CAMINO INTRACRANIAL PRESSURE MONITORING CATHETER WITH LICOX IMC BOLT FITTING, MODEL 110-4L

K022553 · Integra Neurosciences · GWM · Aug 20, 2002 · Neurology

Device Facts

Record IDK022553
Device NameCAMINO INTRACRANIAL PRESSURE MONITORING CATHETER WITH LICOX IMC BOLT FITTING, MODEL 110-4L
ApplicantIntegra Neurosciences
Product CodeGWM · Neurology
Decision DateAug 20, 2002
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 882.1620
Device ClassClass 2

Indications for Use

The Camino® Intracranial Pressure for Use Monitoring Catheter with Licox IMC Bolt Fitting, Model 110-4L is indicated for use by a qualified Neurosurgeon when direct measurement of intracranial pressure in the parenchyma or subarachnoid space is clinically important. The Camino® Intracranial Pressure Monitoring Catheter with Licox IMC Bolt fitting, Model 110-4L is intended to be used only through a Licox Brain Oxygen Monitoring System Bolt.

Device Story

Camino 110-4L is a sterile, transducer-tipped catheter for rapid, continuous intracranial pressure (ICP) monitoring. Used in clinical settings by neurosurgeons; requires insertion through a Licox Brain Oxygen Monitoring System Bolt. Device connects to Camino ICP monitors (models MPM-1 or V420) via cables to provide real-time pressure data. Enables direct parenchymal or subarachnoid pressure measurement; assists clinicians in managing neurological patients where ICP monitoring is critical. Benefits include accurate, continuous diagnostic data for clinical decision-making.

Clinical Evidence

Bench testing only. Biocompatibility studies conducted per FDA G95-1 and ISO 10993. Mechanical testing included pressure, pull, and bend tests to verify design integrity.

Technological Characteristics

Sterile, transducer-tipped pressure monitoring catheter. Materials verified via ISO 10993 biocompatibility standards. Designed for integration with Licox Brain Oxygen Monitoring System Bolt and Camino MPM-1/V420 monitors. Mechanical design validated through pressure, pull, and bend testing.

Indications for Use

Indicated for use by qualified neurosurgeons for direct intracranial pressure measurement in the parenchyma or subarachnoid space when clinically important. Intended for use only through a Licox Brain Oxygen Monitoring System Bolt.

Regulatory Classification

Identification

An intracranial pressure monitoring device is a device used for short-term monitoring and recording of intracranial pressures and pressure trends. The device includes the transducer, monitor, and interconnecting hardware.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ 510(k) Premarket Notification Camino® Model 110-4L Integra NeuroSciences # AUG 2 0 2002 CONFIDENTIAL # Ko22553 # Camino® Intracranial Pressure Monitoring Catheter with LICOX® IMC Bolt Fitting # Model 110-4L ### 510(k) SUMMARY #### Submitter's name and address: Integra NeuroSciences 5955 Pacific Center Blvd. San Diego, CA 92121, USA #### Contact person and telephone number: Nancy A. Mathewson, Esq. Director, Regulatory Affairs (858) 622-2737 #### Date summary was prepared: August 1, 2002 #### Name of the device: | Proprietary Name: | Camino® Intracranial Pressure Monitoring Catheter with<br>LICOX® IMC Bolt Fitting, Model 110-4L | |-----------------------|-------------------------------------------------------------------------------------------------| | Common Name: | Intracranial Pressure Monitoring Kit | | Classification Name: | Intracranial Pressure Monitoring Device<br>Product Code GWM, 21 CFR 882.1620 | | Classification Panel: | Neurology Device Panel | #### Substantial Equivalence: The modified Camino® Intracranial Pressure Monitoring Catheter with LICOX® IMC Bolt Fitting, Model 1104L is substantially equivalent in function and intended use to the original Camino® Intracranial Pressure Monitoring Catheter with LICOX® IMC Bolt Fitting, Model 110-4L. ## Device Description: The Camino® 110-4L Catheter is a sterile transducer-tipped pressure monitoring catheter with accessory items to be used as a diagnostic tool for rapidly determining and continually monitoring intracranial pressure. The 110-4L is to be {2}------------------------------------------------ 510(k) Premarket Notification Camino® Model 110-4L Integra NeuroSciences > used in conjunction with LICOX® IM2 or IM3 bolt and Camino® Intracranial Pressure Monitors model MPM-1 or V420 and their associated cables. #### Statement of Intended Use: The Camino® Intracranial Pressure for Use Monitoring Catheter with Licox IMC Bolt Fitting, Model 110-4L is indicated for use by a qualified Neurosurgeon when direct measurement of intracranial pressure in the parenchyma or subarachnoid space is clinically important. The Camino® Intracranial Pressure Monitoring Catheter with Licox IMC Bolt fitting, Model 110-4L is intended to be used only through a Licox Brain Oxygen Monitoring System Bolt. #### Safety: Biocompatibility studies were conducted per FDA G95-1 and ISO 10993 and have demonstrated that the materials used to manufacture the Camino® 110-4L catheter are safe for its intended use. In addition, the Camino® 110-4L catheter was subjected to extensive mechanical testing, which included pressure, pull and bend tests. Results of the testing showed that the catheter design was mechanically sound and safe for its intended use. The Camino® 110-4L catheter manufacturing process complies with the United States Food and Drug Administration and European Standards for the manufacturing of medical devices. ## Conclusion: The modified Camino® 110-4L catheter is substantially equivalent to the original Camino® 110-4L catheter. The modifications do not affect the intended use or the fundamental scientific technology of the device. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around it. Inside the circle is an abstract symbol that resembles an eagle or a bird in flight, with three human profiles incorporated into the design. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG 2 0 2002 Integra NeuroSciences Nancy A. Mathewson, Esq. Director, Regulatory Affairs 5955 Pacific Center Boulevard San Diego, California 92121 Re: K022553 Trade/Device Name: Camino® Intracranial Pressure Monitoring Catheter with LICOX® IMC Bolt Fitting, Model 110-4L Regulation Number: 882.1620 Regulation Name: Intracranial Pressure Monitoring Devicce Regulatory Class: Class II Product Code: GWM Dated: August 1, 2002 Received: August 2, 2002 Dear Ms. Mathewson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ Page 2 – Ms. Nancy A. Mathewson This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Mark N Milburn Celia M. Witten, Ph.D., M.D. Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # APPENDIX B ## Indications for Use Statement #### 510(K) K022553 Number Device Name: Camino® Intracranial Pressure Monitoring Catheter with LICOX® IMC Bolt Fitting, Model 110-4L The Camino® Intracranial Pressure for Use Monitoring Catheter with Indications Licox IMC Bolt Fitting, Model 110-4L is indicated for use by a qualified Neurosurgeon when direct measurement of intracranial pressure in the parenchyma or subarachnoid space is clinically important. The Camino® Intracranial Pressure Monitoring Catheter with Licox IMC Bolt fitting, Model 110-4L is intended to be used only through a Licox Brain Oxygen Monitoring System Bolt. # PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE) Counter Prescription Use (Per 21 CFR 801. 109) vision of General, Restorative and Neurological Devices 11 510(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...