NeuroWave NeuroFAST MONITORING SYSTEM, MODEL NF-701
K092477 · Neurowave Systems, Inc. · OMC · Oct 29, 2010 · Neurology
Device Facts
Record ID
K092477
Device Name
NeuroWave NeuroFAST MONITORING SYSTEM, MODEL NF-701
Applicant
Neurowave Systems, Inc.
Product Code
OMC · Neurology
Decision Date
Oct 29, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1400
Device Class
Class 2
Indications for Use
The NeuroFAST Monitoring System, Model NF-201 is intended to be used for measuring and recording the electrical activity of a patient's brain, obtained by placing electrodes on the head. The NeuroFAST NF-201 is indicated for use in acquiring electroencephalographic (EEG) signals in the OR, ICU, ER, clinical settings and for clinical research. The NeuroFAST Monitor is to be used under the direction and interpretation of a licensed medical professional. The NeuroFAST Monitoring System, Model NF-201 does not provide any diagnostic conclusion about the patient's condition.
Device Story
NeuroFAST NF-201 is an EEG monitor acquiring analog signals via noninvasive disposable electrodes (EasyPrep Sensor Kit). Patient Module (PM-201) converts analog signals to digital; USB cable transmits data to Display Module (DM-201). System displays raw EEG waveforms and processed variables (power spectrum, frequency bands, MEF, SEF) for clinician interpretation. Used in OR, ICU, ER, and clinical research settings by medical professionals. Provides continuous monitoring and data archiving. Does not provide diagnostic conclusions. Benefits include real-time visualization of brain electrical activity to assist clinical decision-making.
Clinical Evidence
No clinical data. Bench testing only. Device performance evaluated against design specifications and IFCN guidelines for electroencephalographs. Compliance with IEC60601-1, IEC60601-1-4, IEC60601-2-26, and IEC60601-1-2 confirmed via laboratory testing.
Indicated for acquiring EEG signals in OR, ICU, ER, clinical settings, and clinical research. For use under direction and interpretation of licensed medical professional. No patient age or gender restrictions specified.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}------------------------------------------------
K092477
# 510(k) SUMMARY
### 1. Contact Details
2.
| Applicant Name: | Neuro Wave Systems Inc.<br>2490 Lee Blvd., Suite 300<br>Cleveland Heights, OH 44118<br>OCT 29 2010<br>(216) 361-1591 (telephone)<br>(216) 361-1554 (fax) |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Tracie Capozzio,<br>Quality Assurance and Regulatory Compliance<br>tcapozzio@neurowavesystems.com |
| Date Prepared: | October 19, 2010 |
| Device Name | |
| Trade Name: | NeuroFAST Monitoring System, Model NF-201 |
| Common/Usual Name: | EEG Monitor |
| Classification Name: | Electroencephalograph |
| Product Code: | OMC |
21 CFR 882.1400 Class II Neurology
#### Legally Marketed Predicate Device 3.
Regulation Number:
Classification Panel:
Device Class:
| 510(k)<br>Number | Product<br>Code | Trade Name | Manufacturer |
|------------------|-----------------|----------------------|----------------------------------|
| K082886 | OLT | ZOOM-100DC | Brainscope Company, Inc. |
| K072382 | GWQ | Model I-2000 Monitor | Infinite Biomedical Technologies |
#### 4. Description of Device
The NeuroFAST Monitoring System, Model NF-201, is an EEG monitor that records and displays electroencephalograms (EEGs) obtained from noninvasive electrodes placed on a patient's head. The acquired EEG waveforms and processed EEG variables are continuously displayed by the system for interpretation by the licensed medical professional.
The NeuroFAST Monitoring System, Model NF-201 has three main components:
- Display Module (DM-201) processes acquired digital EEG signals, displays . EEG waveforms and processed EEG variables, and archives them for later review
{1}------------------------------------------------
- Patient Module (PM-201) acquires analog EEG signals through the Patient . Cable, and converts them into digital EEG signals
- . EasyPrep Sensor Kit (EK-701) - Noninvasive, disposable, single patient electrodes for acquiring the EEG signal
The NeuroFAST NF-201 system also includes the following cables:
- Patient Cable (PC-201) connects the Patient Module to four (4) noninvasive . electrodes
- USB Data Cable (DC-201) sends acquired digital EEG signals from the . Patient Module to the Display Module
#### 5. Intended Use
The intended use of NeuroFAST Monitoring System, Model NF-201 is consistent with the classification 21 CFR 882.1400, Electroencephalograph:
"An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head."
The NeuroFAST Monitoring System, Model NF-201 is intended to be used for measuring and recording the electrical activity of a patient's brain, obtained by placing electrodes on the head. The NeuroFAST NF-201 is indicated for use in acquiring electroencephalographic (EEG) signals in the OR, ICU, ER, clinical settings and for The NeuroFAST Monitor is to be used under the direction and clinical research. interpretation of a licensed medical professional. The NeuroFAST Monitoring System, Model NF-201 does not provide any diagnostic conclusion about the patient's condition.
#### Substantial Equivalence Comparison 6.
The technology of the NeuroFAST NF-201 device is equivalent to other EEG monitor devices. The EEG signal is acquired in analog format, digitized, processed and presented to the user for interpretation. An EasyPrep Sensor Kit is provided for the convenience of the user in applying the monitoring electrodes. The EasyPrep Sensor Kit is made of the same materials used in the manufacture of several medical sensors listed under 510(k) #781430 (refer to section 15 for addition details).
{2}------------------------------------------------
| Feature | NeuroFAST Monitoring<br>System,<br>Model NF-201<br>(K092477) | ZOOM-100DC<br>(082886) | Model I-2000 Monitor<br>(072382) |
|-----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The NeuroFAST Monitor NF-<br>201 is intended to be used for<br>measuring and recording the<br>electrical activity of a<br>patient's brain, obtained by<br>placing electrodes on the<br>head. The NeuroFAST NF-<br>201 is indicated for use in<br>acquiring<br>electroencephalographic<br>(EEG) signals in the OR, ICU,<br>ER, clinical settings and for<br>clinical research. The<br>NeuroFAST Monitor is to be<br>used under the direction and<br>interpretation of a licensed<br>medical professional. The<br>NeuroFAST Monitor NF-201<br>does not provide any<br>diagnostic conclusion about<br>the patient's condition. | The ZOOM-100DC is used<br>to measure and record the<br>electrical activity of a<br>patient's brain. The ZOOM-<br>100DC is intended to monitor<br>the state of the brain by<br>acquisition and display of<br>electroencephalogram (EEG)<br>signals and by the calculation<br>of standard quantitative EEG<br>(qEEG) parameters. | The I-2000 Monitor is<br>intended to be used for<br>measuring and recording the<br>electrical activity of a<br>patient's brain, obtained by<br>placing electrodes on the<br>head.<br>The I-2000 Monitor is<br>indicated for use in<br>monitoring the state of the<br>brain by acquisition of<br>electroencephalogram<br>(EEG) signals, in research<br>and clinical environments. |
| Modalities | EEG | EEG | EEG |
| Environment<br>of Use | Operating room, intensive<br>care unit, emergency room,<br>and clinical settings where<br>EEG monitoring is used. | Hospitals and Clinics | Hospitals and Clinics |
| Power Source | 120 Volt 60Hz AC power | Li Ion Battery | Battery<br>120 Volt 60Hz AC power |
| System<br>Components | Patient Module, Patient Cable,<br>Display Module, Data Cable<br>and EasyPrep Electrode Kit | ZOOM-100DC, External<br>Patient Interface Cable,<br>External Audio Cable,<br>Compact Flash Card, Battery<br>Charger | Table PC, Battery Powered<br>wireless headbox, headbox<br>charger, power conditioner |
| Sensing<br>Electrodes | Silver-silver chloride<br>disposable EEG electrodes | Standard off-the-shelf EEG<br>electrodes | Unknown type of EEG<br>electrodes |
| Feature | NeuroFAST Monitoring System, Model NF-201 (K092477) | ZOOM-100DC (082886) | Model I-2000 Monitor (072382) |
| Screen Display Details | Displays:<br>1) Raw EEG Waveforms<br>2) Spectral Parameters: EEG power spectrum and frequency bands, 50% 50% Median Edge Frequency (MEF), and 95% Spectral Edge Frequency (SEF) | Displays:<br>1) Raw EEG Waveform | Displays:<br>1) Raw EEG waveforms<br>2) Spectral Parameters: EEG power spectrum 95% Spectral Edge Frequency (SEF) |
| Stored EEG data available | Yes - through removable storage | Yes - through Compact Flash card | Yes - through removable storage |
| EEG Channels/ Montage | 2 bilateral frontal channels viewed concurrently | Up to 5 channels viewed concurrently<br>8 single-ended channels corresponding to 8 electrodes placed anywhere on the head, including but not limited to, all locations defined by the International 10/20 system. (5 different channels can be viewed concurrently) | 2 channels |
| Real Time EEG Display | Yes | Yes | Yes |
| Processed EEG Bandwidth | User Selectable<br>Low Filter: 0.125 or 0.5 Hz<br>High Filter: 30 or 70 Hz | 0.5-50 Hz | Unknown |
| Automatic Artifacting | Yes | Yes | Unknown |
| Amplifier Common Mode Rejection Ration (CMRR) | ≥100 dB | ≥ 100 dB | Unknown |
| Amplifier Input Impedance | ≥50 Meg Ohm | ≥ 100 Meg Ohms | Unknown |
| Electrode Impedance Test | Yes<br>Continuous and on demand by | Yes | Yes |
| Feature | NeuroFAST Monitoring<br>System,<br>Model NF-201<br>(K092477) | ZOOM-100DC<br>(082886) | Model I-2000 Monitor<br>(072382) |
| Derived /<br>Processed<br>EEG Measures | Power spectrum parameters<br>derived from FFT:<br>1) Power spectrum<br>displayed as Density<br>Spectral Array (DSA)<br>2) 95% Spectral Edge<br>Frequency (SEF)<br>3) Median Edge Frequency<br>(MEF)<br>4) Spectral Powers in EEG<br>frequency bands<br>traditionally used to<br>quantify EEG signals ( $\alpha$ ,<br>$\beta$ 1, $\beta$ 2, $\delta$ , $\theta$ and $\gamma$ ) | Power spectrum parameters<br>derived from FFT:<br>1) Absolute Power<br>a. Monopolar Power<br>b. Bipolar Power<br>2) Relative Power<br>a. Relative Monopolar Power<br>b. Relative Biopolar Power<br>3) Mean frequency variables<br>(univariate and multivariate)<br>a. Monopolar Mean<br>Frequency<br>b. Biopolar Mean Frequency<br>4) Coherence<br>a. Monopolar Coherence<br>b. Bipolar Coherence<br>5) Asymmetry<br>a. Monopolar Asymmetry<br>b. Bipolar Asymmetry | Power spectrum parameters<br>derived from FFT:<br>1) Power spectrum<br>displayed as a line<br>graph for each of the<br>four frequency bands<br>traditionally used to<br>quantify EEG signals<br>[Delta (0-4 Hz), Theta<br>(4-8Hz), Alpha (8-<br>13Hz) and Beta (13-30<br>Hz)]<br>2) 95% spectral edge<br>frequency (SEF) |
| Contains<br>patient<br>isolation | Yes - Patient module<br>(analog-to digital converter)<br>provides 6kV electrical<br>isolation of the patient from<br>the monitor | Battery powered, thus no<br>connection between patient<br>and mains | Yes - Wireless Bluetooth<br>communication provides<br>electrical isolation of the<br>patient from the monitor |
| Display<br>Screen | Yes - High-resolution, color,<br>graphical user interface and<br>touch screen | Yes | Yes - Graphical user<br>interface and touch screen<br>tablet PC |
| Event Markers | Yes, user selectable | Unknown | Yes |
-
,
.
:
.
:
.
•
{3}------------------------------------------------
"
,
·
ເ
the control control control of the control of
{4}------------------------------------------------
#### Non-clinical Testing 7.
Laboratory testing, performed on identical hardware to the NeuroFAST subject of this submission, demonstrated that the NeuroFAST Monitoring System, Model NF-201 meets its design and functional requirements, including IFCN Guidelines for electroencephalographs. Actual device functions and features were evaluated against the device specifications and in all instances the NeuroFAST NF-201 performed as expected and no unexpected behavior was observed. The device will meet the requirements of UL medical electrical equipment standards for safety and the IEC particular standard for electroencephalographs.
To ensure safety, the NeuroFAST Monitoring System, Model NF-201 will comply with all applicable requirements of IEC60601-1 and IEC60601-1-4, and will also comply
{5}------------------------------------------------
with the particular requirements for the safety of electroencephalographs established in IEC60601-2-26.
To ensure electromagnetic compatibility, the NeuroFAST NF-201 will comply with all applicable requirements of IEC60601-1-2, and will also comply with the particular requirements for the safety of electroencephalographs established in IEC60601-2-26.
#### 8. Clinical Testing
The NeuroFAST NF-201 device is an electroencephalographic device comprised of hardware that has been bench tested to assess safety and effectiveness and to establish substantial equivalence with the predicate devices. We believe further clinical data is not required to demonstrate performance for the NeuroFAST NF-201 for the indication for use subject to this submission.
#### 9. Conclusion
The NeuroFAST NF-201, when compared to its predicate devices, has the same intended use and similar technological characteristics and principles of operation. The nonclinical tests demonstrate that the device is safe, as effective, and performs at least as safely and effectively as the legally marketed devices. Thus, in indications, intended use and technology, the NeuroFAST NF-201 is substantially equivalent to the predicated devices. Minor technological difference as documented in the Substantial Equivalence Comparison table above, raise no new questions of safety or effectiveness.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare, to the right of the text. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement around the caduceus. The caduceus is depicted with a simple, flowing design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
NeuroWave Systems, Inc. c/o Ms. Tracie Capozzio Quality Assurance and Regulatory Compliance 2490 Lee Blvd., Suite 300 Cleveland Heights, Ohio 44118
OCT 2 9 2010
Re: K092477
> Trade/Device Name: NeuroWave NeuroFAST Monitoring System, Model NF-201 Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph (EEG) Regulatory Class: Class II Product Code: OMC Dated: October 19, 2010 Received: October 20, 2010
Dear Ms. Capozzio:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 - Ms. Tracie Capozzio
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
In A. Ramm
Melvin R. Felder, M.D.
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
## INDICATIONS FOR USE
510(k) Number (if known): K092477
OCT 2 9 2010
Device Name: NeuroFAST Monitoring System, Model NF-201
Indications For Use:
The NeuroFAST Monitoring System, Model NF-201 is intended to be used for measuring and recording the electrical activity of a patient's brain, obtained by placing electrodes on the head. The NeuroFAST NF-201 is indicated for use in acquiring electroencephalographic (EEG) signals in the OR, ICU, ER, clinical settings and for clinical research. The NeuroFAST Monitor is to be used under the direction and interpretation of a licensed medical professional. The NeuroFAST Monitoring System, Model NF-201 does not provide any diagnostic conclusion about the patient's condition.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K092477
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.