K072964 · Cardinal Health, Inc. · GYC · Dec 30, 2008 · Neurology
Device Facts
Record ID
K072964
Device Name
NICOLET CORTICAL STIMULATOR
Applicant
Cardinal Health, Inc.
Product Code
GYC · Neurology
Decision Date
Dec 30, 2008
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 882.1310
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cortical Stimulator is intended for use in functional brain mapping procedures during treatment of patients with seizure disorder, providing stimulation via electrode pairs or a hand held bipolar probe.
Device Story
Nicolet Cortical Stimulator is a constant current electrical stimulator used for functional brain mapping. It consists of a Stimulus Switching Unit (SSU) and a Cortical Stimulator Control Unit (CSCU). The device delivers intermittent electrical stimulation to the cerebral cortex via temporary intracranial strip/grid electrodes or a hand-held bipolar probe. Used in hospitals, operating rooms, neurology clinics, and epilepsy labs by clinicians. The CSCU provides the stimulation source; the SSU directs current to selected electrodes, interfacing with EEG amplifiers and headboxes. During procedures, clinicians observe the patient and record EEG to identify eloquent brain areas, helping surgeons avoid critical tissue during resection. The device helps manage patients with seizure disorders by facilitating safe surgical planning.
Clinical Evidence
No clinical data provided. Safety and effectiveness were demonstrated through bench testing, including software and hardware verification and validation, safety standard compliance, and a review of scientific literature.
Technological Characteristics
Constant current stimulator; 0.1 to 15 mA peak current; 1 to 100 Hz frequency; 0.1 to 1.0 msec pulse width per phase; 15 micro-Coulomb max charge. Components: Stimulus Switching Unit and Cortical Stimulator Control Unit. Interfaces with standard cortical electrodes (min 2.3 mm diameter) and EEG amplifiers. Powered by external supply.
Indications for Use
Indicated for functional brain mapping in patients with seizure disorders to identify eloquent brain areas prior to surgical resection of epileptic foci or tumors.
Regulatory Classification
Identification
A cortical electrode is an electrode which is temporarily placed on the surface of the brain for stimulating the brain or recording the brain's electrical activity.
{1}------------------------------------------------
XO7296Y
# 510(k) Summary
# DEC 3 0 2008
| Company Name: | Cardinal Health, Inc.<br>NeuroCare<br>5225 Verona Road, Building 2<br>Madison, WI 53711 |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Nicolet Cortical Stimulator |
| 510(k) Sponsor, Contact: | Cardinal Health, Inc.<br>NeuroCare<br>5225 Verona Road<br>Madison, WI 53711<br>Robert Burdge<br>Director Quality Assurance and Regulatory Affairs<br>Phone: (608) 441-153<br>Fax: (608) 426-5172 |
| Summary Date: | January 2, 2009 |
| Common Name: | Cortical Stimulator |
| Classification Name: | Cortical Electrode, CFR 882.1310, Product Code: GYC, Class II |
| Predicate Device(s): | K924226 Ojemann Cortical Stimulator<br>K040360 XLTEK EMU128 Switch Matrix |
#### 1.0 Description of Device
The Nicolet Cortical Stimulator supports clinician guided, intermittent electrical stimulation of the brain cortex to support diagnostic and surgical procedures. The cortical stimulation energy is applied through commercially available cortical electrodes (strip and grid electrodes) of a minimum 2.3 mm diametor and a Cardinal Health, NeuroCare stimulation probe. All stimulation is temporary and intermittent, applied in support of functional brain mapping procedures during treatment of patients with seizure disorder.
The Nicolet Cortical Stimulator is a constant current stimulator. The stimulation current and frequency of stimulation is selected by the user. The Nicolet Cortical Stimulator (Cortical Stimulator) consists of two components:
- 1) Stimulus Switching Unit and
- 2) Cortical Stimulator Control Unit.
{2}------------------------------------------------
The Cortical Stimulator Control Unit (CSCU) is powered by an external power supply. The CSCU provides the current stimulation source. The Stimulus Switching Unit (SSU) directs the stimulation to selected electrodes.
The Cortical Stimulation Control Unit can be used in a stand-alone mode with stimulus provided through the Cardinal Health, NeuroCare stimulation probe. Alternatively the Cortical Stimulation Control Unit can be connected to the Stimulation Switching Unit. When the Stimulus Switching Unit is attached, the Stimulus Switching Unit interfaces between the Nicolet C64 Electroencephalography (EEG) amplifier and the amplifier's associated headbox. The headbox provides the interface between the amplificr and the brain contacting electrodes.
#### 1.1 Clinical Application
Prior to and during a surgical procedure to remove epileptic foci or a tumor of the brain, a cortical mapping procedure may be applied. Cortical stimulation may be applied to determine epileptic scizure foci that need to be surgically removed, tissue excised, as the foci are not managed with medications. The cortical stimulator is used with intra cranial cortical electrodes in the form of strips and/or grids temporarily placed on the surface of the patient's cerebral cortex. Having established a "focus" for the patient's epilepsy, the physician performs a number of procedures to establish if surgery is a viable option. Cortical mapping is one of those procedures. Cortical stimulation in support of cortical mapping may be applied during the monitoring period, in the patient epileptic monitoring area.
During cortical mapping electrical stimulation of the patient's cortex is performed through pairs of electrodes temporarily placed on the surface of the brain or by a hand held bipolar stimulator probe. The electroencephalograph (EEG) is recorded during the procedure and the patient is observed in person and through video so as to note and document clinical signs, bchavioral changes and other possible effects, such as speech changes. Thus, the surgeon is able to identify eloquent brain areas and can avoid these if resection of brain tissue is performed.
File: Cortcal Stun 510k reply 1-2-2009
{3}------------------------------------------------
The Cortical Stimulator is used in hospitals and clinical environments such as operating rooms, Neurology Clinics and Epilepsy Labs to support clinical evaluation of brain function.
### 2.0 Intended use of Device
The intended use of the Cortical Stimulator Unit is:
The Cortical Stimulator is intended for use in functional brain mapping procedures during treatment of patients with seizure disorder, providing stimulation via electrode pairs or a hand held bipolar probe.
# 3.0 Technological Characteristics
The technical characteristics of the Cortical Stimulator are equivalent to those of the predicate devices. The following table summarizes equivalence.
| Feature | Cortical<br>Stimulator<br>Under Review | Predicate<br>Ojemann Cortical<br>Stimulator<br>(K924226) | Predicate<br>XLTEK EMU128<br>Switch Matrix<br>(K040360) | Comments |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use,<br>Indications for Use | The Cortical<br>Stimulator is<br>intended for use in<br>functional brain<br>mapping procedures<br>during treatment of<br>patients with seizure<br>disorder, providing<br>stimulation via<br>electrode pairs or a<br>hand held bipolar<br>probe. | The OCS-I is<br>intended for use<br>during surgical<br>procedures such as<br>placement of<br>electrodes and brain<br>mapping during<br>treatment of patients<br>with seizure disorder. | The EMU128 Switch<br>Box used with the<br>EMU128S<br>electroencephalograp<br>h system to support<br>electrode switching<br>for brain mapping<br>studies. | Equivalent<br>The predicate<br>Ojemann Cortical<br>Stimulator is a<br>constant current<br>stimulation source.<br>The predicate<br>EMU128 Switch Box<br>provides switching of<br>an external<br>stimulation source to<br>electrodes. |
| Constant Current<br>Stimulator | Yes | Yes | Not Applicable<br>An external<br>stimulator is applied<br>to this switching<br>device to support<br>stimulation. | Same |
| Maximum<br>Stimulation Charge | 15 micro-Coulomb | 20 micro-Coulomb | Not Applicable<br>An external<br>stimulator is applied<br>to this switching<br>device to support<br>stimulation. | Equivalent<br>The maximum<br>stimulation charge<br>applied is lower than<br>the predicate<br>Ojemann Cortical<br>Stimulator |
{4}------------------------------------------------
| Feature | Cortical<br>Stimulator<br>Under Review | Predicate<br>Ojemann Cortical<br>Stimulator<br>(K924226) | Predicate<br>XLTEK EMU128<br>Switch Matrix<br>(K040360) | Comments |
|-------------------------------------|-----------------------------------------|----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Current Stimulation<br>Range | 0.1 to 15 mA (peak)<br>Constant Current | 0 to 10 mA (peak)<br>Constant Current | Not Applicable<br>An external<br>stimulator is applied<br>to this switching<br>device to support<br>stimulation. | Equivalent<br>The total stimulation<br>energy applied is<br>limited to a<br>maximum of 15<br>micro-Coulomb (15<br>mA at 1 msec). This<br>is less than the<br>predicate Ojemann<br>Cortical Stimulator<br>20 micro-Coulomb<br>(10 mA at 2 msec). |
| Stimulation<br>Frequency | 1 to 100 Hz | 5, 10, 20, 50, 75, 100<br>Hz | Not Applicable<br>An external<br>stimulator is applied<br>to this switching<br>device to support<br>stimulation. | Equivalent |
| Stimulation Pulse<br>Width Duration | 0.1 to 1.0 msec per<br>phase | 0.1 to 2.0 msec per<br>phase | Not Applicable<br>An external<br>stimulator is applied<br>to this switching<br>device to support<br>stimulation. | Equivalent |
## 4.0 Data Summary
Testing of the Nicolet System with the Cortical Stimulator was performed in compliance with the Cardinal Health, Inc. NeuroCare design control process. Testing included:
- I. Software and hardware verification and validation, and
- 2. Safety standard compliance.
A summary of scientific literature supporting the safety of the Nicolet Cortical Stimulator was provided.
#### 5.0 Conclusions
The safety and effectiveness of the Nicolet Cortical Stimulator was demonstrated by testing in compliance with the Design Control process. The intended use and technology of the Nicolet Cortical Stimulator is the same as the predicate devices. No new questions of safety or effectiveness are raised.
File: Cortical Stim 510k reply 1-2-2009
{5}------------------------------------------------
Image /page/4/Picture/12 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized emblem that resembles a caduceus or a staff with a serpent entwined around it. The emblem is rendered in black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Cardinal Health, Inc. % Quality & Regulatory Associates, Inc. Mr. Gary Syring 800 Levanger Lane Stoughton, Wisconsin 53589
Re: K072964
Trade/Device Name: Nicolet Cortical Stimulator Regulation Number: 21 CFR 882.1310 Regulation Name: Cortical electrode. Regulatory Class: Class II Product Code: GYC, ETN Dated: December 23, 2008 Received: December 29, 2008
DEC 3 0 2008
Dear Mr. Syring:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the includions for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisiopp of the 1.0. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misburanti, isamd adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, TDAcova publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2 - Mr. Gary Syring
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicated device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket survillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmanlet Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or Internet oddress http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark H. Wilkerson
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Indications for Use
510(k) Number (if known): ____________________________
Device Name: Nicolet Cortical Stimulator
Indications for Use:
The Cortical Stimulator is intended for use in functional brain mapping procedures during treatment of patients with seizure disorder, providing stimulation via electrode pairs or a hand held bipolar probe.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Signature of Sign Off
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
**510(k) Number** K072964
Page F-1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.