K190010 · Penumbra, Inc. · NRY · Jun 16, 2019 · Cardiovascular
Device Facts
Record ID
K190010
Device Name
Penumbra System Reperfusion Catheter JET 7
Applicant
Penumbra, Inc.
Product Code
NRY · Cardiovascular
Decision Date
Jun 16, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Penumbra Reperfusion Catheters and Separators As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment. Penumbra 3D Revascularization Device As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment. Penumbra Aspiration Tubing As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump. Penumbra Aspiration Pump The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.
Device Story
The Penumbra System Reperfusion Catheter JET 7 (modified) is a neurovascular catheter used to remove thrombus in acute ischemic stroke. It connects to the Penumbra Aspiration Pump via sterile tubing to deliver vacuum pressure directly to the occlusion site. Operated by physicians in a clinical setting, the device is navigated through the neurovasculature to the clot. The vacuum aspiration facilitates mechanical thrombus removal. The device features a hydrophilic coating and a distal tip designed for trackability in tortuous anatomy. By restoring blood flow, the device aims to improve patient outcomes in stroke cases. The modified JET 7 maintains the same dimensions as the predicate JET 7 but incorporates material changes to reduce distal tip stiffness, enhancing deliverability.
Clinical Evidence
No clinical study was conducted. Substantial equivalence is supported by bench-top design verification (dimensional, simulated use, mechanical, and biocompatibility testing) and a literature review of clinical outcomes for predicate devices with similar dimensions used for direct aspiration.
Technological Characteristics
Percutaneous catheter; materials include Grilamid (hub/sleeve), 304 SS (strain relief), PTFE (liner), NiTi/SS wire (reinforcement), and Pt/Ir (markerband). Features hydrophilic coating. Dimensions: 0.085" max OD, 0.072" min ID. Sterilized via EO. Single-use, disposable.
Indications for Use
Indicated for revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (internal carotid, M1/M2 middle cerebral, basilar, and vertebral arteries) within 8 hours of symptom onset. Candidates include patients ineligible for or failing IV t-PA therapy.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
Penumbra System – JET 7 Reperfusion Catheter (K173761)
Reference Devices
Penumbra System – ACE 68 Reperfusion Catheter (K161640)
Related Devices
K173761 — Penumbra System Reperfusion Catheter JET 7 · Penumbra, Inc. · Aug 17, 2018
K202251 — Penumbra System (Penumbra JET 7 Reperfusion Catheter with Xtra Flex Technology; Penumbra JET 7MAX) · Penumbra, Inc. · Aug 31, 2020
K191946 — Penumbra System Penumbra JET 7X · Penumbra, Inc. · Feb 27, 2020
K211476 — ZOOM 71 Reperfusion Catheter; ZOOM Aspiration Tubing · Imperative Care, Inc. · Jun 8, 2021
K183043 — 0.071 ID MantaRay Reperfusion Catheter, 0.055 ID MantaRay Reperfusion Catheter, 0.045 ID MantaRay Reperfusion Catheter, 0.035 ID MantaRay Reperfusion Catheter · Imperative Care, Inc. · Apr 17, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.
June 16, 2019
Penumbra, Inc. Micaela Victoria Regulatory Affairs Specialist III One Penumbra Place Alameda, California 94502
Re: K190010
Trade/Device Name: Penumbra System® (Reperfusion Catheter JETTM 7) Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NRY Dated: May 16, 2019 Received: May 17, 2019
Dear Micaela Victoria:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Xiaolin Zheng, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190010
Device Name Penumbra System (Reperfusion Catheter JET 7)
### Indications for Use (Describe)
Penumbra Reperfusion Catheters and Separators
As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
Penumbra 3D Revascularization Device
As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
### Penumbra Aspiration Tubing
As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump.
Penumbra Aspiration Pump The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Penumbra, a healthcare company. The logo consists of the word "Penumbra" in a red sans-serif font, followed by a circular symbol. The symbol is red with a white "P" inside, and a white line extending from the "P" to the edge of the circle. The logo is simple and modern, and the red color gives it a sense of energy and urgency.
#### 1 510(k) Summary
(as required by 21 CFR 807.92)
Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Penumbra Inc. is providing the summary of Substantial Equivalence for the Penumbra System® Reperfusion Catheter JET™ 7.
#### 1.1 Sponsor/Applicant Name and Address
Penumbra, Inc. One Penumbra Place Alameda, CA 94502 USA
#### 1.2 Sponsor Contact Information
Micaela Victoria Regulatory Affairs Specialist III Phone: (510) 748-2082 FAX: (510) 217-6414 Email: mvictoria@penumbrainc.com
#### 1.3 Date of Preparation of 510(k) Summary
May 16, 2018
#### 1.4 Device Trade or Proprietary Name
Penumbra System® (Reperfusion Catheter JETTM 7)
#### 1.5 Device Classification
Regulatory Class: II Classification Panel: Neurology Classification Name: Percutaneous Catheter Regulation Number: 21 CFR §870.1250 Product Code: NRY (Catheter, Thrombus Removal)
#### 1.6 Predicate Devices
| 510(k) Number / Clearance Date | Name of Device | Name of Manufacturer |
|----------------------------------------------------------------------------------|--------------------------------------------------|---------------------------------------------------------------|
| Primary Predicate Device | | |
| K173761 cleared on August 8, 2018 | Penumbra System – JET 7<br>Reperfusion Catheter | Penumbra, Inc.<br>One Penumbra Place<br>Alameda, CA 94502 USA |
| Reference Device | | |
| K161640 cleared on July 12, 2016<br>(applicable to packaging validation<br>only) | Penumbra System – ACE 68<br>Reperfusion Catheter | Penumbra, Inc.<br>One Penumbra Place<br>Alameda, CA 94502 USA |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the word "Penumbra" in a bold, red font. To the right of the word is a circular logo, also in red, with a white "P" inside. The logo has a white line extending from the "P" to the edge of the circle, giving the impression of motion. The overall design is clean and modern.
#### Predicate Comparison 1.7
| System Name | Penumbra System® | | |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|--|
| Device Name | JET 7<br>(Predicate) | Modified JET 7<br>(Subject) | |
| 510(k) No. | K173761 (applicable for JET 7)<br>K161640 (applicable for packaging validation<br>only - Reference Device) | K190010 | |
| Classification | Class II, NRY | SAME | |
| Indication | Penumbra Reperfusion Catheters and Separators<br>As part of the Penumbra System, the<br>Reperfusion Catheters and Separators are<br>indicated for use in the revascularization of<br>patients with acute ischemic stroke secondary to<br>intracranial large vessel occlusive disease<br>(within the internal carotid, middle cerebral -<br>M1 and M2 segments, basilar, and vertebral<br>arteries) within 8 hours of symptom onset.<br>Patients who are ineligible for intravenous tissue<br>plasminogen activator (IV t-PA) or who fail IV<br>t-PA therapy are candidates for treatment.<br>Penumbra 3D Revascularization Device<br>As part of the Penumbra System, the Penumbra<br>3D Revascularization Device is indicated for use<br>in the revascularization of patients with acute<br>ischemic stroke secondary to intracranial large<br>vessel occlusive disease (within the internal<br>carotid, middle cerebral - M1 and M2 segments)<br>within 8 hours of symptom onset. Patients who<br>are ineligible for intravenous tissue plasminogen<br>activator (IV t-PA) or who fail IV t-PA therapy<br>are candidates for treatment.<br>Penumbra Aspiration Tubing<br>As part of the Penumbra System, the Penumbra<br>Sterile Aspiration Tubing is indicated to connect<br>the Penumbra Reperfusion Catheters to the<br>Penumbra Aspiration Pump.<br>Penumbra Aspiration Pump<br>The Penumbra Aspiration Pump is indicated as a<br>vacuum source for Penumbra Aspiration<br>Systems. | SAME | |
| System Name | Penumbra System® | | |
| Device Name | JET 7<br>(Predicate) | Modified JET 7<br>(Subject) | |
| Materials | | | |
| Proximal hub | Grilamid (TR55-LX) | SAME | |
| Strain Relief [Hub Sleeve] | Grilamid (TR55) | SAME | |
| Strain Relief | 304 Stainless Steel (SS) | SAME | |
| ID Band | Polyolefin, PET black [white foil] | SAME | |
| Liner | PTFE | SAME | |
| Catheter Shaft | | | |
| Extrusions | Polyurethane | Equivalent | |
| | Polyether Block Amide | SAME | |
| | Nylon 12 | SAME | |
| Distal Coil Reinforcement | NiTi wire | SAME | |
| Proximal Coil<br>Reinforcement | SS wire and NiTi wire | SAME | |
| Extrusion Colorants | Clear/ Natural or Purple | SAME | |
| Tip Shape | Straight | SAME | |
| Markerband | Platinum/Iridium (90% Pt, 10% Ir) | SAME | |
| Coating | Hydrophilic (proprietary) | Equivalent | |
| Dimensions | | | |
| Proximal OD | 0.085 in Max | SAME | |
| Proximal ID | 0.072 in Min | SAME | |
| Distal OD | 0.085 in Max | SAME | |
| Distal ID | 0.072 in Min | SAME | |
| Effective Length | 115, 120, 125, 127, 132 cm | SAME | |
| Distal Flex Length | 30 cm | SAME | |
| Coating Length | 30 cm | SAME | |
| Accessories | | | |
| Peelable Sheath | PTFE | SAME | |
| Rotating Hemostasis Valve | Polycarbonate, silicone o-ring | SAME | |
| Shaping Mandrel | 0.038in OD stainless steel | SAME | |
| Packaging Materials | | | |
| System Name | Penumbra System® | | |
| Device Name | JET 7<br>(Predicate) | Modified JET 7<br>(Subject) | |
| Pouch | Polyester/Polyethylene/Tyvek | SAME | |
| Packaging Hoop | Polyethylene | SAME | |
| Packaging Tray<br>(Kit Configuration) | Polyethylene terephthalate, Polystyrene | SAME | |
| Packaging Card | Polyethylene | SAME | |
| Display Carton | SBS Paperboard | SAME | |
| Packaging Configuration | Hoop: Hoop/Packaging Card/Pouch/Box | SAME | |
| | Kit: Tray/Retainer/Lid/Aspiration<br>Tubing/Accessory Pouch/Pouch/Box…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.