K110410 · Nihon Kohden Corp. · GYC · Aug 4, 2011 · Neurology
Device Facts
Record ID
K110410
Device Name
MS-120BK ELECTRIC STIMULATOR (EXTENSION UNIT)
Applicant
Nihon Kohden Corp.
Product Code
GYC · Neurology
Decision Date
Aug 4, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1310
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
When the Low output is selected the MS-120BK is used as a nerve stimulator for surgical procedures and brain mapping during treatment of patients with seizure disorders. When High output is selected the MS-120BK is used for the intraoperative diagnosis of acute dysfunction in corticospinal axonal conduction brought about by mechanical trauma (traction, shearing, laceration, or compression) or vascular insufficiency. The system is intended for use by qualified medical personnel within a hospital or clinical environment. The stimulator is available for use on any patients as determined by the qualified medical personnel.
Device Story
The Nihon Kohden MS-120BK is a dual-output constant current electric stimulator used as an extension unit for the MEE-1000A system. It functions as a nerve stimulator for surgical brain mapping or intraoperative diagnosis of corticospinal conduction dysfunction. The device receives control signals from the MEE-1000A via an amplifier (JB-116BK or JB-132BK). It delivers electrical stimulation pulses through a stimulation pod (JS-102B) or extension cord (BM-121B) to external electrodes. Operated by qualified medical personnel in clinical/hospital environments, the device allows selection of stimulation current, frequency, and pulse parameters. Output affects clinical decision-making by enabling real-time mapping and diagnostic monitoring during surgery, potentially benefiting patients by identifying functional brain areas or assessing nerve conduction integrity during trauma or vascular events.
Clinical Evidence
Bench testing only. No clinical data provided. Verification and validation testing performed in compliance with design control processes, including adherence to IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-40 standards for medical electrical equipment and safety.
Technological Characteristics
Dual-output constant current stimulator; 350V max voltage; 200mA max current; 50 Ω output impedance. Supports monophasic, biphasic, and alternate pulse modes. Pulse duration 50-1000 μs; frequency 0.1-50Hz. Connectivity via MEE-1000A interface. Complies with IEC 60601-1, 60601-1-2, and 60601-2-40 standards.
Indications for Use
Indicated for surgical nerve stimulation and brain mapping in patients with seizure disorders (Low output) and intraoperative diagnosis of acute corticospinal axonal conduction dysfunction due to trauma or vascular insufficiency (High output). Use by qualified medical personnel in hospital/clinical settings on any patient population.
Regulatory Classification
Identification
A cortical electrode is an electrode which is temporarily placed on the surface of the brain for stimulating the brain or recording the brain's electrical activity.
Nihon Kohden SEN-4100 Electric Stimulator (K071969)
Submission Summary (Full Text)
{0}------------------------------------------------
Nihon Kohden America, Inc. July 25, 2011
Traditional 510(k) Notification MS-120 Electric Stimulator
# K110410
### 510(k) Summary
ﺒﻪ .
| Company Name: | Nihon Kohden Corporation<br>90 Icon Street<br>Foothill Ranch, CA 92610 | AUG - 4 2011 |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------|--------------|
| Device Name: | Nihon Kohden MS-120BK Electric Stimulator (Extension Unit) | |
| 510(k) Sponsor,<br>Contact: | Nihon Kohden America, Inc.<br>90 Icon Street<br>Foothill Ranch, CA 92610 | |
| | Steve Geerdes<br>Director Quality Assurance and Regulatory Affairs<br>Phone: (949) 580-1555 Ext. 3325<br>Fax: (949) 580-1550 | |
| Summary Date: | 7/25/2011 | |
| Common Name: | MS-120BK Electric Stimulator (Extension Unit) | |
## Classification Names:
| Evoked Response Electrical Stimulator | 882.1870 | GWF |
|---------------------------------------|----------|-----------------|
| Cortical Electrode | 882.1310 | GYC<br>Class II |
## Predicate Device(s):
| Nicolet Cortical Stimulator | K072964 |
|-----------------------------------------------|---------|
| Ojemann Cortical Stimulator | K924226 |
| Digitimer D185 Multipulse Cortical Stimulator | K020400 |
| Axon Eclipse Neurological workstation | K050798 |
| Nihon Kohden SEN-4100 Electric Stimulator | K071969 |
{1}------------------------------------------------
, l
#### . 1.0 Description of Device
The Nihon Kohden MS-120BK is a dual (high & low) output constant current stimulator. The high and low setting, stimulation current and frequency of stimulation is selected by the user. The Nihon Kohden MS-120BK is connected to the MEE 1000A through the JB-116BK or JB-132BK amplifier.
In the Low output setting the Nihon Kohden MS-120BK applies cortical stimulation energy through the Nihon Kohden stimulation pod (JS-102B) which is connected to commercially available cortical electrodes (strip and grid electrodes).
In the High output setting, the MS-120BK outputs electrostimulation pulse through the Nihon Kohden extension cord (BM-121B) which is connected to commercially available stimulation electrode(s).
### 2.0 Intended Use of Device
When the Low output is selected the MS-120BK is used as a nerve stimulator for surgical procedures and brain mapping during treatment of patients with seizure disorders.
When High output is selected the MS-120BK is used for the intraoperative diagnosis of acute dysfunction in corticospinal axonal conduction brought about by mechanical trauma (traction, shearing, laceration, or compression) or vascular insufficiency.
The system is intended for use by qualified medical personnel within a hospital or clinical environment. The stimulator is available for use on any patients as determined by the qualified medical personnel.
#### 3.0 Technical Characteristics
The technical characteristics of the MS-120BK are equivalent to those of the predicate devices. The following tables summarizes equivalence.
| Stimulator Comparison | Predicate Device<br>Nihon Kohden<br>SEN-4100A (K#071969) | Digitimer D185 | New Device<br>Nihon Kohden<br>MS-120BK |
|------------------------|----------------------------------------------------------|------------------|----------------------------------------|
| Stimulator Output | | | |
| Voltage | 1000V | 1000V | 350V |
| Current | 1A (1k Ω load) | 1A (1k Ω load) | 200mA (1k Ω load) |
| Maximum Stimulation | | | |
| Energy | 50mJ | 50mJ | 40mJ |
| Output Impedance | less than 120 Ω | less than 120 Ω | 50 Ω |
| Stimulation Mode | Constant Voltage | Constant Voltage | Constant Current |
| Stimulation Parameters | | | |
{2}------------------------------------------------
స
| Pulse Duration | 50 μs | 50 μs | 50-1000 μs |
|-----------------------------------------------|-----------------------------------------------------|-----------------------------------------------------|-------------------------------------------------------------|
| Output Frequency Range | 1 Hz | 1 Hz | 0.1-1Hz |
| Number of Pulse Trains | 1-9 | 1-9 | 1-9 |
| Interpulse Interval | 1.0-9.9ms | 1.0-9.9ms | 1.0-1000ms |
| Limitation of Voltage /<br>Current and Pulses | 1000V: 3 Pulses<br>510V: 6 Pulses<br>350V: 9 Pulses | 1000V: 3 Pulses<br>510V: 6 Pulses<br>350V: 9 Pulses | 200mA: 3 Pulses*1<br>142mA: 6 Pulses*1<br>115mA: 9 Pulses*1 |
| Polarity Change | OK | OK | OK |
| Monophasic | OK | OK | OK |
| Biphasic | NO | NO | OK |
| Alternate | NO | NO | OK |
| Maximum Charge | 50 micro-Coulomb | 50 micro-Coulomb | 200 micro-Coulomb |
| Pulse Shape | Rectangular | Rectangular | Rectangular |
| Maximum Energy Per Pulse | 50mJ | 50mJ | 40mJ |
| Maximum Energy Per<br>Second | 150mJ | 150mJ | 40mJ |
| Electrode Minimum Size | 0.5cm2 | 0.5cm2 | 0.5cm2 |
| Maximum Charge Density | 100 Micro-C/cm2 | 100 Micro-C/cm2 | 400 Micro-C/cm2 |
| Interface<br>Control From EP/EMG<br>Machine | MEE-1000A | NO | Only MEE-1000A |
| Trigger Input | OK | OK | OK*2 |
| Trigger Output | OK | OK | OK*2 |
| Foot Switch Control | OK | OK | OK*2 |
*1: The pulse width in the case
<sup>. *2:</sup> Depends on interface of MEE-1000A
| Stimulator Comparison | Predicate Device Nicolet<br>Cortical Stimulator<br>(K#072964) | Ojemann<br>Cortical Stimulator<br>(K#924226) | New Device<br>Nihon Kohden<br>MS-120BK |
|------------------------------------------|---------------------------------------------------------------|----------------------------------------------|----------------------------------------|
| Constant Current Stimulator | Yes | Yes | Yes |
| Maximum Stimulation Charge | 15 micro-Coulomb | 20 micro-Coulomb | 4.5 micro-Coulomb |
| Current Stimulation Range | 0.1 to 15mA (peak) | 0 to 10mA (peak) | 0 to 15mA (peak) |
| Stimulation Frequency | 1 to 100Hz | 5, 10, 20, 50, 75,<br>100Hz | 0.1 to 50Hz |
| Stimulation Pulse Width Duration | 0.1 to 1.0 msec Per<br>Phase | 0.1 to 2.0 msec Per<br>Phase | 0.05 to 0.3 msec Per<br>Phase |
| Pulse Shape | Rectangular | Rectangular | Rectangular |
| Electrode Minimum Size | 0.04cm2 | 0.5cm2 | 0.04cm2 |
| Maximum Charge Density | 375 micro-C/cm2 | 40 micro-C/cm2 | 113 micro-C/cm2 |
| Maximum RMS Current Density<br>Per Pulse | 119 mA RMS/cm2 | 8.9 mA RMS/cm2 | 46 mA RMS/cm2 |
of 1ms
{3}------------------------------------------------
#### 4.0 Data Summary
Testing of the Nihon Kohden System with the MS-120BK was performed in compliance with the Nihon Kohden Corporation design control process. Testing included:
> Software and hardware verification and validation, and the device is in compliance with the following voluntary industrial standards:
Medical Electrical Equipment Par1: General requirements for safety IEC 60601-J 1988 IEC 60601-1, Amendment 1 Part 1: General Requirements for safety, Amendment 1, 1991 IEC 60601-1, Amendment 2 Part 1: General Requirements for safety, Amendment 2, 1995 IEC 60601-1-2-40 * Part 2-40: Particular Requirements for the safety of electromyographs and evoked response equipment, 1998 IEC 60601-1-1 200 edition: Part 1-1: General requirements for safety -Collateral standard Safety requirements for medical electrical systems, 2000 IEC 60601-1-2 2" edition: Medical electrical equipment- Part 1: General requirements for safety - 2. Collateral standard: Electromagnetic compatibility, 2001 EN 60601-1: 1990 Equivalent to IEC 60601-1: 1998* EN 60601-1 Amendment 1 Medical equipment general requirements for safety, 1993 EN 60601-1 Amendment 2 Medical equipment general requirements for safety, 1995 EN 60601-1-1: 2001 Equivalent to IEC 60601-1:2001 Electromagnetic compatibility requirement test EN 60601-1-2:2001 CAN/CSA-C22.2 No. 601.1-M90 Medical electrical equipment, Part 1: General requirements for safety. CAN/CSA-C22.2 No. 601.1S1-94 Supplement No. 1-94 to CAN/CSA-C22.2 No. 601-1-M90 Medical Equipment- Part 1:General requirements for safety. CAN/CSA-C22.2 No. 60601-1-1-02 Medical electrical equipment, Part 1-1: General requirements for safety- Collateral: Safety requirements for medical electrical systems, 2006 CAN/CSA-C22.2 No. 60601-2-40-01 Medical electrical equipment, Part 2-40: Particular requirements for safety of electromyographs and evoked response equipment (adopted 60601-2-40: 1998 CAN/CSA-C22.2 No. 601.1B-90 Amendment 2 to CAN/CSA-C22.2 No. 601.1-M90 Medical equipment Part 1: General requirements for safety: 2002
5.0 Conclusions
{4}------------------------------------------------
Substantial equivalence of the Nihon Kohden MS-120BK was demonstrated by testing in compliance with the Design Control process. The intended use and technology of the Nihon Kohden MS-120BK is equivalent to the predicate devices. No new questions of safety or effectiveness are raised.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows a logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & H." are arranged in a semi-circle around the caduceus. The image is in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Steve Geerdes Director Regulatory Affairs and Quality Assurance Nihon Kohden America, Inc. 90 Icon St. Foothill Ranch. CA 92610
AUG - 4 2011
Re: K110410
Trade/Device Name: Nihon Kohden MS-120BK Electric Stimulator (Extension Unit) Regulation Number: 21 CFR 882.1310 Regulation Name: Cortical Electrode Regulatory Class: Class II Product Code: GYC and GWF Dated: June 27, 2011 Received: June 28, 2011
Dear Mr. Geerdes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{6}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm11.5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Keoia Alexander
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Indications for Use Form
510(k) Number (if known):
Device Name: _ _ MS-120BK Electric Stimulator (Extension Unit)
Indications for Use:
When the Low output is selected the MS-120BK is used as a nerve stimulator for surgical procedures and brain mapping during treatment of patients with seizure disorders.
When High output is selected the MS-120BK is used for the intraoperative diagnosis of acute dysfunction in corticospinal axonal conduction brought about by mechanical trauma (traction, shearing, laceration, or compression) or vascular insufficiency.
The system is intended for use by qualified medical personnel within a hospital or clinical environment. The stimulator is available for use on any patients as determined by the qualified medical personnel.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Kristen Bowsher
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
Page 1 of ।
510(k) Number_K 110410
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.