K191946 · Penumbra, Inc. · NRY · Feb 27, 2020 · Cardiovascular
Device Facts
Record ID
K191946
Device Name
Penumbra System Penumbra JET 7X
Applicant
Penumbra, Inc.
Product Code
NRY · Cardiovascular
Decision Date
Feb 27, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Penumbra Reperfusion Catheters and Separators As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment. Penumbra 3D Revascularization Device As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment. Penumbra Aspiration Tubing As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump. Penumbra Aspiration Pump The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.
Device Story
The Penumbra JET 7MAX consists of a JET 7 Reperfusion Catheter and an optional MAX Delivery Device (Access Assist Tool). The device is used in neurointerventional procedures to remove thrombus from intracranial vessels. The MAX Delivery Device is inserted into the catheter to assist in navigating tortuous neurovasculature and is removed prior to aspiration. Aspiration is delivered from an external Penumbra Aspiration Pump through the catheter to the occlusion site. The device is operated by a physician in a clinical setting. The output is the mechanical removal of thrombus, which restores blood flow to the brain, potentially improving clinical outcomes for acute ischemic stroke patients.
Clinical Evidence
No clinical study was conducted. Substantial equivalence is supported by bench-top design verification testing, biocompatibility testing, and a literature review of clinical outcomes for technologically equivalent devices.
Indicated for revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (internal carotid, M1/M2 middle cerebral, basilar, vertebral arteries) within 8 hours of symptom onset. Candidates include patients ineligible for or failing IV t-PA therapy.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Neuron Intracranial Access System – Neuron Select Catheter 070 (K083125)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 27, 2020
Penumbra, Inc. Micaela Victoria Regulatory Affairs Specialist III One Penumbra Place Alameda, California 94502
Re: K191946
Trade/Device Name: Penumbra System® JET™ 7 Reperfusion Catheter with MAX Delivery Device (JETTM 7MAX)] Regulation Number: 21 CFR 870.1250
Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NRY Dated: January 27, 2020 Received: January 28, 2020
Dear Micaela Victoria:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Xiaolin Zheng, Ph.D., M.S. Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191946
### Device Name
Penumbra System® [JET™ 7 Reperfusion Catheter with MAX Delivery Device (JET™ 7MAX)]
### Indications for Use (Describe)
Penumbra Reperfusion Catheters and Separators
As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
Penumbra 3D Revascularization Device
As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
Penumbra Aspiration Tubing
As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump.
Penumbra Aspiration Pump The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.
Type of Use (Select one or both, as applicable)
| <span style="font-size:10px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
| <span style="font-size:10px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "Penumbra" in a red font. To the right of the word is a red circle with a white "P" inside. The logo is simple and modern. The registered trademark symbol is visible in the bottom right corner of the circle.
#### 1 510(k) Summary
(as required by 21 CFR 807.92)
Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Penumbra, Inc. is providing the summary of Substantial Equivalence for the Penumbra System® Penumbra JET™ 7MAX (Penumbra JETTM 7 Reperfusion Catheter with MAX Delivery Device)
#### 1.1 Sponsor/Applicant Name and Address
Penumbra, Inc. One Penumbra Place Alameda, CA 94502 USA
#### 1.2 Sponsor Contact Information
Micaela Victoria Regulatory Affairs Specialist III Phone: (510) 748-2082 FAX: (510) 217-6414 Email: mvictoria@penumbrainc.com
#### 1.3 Date of Preparation of 510(k) Summary
January 24, 2020
#### 1.4 Device Trade or Proprietary Name
Penumbra System® [JET™ 7 Reperfusion Catheter with MAX Delivery Device (JET™ 7MAX)]
#### 1.5 Device Classification
| Regulatory Class: | II |
|-----------------------|----------------------------------|
| Classification Panel: | Neurology |
| Classification Name: | Percutaneous Catheter |
| Regulation Number: | 21 CFR §870.1250 |
| Product Code: | NRY (Catheter, Thrombus Removal) |
#### 1.6 Predicate Devices
| 510(k) Number / Clearance Date | Name of Device | Name of Manufacturer |
|----------------------------------|----------------------------------------------------------------|---------------------------------------------------------------|
| Primary Predicate Device | | |
| K190010 cleared on June 16, 2019 | Penumbra System® -<br>Penumbra JETTM 7 Reperfusion<br>Catheter | Penumbra, Inc.<br>One Penumbra Place<br>Alameda, CA 94502 USA |
| Reference Device | | |
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "Penumbra" in a red font, followed by a red circle with a white "P" inside. The "P" is formed by a white line that extends from the left side of the circle to the right side, creating the shape of the letter. The logo is simple and modern, with a focus on the company name and a stylized initial.
| K173761 cleared on August 8, 2018<br>(Animal Testing only) | Penumbra System® –<br>Penumbra JETT™ 7 Reperfusion<br>Catheter | Penumbra, Inc.<br>One Penumbra Place<br>Alameda, CA 94502 USA |
|----------------------------------------------------------------------------|----------------------------------------------------------------------|---------------------------------------------------------------|
| K083125 cleared on November 21,<br>2008 (Biocompatibility Testing<br>only) | Neuron Intracranial Access<br>System – Neuron Select Catheter<br>070 | Penumbra, Inc.<br>One Penumbra Place<br>Alameda, CA 94502 USA |
#### 1.7 Predicate Comparison
| System Name | Penumbra System® | |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Penumbra JET 7<br>(Predicate) | Penumbra JET 7MAX<br>(Subject) |
| 510(k) No. | K190010 (applicable for Penumbra JET 7 with<br>Access Assist Tool)<br>K173761 (applicable for Animal Testing -<br>Reference Device) | K191946 |
| Classification | Class II, NRY | SAME |
| System Name | Penumbra System® | |
| Device Name | Penumbra JET 7<br>(Predicate) | Penumbra JET 7MAX<br>(Subject) |
| Indication | Penumbra Reperfusion Catheters and Separators<br>As part of the Penumbra System, the<br>Reperfusion Catheters and Separators are<br>indicated for use in the revascularization of<br>patients with acute ischemic stroke secondary to<br>intracranial large vessel occlusive disease<br>(within the internal carotid, middle cerebral –<br>M1 and M2 segments, basilar, and vertebral<br>arteries) within 8 hours of symptom onset.<br>Patients who are ineligible for intravenous tissue<br> | SAME |
| Reperfusion Catheter | Penumbra JET™ 7 | SAME |
| Accessories | | |
| Peelable Sheath | PTFE | SAME |
| Rotating Hemostasis Valve | Polycarbonate, silicone o-ring | SAME |
| System Name | Penumbra System® | |
| Device Name | Penumbra JET 7<br>(Predicate) | Penumbra JET 7MAX<br>(Subject) |
| Shaping Mandrel | Stainless Steel | SAME |
| Access Assist Tool | N/A – not included | Materials:<br>Nylon 12, Copolyester, Polyolefin,<br>Polyurethane, Polyether Block<br>Amide, PTFE, Platinum/Tungsten,<br>Hydrophilic Coating<br>Dimensions:<br>ID: 0.018in<br>OD: 0.071in<br>Extension Length: 1.5cm |
| Packaging Materials | | |
| Pouch | Polyester/Polyethylene/Tyvek | SAME |
| Packaging Tray<br>(Kit Configuration) | Polyethylene terephthalate, Polystyrene | SAME |
| Display Carton | SBS Paperboard | SAME |
| Kit Packaging<br>Configuration | Tray/Retainer/Lid/Aspiration Tubing/Accessory<br>Pouch/Pouch/Box | SAME |
| Sterilization | EO | SAME |
| Shelf-Life | 36 Months | 12 Months |
| Use | Single use, disposable | SAME |
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "Penumbra" in a red, sans-serif font. To the right of the word is a circular logo, also in red, with a white "P" inside. The logo has a white line extending from the left side of the "P" to the edge of the circle, creating a visual connection between the text and the logo.
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the Penumbra logo. The word "Penumbra" is written in a red sans-serif font. To the right of the word is a red circle with a white "P" inside. The logo is simple and modern.
#### Device Description 1.8
The Penumbra JET 7 with MAX Delivery Device, known as Penumbra JET 7MAX, is an additional configuration being added to the currently available Penumbra System. The MAX Delivery Device is an optional accessory for use with the Penumbra JET 7 Reperfusion Catheter and is removed prior to aspiration. The Reperfusion Catheter Penumbra JET 7 delivers aspiration from the Aspiration Pump directly to the site of occlusion to assist in the removal of thrombus from the neurovasculature. The devices are provided sterile, nonpyrogenic, and intended for single use only.
{8}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the Penumbra company logo. The word "Penumbra" is written in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The logo is simple and modern.
#### 1.9 Indications for Use
# Penumbra Reperfusion Catheters and Separators
As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
## Penumbra 3D Revascularization Device
As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
## Penumbra Aspiration Tubing
As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump.
## Penumbra Aspiration Pump
The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.
## 1.10 Summary of Non-Clinical Data
As required under Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, a summary of any information regarding substantial equivalence of the device follows.
Included in this section are summary descriptions of the testing, which substantiates the safe and effective performance of the subject Penumbra JET 7 with MAX Delivery Device as well as its substantial equivalence to the predicate and reference devices:
- Biocompatibility ●
- Design Verification (Bench-Top Testing) ●
- Design Validation (GLP Animal Testing) ●
{9}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the word "Penumbra" in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The "P" is stylized with a horizontal line extending from the left side of the "P" to the edge of the circle. The logo is simple and modern.
The subject Penumbra JET 7 with MAX Delivery Device met all established requirements.
# 1.10.1 Biocompatibility Testing
Biocompatibility was conducted on the subject Penumbra JET 7 with MAX Delivery Device. The studies were selected in accordance with ISO 10993-1 guidelines (Biological Evaluation of Medical Devices) for a limited exposure (< 24 hours), externally communicating device with circulating blood contact. All studies were conducted pursuant to 21 CFR, Part 58, Good Laboratory Practices (GLP). Reperfusion Catheter testing was previously performed on the predicate Penumbra JET 7 [K190010]; therefore, the scope of the testing (in the table below) was only performed on the MAX Delivery Device.
| Tests | Acceptance Criteria | Results | Conclusion |
|-----------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| Cytotoxicity:<br>MEM Elution<br>(10993-5) | Sample extracts must have a<br>cytotoxic reactivity score of grade<br>2 or lower | Grade = 0 (Reactivity None) | Pass<br>Non-cytotoxic |
| Sensitization:<br>Magnusson-<br>Kligman Method<br>(10993-10) | Test Group shall yield Grade < 1<br>score on Magnusson and Kligman<br>scale (provided Control Group<br>yields Grade < 1) | NaCl Extract<br>Grade = 0<br>CSO Extract<br>Grade = 0 | Pass<br>Non-sensitizing |
| Irritation:<br>Intracutaneous<br>Reactivity<br>(10993-10) | The difference between the<br>average scores for the extract of<br>the test article and the control is<br>≤ 1.0 | NaCl Extract<br>Difference = 0.0<br>CSO Extract<br>Difference = 0.0 | Pass<br>Non-irritating |
| Systemic Toxicity:<br>Acute Systemic<br>Injection<br>(10993-11) | Sample extracts must not cause<br>significant biological reaction<br>greater than control. That is:<br>• Death in 2 or more animals<br>• Signs of toxicity in 2 or more<br>animals (i.e. convulsions,<br>prostration)<br>• Weight loss > 10% in 3 or<br>more animals | No evidence of systemic<br>toxicity from sample extracts<br>(both NaCl and CSO extracts).<br>That is:<br>No deaths<br>• No signs consistent with<br>toxicity<br>No weight loss > 10% | Pass<br>Non-toxic |
| Systemic Toxicity:<br>Material Mediated<br>Pyrogen<br>(10993-11, USP) | Sample extracts must not cause a<br>total rise in body temperature of<br>≥ 0.5 °C | Non-pyrogenic: no single<br>animal had an individual rise<br>in body temperature ≥ 0.5 °C | Pass<br>Non-pyrogenic |
| Hemocompatibility:<br>Prothrombin Time<br>(PT) | Clotting times of test article must<br>be similar to predicate values using<br>analysis of variance. | Test article coagulation times<br>are statistically similar to<br>predicate | Pass<br>Hemocompatible |
| Biocompatibility Test Results | | | |
|-------------------------------|--|--|--|
|-------------------------------|--|--|--|
{10}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the word "Penumbra" in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The logo is simple and modern, and the colors are bold and eye-catching. The wordmark and the symbol are well-balanced and create a strong visual identity.
| Tests | Acceptance Criteria | Results | Conclusion |
|----------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
| (10993-4) | | | |
| Hemocompatibility:<br>Partial<br>Thromboplastin<br>Time (PTT)<br>(10993-4) | Clotting times of test article must<br>be similar to predicate values using<br>analysis of variance | Test article coagulation times<br>are statistically similar to<br>predicate | Pass<br>Hemocompatible |
| Hemocompatibility:<br>Complement<br>Activation<br>(10993-4) | The concentration of SC5b-9 of<br>test article must be similar to<br>predicate values using analysis of<br>variance | SC5b-9<br>Test article concentrations are<br>statistically similar to predicate<br>at all exposure time points:<br>• 30min<br>• 60min<br>• 90min | Pass<br>Hemocompatible |
| Hemocompatibility:<br>Hemolysis (indirect<br>contact)<br>(10993-4) | Sample extracts must be non-<br>hemolytic (≤ 2% hemolytic<br>index) | Hemolytic Index = 0.22% | Pass<br>Non-hemolytic |
| Hemocompatibility:<br>Hemolysis (direct<br>contact)<br>(10993-4) | Sample must be non-<br>hemolytic (≤ 2% hemolytic<br>index) | Hemolytic Index = 0.00% | Pass<br>Non-hemolytic |
| Hemocompatibility:<br><i>In vitro</i><br>Thrombogenicity<br>(10993-4) | Device must be non-thrombogenic<br><i>in vitro</i> when compared to<br>predicate device | Test article performed equal or<br>better than predicate in three<br>separate <i>in vitro</i> assays | Pass<br>Non-thrombogenic |
In summary, non-clinical testing substantiates that the Penumbra System Penumbra JET 7 with MAX Delivery Device is non-cytotoxic, non-sensitizing, non-irritating, non-toxic, non-pyrogenic, non-hemolytic, and non-thrombogenic.
# 1.10.2 Design Verification - Bench Top Testing
The physical and mechanical properties of the subject Penumbra JET 7 with MAX Delivery Device was assessed using standard test methods and pre-determined acceptance criteria. The following tests were performed and all tests passed successfully:
{11}------------------------------------------------
Image /page/10/Picture/1 description: The image features the word "Penumbra" in a bold, red font. To the right of the word is a circular logo, also in red, with a white line running horizontally through the center. Inside the circle is a stylized letter "P" formed by white lines.
| Attribute | Specification | Results |
|--------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Dimensional/<br>Visual Inspection | These evaluations confirm that the units used in this Design<br>Verification testing meet all product specifications. | Pass |
| Simulated Use<br>[Intracranial<br>Access & Vessel<br>Access Entry<br>Performance.<br>Delivery/Retrieval<br>Forces] | Simulated use testing of the Access Assist Tool with Penumbra JET<br>7 Reperfusion Catheter and relevant accessory devices in an<br>anatomical model which simulated the tortuosity of the<br>neurovasculature. Devices were delivered through the tortuous<br>anatomical model to evaluate the effectiveness of the device to<br>assist in the delivery of the Reperfusion Catheter to the target site. | Pass |
| Reperfusion<br>Catheter / Access<br>Assist Tool<br>compatibility<br>(Friction Force) | Maximum value per specification | Pass |
| Access Assist<br>Tool / 0.016"<br>Guidewire<br>compatibility<br>(Friction Force) | Maximum value per specification | Pass |
| Markercoil<br>Visibility | The markercoil is fluoroscopically visible | Pass |
| Torsion | Number of turns will be recorded for informational purposes only<br>[FIPO]. | FIPO |
| Corrosion | No visible corrosion immediately after Corrosion Testing procedure | Pass |
| Particulate<br>Testing | ≥ 10 um will be ≤ 6000 particles | Pass |
| | ≥ 25 um will be ≤ 600 particles | Pass |
| | ≥ 75 um particles will be recorded for informational purposes only<br>[FIPO] | FIPO |
| | ≥ 125 um particles will be recorded for informational purposes only<br>[FIPO] | FIPO |
| Coating Integrity<br>(Pre-Inspection) | Coating has not delaminated, peeled, or flaked prior to simulated<br>use particulate testing | Pass |
| Coating Integrity<br>(Post-Inspection) | Coating has not delaminated, peeled, or flaked after simulated use<br>particulate testing | Pass |
| Attribute | Specification | Results |
| Hub/Air<br>Aspiration | When negative pressure is pulled, no air may leak into hub | Pass |
| Bond Strength<br>Distal Joint 1 | Minimum value per specification | Pass |
| Bond Strength<br>Distal Joint 2 | Minimum value per specification | Pass |
| Bond Strength<br>Midjoint 1 | Minimum value per specification | Pass |
| Bond Strength<br>Midjoint 2 | Minimum value per specification | Pass |
| Proximal Joint | Minimum value per specification | Pass |
| Hub to Shaft<br>Bond Strength | Minimum value per specification | Pass |
| Elongation to<br>Failure - Access<br>Assist Tool | Meets value per specification | Pass |
| Pressure Test | Minimum value per specification | Pass |
{12}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the word "Penumbra" in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The logo is simple and modern, with the wordmark and symbol working together to create a recognizable brand identity. The red color is bold and eye-catching, while the white "P" provides contrast and visual interest.
# 1.10.3 Design Validation - Animal Study
The safety and efficacy of the reference device, Penumbra JET 7 Reperfusion Catheter (Penumbra JET 7) was evaluated in the accepted porcine model [K173761]. The purpose of this study was to evaluate the aspiration vascular response of the reference device Penumbra JET 7. The subject Penumbra JET 7 with MAX Delivery Device consists of a Penumbra JET 7 Reperfusion Catheter and optional catheter accessory, which is removed from the Reperfusion Catheter prior to aspiration. The subject, Penumbra JET 7 with MAX Delivery Device, Reperfusion Catheter dimensions are identical to that of the reference device Penumbra JET 7. As a result, no additional animal testing was required.
{13}------------------------------------------------
Image /page/12/Picture/1 description: The image shows the word "Penumbra" in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The logo is simple and modern.
# 1.11 Performance Data – Clinical:
No clinical study was conducted as bench and previously performed animal testing was determined sufficient for verification and validation purposes. A review was conducted considering published clinical study articles for devices with similar dimensions used for direct aspiration. The literature review was used to support the determination of substantial equivalence by using clinical outcomes from devices that are considered technologically equivalent.
# 1.12 Summary of Substantial Equivalence
The Penumbra JET 7 with MAX Delivery Device is substantially equivalent to the predicate and reference devices, provided in Section 1.6, with regard to indications, intended use, design, performance, materials, sterilization and packaging.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.