K122550 · Gn Otometrics · GWN · Feb 1, 2013 · Neurology
Device Facts
Record ID
K122550
Device Name
ICS IMPULSE
Applicant
Gn Otometrics
Product Code
GWN · Neurology
Decision Date
Feb 1, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1460
Device Class
Class 2
Indications for Use
The ICS Impulse System is used in the assessment of the vestibular-ocular reflex (VOR) by measuring, recording, displaying, and analyzing eye and head movements.
Device Story
ICS Impulse is a vestibular testing device consisting of lightweight goggles with an integrated high-speed FireWire video camera, infrared LED for eye illumination, and a gyroscope sensor for head movement tracking. During a 'head impulse test,' a clinician rotates the patient's head in brief, unpredictable, horizontal movements while the patient fixates on a target. The system captures head velocity (stimulus) and eye movement (response) data. OTOsuite Vestibular software processes these inputs to display synchronized head and eye movement data. Clinicians use these displays to assess VOR function and determine if responses are within normal limits. The device aids in diagnosing vestibular disorders by quantifying eye-head coordination, providing objective data to support clinical decision-making regarding balance and dizziness.
Clinical Evidence
Clinical validation testing compared ICS Impulse test results to Scleral Search Coils test results. The device met design requirements and acceptance criteria. No specific sensitivity/specificity metrics provided.
Technological Characteristics
Goggles with integrated high-speed FireWire camera, infrared LED, and gyroscope. Calibration via two built-in Class II lasers at +/- 7.5 degrees. Connectivity: FireWire (camera), USB 2.0 (data). Electrical safety: UL 60601-1, IEC 62471, IEC 60825-1. EMC: IEC 60601-1-2: 2007. Software: OTOsuite Vestibular.
Indications for Use
Indicated for assessment of the vestibular-ocular reflex (VOR) in patients requiring evaluation of eye and head movements for dizziness or balance problems. Prescription use only.
Regulatory Classification
Identification
A nystagmograph is a device used to measure, record, or visually display the involuntary movements (nystagmus) of the eyeball.
Micromedical Technologies Inc. VisualEyes (K964325)
Submission Summary (Full Text)
{0}------------------------------------------------
510(K) Summary, 510(k) K122550 Submitter: GN Otometrics A/S Hoerskaetten 9 Taastrup, DENMARK DK-2630 Registration number: 9612197 C/O GN Otometrics North America 50 Commerce Dr Ste 180 Schaumburg. IL 60173 (US) Phone: 847-534-2150 (US) Fax: 847-534-2153 Contact: Dan Sansonetti, Manager of Research and Development Date Prepared: January 13, 2012
FEB 0 1 2013
Identification of the Device: 1. Proprietary-Trade Name: ICS Impulse Classification Name: Class II, Product Codes: GWN and LXV, Device: Nystagmograph Common/Usual Name: Vestibular testing device
- Equivalent legally marketed devices: Micromedical Technologies Inc. Vorteq, K891008 and 2. Micromedical Technologies Inc. VisualEyes K964325.
- 3. Description of the Device: The device is a combination of hardware and software. The patient wears a pair of lightweight, tightly-fitting goggles on which is mounted a very small, very light, very fast, fire-wire video camera and a half silvered mirror. This transparent mirror reflects the image of the patient's eye into the camera. The eye is illuminated by a low-level infra-red light emitting diode which is not visible to the patients. A small sensor on the goggles measures the head movement. The whole goggle system is lightweight but it must be secured tightly to the head to minimize goggle slippage. The software records and displays the information obtained during what is known as a "head impulse test" The basic head impulse test starts with the tester standing behind the patient who is wearing the goggles. While the patient is asked to stare at the fixation dot placed on a projection surface in front of them, the tester rotates the patient's head horizontally through a small angle (about 10-20 degrees) in a brief, abrupt and unpredictable manner, varying the direction and the velocity. The goggles collect both head and eye data. The gyroscope measures the velocity of the head movement (the stimulus). The high-speed camera captures the image of the eve. The OTOsuite Vestibular software processes the head velocity data and velocity data for eye movement (the response). Simultaneous displays of the data for head movement and for eye movement allow the clinician to determine if the response is within normal limits or not.
- 4. Indications for Use (intended use): The ICS Impulse System is used in the assessment of the vestibular-ocular reflex (VOR) by measuring, recording, displaying, and analyzing eve and head movements. (Prescription use).
- 5. Safety and Effectiveness, comparison to predicate device. This device has the same indications for use as the predicate device but employs different technology to accomplish the same tasks.
- 6. Description of Testing: The device passed UL Electrical Safety testing and EMC testing. Software validation and risk analysis was performed. Clinical testing compared test results to Scleral Search Coils test results. ICS Impulse adequately meets the design requirements and acceptance criteria.
{1}------------------------------------------------
## 7. Substantial Equivalence Chart
| Characteristic | Micromedical Technologies<br>Inc. Vorteq, K891008 and<br>VisualEyes K964325. | ICS Impulse |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: | VORTEQ® is designed to<br>provide information about the<br>Vestibular Ocular Reflex<br>(VOR) in patients with<br>dizziness or balance problems. | The ICS Impulse System is used in the assessment of<br>the vestibular-ocular reflex (VOR) by measuring,<br>recording, displaying, and analyzing eye and head<br>movements. |
| Configuration | VORTEQ® utilizes an angular<br>velocity sensor mounted<br>directly to the VisualEyes™<br>FireWire Binocular Goggles.<br>With the VisualEyes™<br>Monocular Goggles, the<br>angular velocity sensor is<br>attached to the back of the<br>goggles headband for<br>VORTEQ® testing | The patient wears a pair of lightweight, tightly-fitting<br>goggles on which is mounted a very small, very light,<br>very fast, fire-wire video camera and a half silvered<br>mirror. This transparent mirror reflects the image of<br>the patient's eye into the camera. The eye is<br>illuminated by a low-level infra-red light emitting<br>diode which is not visible to the patients. A small<br>sensor on the goggles measures the head movement.<br>The whole goggle system is lightweight (only about<br>60g) but it must be secured tightly to the head to<br>minimize goggle slippage |
| Photo | Image: Person wearing goggles with camera attached | Image: Goggles with camera attached |
| Interfaces | Firewire for Camera<br>Data: Not specified | Firewire for Camera<br>USB 2 for Data |
| Electrical<br>safety | Electrical Safety per UL2601 -<br>IEC-60601. | Complies with UL 60601-1, IEC 62471, 1st.ed., IEC<br>60825-1, 2.ed. UL Listed |
| EMC | Not specified | IEC 60601-1-2: 2007 |
| Calibration | Performed using a Digital<br>Lightbar, LCD projector or<br>Secondary monitor. Stimulus +/-<br>15 degrees for horizontal and +/-<br>10 degrees for vertical | Performed using 2 Built-In Laser (2) Class II @ +/-7.5<br>degrees. |
- 8. Conclusion: After analyzing bench testing, safety, EMC, software, and clinical validation testing we conclude that the ICS Impulse is as safe and effective as the predicate device, and has essentially the same indications for use, thus rendering it substantially equivalent to the predicate device.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS) of the United States. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
February 1, 2013
GN Otometrics A/S % Mr. Daniel Kamm, P.E. Principal Engineer Kamm & Associates 8870 Ravello Court Naples, FL 34114
Re: K122550
Trade/Device Name: ICS Impulse Regulation Number: 21 CFR 882.1460 Regulation Name: Nystagmograph Regulatory Class: II Product Code: GWN, LXV Dated: January 21, 2013 Received: January 23, 2013
Dear Mr. Daniel Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
1f your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r hat FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Mr. Daniel Kamm, P.E.
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Joyce M. Whang
for Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): K122550
Device Name: ICS Impulse
Indications For Use:
The ICS Impulse System is used in the assessment of the vestibular-ocular reflex (VOR) by measuring, recording, displaying, and analyzing eye and head movements.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
## Joyce M. Whanq
(Division Sign Off) Division of Neurological and Physical Medicine Devices (DNPMD)
510(k) Number K122550
Page 1 of 1
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.