Wandy Self-adhesive Electrode is intended to transmit electrical current to patient skin, for TENS (Transcutaneous Electrical Nerve Stimulation) and EMS (Electrical Muscular Stimulation) applications, for OTC (Over-The -Counter) or Prescription use. The Electrode are used for adults only.
Device Story
Wandy Self-adhesive Electrodes are cutaneous electrodes used as accessories for TENS or EMS devices. The device receives electrical current from a TENS/EMS unit via a lead wire or snap button connection; it then transmits this current to the patient's skin through an integrated conductive gel layer. Used in clinical or home settings by adults; operated by patients or healthcare providers. The device facilitates electrical stimulation therapy for nerve or muscle applications. Benefit includes providing a conductive interface for therapeutic electrical stimulation.
Clinical Evidence
No clinical data. Substantial equivalence based on bench testing (impedance, wire pull, and adhesiveness) and comparison to predicate devices with established marketing history.
Technological Characteristics
Cutaneous electrode consisting of a conductive gel layer, snap button or lead wire interface. Designed for TENS/EMS applications. Bench testing performed for impedance, wire pull, and adhesiveness.
Indications for Use
Indicated for adult patients requiring TENS or EMS therapy. Used for OTC or prescription applications to transmit electrical current to the skin.
Regulatory Classification
Identification
A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.
K102842 — ADHESIVE ELECTRODES · Reno Medical Limited · Dec 2, 2010
K170744 — Adhesive Electrodes · Reno Medical , Ltd. · May 8, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
DEC 2 0 2013
#### K132998
#### Chapter 5- 510(K) Summary
| Submitter | Janis Yang |
|---------------------|--------------------------------------------------------|
| | Wandy Rubber Industrial Co., Ltd |
| | No.48, Lane 392, Fu Teh 1Rd, Shi-Chih City 221, Taipei |
| | Taiwan |
| | Tel: +886-2-26943185 |
| | Fax: +886-2-26944574 |
| Proprietary Name | Wandy Self-adhesive Electrodes |
| Common Name | Cutaneous electrode. |
| Classification Name | Cutaneous electrode. |
- Panel
Neurology
Classification
| classification name | 21 CFR<br>section | Product code | Class |
|---------------------|-------------------|--------------|-------|
| Cutaneous electrode | 882.1320 | GXY | II |
### Predicate Device_
Wandy Self adhesive Neurostimulation electrode (K002219) Wandy Silicon Conductive Rubber Pad (K002227) Jiajian Self- adhesive Electrode (K090198)
#### Description and Indication for Use
Wandy Self-adhesive Electrode is intended to transmit electrical current to patient skin for TENS (Transcutaneous Electrical Nerve Stimulation) and EMS (Electrical Muscular Stimulation) applications, for OTC (Over-The -Counter) or Prescription use. The Electrode are used for adults only.
The Wandy Self-adhesive Electrodes transmits electrical current to patient skin, the electrical current is first transmitted via the lead wire or snap button then transmitted to the conductive gel which is adhered to patient skin.
{1}------------------------------------------------
The Wandy Self-adhesive Electrodes is used as an accessory with the TENS or EMS device units.
,
. .
.
and the comments of the comments of
and the comments of the comments of the comments of the comments of
and the comments of the comments of the comments of
.
.
. .
{2}------------------------------------------------
#### Substantial Equivalency
After analyzing both bench and clinical testing data, it is the conclusion of Wandy Self-adhesive Electrodes are substantial equivalent to the predicate device, Wandy Self adhesive Neurostimulation electrode (K002219), Wandy Silicon Conductive Rubber Pad (K002227) and Jiajian Self- adhesive Electrode (K090198).
#### Performance
The Wandy Self-adhesive Neurostimulation Electrode passed self-evaluation tests for Impedance, Wire pull and adhesiveness.
#### Discussion of Clinical Tests Performed:
All design specification are exactly equivalent between new devices (Wandy Self-adhesive Electrodes) and predicate devices (Wandy Self-Adhesive Neurostimulation Electrode and Wandy Silicon Conductive Rubber Pad), except claimed indication for use and proposed labeling. Two predicate devices have thousands of marketing clinical experience and didn't have adverse events occurred in the world.
#### Conclusions:
Wandy Self-adhesive Electrodes have the same intended use and similar technological characteristics as the above predicate devices. Moreover, information contained in this submission supplied demonstrates that any differences in their characteristics do not raise any new questions of safety or effectiveness. Thus, Wandy Self-adhesive Electrodes are substantially equivalent to the predicate devices
{3}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
## December 20, 2013
Wandy Rubber Industrial Co., Ltd. Jack Yu No.48, Lane 392, Fu Teh 1 Rd, Xi Zhi Dist New Taipei City Taiwan 221
Re: K132998
Trade/Device Name: Wandy Self-Adhesive Electrodes Regulation Number: 21 CFR 882.1320 Regulatory Class: Class II Product Code: GXY Dated: September 9, 2013 Received: September 24, 2013
Dear Mr. Yu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
> Sincerely yours, Joyce M. Whang -S
Carlos L. Peña. PhD for Director Division of Neurological and Physical Medicine Devices
Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K132998
Device Name: Wandy Self-Adhesive Electrodes
Indications For Use:
Wandy Self-adhesive Electrode is intended to transmit electrical current to patient skin, for TENS (Transcutaneous Electrical Nerve Stimulation) and EMS (Electrical Muscular Stimulation) applications, for OTC (Over-The -Counter) or Prescription use. The Electrode are used for adults only.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use × (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
# Joyce M. Whang -S
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.