The CGX Quick-20m is intended to be used to acquire the electroencephalogram (EEG) and transmit it wirelessly to a computer.
Device Story
The Quick-20m is a wireless, battery-operated EEG headset using dry sensor technology in a 10-20 montage. It captures EEG signals from a seated patient, amplifies them, and transmits the data wirelessly to a computer for review by a healthcare professional. The device is designed for use in clinics or homes, allowing patient movement during data collection. It features 24-bit resolution at 500 Hz and shielding to reject ambient electrical noise. Healthcare providers use the transmitted EEG data to support clinical decision-making. The device benefits patients by providing high-quality EEG acquisition with minimal scalp preparation.
Clinical Evidence
No clinical data was required. Evidence consists of bench testing (Intertek certification to IEC 60601-1, 60601-1-2, 60601-2-26, 80601-2-26), internal design verification, and a human factors usability program (in-field testing, academic studies, and service/repair data).
Technological Characteristics
Wireless EEG headset with 20 monopolar channels in 10-20 layout. Uses active dry Ag/AgCl electrodes. 24-bit Delta-Sigma A/D conversion at 500 Hz/channel. Input impedance >100 MOhm; CMRR >90 dB at 60 Hz. Powered by two 1.5V AAA batteries. Cap material: HP Multi-Jet Fusion PA11 and PA12. Connectivity: Wireless to receiver, USB to computer. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-26, 80601-2-26, ISO 14971, IEC 62304, ISO 10993-1.
Indications for Use
Indicated for the acquisition of electroencephalogram (EEG) signals in patients, suitable for general-purpose EEG in home or healthcare facility settings.
Regulatory Classification
Identification
A physiological signal amplifier is a general purpose device used to electrically amplify signals derived from various physiological sources (e.g., the electroencephalogram).
Predicate Devices
g.tec medical engineering GmbH g.Nautilus PRO (K171669)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
CGX, LLC % Prithul Bom Regulatory Technology Services, LLC 1000 Westgate Drive, Suite #510k Saint Paul, Minnesota 55114
Re: K203331
Trade/Device Name: Quick-20m Regulation Number: 21 CFR 882.1835 Regulation Name: Physiological signal amplifier Regulatory Class: Class II Product Code: GWL, GXY Dated: May 4, 2021 Received: May 5, 2021
Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203331
Device Name Quick-20m
Indications for Use (Describe)
The Quick-20m is intended to be used to acquire the electroencephalogram (EEG) and transmit it wirelessly to a computer.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a logo with the letters "CGX" inside a circular shape. The circle is divided vertically into two halves, with the left half in a gradient of light to dark teal and the right half in a solid blue color. The letters "CGX" are white and positioned in the center of the circle, with each letter taking up roughly one-third of the space.
### 510(k) Summary
| Submitter: | CGX, LLC |
|------------------------|------------------------------------------------------|
| Address: | 8445 Camino Santa Fe, Suite 104, San Diego, CA 92121 |
| Phone number: | 858-864-9400 |
| Contact person: | Spencer Linton |
| Phone number: | (858) 864-9400 |
| Email: | Spencer@CGXSystems.com |
| Date prepared: | June 23, 2020 |
| Trade name: | Quick-20m |
| Common name: | Physiological Signal Amplifier |
| Product Code, Primary: | GWL, GXY |
| Regulation: | 21 CFR 882.1835 |
Substantial equivalence claimed to: g.tec medical engineering GmbH g.Nautilus PRO, K171669, a Class II device.
## Device Description:
The Quick-20m is a wireless, battery-operated 10-20 montage EEG headset utilizing dry sensor technology. The headset provides an integrated approach to the wireless acquisition of EEG signals. A seated patient is free to exhibit natural movements while real-time data is collected. The Quick-20m includes advanced amplification and shielding to reject ambient electrical noise.
The headset obtains high-quality EEG with minimal scalp preparation. Patented mechanisms and a range of replaceable dry sensors align to various head shapes and sizes, maintaining sensor positions in a standard 10-20 layout. EEG channels are digitized with 24 bits of resolution at 500 Hz. The Quick-20m is suitable for general-purpose EEG.
## Principle of Operation
The Quick-20m headset is a wireless physiological signal amplifier that integrates cutaneous electrodes in a 10-20 montage. The headset captures EEG from the electrodes with 24 bits of resolution at 500 Hz, amplifies the signal and sends the data to a computer for further analysis by a healthcare professional.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a logo with the letters 'CGX' inside a circle that is split vertically into two halves. The left half of the circle is a light teal color, while the right half is a darker blue. The letters 'CGX' are white and are positioned in the center of the circle, with 'C' on the left half and 'GX' on the right half. The overall design is simple and modern.
The environment of use for this device is in a home or healthcare facility setting, such as a doctor's office. The device is not sterile.
The CGX Quick-20m works in the same manner as the predicate device, the g.tec medical engineering GmbH g.Nautilus PRO (K171669), with no safety concerns.
## Intended Use(s)
The CGX Quick-20m is intended to be used to acquire the electroencephalogram (EEG) and transmit it wirelessly to a computer.
## Technological Characteristics
The CGX Quick-20m is substantially equivalent to the predicate device, the g.tec medical engineering GmbH g.Nautilus PRO for wireless transmission of EEG to a computer. It utilizes dry sensors on 20 monopolar channels to acquire EEG data via a headset worn on the head and transmits this data, via 24-bit Delta-Sigma A/D conversion at a sampling rate of 500 Hz/channel, wirelessly to a computer. The Quick-20m is battery powered with exactly the same CMRR, noise and power-on LED indicators as the predicate. The CGX Quick-20m meets all the applicable safety standards as the predicate, as listed in the Substantial Equivalence Table below with the same system components, patient connections/inputs and firmware/driver software as the predicate.
| Item | CGX Quick-20m<br>This Submission | g.tec medical<br>engineering GmbH<br>g.Nautilus PRO | Substantial<br>Equivalence<br>Comments |
|--------------|-------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|
| Intended Use | The Quick-20m is<br>intended to be used to<br>acquire the<br>electroencephalogram<br>(EEG) and transmit it<br>wirelessly to a<br>computer | The g.Nautilus PRO is<br>intended to be used to<br>acquire the<br>electroencephalogram<br>(EEG) and transmit it<br>wirelessly to a<br>computer | Same as predicate |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows a logo with the letters "CGX" inside of a circle. The circle is divided vertically into two halves. The left half is a light teal color, and the right half is a darker blue color. The letters "CGX" are white and are placed in the center of the circle, overlapping both halves.
| EEG/Polygraphic<br>Channels | 20 monopolar<br>channels. Electrode<br>grid is standard 10-20<br>layout | Up to 32 monopolar<br>channels, different<br>electrode grids<br>available providing 8,<br>16 or 32 channels and<br>32 or 16 channels CSP<br>layout on predefined<br>positions | Fewer maximum<br>number of channels<br>but equivalent in<br>safety and<br>effectiveness |
|-------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
| DC Channel | All | All | Same as predicate |
| Full Scale Input<br>Range | ± 300 mV | ± 187.5 mV to<br>± 2.25 V | Fixed range but<br>equivalent in safety<br>and effectiveness |
| A/D Conversion | 24-Bit Delta-Sigma | 24 Bit Delta-Sigma | Same as predicate |
| Sampling Rate | 500 Hz/channel | User selectable (250,<br>500 Hz/channel) | Similar to predicate<br>but equivalent in<br>safety and<br>effectiveness |
| CMRR | >90 dB at 60 Hz | >90 dB at 60 Hz | Same as predicate |
| Noise | <0.6 µV RMS,<br>1-30 Hz | <0.6 µV RMS | Same as predicate |
| Power Supply | Battery, 3.0 V DC<br>(Two 1.5 V AAA) | Battery, 3.7 V DC | Different battery<br>chemistry but<br>equivalent in safety<br>and effectiveness |
| Rated Power<br>Consumption | 0.3 VA | 0.5 VA | Less power<br>consumption but<br>equivalent in safety<br>and effectiveness |
| Amplifier-PC<br>Interface | Wireless to receiver,<br>USB to computer | Wireless to receiver,<br>USB to computer | Same as predicate |
| Other Interfaces | Power on LED | Power on LED | Same as predicate |
| Use Standard<br>Sensors And<br>Electrodes | Electrodes are<br>included (Ag/AgCl) -<br>Commercially<br>available sensors | Electrodes and cap are<br>included (Ag/AgCl or<br>golden dry electrodes,<br>elastic cap with certain<br>electrode positions) | Similar to predicate<br>but equivalent in<br>safety and<br>effectiveness |
| Dimension | 20 (L) x 18 (W) x 19<br>(H) cm | 80 (L) x 60 (W) x 27<br>(H) mm | Larger dimensions<br>but similar in safety<br>and effectiveness |
| Weight | <600 g | <200 g | More weight but<br>equivalent in safety<br>and effectiveness |
| Isolation | wireless, patient<br>isolation | wireless, patient<br>isolation | Same as predicate |
| System<br>Components | Amplifier/Digitization/<br>Electrodes/Cap<br>Charging Device +<br>USB cable Receiver +<br>USB cable | Amplifier/Digitization/<br>Electrodes/Cap<br>Charging Device +<br>USB cable Receiver +<br>USB cable | Same as predicate |
| Firmware / Driver<br>Software | Resident | Resident / CD | Same as predicate |
| Digital<br>Inputs/Outputs | 16 inputs, all patient<br>separated, no outputs | 8 inputs, all patient<br>separated, no outputs | More digital inputs<br>but equivalent in<br>safety and<br>effectiveness |
| Stimulation Unit<br>Input/Output | Not available | Not available | Same as predicate |
| Patient Connection<br>And Inputs | Receiver:<br>USB - 1 connector<br>DIGITAL IN - 1<br>connector | Receiver:<br>USB - 1 connector<br>DIGITAL IN - 1<br>connector | Same as predicate |
| Type Of Applied<br>Part | BF | BF | Same as predicate |
| Impedance<br>Measurement | Performed with<br>125 Hz | Performed with 10 Hz | Impedance check<br>performed at<br>different frequency<br>but similar in safety<br>and effectiveness |
| Input Impedance | >100 MOhm | >100 MOhm | Same as predicate |
| Filters | DC up to 131 Hz | DC up to 200 Hz<br>(depending on<br>sampling frequency) | Fewer digital filters<br>but equivalent in<br>safety and<br>effectiveness |
| Frequency<br>Response | Linear between 0.1<br>and 100 Hz | Linear between 0.1<br>and 100 Hz | Same as predicate |
| Environment Of<br>Use | Electrophysiological | Electrophysiological | Same as predicate |
| Where Used | On the head | On the head | Same as predicate |
| Number Of<br>Possible Electrodes | 20 Channels in<br>standard 10-20 head<br>map | Different electrode<br>grids available<br>providing 8, 16 or 32<br>channels and 32 or 16<br>channel CSP layout on<br>predefined positions | Fewer number of<br>possible electrodes<br>but equivalent in<br>safety and<br>effectiveness |
| Size Of Caps | One size for adults:<br>52-62 cm | Various for infants,<br>children (mini, midi,<br>maxi) and adults<br>(small, medium,<br>large), head<br>circumference: 32-62<br>cm | Smaller range of<br>sizes but equivalent<br>in safety and<br>effectiveness |
| Style Of Caps | Cap is full mechanical<br>headset | Full head cap | Different style of<br>construction but<br>similar in safety and<br>effectiveness |
| Performance<br>Requirements | Needs to transmit<br>electrophysiological<br>signals from an<br>individual to data<br>collection device | Needs to transmit<br>electrophysiological<br>signals from an<br>individual to data<br>collection device | Same as predicate |
| Cap Material | HP Multi-Jet Fusion<br>PA11 and PA12 | Oeko-Tex certificate | Other material but<br>equivalent in safety<br>and effectiveness |
| Biocompatibility<br>Testing On Patient<br>Contact Materials | Evaluation done | Evaluation done | Same as predicate |
| Type Of Electrodes | Active dry | Active, wet and dry | Same as predicate<br>for dry |
| Reprocessing | Cleaning of all<br>components with<br>isopropyl based wipes | Cleaning with<br>multistage enzymatic<br>cleaner of cap and wet<br>electrodes | Similar<br>reprocessing but<br>equivalent in safety<br>and effectiveness |
| | | Disinfection of cap<br>and wet electrodes<br>with low level<br>disinfectant based on<br>Glucoprotamin | |
| | | Cleaning of dry<br>electrodes with<br>cleaning wipe | |
| | | Disinfection of dry<br>electrodes with<br>disinfection wipe | |
| Safety Standards | IEC 60601-1 / AAMI<br>ANSI ES60601-1 | IEC 60601-1 / AAMI<br>ANSI ES60601-1 | Similar to predicate<br>but equivalent in<br>safety and<br>effectiveness |
| | IEC 60601-1-2 | IEC 60601-1-2 | |
| | IEC 60601-2-26 | IEC 60601-2-26 | |
| | IEC 80601-2-26 | | |
| | IEC 60601-2-40 (not<br>applicable) | IEC 60601-2-40 | |
| | ISO 14971 | ISO 14971 | |
| | IEC 62304 / AAMI<br>ANSI IEC 62304 | IEC 62304 / AAMI<br>ANSI IEC 62304 | |
| | AAMI ANSI ISO<br>10993-1 | AAMI ANSI ISO<br>10993-1 | |
| | AAMI ANSI IEC<br>62366 | AAMI ANSI IEC<br>62366 | |
| | IEEE 2010-2012 | IEEE 2010-2012 | |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows a logo with the letters "CGX" inside of a circle. The circle is split vertically into two halves. The left half is a light teal color, and the right half is a darker blue color. The letters "CGX" are white and are placed in the center of the circle, overlapping both halves.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows a logo with the letters "CGX" inside a circle. The circle is divided vertically into two halves, with the left half being a light teal color and the right half being a darker blue color. The letters "CGX" are white and are positioned in the center of the circle, spanning both halves.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows a logo with the letters "CGX" inside of a circle. The circle is divided vertically into two halves. The left half is a light teal color, and the right half is a darker blue color. The letters "CGX" are white and are placed in the center of the circle, with each letter appearing to be in its own quadrant.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows a logo with the letters "CGX" inside of a circle. The circle is divided vertically into two halves. The left half is a light teal color, and the right half is a darker blue color. The letters "CGX" are white and are placed in the center of the circle, with the "C" on the left side and the "GX" on the right side.
### Substantial Equivalence Discussion
The CGX Quick-20m is substantially equivalent to the g.Nautilus PRO based on indications for use and comparison of functional and technological characteristics. Both devices are intended to provide the healthcare professional with an electroencephalogram (EEG) and transmit this data wirelessly to a computer. Both products fit on the head, utilize dry sensors and have the same amplifier/PC interface.
The minor differences between the CGX Quick-20m and the g.Nautilus PRO do not raise new safety or effectiveness concerns or questions. These differences are discussed below and in the substantial equivalence table.
The CGX Quick-20m is slightly larger than the g.Nautilus PRO (predicate), has slightly less input range (+/- 300 mV), a 500 Hz/channel sampling rate versus the user selectable sampling rate of 250 or 500 Hz/channel on the predicate, a battery chemistry of 3.0V DC versus the 3.7V DC of the predicate, and consumes slightly less power than the predicate. None of the aforementioned differences impact safety nor effectiveness. Further, the g.Nautilus PRO (predicate) offers both wet and dry sensors, while the Quick-20m offers dry sensors.
The CGX Quick-20m device was evaluated by a third party (Intertek) for all applicable safety standards and passed without anomalies.
Further, the company completed human factors and usability evaluation according to applicable versions of IEC 62366 and applicable FDA guidance, and the device demonstrated usability across sizes and positioning, resulting in quality, usable EEG data.
Specifically, the human factors usability program consisted of three components. An In-Field Human Factors test, six published academic studies, and service and repair data.
Bench testing on the Ouick-20m includes Intertek certifications to 80601-2-26. in addition to internal company design verification testing that successfully demonstrated compliance to user inputs, and in accordance with 21 CFR 820.40. Clinical testing on the Quick-20m was not applicable.
## Animal Studies
This submission does not require nor include data on animal testing.
## Clinical Studies
Clinical studies were not required to demonstrate that this device is as effective and at least as safe as the predicate device.
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows a logo with the letters "CGX" inside of a circle. The circle is divided vertically into two halves. The left half is a light blue-green color, and the right half is a darker blue color. The letters "CGX" are white and are placed in the center of the circle, overlapping both halves.
## Conclusion:
CGX has been commercializing this EEG technology for research purposes (i.e. not clinical use) since 2015 and in over tens of thousands of uses there have been no safety related customer complaints or quality concerns. This customer experience was leveraged as the platform for the development of the medical device version - the CGX Quick-20m.
A majority of the characteristics of the CGX Quick-20m are identical to those of the g.Nautilus PRO (predicate). The minor differences discussed above and in the Substantial Equivalence Table raise no safety concerns and do not impact effectiveness of the acquisition of EEG. The CGX Quick-20m performs as effectively as the predicate device for the same intended use with no safety concerns.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.