The BNA™ Platform is to be used by qualified medical professionals for the post-hoc statistical analysis of the human electroencephalogram ("EEG"), including event-related potentials ("ERPs"). This device is indicated for use in individuals 12 to 85 years of age. The BNA™ Platform is to be used with the Auditory Oddball, Visual Go No-Go (age range of 25 to 85 years), and Eyes-Closed tasks.
Device Story
Software-only platform for post-hoc statistical analysis of human EEG/ERP data; inputs include artifact-free digital EEG recordings from FDA-cleared EEG systems during Auditory Oddball, Visual Go No-Go, and Eyes-Closed tasks. Algorithm performs frequency decomposition (Delta, Theta, Alpha, Beta), time-domain peak analysis, and calculates Z-scores by comparing patient data against an age-matched normative database. Outputs include ERP, Behavioral, Summary, and Resting-State EEG reports with topographical maps and spectral analysis scores (e.g., absolute/relative power, hemispheric asymmetry). Used by clinicians in clinical settings to evaluate brain activity; results assist in clinical interpretation of patient status relative to reference groups. Cloud-based (AWS) processing; provides objective metrics to support clinical decision-making.
Clinical Evidence
Clinical study validated the Reference-Group Database and test-retest reliability using healthy volunteers (12-85 years). Normality of EEG/ERP scores confirmed using Anderson-Darling, Cramer-von Mises, Kolmogorov-Smirnov, and Shapiro-Wilk tests (success rates >97-98%). Poolability of data validated via linear regression (success rates 99-100%). Test-retest reliability established via Bland-Altman analysis.
Technological Characteristics
Software-only device; cloud-based (AWS) architecture. Performs frequency decomposition (Delta, Theta, Alpha, Beta) and time-domain peak analysis. Generates Z-scores based on normative database (1900+ subjects). Connectivity: accepts data from external FDA-cleared EEG systems. No patient-contacting components; no electrical/EMC requirements.
Indications for Use
Indicated for post-hoc statistical analysis of human EEG and ERPs in individuals aged 12-85 years. Used with Auditory Oddball, Visual Go No-Go (ages 25-85), and Eyes-Closed tasks.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
elminda, Ltd % Donna-Bea Tillman Senior Consultant Biologics Consulting 1555 King Street. Suite 300 Alexandria, Virginia 22314
Re: K202588
Trade/Device Name: BNA Platform Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OLU Dated: September 4, 2020 Received: September 8, 2020
Dear Donna-Bea Tillman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K202588
Device Name BNA™ Platform
Indications for Use (Describe)
The BNA™ Platform is to be used by qualified medical professionals for the post-hoc statistical analysis of the human electroencephalogram ("EEG"), including event-related potentials ("ERPs').
This device is indicated for use in individuals 12 to 85 years of age.
The BNA™ Platform is to be used with the Auditory Oddball, Visual Go No-Go (age range of 25 to 85 years), and Eyes-Closed tasks.
Type of Use (Select one or both, as applicable)
| <div><span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div><span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
|---------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the BNA™ Platform is provided below.
#### 1. SUBMITTER
| Applicant: | elminda Ltd. |
|---------------------------|----------------------------------|
| | 1 Arie Shenkar Street |
| | Herzliya, 4672501, Israel |
| Contact: | Keren Elghouzzi-Kazachinsky |
| | QA/RA Manager |
| | Elminda |
| | +972-53 7739780 |
| | quality@elminda.com |
| Submission Correspondent: | Donna-Bea Tillman, Ph.D. |
| | Senior Consultant |
| | Biologics Consulting |
| | 1555 King St, Suite 300 |
| | Alexandria, VA 22314 |
| | 410-531-6542 |
| | dtillman@biologicsconsulting.com |
| Date Prepared: | December 3, 2020 |
#### 2. DEVICE
| Device Trade Name: | BNA™ Platform |
|----------------------|------------------------------------------------------------|
| Device Common Name: | Normalizing Quantitative Electroencephalograph<br>Software |
| Classification Name: | 21 CFR 882.1400 Electroencephalograph, |
| Regulatory Class: | II |
| Product Code: | OLU |
#### PREDICATE DEVICE 3.
| Predicate Device: | K121119 - BNA™ Analysis System (Elminda) |
|-----------------------------|-----------------------------------------------------|
| Secondary Predicate Device: | K171414 - qEEG-Pro (BrainMasters Technologies, Inc) |
{4}------------------------------------------------
#### DEVICE DESCRIPTION 4.
The BNATM Platform is intended for the post-hoc statistical analysis of the human electroencephalogram ("EEG"), utilizing both resting-state EEG and Event-Related Potentials ("ERP") in a patient's response to outside stimuli during various states of alertness, disease, diagnostic testing, treatment, surgery, or drug related dysfunction. An Event-Related Potential (or "evoked response") is an electrical potential recorded from the nervous system following the presentation of a stimulus (e.g., as part of a cognitive task). An ERP signal consists of typical ERP components - positive or negative voltage spatiotemporal peaks within the ERP waveform that are measured within one second post-stimulus presentation. The BNATM Platform is intended to analyze EEG data recorded at rest and during the performance of two conventionally used ERP tasks, the Auditory Oddball (AOB) and the Visual Go No-Go (VGNG).
The EEG is recorded continuously while the patient is at rest with eyes-closed (hereby Eyes-Closed) or performs one of the ERP tasks (hereby ERP tasks). The acquisition site is asked to provide reliable samples of artifact-free digital EEG for purposes of analysis. After the recording, the artifact-free EEG data is imported into the BNATM Platform and is automatically analyzed by the algorithm and the results of the processed data are compiled into individualized Reports:
- . ERP Report
- Behavioral Report ●
- . Summary Report
- Resting-State EEG Report ●
Scores are presented as Z-Scores based on comparing the patient to an age-matched relevant reference group based on elminda's normative database. This presentation expresses the differences between the patient and the reference group.
The BNA™ Reports are intended to be used by clinicians to enable the evaluation of the patient's brain activity during a specific task compared to an age-range matched reference group.
The system consists of the following components: a computer environment; EEG data input software algorithms for BNA™ calculations; a report generator and a functionality for data transfer and storage.
The device processes and analyzes data received from a dedicated, commercially available, and FDA cleared EEG system, which complies with the BNATM Platform specifications.
### 5. INTENDED USE/INDICATIONS FOR USE
The BNA™ Platform is to be used by qualified medical professionals for the post-hoc statistical analysis of the human electroencephalogram ("EEG"), including event-related potentials ("ERPs").
This device is indicated for use in individuals 12 to 85 years of age.
{5}------------------------------------------------
The BNA™ Platform is to be used with the Auditory Oddball, Visual Go No-Go (age range of 25 to 85 years), and Eyes-Closed tasks.
#### 6. SUBSTANTIAL EQUIVALENCE
### Comparison of Indications
The subject device and the two predicate devices all have the same intended use, namely, to provide analyses of human EEG data. The features of the subject device have been expanded from those of the primary predicate BNA device to include additional age ranges and tasks, similar to the capabilities of the predicate qEEG-Pro device. The differences in the wording of the Indications for use reflect this expansion in features as well as a minor modification to reflect a change to the currently accepted terminology of "event-related potentials" as opposed to "evoked response potentials". These differences do not impact the intended use, and therefore the subject device has the same intended use as the predicate device.
# Technological Comparisons
The table below compares the key technological feature of the subject devices to the predicate device K121119 - BNATM Analysis System (K121119) and qEEG-Pro (K171414).
| | BNA™ Platform | BNA™ Analysis<br>System | qEEG-Pro | Discussion |
|------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Subject Device | Primary Predicate<br>Device | Secondary Predicate<br>Device | |
| Manufacturer | elminda | elminda | BrainMasters<br>Technologies, Inc | N/A |
| 510k Number | N/A | K121119 | K171414 | N/A |
| Product Code | OLU | OLU | OLU | Same as Predicate |
| Intended<br>patient<br>population | 12-85 years (for AOB<br>and Eyes closed tasks)<br>25-85 years (for VGNG) | 14-24 years | 4-82 years | Extended age range;<br>Clinical data<br>demonstrates equivalent<br>performance. |
| Tasks | AOB, VGNG, and Eyes-<br>Closed Resting-State | Auditory Oddball<br>(AOB) | Eyes-Closed and<br>Eyes-Open Resting-<br>State | Additional Tasks:<br>VGNG and<br>Resting-State EEG;<br>clinical testing<br>demonstrates equivalent<br>in safety and<br>effectiveness. |
| Intended User | Used by a Clinician /<br>Prescription device | Used by a Clinician /<br>Prescription device | Used by a Clinician<br>/ Prescription device | Same as predicate |
| | BNA™ Platform | BNA™ Analysis<br>System | qEEG-Pro | Discussion |
| | Subject Device | Primary Predicate<br>Device | Secondary Predicate<br>Device | |
| Band Passing<br>(ERP) | Frequency<br>decomposition of the<br>EEG data into: Delta,<br>Theta and Alpha<br>frequency bands. | Frequency<br>decomposition of the<br>EEG data into: Delta,<br>Theta and Alpha<br>frequency bands | Not relevant - does<br>not cover ERP<br>functions | Same as Predicate |
| Time-<br>Analysis<br>(ERP) | A time domain peak<br>analysis is performed | A time domain peak<br>analysis is performed | Not relevant (ERP<br>only) | Same as predicate |
| Analysis of<br>Band-passed<br>(ERP) | Peak analysis is applied<br>for the individual band-<br>passed time-series (i.e.,<br>peak analysis for Alpha,<br>Delta, Theta) | Peak analysis is applied<br>for the individual band-<br>passed time-series (i.e.,<br>peak analysis for Alpha,<br>Delta, Theta) | Not relevant (ERP<br>only) | Same as Predicate |
| Peak<br>Analysis<br>(ERP) | Peak-detection at the<br>level of the decomposed<br>ERP (post frequency<br>decomposition) is<br>performed for each<br>electrode separately<br>followed by selection of<br>the highest peak. | Peak-detection at the<br>level of the<br>decomposed ERP<br>(post frequency<br>decomposition) is<br>performed for each<br>electrode separately in a<br>specific time and space<br>range | Not relevant (ERP<br>only) | Performance testing<br>demonstrates correct<br>implementation. The<br>results of poolability<br>and normality<br>performance validation<br>test demonstrated<br>equivalence to the<br>success rate of<br>predicate and do not<br>raise new issue of<br>safety and effectiveness |
| Neural-<br>Consistency | The algorithm calculates<br>a new score – 'Neural<br>Consistency' based on<br>the similarity of the<br>amplitude activation<br>between single-ERP<br>trials | No; new score | Not relevant (ERP<br>only) | Performance testing<br>demonstrates correct<br>implementation. The<br>results of poolability<br>and normality<br>performance validation<br>test demonstrated<br>equivalence to the<br>success rate of<br>predicate and do not<br>raise new issue of<br>safety and effectiveness |
| Reported<br>ERP peaks | The highest ERP peaks<br>within literature-based<br>time constraints, which<br>show the highest spatio-<br>temporal similarity to the | The ERP peaks at<br>network-based, pre-<br>defined electrodes<br>appear in the report | Not relevant (ERP<br>only) | Performance testing<br>demonstrates correct<br>implementation. The<br>results of poolability<br>and normality<br>performance validation<br>test demonstrated<br>equivalence to the<br>success rate of |
| | BNATM Platform | BNATM Analysis<br>System | qEEG-Pro | Discussion |
| | Subject Device | Primary Predicate<br>Device | Secondary Predicate<br>Device | |
| | group ERP appear in the<br>report | | | equivalence to the<br>success rate of<br>predicate and do not<br>raise new issue of<br>safety and effectiveness |
| Network<br>Analysis | No network analysis | Network analysis based<br>on the ERP-peaks as<br>the network nodes | Not relevant (ERP<br>only) | Feature removed |
| Topographica<br>l maps of<br>ERP | Subject ERP-peak<br>position is displayed on a<br>topographical map<br>together with the group<br>ERP-peak position. | Z-scores are presented<br>as topographical-maps<br>(electrode coordinates).<br>Networks activity over<br>time is presented in<br>time evolution maps | Not relevant (ERP<br>only) | Performance testing<br>demonstrates correct<br>implementation. The<br>results of poolability<br>and normality<br>performance validation<br>test demonstrated<br>equivalence to the<br>success rate of<br>predicate and do not<br>raise new issue of<br>safety and effectiveness |
| ERP<br>Waveforms | Broadband (0.5-30Hz)<br>and band-pass (Delta,<br>Theta and Alpha) ERP<br>waveforms of the subject<br>are visualized in<br>comparison to the<br>averaged ERP waveform<br>of the age-matched<br>reference group. | Broadband (0.5-30Hz)<br>ERP waveforms of the<br>subject are visualized in<br>comparison to the<br>averaged ERP<br>waveform of the age-<br>matched reference<br>group. | Not relevant (ERP<br>only) | Performance testing<br>demonstrates correct<br>implementation. The<br>results of poolability<br>and normality<br>performance validation<br>test demonstrated<br>equivalence to the<br>success rate of<br>predicate and do not<br>raise new issue of<br>safety and effectiveness |
| ERP<br>waveforms<br>display by<br>stimulus type | The ERP waveform<br>display is divided into<br>sections by stimulus type | The ERP waveform<br>display is divided into<br>sections by stimulus<br>type | Not relevant (ERP<br>only) | Same as Predicate |
| Display of<br>representative<br>ERPs | Within each stimulus-<br>section, 19 panels<br>represent 19 selected<br>electrode locations on<br>the scalp. These are<br>electrode positions<br>described in the<br>international 10-20 | Within each stimulus-<br>section 20 panels<br>represent 20 selected<br>electrode locations on<br>the scalp. These are<br>electrode positions<br>described in the<br>international 10-20 | Not relevant (ERP<br>only) | Performance testing<br>demonstrates correct<br>implementation. |
| | BNA™ Platform | BNA™ Analysis System | qEEG-Pro | Discussion |
| | Subject Device | Primary Predicate Device | Secondary Predicate Device | |
| | system of electrode placement. | system of electrode placement. | | |
| Comparison to Normative Database | Yes; 1900 and 1407 subjects covering the age-range 12-85 for the tasks AOB and Eyes-Closed Resting-State EEG, respectively. In addition, 1005 subjects covering the age-range 25-85 for the VGNG task | Yes; 120 subjects in the age-range 14-24 AOB task only | Yes: Z-scores are calculated based on reference-group mean and std. (Not for ERP) | Reference database expanded to include additional subjects; performance testing demonstrates validity of revised database |
| Resting-State EEG data Comparison against the Normative Database | Yes; 1407 subjects covering the age-range 12-85 for Eyes-Closed Resting-state EEG | No | Yes; 1482 samples (eyes-closed); 1231 subjects (eyes-open) | Clinical data demonstrates equivalent performance |
| Resting-State EEG Spectral-Analysis | Yes; 4 frequency bands (delta, theta, alpha, and beta) | No | Yes; 4 frequency bands (delta, theta, alpha, and beta) | Same as predicate |
| Spectral Analysis scores | Yes, the following scores are extracted from PSD: Average Absolute Power, Relative Power, Individual Alpha Frequency, Hemispheric Asymmetry and Frequency Ratios | No | Yes; The following scores are extracted from PSD: Absolute Power, Relative Power, Alpha Peak Frequency, Hemispheric Asymmetry, Coherence and Frequency Ratios | The Average Absolute Power score is well-known in the scientific literature and was statistically validated for normality. |
| Age Range Included in the Normative Database | 12-85 years | 14-24 years | 4-82 years | The existing normative database has been extended to include additional age-ranges. The additional age-bins and tasks were clinically validated. |
| | BNA™ Platform | BNA™ Analysis<br>System | qEEG-Pro | Discussion |
| | Subject Device | Primary Predicate<br>Device | Secondary Predicate<br>Device | |
| Visual<br>Display of<br>EEG | Yes: topographical maps of Average Absolute and Relative Power, Individual Alpha Frequency, Hemispheric Asymmetry and Frequency Ratios | No | Yes: topographical maps of Absolute and Relative Power, Power Asymmetry, and Coherence for 19 monopolar and 171 selected bipolar derivations of the EEG | The features are equivalent and the results of performance testing demonstrate equivalent performance |
| Hardware | Runs on the AWS cloud and accepts EEG data from all sites | Standalone computer running the BNA™ engine locally and contains the EEG/ERP data locally | | The results of System and Software Test Verification demonstrate that the subject device meets the defined requirements |
Table 1: Technological Comparison
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
In conclusion, the differences in technological characteristics do not raise new questions of safety and effectiveness and the BNA™ Platform can be found substantially equivalent to the predicate devices.
#### 7. PERFORMANCE DATA
### Biocompatibility Testing
BNA™ Platform is a software only device. There are no direct or indirect patient-contacting components of the subject device. Therefore, patient contact information is not needed for this device.
### Sterilization and Shelf Life
BNA™ Platform is a software only device. Therefore, sterilization and shelf life are not applicable.
# Electrical safety and electromagnetic compatibility (EMC)
Not applicable. BNA™ Platform is a software only device. The device contains no electric components, generates no electrical emissions, and uses no electrical energy of any type.
### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this
{10}------------------------------------------------
device was considered as a moderate level of concern. Testing was conducted to ensure the product works as designed and to validate the performance of the device.
## Bench Testing
Not applicable. Bench testing was not necessary to establish the substantial equivalence of this device.
## Animal Testing
Not applicable. Animal studies were not necessary to establish the substantial equivalence of this device.
# Clinical Data
The objective of the clinical study was to establish and validate the Reference-Group Database and to evaluate its test-retest reliability.
Healthy volunteers aged 12-85 years old were recruited from 13 clinical sites, with 12 sites located within the US. Depending on the site's protocol, the volunteers either participated at a single visit, or at two successive visits, separated by approximately 1 week to provide data for the re-test database. Brain electrophysiological activity was collected using the BNA 100 Platform while the participants performed the Eyes-Closed (EC) task and one or more cognitive task, including the Auditory Oddball (AOB) task (using the hearing adjusted version for participants older than 50 years) and the Visual Go-No-Go (VGNG) task.
The Reference-Group Database for the AOB and VGNG tasks is divided into the following age groups: 25-35, 35-50, 50-65, 65-75, and 75-85 years for VGNG, and 12-14, 14-16, 16-18, 18-25, 25-35, 35-50, 50-65, 65-75, and 75-85 years for AOB. For the Resting-EEG, reference groups were created at 0.5-year resolution by taking 133 overlapping bins covering the range 12-85.
EEG/ERP scores were transformed and validated for their normal distribution by using the same methodology in accordance with the predicate device: using four normality tests Anderson-Darling, the Cramer-von Mises, the Kolmogorov-Smirnov, and the Shapiro-Wilk tests. A score was considered to be normally distributed if at least 2 of the four tests were not statistically significant (p-value > 0.05) Normality of the EEG Scores was also tested using a Gaussian leaveone-out sensitivity test.
The normality test results are in accordance with the success rates presented in the predicate device statistical performance and, from a clinical perspective, allow for an accurate clinical interpretation of z-scores.
All Resting-EEG and ERP scores pass the 'two out of four' method tests (i.e., passing at least 2 out of 4 different normality tests) with a success rate above 97% and 98%, respectively, with Resting-EEG scores also passing the Gaussian leave-one-out sensitivity tests with a success rate larger than 97.5% an acceptable percentage of failures, given the large number of scores tested. These results are in accordance with the success rates presented in the predicate device statistical performance and allow for an accurate clinical interpretation of z-scores.
Poolability of the Reference Group data was tested with a linear regression model for each combination of ERP score or behavioral performance score and age-bin and checked for cases
{11}------------------------------------------------
with a significant age-effect (p<0.05) following False Discovery Rate (FDR) correction. All AOB and VGNG ERP scores pass the poolability test with a success rate of 99% and 100%, respectively. This is an acceptable percentage of failures, given the large amount of scores tested, allowing for a reliable division of the entire dataset into age groups.
The Bland-Altman analysis of the Test-Retest database provides test-retest reliability statistical estimates, for all combinations of tasks, conditions and parameters. These test-retest reliability results can be used by clinicians to estimate changes in BNA™ Scores between successive visits.
#### CONCLUSION 8.
Based on the detailed comparison to the predicate devices, the performance testing, and the clinical testing, the BNA™ Platform can be found substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.