K151504 · Gn Otometrics · GWN · Nov 16, 2015 · Neurology
Device Facts
Record ID
K151504
Device Name
ICS Impulse
Applicant
Gn Otometrics
Product Code
GWN · Neurology
Decision Date
Nov 16, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1460
Device Class
Class 2
Indications for Use
The ICS Impulse System Model 1085 is used in the assessment of the vestibular-ocular reflex (VOR) and nystagmus by measuring, recording, displaying, and analyzing eye and head movements.
Device Story
ICS Impulse System Model 1085 consists of lightweight goggles with integrated high-speed USB video camera, infrared LED for eye illumination, and motion sensor (gyroscope/accelerometer). Used by clinicians to perform vestibular testing, including Video Head Impulse Test (vHIT), Oculomotor, and Positional tests. During testing, clinician rotates patient's head while patient fixates on target; goggles capture eye video and head movement data. OTOsuite Vestibular software processes inputs to calculate slow phase velocities (SPV) and eye position shifts. System provides simultaneous displays of head and eye movement data, allowing clinicians to assess VOR function and nystagmus. Supports both vision and vision-denied testing environments. Output aids clinical decision-making regarding vestibular system health. Device operates via PC connection.
Clinical Evidence
Clinical evidence based on literature review. Bench testing included electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and biocompatibility of patient-contact materials. Software validation and risk analysis were performed. No primary clinical trial data reported.
Technological Characteristics
Lightweight goggles with USB video camera, infrared LED, and motion sensor (gyroscope/accelerometer). Half-silvered mirror for eye imaging. Connectivity via USB to PC. Software-based analysis of eye/head movement. Biocompatible patient-contact materials. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2.
Indications for Use
Indicated for the assessment of vestibular-ocular reflex (VOR) and nystagmus in patients requiring clinical evaluation of eye and head movements.
Regulatory Classification
Identification
A nystagmograph is a device used to measure, record, or visually display the involuntary movements (nystagmus) of the eyeball.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of three human profiles facing right, stacked on top of each other. The profiles are rendered in black and have a flowing, abstract design. Surrounding the graphic is text arranged in a circular pattern, reading "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 16, 2015
GN Otometrics % Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct Naples, Florida 34114
Re: K151504
Trade/Device Name: ICS Impulse System Model 1085 Regulation Number: 21 CFR 882.1460 Regulation Name: Nystamograph Regulatory Class: Class II Product Code: GWN Dated: May 29, 2015 Received: October 8, 2015
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# William J. Heetderks -S
for
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151504
Device Name ICS Impulse System Model 1085
Indications for Use (Describe)
The ICS Impulse System Model 1085 is used in the assessment of the vestibular-ocular reflex (VOR) and nystagmus by measuring, recording, displaying, and analyzing eye and head movements.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div style="display:inline-block;"> <input checked="" type="checkbox" value=""/></div> Residential Use (Rentals of 30+ Consecutive Days) |
|------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:inline-block;"> <input type="checkbox" value=""/></div> Commercial Use (Rentals of 1-29 Consecutive Days) |
X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(K) Summary, 510(k) K15 Submitter: GN Otometrics A/S Hoerskaetten 9, Taastrup, DENMARK DK-2630 Registration number: 9612197 C/O GN Otometrics North America 50 Commerce Dr Ste 180 Schaumburg, IL 60173 (US) Phone: 847-534-2150 (US) Fax: 847-534-2153 Contact: Anders Rasmussen, Manager of Research and Development Date Prepared: May 15, 2015
## 1. Identification of the Device:
Proprietary-Trade Name: ICS Impulse Model 1085 Classification Name: Code GWN, Nystagmograph, Regulation 882.1460 Common/Usual Name: Nystagmograph
### 2. Equivalent legally marketed devices:
| 510(k) # | Name of Device | Manufacturer |
|----------|------------------------------|---------------|
| K061791 | Type 1068 ICS Chartr 200 VNG | GN Otometrics |
| K122550 | ICS Type 1085 Impulse | GN Otometrics |
- 3. Description of the Device: The device is a combination of hardware and software. The patient wears a pair of lightweight, tightly-fitting goggles on which is mounted a very small, very light, very fast, USB video camera and a half silvered mirror. This transparent mirror reflects the image of the patient's eye into the camera. The eye is illuminated by a low-level infra-red light emitting diode which is not visible to the patients. A small sensor on the goggles measures the head movement. The whole goggle system is lightweight but it must be secured tightly to the head to minimize goggle slippage. The software records and displays the information obtained during what is known as a "head impulse test" The basic head impulse test starts with the tester standing behind the patient who is wearing the goggles. While the patient is asked to stare at the fixation dot placed on a projection surface in front of them, the tester rotates the patient's head horizontally through a small angle (about 10-20 degrees) in a brief, abrupt and unpredictable manner, varying the direction and the velocity. The goggles collect both head and eye data. The gyroscope measures the velocity of the head movement (the stimulus). The high-speed camera captures the image of the eye. The OTOsuite Vestibular software processes the head velocity data and velocity data for eye movement (the response). Simultaneous displays of the data for head movement and for eye movement allow the clinician to determine if the response is within normal limits or not. The software also records and displays the information obtained during Positional and Oculomotor tests. A Positional test is performed by moving that patient from one position to another position. In the example of Dix-Hallpike, the patient is sitting and the patient's head is turned 45 degrees to one side and then the patient is moved from the sitting position to the supine position. An Oculomotor test is performed by having the patient stare in various directions or under various environments. In the example of Gaze, the patient is sitting and the patient is asked to stare left, right, up, down or center. The Gaze test can be performed with vision or with vision denied. For both Positional and Oculomotor the goggles collect head and eye data. The accelerometer in the small sensor measures where the head is in space. That information is taken and the patient's head position or any movement during testing is displayed in the software. The high speed camera captures the image of the eye. The OTOsuite Vestibular software processes the eye velocity data (the response). The eye movement is
{4}------------------------------------------------
analyzed to determine the slow phase velocities (SPV). The head data is only used during collection to display if the patient's head is moving and to guide the tester to position the patient's head appropriately for the test. This is what we refer to as "Head Position Feedback". Tests where slow phase velocity is measured display the eve position trace and slow phase velocity beats in a graph. In Oculomotor there are 2 tests VOR and Skew Deviation that are not SPV tests. VOR (vestibular ocular reflex) which allows for both visual VOR (VVOR) and VOR suppression (VORS). This test is very similar to the head impulse test but the head movement is slow (0.5 Hz) and small (10 degrees). In VVOR the patient is sitting and the examiner moves the head from side to side (like a sinusoid) while the patient stares at a fixed target. In VORS the patient is sitting and the examiner moves the head from side to side (like a sinusoid) while the patient stares at a moving dot projected from the goggle using one of the lasers. The analysis is similar to head impulse, simultaneous displays of the data for head movement and for eye movement allow the clinician to determine if the response is within normal limits or not. In Skew Deviation (also known as cover test or alternate cover test) the patient is sitting and the tester covers and uncovers. The OTOsuite Vestibular software measures the eye position trace during the cover and uncover environments and displays an average eye position shift.
- 4. Indications for Use (intended use): The ICS Impulse Model 1085 is used in the assessment of the vestibular-ocular reflex (VOR) and nystagmus by measuring, displaying, and analyzing eye and head movements.
- 5. Safety and Effectiveness, comparison to predicate device. This device has the same core hardware and software as our K122550 and adds the ability to record nystagmus as in our other predicate K061791. Video camera interface has been changed from firewire to USB. This is not a meaningful difference. A "vision denied" solution (see table below) has been added which is functionally the same as in K061791.
- 6. Description of Testing: The device platform (1085) had previously passed UL Electrical Safety testing and EMC testing. Software validation and risk analysis was performed. Biocompatibility testing was successfully performed on the patient contact materials. A clinical analysis based on published literature. The ICS Impulse adequately meets the design requirements and acceptance criteria.
| Characteristic | GN Otometrics A/S<br>Type 1068 ICS Chartr<br>200 VNG K061791 | GN Otometrics A/S ICS<br>Type 1085 Impulse<br>K122550 | GN Otometrics A/S<br>Type 1085 ICS Impulse |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Focus VNG/ENG<br>is a nystagmograph<br>that is intended to<br>measure, record and<br>display involuntary<br>movements<br>(nystagmus) of the<br>eyeball. | The ICS Impulse System<br>is used in the assessment<br>of the vestibular-ocular<br>reflex (VOR) by<br>measuring, recording,<br>displaying, and analyzing<br>eye and head<br>movements. | The ICS Impulse System is<br>used in the assessment of the<br>vestibular-ocular reflex<br>(VOR) and nystagmus by<br>measuring, recording,<br>displaying, and analyzing<br>eye and head movements. |
| Product Code | GWN | GWN/LXV | GWN |
| Characteristic | GN Otometrics A/S<br>Type 1068 ICS Chartr<br>200 VNG K061791 | GN Otometrics A/S ICS<br>Type 1085 Impulse<br>K122550 | GN Otometrics 1068 A/S<br>Type 1085 ICS Impulse |
| Configuration | ICS Chartr 200 VNG<br>consists of a hardware<br>box and VG-40<br>binocular video<br>goggles. The goggles<br>can record one eye<br>(monocular) or two<br>eye binocular. The<br>ICS Chartr 200 cannot<br>measure head<br>movement | The patient wears a pair<br>of lightweight, tightly-<br>fitting goggles on which<br>is mounted a very small,<br>very light, very fast,<br>firewire video camera<br>and a half silvered<br>mirror. This transparent<br>mirror reflects the image<br>of the patient's eye into<br>the camera. The eye is<br>illuminated by a low-<br>level infra-red light<br>emitting diode which is<br>not visible to the<br>patients. A small sensor<br>on the goggles measures<br>the head movement. The<br>whole goggle system is<br>lightweight but it must be<br>secured tightly to the<br>head to minimize goggle<br>slippage. | The patient wears a pair of<br>lightweight, tightly-fitting<br>goggles on which is mounted<br>a very small, very light, very<br>fast, USB video camera and<br>a half silvered mirror. This<br>transparent mirror reflects<br>the image of the patient's<br>eye into the camera. The eye<br>is illuminated by a low-level<br>infra-red light emitting diode<br>which is not visible to the<br>patients. A small sensor on<br>the goggles measures the<br>head movement. The whole<br>goggle system is lightweight<br>but it must be secured tightly<br>to the head to minimize<br>goggle slippage. |
| Test Battery | The ICS Chartr 200 is<br>a full VNG system<br>which allows for the<br>following types of<br>tests: Oculomotor,<br>Positional, and<br>Calorics. The system<br>allows for recording of<br>the eye with vision and<br>vision denied. The<br>software includes a<br>slow phase velocity<br>algorithm developed in<br>collaboration with Dr<br>Stockwell. | The ICS Impulse system<br>allows for the following<br>types of tests: recording<br>the eye video and room<br>video as known as<br>Monocular Video<br>Frenzel, and Video Head<br>Impulse Test (vHIT).<br>The system allows for<br>recording of the eye with<br><b>vision only</b> . The system<br>also allows for recording<br>of the room through the<br>use of a webcam. A<br>monocular video frenzel<br>(only allows recording of<br>the eye and room with no<br>data analysis). | The ICS Impulse system<br>allows for the following<br>types of tests: Monocular<br>Video Frenzel, Video Head<br>Impulse Test (vHIT)<br>Oculomotor and Positional.<br>The system allows for<br>recording of the eye with<br><b>vision and vision denied</b> .<br>The system also allows for<br>recording of the room<br>through the use of a<br>webcam. A monocular video<br>frenzel (only allows<br>recording of the eye and<br>room with no data analysis).<br>The software includes a slow<br>phase velocity algorithm<br>developed in collaboration<br>with Dr Barin a protégé of<br>Dr Stockwell. |
| Characteristic | GN Otometrics A/S<br>Type 1068 ICS Chartr<br>200 VNG K061791 | GN Otometrics A/S ICS<br>Type 1085 Impulse<br>K122550 | GN Otometrics 1068 A/S<br>Type 1085 ICS Impulse |
| Photo | Image: GN Otometrics A/S Type 1068 ICS Chartr 200 VNG K061791 | Image: GN Otometrics A/S ICS Type 1085 Impulse K122550 | Image: GN Otometrics 1068 A/S Type 1085 ICS Impulse |
| Vision Denied<br>Solution | The VG-40 goggles<br>have a shield that can<br>be opened and closed. | There was no vision<br>denied solution | The ICS Impulse has a<br>solution with a cup and a<br>patch. The cup adheres to<br>the right (test) eye. The cup<br>has a window that allows the<br>IR camera to see through in<br>order to record eye<br>movement but the patient<br>can not see through it. The<br>patch covers the left (non-<br>test) eye and the patient can<br>not see through it. |
| Electrical safety | IEC 60601-1 | SAME | SAME |
| EMC | IEC 60601-1-2 | SAME | SAME |
| Weight | 15.2 oz (430 g) | 2.1 oz (60g) - goggles | 2.1 oz (60g) - goggles |
| Power Source | USB via PC | SAME | SAME |
| Computer | Microsoft® XP<br>Professional - Service Pack<br>2, Vista® Business or<br>Windows 7 Professional 32<br>or 64 bit | Windows XP 32-bit<br>Professional SP3 or Windows<br>7 32-bit Professional or<br>Windows 7 64-bit<br>Professional | 64-bit: Windows 8 Pro, Windows<br>7 Professional<br>32- bit: Windows 8 Pro, Windows<br>7 Professional |
| Operator | Trained personnel required | SAME | SAME |
## 7. Substantial Equivalence Chart
{5}------------------------------------------------
{6}------------------------------------------------
- Conclusion: After analyzing software validation, risk analysis, bench testing (including 8. biocompatibility), safety, EMC, and clinical validation (literature review) testing we conclude that the ICS Impulse is as safe and effective as the predicate devices, and has essentially the same indications for use and technology, thus rendering it substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.