Ennova Self-adhesive Electrode is intended to transmit electrical current to patient skin for use with transcutaneous electrical stimulation devices. Some common types of the stimulation devices include, but are not limit to TENS (Transcutaneous Electrical Nerve Stimulation) and EMS (Electrical Muscular Stimulation). The electrode is for OTC (Over-The -Counter) or Prescription use.
Device Story
Accessory for TENS/EMS devices; transmits electrical current from stimulator to patient skin. Composed of cover, connector (lead wire or snap button), conductive carbon film, and conductive hydrogel. Used by adult patients in clinical or home settings. Current flows from stimulator through lead wire/snap button to conductive hydrogel adhered to skin. Healthcare providers or patients apply electrodes to skin; stimulator settings determine current delivery. Benefits include facilitating electrical stimulation for nerve/muscle therapy.
Clinical Evidence
No clinical data included. Bench testing only: biocompatibility (ISO 10993-5, ISO 10993-10), accelerated aging (ASTM F1980), electrical safety (AAMI/ANSI ES60601-1, IEC 60601-2-2), impedance testing, and adhesive testing.
Technological Characteristics
Materials: PU cover, conductive carbon film, conductive hydrogel, PET release liner. Connectivity: Lead wire (female socket) or snap button (male). Energy: Passive conductor for external TENS/EMS stimulators. Standards: ASTM F1980, ISO 10993-5/10, AAMI/ANSI ES60601-1, IEC 60601-2-2. Non-sterile, single-patient multiple-use.
Indications for Use
Indicated for adult patients (age 18+) requiring transcutaneous electrical nerve stimulation (TENS) or electrical muscular stimulation (EMS) via an external stimulation device.
Regulatory Classification
Identification
A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three figures facing right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 18, 2016
Suzhou Ennova Electronics Technology Co.,Ltd % Doris Dong Manager Shanghai CV Technology Co., Ltd Room 1706, No.128 Songle Rd., Songjiang Area Shanghai, 201600 CHINA
Re: K152648
Trade/Device Name: Ennova Self-adhesive Electrode Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode Regulatory Class: Class II Product Code: GXY Dated: January 12, 2016 Received: January 19, 2016
Dear Doris Dong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J. Hoffmann -A
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K152648
Device Name Ennova Self-adhesive Electrode
#### Indications for Use (Describe)
Ennova Self-adhesive Electrode is intended to transmit electrical current to patient skin for use with transcutaneous electrical stimulation devices. Some common types of the stimulation devices include, but are not limit to TENS (Transcutaneous Electrical Nerve Stimulation) and EMS (Electrical Muscular Stimulation). The electrode is for OTC (Over-The -Counter) or Prescription use.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------|----------------------------------------------------------------|
| <div> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### Section 5 510(k) Summary
[As required by 21 CFR 807.92]
#### 1. Submission Information:
| 510(k) Number: | K152648 |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | January 12, 2016 |
| Type of 510(k) Submission: | Traditional |
| Basis for 510(k) Submission: | New device |
| Submitter/Manufacturer: | Suzhou Ennova Electronics Technology Co., Ltd.<br>Jinsigang Industrial Park, Guoxiang, Wuzhong District, Suzhou City, Jiangsu<br>Province, 215124 China |
| Contactor: | Doris Dong<br>Shanghai CV Technology Co., Ltd.<br>Room 1706, No. 128 Songle Rd., Songjiang Area, Shanghai, 201600 China<br>E-mail: doris_d@126.com<br>Tel: 86 21-31261348 / Fax: 86 21-37824346 |
#### 2. Device Description:
| Proprietary Name: | Ennova Self-adhesive Electrode |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Cutaneous electrode |
| Classification Name: | Cutaneous electrode |
| Product Code: | GXY |
| Device Class: | II |
| Regulation Number: | 882.1320 |
| Review Panel: | Neurology |
| Indications for use: | Ennova Self-adhesive Electrode is intended to transmit electrical current to patient skin for use with transcutaneous electrical stimulation devices. Some common types of the stimulation devices include, but are not limit to TENS (Transcutaneous Electrical Nerve Stimulation) and EMS (Electrical Muscular Stimulation). The electrode is for OTC (Over-The -Counter) or Prescription use. |
| Device Description: | Ennova Self-adhesive Electrode is used as an accessory to the TENS or EMS device unit, which transmits electrical current to patient skin. The electrical current is first transmitted via the lead wire or snap button then transmitted to the conductive gel which is adhered to patient skin. Ennova Self-adhesive Electrode is composed of a cover, connector lead wire or snap button, conductive carbon film, conductive hydrogel, and an electrode carrier liner. It is non-sterile and intended for single adult patient(age=18) multiple application use. Ennova Self-adhesive Electrode has various shapes and sizes. To connect with a nerve or muscle stimulator, this Electrode has lead wire type and snap button type. For lead wire type electrode, the wire is at least 40mm long, with 2.0~3.5mm diameter female socket. For snap button type electrode, the connector is male snap button. The individual electrodes are electrically connected in pairs. When not in use, the hydrogel face is covered by a PET (polyethylene terephthalate) release liner. |
{4}------------------------------------------------
The conductive hydrogel is imported from Axelgaard, USA, which was 510(k) cleared with number of K130987. And the whole Ennova Self-adhesive Electrode together with the conductive hydrogel has again passed the Biocompatibility test, Shelf life test, Impedance test, Adhesive test, and so on.
#### 3. Predicate Device Identification
| 510k number | Device name | Manufacturer | Date cleared |
|-------------|---------------------------------|---------------------------------------------|--------------|
| K090198 | Jiajian Self-adhesive Electrode | Wuxi Jiajian Medical Instrument Co., Ltd | 2009 |
| K150971 | Zekang Self-adhesive Electrode | Jiangsu Zekang Medical Technology Co., Ltd. | 2015 |
#### 4. Non-Clinical Test Conclusion
Bench tests were conducted on the proposed Self-adhesive Electrode to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- ASTM F1980 (Reapproved 2011), Standard guide for accelerated aging of sterile barrier systems for medical devices. (Sterility)
- ISO 10993-5:2009/(R) 2014, Biological evaluation of medical devices -- Part 5: Tests for In Vitro
Cytotoxicity. (Biocompatibility)
- ISO 10993-10 Third Edition 2010-08-01, Biological evaluation of medical devices - Part 10: Tests for Irritation and Skin Sensitization. (Biocompatibility)
- AAMI / ANSI ES60601-1:2005/(R) 2012, Medical electrical equipment -- Part 1: General requirements for basic safety and essential performance
- IEC 60601-2-2 Edition 5.0 2009-02, Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
#### 5. Substantial Equivalent Based on Assessment of Clinical Performance Data:
Clinical data was not including in this submission.
#### 6. Substantially Equivalent Comparison Conclusion
Detailed comparison data is included in Section 9 of "Substantial Equivalence Discussion" of this 510(k) submission.
| | New Device | Predicate Device 1 | Predicate Device 2 | | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| 501(k) number | K152648 | K150971 | K090198 | | | |
| Trade Name: | Ennova Self-adhesive Electrode | Zekang Self-adhesive Electrode | Jiajian Self-adhesive Electrode | | | |
| Common Name: | Cutaneous electrode | Cutaneous electrode | Cutaneous electrode | | | |
| Classification | Cutaneous electrode | Cutaneous electrode | Cutaneous electrode | | | |
| Name: | | | | | | |
| Product Code: | GXY | GXY | GXY | | | |
| Regulation | 882.1320 | 882.1320 | 882.1320 | | | |
| Number: | | | | | | |
| Medical | Neurology | Neurology | Neurology | | | |
| Specialty: | | | | | | |
| Device Class: | II | II | II | | | |
| Indications for<br>use: | | Ennova Self-adhesive Electrode | | Zekang Self-adhesive Electrode is | | The self-adhesive electrode is |
| | | is intended to transmit electrical | | intended to transmit electrical current | | intended to be used to apply |
| | | current to patient skin for use | | to patient skin for use with | | electrical stimulation current to |
| | | with transcutaneous electrical | | transcutaneous electrical stimulation | | the patient's skin. |
| | | stimulation devices. Some | | devices. Some common types of the | | Example electrical stimulations |
| | | common types of the stimulation | | stimulation devices include, but are | | for current applications of the |
| | | devices include, but are not limit | | not limit to TENS (Transcutaneous | | electrodes are: TENS and EMS. |
| | | to TENS (Transcutaneous | | Electrical Nerve Stimulation) and | | |
| | | Electrical Nerve Stimulation) | | EMS (Electrical Muscular | | |
| | | and EMS (Electrical Muscular | | Stimulation). The electrode is for | | |
| | | Stimulation). The electrode is for | | OTC (Over-The -Counter) or | | |
| | | OTC (Over-The -Counter) or | Prescription use. | | | |
| | Prescription use. | | | | | |
| Target<br>population: | | Single patient use and multiple<br>application | | Single patient use and multiple<br>application | | Single patient use and multiple<br>application |
| Design (shape): | | Round, rectangle, oval and<br>butterfly, according to | | Round, rectangle, oval, gourd,<br>butterfly and saddle, according to | | Round, rectangle, oval, gourd,<br>butterfly and saddle, according to |
| | | customized specification. | | customized specification. | | customized specification. |
| Size of electrodes | Round | Min.Ø32mm<br>Max.Ø70mm | Round | Min.Ø20mm<br>Max.Ø70mm | Round | Min.Ø20mm<br>Max.Ø80mm |
| | Rectangle | Min.40×40mm<br>Max.100×130mm | Rectangle | Min.20×20mm<br>Max.150×150mm | Rectangle | Min.30×10mm<br>Max.80×130mm |
| | Oval | Min.50×120mm<br>Max.100×240mm | Oval | Min.50×120mm<br>Max.100×240mm | Oval | Min.50×120mm<br>Max.100×240mm |
| | Butterfly | Min.55×75mm<br>Max.95×165mm | Butterfly | Min.55×75mm<br>Max.95×165mm | Butterfly | Min.55×75mm<br>Max.95×165mm |
| | | | Gourd | Min.85×50mm<br>Max.170×100mm | Gourd | Min.85×50mm<br>Max.170×100mm |
| | | | Saddle | Min.75×110mm<br>Max.150×220mm | Saddle | Min.75×110mm<br>Max.150×220mm |
| Patient contact<br>area | Round | Min.804mm²<br>Max.3846.5mm² | Round | Min.314mm²<br>Max.3846.5mm² | Round | Min.314mm²<br>Max.5024mm² |
| | Rectangle | Min.1600 mm²<br>Max.13000mm² | Rectangle | Min.400 mm²<br>Max.22500mm² | Rectangle | Min.300 mm²<br>Max.10400mm² |
| | Oval | Min.4710 mm²<br>Max.18840mm² | Oval | Min.4710 mm²<br>Max.18840mm² | Oval | Min.4710 mm²<br>Max.18840mm² |
| | Butterfly | Min.4125 mm²<br>Max.15675mm² | Butterfly | Min.4125 mm²<br>Max.15675mm² | Butterfly | Min.4125 mm²<br>Max.15675mm² |
| | | | Gourd | Min.4250 mm²<br>Max.17000mm² | Gourd | Min. 4250 mm²<br>Max.17000mm² |
| | | | Saddle | Min.8250 mm²<br>Max.33000mm² | Saddle | Min.8250 mm²<br>Max.33000mm² |
| Impedance<br>parameters | Round | $397\text{~}612 \Omega$ | Round | $300\text{~}500 \Omega$ | Round | $300\text{~}1000 \Omega$ |
| | Rectangle | $432\text{~}624 \Omega$ | Rectangle | $300\text{~}700 \Omega$ | Rectangle | |
| | Oval | $412\text{~}642 \Omega$ | Oval | $500\text{~}700 \Omega$ | Oval | |
| Jinsigang Industrial Park, Guoxiang, Wuzhong District, Suzhou City, Jiangsu Province, 215124 China | | | | | | |
| | Butterfly | 407~616 Ω | Butterfly | 400~600 Ω | Butterfly | 400~600 Ω |
| | | | Gourd | 400~600 Ω | Gourd | |
| | | | Saddle | 400~800 Ω | Saddle | |
| Current<br>density(Use<br><span style="font-style:italic">I</span> RMS= 13.7mA for<br>calculation ) | Round | Min.0.36mA/cm2<br>Max.1.70mA/cm2 | Round | Min.0.36mA/cm2<br>Max.4.36mA/cm2 | Round | Min.0.27mA/cm2<br>Max.4.36mA/cm2 |
| | Rectangle | Min.0.11mA/cm2<br>Max.0.86mA/cm2 | Rectangle | Min.0.06mA/cm2<br>Max.3.43mA/cm2 | Rectangle | Min.0.13mA/cm2<br>Max.4.57mA/cm2 |
| | Oval | Min.0.07mA/cm2<br>Max.0.29mA/cm2 | Oval | Min.0.07mA/cm2<br>Max.0.29mA/cm2 | Oval | Min.0.07mA/cm2<br>Max.0.29mA/cm2 |
| | Butterfly | Min.0.09mA/cm2<br>Max.0.33mA/cm2 | Butterfly | Min.0.09mA/cm2<br>Max.0.33mA/cm2 | Butterfly | Min.0.09mA/cm2<br>Max.0.33mA/cm2 |
| | | | Gourd | Min.0.08mA/cm2<br>Max.0.32mA/cm2 | Gourd | Min.0.08mA/cm2<br>Max.0.32mA/cm2 |
| | | | Saddle | Min.0.04mA/cm2<br>Max.0.17mA/cm2 | Saddle | Min.0.04mA/cm2<br>Max.0.17mA/cm2 |
| Maximum Power<br>Density (W/cm2) | Round | 0.0117W/cm2 | Round | 0.0299W/cm2 | Round | 0.0299W/cm2 |
| | Rectangle | 0.0059W/cm2 | Rectangle | 0.0235W/cm2 | Rectangle | 0.0313W/cm2 |
| | Oval | 0.002W/cm2 | Oval | 0.002W/cm2 | Oval | 0.002W/cm2 |
| | Butterfly | 0.0023W/cm2 | Butterfly | 0.0023W/cm2 | Butterfly | 0.0023W/cm2 |
| | | | Gourd | 0.0022W/cm2 | Gourd | 0.0022W/cm2 |
| | | | Saddle | 0.0011W/cm2 | Saddle | 0.0011W/cm2 |
| (FDA guidance states that a maximum average power density should be less than 0.25 watts per square<br>centimeter of electrode conductive surface area to reduce the risk of thermal burns.) | | | | | | |
| Maximum<br>Current for<br>Power Density @<br>500 Ohms |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.