # 7. General Safety Concerns The electrical and EMC risk are not affected since the new device using the same hardware The removal of the hands electrodes cannot affect the device safety. The new material used for the foot electrodes is 510k cleared. # 8. Conclusions The SudoC device is equivalent in performances, technology, safety and efficacy to the legally marketed (unmodified) predicate device Signature: Albert MAAREK Premarket notification [510K] Number: K152216 Image /page/5/Picture/23 description: The image shows a close-up of a white object, possibly fabric or paper, with a slightly wrinkled or folded texture. The lighting is bright and even, highlighting the subtle variations in the surface. The composition is simple, focusing solely on the texture and form of the white material.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...