K203440 · Penumbra, Inc. · NRY · May 20, 2021 · Cardiovascular
Device Facts
Record ID
K203440
Device Name
Penumbra System (Reperfusion Catheter RED 62)
Applicant
Penumbra, Inc.
Product Code
NRY · Cardiovascular
Decision Date
May 20, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Penumbra Reperfusion Catheters and Separators As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment. Penumbra 3D Revascularization Device As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment. Penumbra Aspiration Tubing As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump. Penumbra Aspiration Pump The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.
Device Story
The Penumbra System is a mechanical thrombectomy device used in clinical settings by physicians to remove thrombus from intracranial vessels. The system consists of a Reperfusion Catheter (RED 62), aspiration pump, tubing, and optional Separator or 3D Revascularization Device. The catheter is introduced via guide catheter or femoral sheath, navigated over a guidewire to the occlusion site under fluoroscopic guidance. The aspiration pump provides continuous vacuum suction through the catheter to aspirate thrombus. If the catheter lumen becomes blocked, the Separator may be deployed to clear the tip. The 3D Revascularization Device facilitates thrombus removal. The system enables rapid revascularization, potentially improving clinical outcomes for acute ischemic stroke patients.
Clinical Evidence
No clinical or animal studies were conducted. Substantial equivalence was established through bench testing, including dimensional, mechanical (tensile, torque, burst, pressure), simulated use, and biocompatibility testing (ISO 10993).
Technological Characteristics
Materials: Stainless steel, PTFE, polyurethane, polyether block amide, nylon 12, nitinol, platinum/iridium. Sensing/Actuation: Mechanical aspiration via vacuum pump. Form factor: Catheter with hydrophilic coating, steam shaping mandrel, and rotating hemostasis valve. Sterilization: Ethylene Oxide (EO).
Indications for Use
Indicated for revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (internal carotid, M1/M2 middle cerebral, basilar, vertebral arteries) within 8 hours of symptom onset. Candidates include patients ineligible for or failing IV t-PA therapy.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
Penumbra System ACE 68 Reperfusion Catheter (K161640)
K211411 — Penumbra System (RED 68 Reperfusion Catheter) · Penumbra, Inc. · Jul 1, 2021
K242104 — Penumbra System (Reperfusion Catheter RED 72) · Penumbra, Inc. · Feb 13, 2025
K222808 — Penumbra System (Reperfusion Catheter RED 43) · Penumbra, Inc. · Dec 20, 2022
K211654 — Penumbra System (Reperfusion Catheter RED 72) · Penumbra, Inc. · Aug 16, 2021
K072718 — PENUMBRA SYSTEM · Penumbra, Inc. · Dec 28, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
May 20, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Penumbra, Inc. Aditi Kolla Regulatory Affairs Program Manager One Penumbra Place Alameda, California 94502
Re: K203440
Trade/Device Name: Penumbra System (Reperfusion Catheter RED 62) Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NRY Dated: April 14, 2021 Received: April 15, 2021
Dear Aditi Kolla:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203440
Device Name Penumbra System (Reperfusion Catheter RED 62)
### Indications for Use (Describe)
Penumbra Reperfusion Catheters and Separators
As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
Penumbra 3D Revascularization Device
As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
## Penumbra Aspiration Tubing
As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump.
Penumbra Aspiration Pump The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <span style="font-size:100%;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "Penumbra" in red font, followed by a red circle with a white "P" inside. The "P" is stylized with a line through the middle. The logo is simple and modern. The text is clear and easy to read.
# 510(k) Summary
(as required by 21 CFR 807.92)
Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Penumbra, Inc. is providing the summary of Substantial Equivalence for the subject Penumbra System® (Reperfusion Catheter RED™ 62).
#### 1.1 Submitter
Penumbra, Inc. One Penumbra Place Alameda, CA 94502 USA
Contact Person: Aditi Kolla Regulatory Affairs Program Manager Tel: (510) 995-2010 Fax: (510) 217-6414 E-mail: akolla@penumbrainc.com
Date of Preparation: May 19, 2021
#### Subject Device 1.2
Penumbra System® (Reperfusion Catheter RED™ 62)
Regulatory Class: II Classification Panel: Neurology Classification Name: Percutaneous Catheter Regulation Number: 21 CFR 870.1250 Product Code: NRY
#### 1.3 Predicate Devices
| 510(k) Number | Name of Device | Name of<br>Manufacturer |
|---------------|---------------------------------------------|-------------------------|
| K161640 | Penumbra System ACE 68 Reperfusion Catheter | Penumbra,<br>Inc. |
| K162901 | Penumbra 3D Revascularization Device | Penumbra,<br>Inc. |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Penumbra, a healthcare company. The word "Penumbra" is written in red, with a stylized "P" in a circle to the right. The logo is simple and modern, with a focus on the company's name.
### Device Description 1.4
The Penumbra System® is comprised of the following devices:
- Penumbra Reperfusion Catheter .
- . Penumbra Aspiration Pump
- Penumbra Aspiration Pump/Canister Tubing .
- . Penumbra Aspiration Tubing
- Penumbra Separator .
- Penumbra 3D Revascularization Device .
The Penumbra System is designed to remove thrombus from the vasculature using continuous aspiration. The Reperfusion Catheter targets aspiration from the pump directly to the thrombus. The 3D Revascularization Device is used with Reperfusion Catheters to facilitate aspiration and removal of the thrombus when needed. The Separator may be used to clear the lumen of the Reperfusion Catheter should it become blocked with thrombus. The use of the Separator may not be necessary when using a Reperfusion Catheter with an ID of 0.054 in. or larger. The Reperfusion Catheter is introduced through a guide catheter or long femoral sheath and into the intracranial vasculature and guided over a neurovascular guidewire to the site of the primary occlusion. The Penumbra Reperfusion Catheter is used with the Aspiration Pump to aspirate thrombus from an occluded vessel. As needed, a Penumbra Separator may be deployed from the Reperfusion Catheter to assist with thrombus removal. The Penumbra Separator is advanced and retracted through the Penumbra Reperfusion Catheter at the proximal margin of the primary occlusion to facilitate clearing of the thrombus from the Reperfusion Catheter tip. For the aspiration source, the Penumbra Reperfusion Catheter is used in conjunction with the Aspiration Pump, which is connected using the Penumbra Aspiration Tubing and the Penumbra Pump/Canister Tubing. The Penumbra Reperfusion Catheter is provided with a steam shaping mandrel and rotating hemostasis valve, and a peelable sheath. The Penumbra 3D Revascularization Device is provided with an introducer sheath. The Penumbra Separator is provided with an introducer and torque device. The Penumbra Reperfusion Catheters, 3D Revascularization Device and Separators are visible under fluoroscopy.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Penumbra" in red font. To the right of the word is a red circle with a white line going through the middle. The logo is simple and modern.
### Indications For Use ાં રે
## Penumbra Reperfusion Catheters and Separators
As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral -M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
## Penumbra 3D Revascularization Device
As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral -M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
## Penumbra Aspiration Tubing
As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump.
## Penumbra Aspiration Pump
The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.
### 1.6 Comparison of Indications for Use and Technological Characteristics with the Predicate Devices
| Device Attribute | Predicate Device | Subject Device |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------|
| Trade Name | Penumbra System ACE 68<br>Reperfusion Catheter | Penumbra System (Reperfusion<br>Catheter RED 62) |
| FDA Product<br>Classification | Class II, NRY, 21 CFR 870.1250 | SAME |
| 510(k) Number | K161640 | K203440 |
| Indications for Use | Penumbra Reperfusion Catheters and<br>Separators<br>As part of the Penumbra System, the<br>Reperfusion Catheters and Separators | Same as Predicate Device with<br>K162901 |
| Device Attribute | Predicate Device | Subject Device |
| | are indicated for use in the | |
| | revascularization of patients with | |
| | acute ischemic stroke secondary to | |
| | intracranial large vessel occlusive | |
| | disease (within the internal carotid, | |
| | middle cerebral - M1 and M2 | |
| | segments, basilar, and vertebral | |
| | arteries) within 8 hours of symptom | |
| | onset. | |
| | Penumbra Aspiration Tubing | |
| | As part of the Penumbra System, the | |
| | Penumbra Sterile Aspiration Tubing | |
| | is indicated to connect the Penumbra | |
| | Reperfusion Catheters to the | |
| | Penumbra Pump MAX. | |
| | Penumbra Pump MAX | |
| | The Penumbra Pump MAX is | |
| | indicated as a vacuum source for<br>Penumbra Aspiration Systems. | |
| Principles of<br>Operation | See Section 1.4 | SAME |
| Device Materials | Stainless Steel, PTFE, Polyurethane,<br>Polyether Block Amide, Nylon 12,<br>Nitinol, Platinum/Iridium | SAME |
| ID Band Color | Yellow [black text] | Black [white text] |
| Coating | Hydrophilic coating (proprietary) | Equivalent |
| Min. ID | 0.068 in. (1.73 mm) | 0.062 in. (1.57 mm) |
| Max. OD | 0.084 in. (2.13 mm) | 0.076 in. (1.93 mm) |
| Distal Flex Length | 30 cm | SAME |
| Coating Length | 30 cm | SAME |
| Effective Lengths | 115, 120, 125, 127, 132 cm | 115, 120, 125, 127, 132, 138,<br>160 cm |
| Accessories | Peelable Sheath, Shaping Mandrel,<br>RHV | SAME |
| Packaging<br>Materials | Polyester/Polyethylene/Tyvek,<br>Polystyrene, SBS Paperboard | SAME |
| Condition Supplied | Sterile and Single Use | SAME |
| Sterilization<br>Method | EO | SAME |
| Device<br>Attribute | Predicate Device | Subject Device |
| Trade Name | Penumbra 3D Revascularization Device | Penumbra System<br>(Reperfusion<br>Catheter RED 62) |
| FDA Product<br>Classification | Class II, NRY, 21 CFR 870.1250 | SAME |
| 510(k)<br>Number | K162901 | K203440 |
| Indications<br>For Use | Penumbra Reperfusion Catheters and Separators<br>As part of the Penumbra System, the Reperfusion<br>Catheters and Separators are indicated for use in the<br>revascularization of patients with acute ischemic stroke<br>secondary to intracranial large vessel occlusive disease<br>(within the internal carotid, middle cerebral - M1 and<br>M2 segments, basilar, and vertebral arteries) within 8<br>hours of symptom onset. Patients who are ineligible for<br>intravenous tissue plasminogen activator (IV t-PA) or<br>who fail IV t-PA therapy are candidates for treatment.<br>Penumbra 3D Revascularization Device<br>As part of the Penumbra System, the Penumbra 3D<br>Revascularization Device is indicated for use in the<br>revascularization of patients with acute ischemic stroke<br>secondary to intracranial large vessel occlusive disease<br>(within the internal carotid, middle cerebral - M1 and<br>M2 segments) within 8 hours of symptom onset. Patients<br>who are ineligible for intravenous tissue plasminogen<br>activator (IV t-PA) or who fail IV t-PA therapy are<br>candidates for treatment.<br>Penumbra Aspiration Tubing<br>As part of the Penumbra System, the Penumbra Sterile<br>Aspiration Tubing is indicated to connect the Penumbra<br>Reperfusion Catheters to the Penumbra Aspiration<br>Pump. | SAME |
| Device<br>Attribute | Predicate Device | Subject Device |
| | Penumbra Aspiration Pump<br>The Penumbra Aspiration Pump is indicated as a<br>vacuum source for Penumbra Aspiration Systems. | |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Penumbra, Inc., a healthcare company. The word "Penumbra" is written in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The logo is simple and modern, and the red color gives it a sense of energy and urgency.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "Penumbra" in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The logo appears to be for a company or organization called Penumbra.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "Penumbra" in a red, sans-serif font. To the right of the word is a red circle with a white outline. Inside the circle is a white "P" with a horizontal line through the middle.
### Performance Data 1.7
The following performance data were provided in support of the substantial equivalence determination:
- Design Verification ●
- Biocompatibility ●
- Shelf Life .
- . Sterilization
- Packaging Validation .
The subject device met all established requirements.
### Design Verification Testing 1.7.1
The following design verification tests were performed on the subject device:
| Test | Test Method Summary | Conclusion |
|------------------------------|--------------------------------------------------------------------------------------------------------|-------------------------|
| Dimensional/Visual<br>Test | Confirms the units meet all dimensional and visual<br>product specifications. | Acceptance Criteria Met |
| Friction Test | Confirms units meet product specification related to<br>friction. | Acceptance Criteria Met |
| Fluoroscopy Test | Confirms the marker band is fluoroscopically visible. | Acceptance Criteria Met |
| Simulated Use Test | Confirms the functionality of units using clinically<br>relevant benchtop model. | Acceptance Criteria Met |
| Particulate Test | Particulates generated during simulated use (including<br>multiple deployment cycling) were evaluated. | Acceptance Criteria Met |
| Hub Air Test | Confirms units have no leaks when tested. | Acceptance Criteria Met |
| Tensile Test | Confirms units meet product specification related to<br>tensile strength. | Acceptance Criteria Met |
| Pressure Test | Confirms units meet product specification related to<br>pressure. | Acceptance Criteria Met |
| Elongation Test | Confirms units meet product specification related to<br>elongation. | Acceptance Criteria Met |
| Corrosion Resistance<br>Test | Confirms there is no visible corrosion on the units<br>when tested. | Acceptance Criteria Met |
| Torque Strength Test | Confirms units have sufficient torque strength. | Acceptance Criteria Met |
| Burst Pressure Test | Confirms units can withstand sufficient pressure. | Acceptance Criteria Met |
| Distal Tip Stiffness<br>Test | Confirms units have appropriate distal tip stiffness. | Acceptance Criteria Met |
| Shelf-Life | Confirms expiration date based on accelerated aging<br>test studies. | Acceptance Criteria Met |
| Packaging Validation<br>Test | Confirms the packaging of the units meet all product<br>specifications | Acceptance Criteria Met |
{9}------------------------------------------------
Image /page/9/…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.